K Number
K122047

Validate with FDA (Live)

Device Name
ALAMO P
Date Cleared
2012-09-04

(54 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Alamo P is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as back pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment prior to treatment with an intervertebral cage. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. The device system must be used with supplemental fixation and autograft to facilitate fusion and is to be implanted via a posterior approach.

Device Description

The Alamo P is used for spinal fusion surgery to provide support and structural stability at the fusion site following discectomy. The device is manufactured from PEEK Optima® LT1 per ASTM F2026 and includes tantalum markers per ASTM F560 for radiographic visualization. The device footprint has a hollow centre to accommodate bone graft to facilitate bone integration and fusion between the end plates from a posterior (PLIF) surgical approach. The device is available in various heights to accommodate variability among patients and the inferior and superior surfaces are designed with ridges to improve fixation and stability and prevent back out and migration.

AI/ML Overview

The manufacturer conducted mechanical testing to demonstrate substantial equivalence to the predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

Test NameAcceptance Criteria (Predicate)Reported Device Performance (Alamo P)
Static CompressionMet requirements of ASTM F2077 and F2267Functioned as intended; substantially equivalent to predicate device
Dynamic CompressionMet requirements of ASTM F2077 and F2267Functioned as intended; substantially equivalent to predicate device
SubsidenceMet requirements of ASTM F2077 and F2267Functioned as intended; substantially equivalent to predicate device
ExpulsionMet requirements of ASTM F2077 and F2267Functioned as intended; substantially equivalent to predicate device

2. Sample Size and Data Provenance for Test Set:

The submission states "worst case device" was used for mechanical testing. No specific sample sizes for repeat tests are provided. Given the nature of mechanical testing, samples would likely be manufactured in the country of origin of the device (not explicitly stated, but the contact person is in the US) and tested in a laboratory. This is a laboratory-based, prospective test.

3. Number of Experts and Qualifications for Ground Truth:

Not applicable. This device is an intervertebral body fusion device and its performance is evaluated through mechanical testing, not through expert-labeled diagnostic data.

4. Adjudication Method for Test Set:

Not applicable. Mechanical testing does not involve human adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

Not applicable. This is a physical device, not an AI or diagnostic tool requiring human reader studies.

6. Standalone (Algorithm Only) Performance:

Not applicable. This is a physical device, not an algorithm. Its performance is assessed through its mechanical properties and structural integrity.

7. Type of Ground Truth Used:

The ground truth is based on the established engineering standards for intervertebral body fusion devices, specifically ASTM F2077 (Standard Test Methods for Static and Dynamic Axial Compression Fatigue Properties of Spinal Arthrodesis Implants) and ASTM F2267 (Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression). The predicate device's performance against these standards serves as the benchmark for substantial equivalence.

8. Sample Size for Training Set:

Not applicable. This is a physical device, not a machine learning model requiring a training set.

9. How Ground Truth for Training Set Was Established:

Not applicable.

{0}------------------------------------------------

510(k) Summary

Summary of 510(k) Safety and Effectiveness

SEP 4 2012

Submitted By:

Alliance Partners. LLC 121 Interpark Blvd. #601 San Antonio, TX 78216

July 6, 2012

Contact Person:

Date:

Contact Telephone:

Device Trade Name: Device Classification Name:

Device Classification: Reviewing Panel: Regulation Number: Product Code: Predicate Device:

Jennifer Palinchik Development and Regulatory Consultant (440) 933-8850

Alamo P Intervertebral Body Fusion Device with Bone Graft, Lumbar Class II Orthopedic 888.3080 MAX Medtronic Sofamor Danek CAPSTONE® Spinal System (K073291)

Device Description:

The Alamo P is used for spinal fusion surgery to provide support and structural stability at the fusion site following discectomy. The device is manufactured from PEEK Optima® LT1 per ASTM F2026 and includes tantalum markers per ASTM F560 for radiographic visualization.

The device footprint has a hollow centre to accommodate bone graft to facilitate bone integration and fusion between the end plates from a posterior (PLIF) surgical approach. The device is available in various heights to accommodate variability among patients and the inferior and superior surfaces are designed with ridges to improve fixation and stability and prevent back out and migration.

Intended Use:

The Alamo P is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as back pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment prior to treatment with an intervertebral cage. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. The device

Page 1 of 2

{1}------------------------------------------------

system must be used with supplemental fixation and autograft to facilitate fusion and is to be implanted via a posterior approach.

Substantial Equivalence Information:

The design features, material, and indications for use of the Alamo P device are substantially equivalent to the predicate device listed above. The safety and effectiveness is adequately supported by the substantial equivalence, material information, and analysis data provided within this Premarket Notification.

ItemAlamo PCAPSTONE® Spinal System
Product CodeMAXMAX
Classification NameIntervertebral Body FusionDeviceSame
Intended Usedegenerative disc disease(DDD) at one or twocontiguous levels of thelumbosacral spine (L2-S1)Same
FootprintBlocks have one width and avariety of heights andlengths with axial and lateralopenings for bone graft androws of teeth.Same
Teeth to preventmigrationLocated on superior andinferior surfaces. LinearpatternSame
Radiographic markersYesYes
Axial Footprintdimensions10 x 22mm, 10 x 25mm, 10x 28mm10 x 22mm, 10 x 26mm, 10 x32mm, 10 x 36mm
MaterialPEEK Optima LT1 andTantalum (markers)Same

Predicate Comparison Summary Table:

Mechanical Testing:

Performance testing was conducted via the following mechanical tests per ASTM F2077 and F2267 using the worst case device: Static Compression, Dynamic Compression, Subsidence, and Expulsion. The device functioned as intended and the performance results show that the Alamo P is substantially equivalent to the predicate device.

Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings outstretched, and a caduceus symbol at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Alliance Partners, LLC % Ms. Jennifer Palinchik Development and Regulatory Consultant 121 Internark Boulevard. Suite 601 San Antonio, Texas 78216

SEP 4 2012

Re: K122047

Trade/Device Name: Alamo P Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 10, 2012 Received: July 12, 2012

Dear Ms. Palinchik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include reguirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Jennifer Palinchik

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melk son Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K122047

Device Name: Alamo P

Indications for Use:

The Alamo P is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as back pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment prior to treatment with an intervertebral cage. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. The device system must be used with supplemental fixation and autograft to facilitate fusion and is to be implanted via a posterior approach.

Prescription Use ਮ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Carl-Affi

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K122047

Page 1 of 1

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.