K Number
K121789

Validate with FDA (Live)

Device Name
SHANGRING
Date Cleared
2012-08-03

(45 days)

Product Code
Regulation Number
884.4530
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ShangRing™ device is indicated for circumcision of adult males, defined as circumferential excision of the foreskin or prepuce at or near the level of the coronal sulcus.

Device Description

The ShangRing is a sterile, single use, disposable circumcision device consisting of two concentric rings made from polycarbonate. The outer ring is formed from two semi-circular halves that are joined together on one end by a hinge, and by interlocking ratchets on the opposite end. The inner ring is lined with a soft silicone band. The ShangRing is available in 14 sizes with inner ring diameters ranging from 26mm to 40mm. A disposable sizing guide is provided to assist in selecting the appropriate size ShangRing. The ShangRing is designed to perform a circumferential crushing of the foreskin (prepuce) between the inner ring and outer ring. The inner ring is placed over the penis, at the level of the coronal sulcus. After local anesthesia, the healthcare practitioner everts the patient's foreskin over the inner ring and the outer ring halves are locked together, sandwiching the foreskin between the two rings which effectively prevents bleeding. Skin forward of the crush line is trimmed away. After the procedure, the ShangRing device remains on the patient for 7 days. After 7 days, the ShangRing is removed by the healthcare practitioner.

AI/ML Overview

The provided text describes the ShangRing™ circumcision device and its regulatory submission. While it mentions clinical evaluations, it does not explicitly define acceptance criteria in a quantitative manner or provide a detailed study report that proves the device meets specific performance metrics in a structured way that would allow for the direct extraction of acceptance criteria and reported performance in a table format.

However, I can infer the "acceptance criteria" and "reported device performance" based on the general conclusions drawn from non-clinical and clinical tests as stated in the document.

Here's an attempt to extract and present the information requested, with necessary inferences and notations where direct data is absent:

Derived Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred)Reported Device Performance (Inferred from study conclusions)
BiocompatibilityMust be biocompatible for prolonged contact with intact skin.Confirmed to be biocompatible for its intended use according to ISO 10993-1 for a surface device in prolonged contact with intact skin.
Mechanical Strength- Outer ring locking mechanism must not disengage during normal use conditions. - Device must remain in place on the patient (skin retention) when exposed to anticipated forces.- Force required to pull apart the locking mechanism was determined.- Force required to pull skin out of the ShangRing in the locked position was measured.Conclusion: Possesses the required mechanical strength to prevent bleeding and remain attached to the foreskin for at least 7 days following closure of the locking ring.
Clinical Safety- Infrequent adverse events. - Mild adverse events, requiring minimal management. - Complication rates lower than conventional methods.- Adverse events were infrequent. - Adverse events were mild, requiring minimal management. - Complication rates were lower than experienced with conventional methods. Overall Conclusion: ShangRing™ is a safe means of performing circumcision.
Clinical Effectiveness- Easily applied. - Short surgical procedure. - No suturing required. - Performs circumcision effectively.- Easily applied in a short surgical procedure. - No suturing required. Overall Conclusion: ShangRing™ is an effective means of performing circumcision on adult males. The device achieves circumferential crushing of the foreskin.
Wear DurationDevice should remain on the patient for 7 days.Mechanical testing and clinical studies support the device remaining attached for at least 7 days, allowing for effective crushing and subsequent removal. (Implicit in the "remains attached for at least 7 days" and "device remains on the patient for 7 days" statements).

Here's the breakdown of the other requested information based on the provided text:

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Approximately 3000 cases.
  • Data Provenance: Clinical evaluations were conducted in China and Kenya. The text does not explicitly state whether these studies were retrospective or prospective, but clinical evaluations for device efficacy and safety are typically prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • The document mentions "clinical evaluations" and "clinical data" but does not specify the number of individual experts (e.g., surgeons, physicians) involved in establishing the ground truth or their specific qualifications (e.g., years of experience, specialization). It refers to healthcare practitioners performing the procedures.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • The document does not describe any specific adjudication method for establishing ground truth or evaluating outcomