(213 days)
The ESA615 Electrical Safety Analyzer is an electronic signal source and measurement device for verifying the electrical safety of medical devices. The ESA615 also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. The ESA615 provides following function categories: ECG Functions, ECG-Performance Testing.
Fluke Biomedical's ESA615 Electrical Safety Analyzer (hereafter referred to as the ESA615) provides a basis for verifying the electrical safety of medical devices. The Product also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. The ESA615 can be controlled by a suitably equipped computer to carry out tests, via either a cable or wirelessly. The ESA615 consists of the following components: 1) Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors. 2) Plastic injection molded case parts. 3) Liquid Crystal Display for user interface. 4) Power cord for powering the unit at 120V and 60Hz.
Here's a summary of the acceptance criteria and study information for the ESA615 Electrical Safety Analyzer, based on the provided text:
Device: ESA615 Electrical Safety Analyzer
1. Table of Acceptance Criteria and Reported Device Performance
The submission claims substantial equivalence to the predicate device (MPS450) based on similar technological characteristics and performance, not against specific, quantifiable acceptance criteria. The "differences" column below highlights how ESA615 performance differs from the predicate.
| Feature | Acceptance Criteria (Predicate: MPS450) | Reported Device Performance (ESA615) | Met Criteria? (Comparison to Predicate) |
|---|---|---|---|
| Intended Use | Test and verify basic operation of patient monitoring devices for ECG, Respiration, Invasive Blood Pressure, and Cardiac Output. | Verify electrical safety of medical devices; provide ECG simulation and performance waveforms. | Differs: Fewer functions (no Respiration, Invasive Blood Pressure, Cardiac Output), but adds Electrical Safety Analysis. |
| Construction | Plastic Case | Plastic Case | Yes |
| Size | 5.5" wide x 7.5" deep x 1.8" high | 6.94" wide x 3.3" deep x 11.2" high | Bigger |
| Weight | 2 lbs | 3.5 lbs | Heavier |
| Display | 4 line x 20-character super twist LCD display | 3" diagonal screen Monochrome STN transflective | Larger screen |
| Function Key | Soft | Soft | Yes |
| ECG Leads | 10 binding posts | 5 applied parts jacks (with optional expander box) | Fewer applied parts |
| Communications Port | RS232 | USB | Change from RS232 to USB |
| Power | 9V alkaline battery & battery eliminator | No Battery - AC line powered only | No battery due to higher power demands |
| Lead Configuration | 12 leads | 5 leads | Fewer leads |
| Amplitude Accuracy | ±2% of setting | ± 5% of 1mV setting | Less accurate |
| Rate Accuracy | ±1% setting | ± 2% of setting | Less accurate |
| Normal Sinus Rhythm | 30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 bpm | 30, 60, 120, 180, 240 bpm | Fewer NSR waves |
| Sine wave | 0.5, 5, 10, 40, 50, 60, 100 Hz | 10, 40, 50, 60, 100 Hz | Fewer frequencies |
| Square wave | 0.125, 2.0 Hz | 0.125, 2.0 Hz | Yes |
| Triangle wave | 2.0, 2.5 Hz | 2.0 Hz | Fewer frequencies |
| Pulse wave | 30, 60 bpm, 60 ms pulse width | 30, 60 bpm, 63 ms pulse width | Wider pulse width |
| Cable Connector | ECG leads, 10 binding postings | ECG leads, 5 binding postings | Fewer binding postings |
2. Sample size used for the test set and the data provenance
The document mentions "Laboratory studies have been conducted with a representative patient monitor..." but does not specify the sample size (number of "representative patient monitors" or tests performed) or the data provenance (country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The device is an analyzer, not a diagnostic tool requiring expert interpretation. The "ground truth" for its performance is its physical and electronic output measurements compared against its published specifications.
4. Adjudication method for the test set
Not applicable, as this is a device performance verification, not a clinical study requiring human interpretation and multi-expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an electrical safety analyzer and ECG simulator, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the testing described appears to be standalone performance verification of the ESA615 itself, against its published specifications. The description "Laboratory studies have been conducted with a representative patient monitor to verify and validate the ESA615 will perform within its' published specifications" indicates this.
7. The type of ground truth used
The ground truth for the test set appears to be its own published specifications, as verified by "Laboratory studies... to verify and validate the ESA615 will perform within its' published specifications."
8. The sample size for the training set
Not applicable. This is a hardware device with firmware, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Fluke Biomedical. The logo is black and white and features the word "FLUKE" in large, bold letters. Below the word "FLUKE" is the word "Biomedical" in smaller letters. In the upper right corner of the image is the fraction "1/4" written in black ink.
Section III -510(k) Summary of Safety and Effectiveness
Submitter:
JAN 2 5 2013
Fluke Biomedical 6920 Seaway Blvd Everett WA. 98203 (440) 498-2579 -Phone (440-349-2307) - Fax John Nelson - Contact Person
Device Name:
- Trade Name - ESA615 Electrical Safety Analyzer
- Common Name Analyzer .
- Classification Name Cardiac monitor (including cardiotachometer and rate alarm), per 21 CFR & ● 870.2300
- Product Codes -DRT .
Devices for Which Substantial Equivalence is Claimed:
- MPS450
Device Description:
Principles of Operation
Fluke Biomedical's ESA615 Electrical Safety Analyzer (hereafter referred to as the ESA615) provides a basis for verifying the electrical safety of medical devices. The Product also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. The ESA615 can be controlled by a suitably equipped computer to carry out tests, via either a cable or wirelessly.
Technological Characteristics
The ESA615 consists of the following components:
-
Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors.
-
Plastic injection molded case parts.
-
- Liquid Crystal Display for user interface.
-
- Power cord for powering the unit at 120V and 60Hz.
Intended Use of the Device:
The Product is an electronic signal source and measurement device for verifying the electrical safety of medical devices. The Product also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications.
{1}------------------------------------------------
The Product provides the following function categories:
- ECG Functions .
- ECG-Performance Testing .
The intended user is a trained biomedical equipment technician who performs periodic preventative maintenance checks on patient monitors in service. Users can be associated with hospitals, clinics, original equipment manufacturers and independent service companies that repair and service medical equipment.
The end user is an individual, trained in medical instrumentation technology. This Product is intended to be used in the laboratory environment, outside of the patient care area, and is not intended for use on patients, or to test devices while connected to patients. This Product is not intended to be used to calibrate medical equipment.
ESA615 is intended for over the counter use.
Summary of Technological Characteristics:
The ESA615 is substantially equivalent to one other legally marketed and FDA approved device in the United States. The ESA615 functions in a manner similar to and is intended for the similar use as the MPS450 manufactured by Fluke Biomedical. The ESA615 is similar to MPS450 in that it uses LCD display, and allows user to simulate physiological parameter to verify the operation of patient monitors. The ESA615 differs from MPS450 in that ESA615 works only using AC power cord and has additional options of performing electrical safety analysis.
| Features | ESA615 | MPS450V | Difference |
|---|---|---|---|
| Intended Use | The Product is an electronicsignal source and measurementdevice for verifying theelectrical safety of medicaldevices. The Product alsoprovides ECG simulation andperformance waveforms toverify patient monitors areperforming within theiroperating specifications.The Product provides thefollowing function categories:• ECG Functions• ECG-PerformanceTestingThe intended user is a trainedbiomedical equipment | The intended use of MPS450 isto test and verify the basicoperation of patient monitoringdevices or systems used tomonitor various physiologicalparameters of patient, includingECG, Respiration, InvasiveBlood Pressure, and CardiacOutput.The intended user is a trainedbiomedical equipment technicianwho is performing periodicpreventative maintenance checkson patient monitors in service.Users can be associated withHospitals, clinics, originalequipment manufacturer orequipment | Fewer functions,Respiration,Invasive BloodPressure, CardiacOutput |
| Biomedical | |||
| technician who performsperiodic preventativemaintenance checks on patientmonitors in service. Users canbe associated with hospitals,clinics, original equipmentmanufacturers and independentservice companies that repairand service medicalequipment.The end user is an individual,trained in medicalinstrumentation technology.This Product is intended to beused in the laboratoryenvironment, outside of thepatient care area, and is notintended for use on patients, orto test devices while connectedto patients. This Product is notintended to be used to calibratemedical equipment. It isintended for over the counteruse. | independent service companiesthat repair and service medicalequipment. The end user istechnically trained individual,specializing in medicalinstrumentation technology.The MPS450 is intended to beused in the laboratoryenvironment and is not intendedfor use on patients, or to testdevices while connected topatients. The MPS450 is notintended to be used to calibratemedical equipment and notintended for over the counteruse. | 3/4 | |
| Construction | Plastic Case | Plastic Case | None |
| Size | 6.94" wide x 3.3" deep x 11.2"high | 5.5" wide x 7.5" deep x 1.8"high | Bigger |
| Weight | 3.5 lbs | 2lbs | Heavier |
| Display | 3" diagonal screenMonochrome STNtransflective | 4 line x 20-character super twistLCD display | Larger screen |
| Function Key | Soft | Soft | None |
| ECG Leads | 5 applied parts jacks and easyECG snap connection withoptional expander box | 10 binding posts; compatiblewith disposable snaps, 3.2 mmor 4 mm electrodes, and bananaplugs | Fewer appliedparts on ESA615 |
| Communications Port | USB | RS232 | Change fromRS232 to USBdata port withadvancement intechnology |
| Power | No Battery - AC line poweredonly | 9V alkaline battery with lowbattery indicator; or batteryeliminator; transformer certifiedto CSA | No battery becauseof higher powerrequirements |
000144
Blomedica
{2}------------------------------------------------
:
- 0001455
.
・・
{3}------------------------------------------------
| FLuke. |
|---|
| Biomedical |
000146
| LeadConfiguration | 5 leads | 12 leads | None |
|---|---|---|---|
| AmplitudeAccuracy | ± 5% of 1mV setting | ±2% of setting | Less accurate |
| Rate Accuracy | ± 2% of setting | ±1% setting | Less accurate |
| Normal SinusRhythm | 30, 60, 120, 180, 240 bpm | 30, 40, 45, 60, 80, 90, 100, 120,140, 160, 180, 200, 220, 240,260, 280, 300 bpm | Fewer NSR waves |
| Sine wave | 10, 40, 50, 60, 100 Hz | 0.5, 5, 10, 40, 50, 60, 100 Hz | Fewer frequencies |
| Square wave | 0.125, 2.0 Hz | 0.125, 2.0 Hz | None |
| Triangle wave | 2.0 Hz | 2.0, 2.5 Hz | Fewer frequencies |
| Pulse wave | 30, 60 bpm, 63 ms pulse width | 30, 60 bpm, 60 ms pulse width | Wider pulse width |
| CableConnector | ECG leads, 5 binding postings | ECG leads, 10 binding postings | Fewer bindingpostings |
Substantial Equivalence:
The ESA615 is substantially equivalent to one other legally marketed device in the United States. The ESA615 functions in a manner similar to and is intended for the same use as the MPS450 manufactured by Fluke Biomedical.
The ESA615 is similar to the MPS450 in that it uses LCD display, and allows user to simulate ECG parameters to verify the operation of patient monitors. The ESA615 differs from the MPS450 in that the ESA615 is not battery operated; it performs Electrical Safety Analysis and does not perform respiration, blood pressure or cardiac output.
Non-Clinical Test Data:
Laboratory studies have been conducted with a representative patient monitor to verify and validate the ESA615 will perform within its' published specifications -- Document: NPI-05012012-00002
The ESA615 software has been successfully validated to confirm the performance of the device. Document: NPI-03012012-00001 and NPI-05042012-00001
Clinical Test Data:
Clinical testing has not been conducted on this product.
Conclusion:
Based upon the laboratory studies, similar technological/performance characteristics as compared to the predicate device, and successful validation of the ESA615 software, the performance of the ESA615 is deemed to be substantially equivalent to the MPS450.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUM AN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
JAN 2 5 2013
Fluke Biomedical c/o Mr. John Nelson Director of Regulatory/Quality Affairs 6045 Cochran Rd. Solon, OH 44139
Re: K121860
Trade/Device Names: ESA615 Regulatory Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (Two) Product Code: DRT Dated: January 7, 2013 Received: January 8, 2013
Dear Mr. Nelson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. John Nelson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Owen P. Faris -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K12if60
Image /page/6/Picture/1 description: The image shows the Fluke Biomedical logo. The logo consists of the word "FLUKE" in large, bold, sans-serif font, with a stylized design above the letters. Below the word "FLUKE" is the word "Biomedical" in a smaller, bold, sans-serif font. The image is black and white.
Indications for Use
510(k) Number (if known): 长しさした 6 0
Device Name: ESA 615 Electrical Safety Analyzer
Indications for Use:
The ESA615 Electrical Safety Analyzer is an electronic signal source and measurement device for verifying the electrical safety of medical devices. The ESA615 also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications.
The ESA615 provides following function categories:
- ECG Functions .
- ECG-Performance Testing
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
000004
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) .
Concurryal; goff CDRJL Office_of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Devices
Kr21860 510(k) Number_
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).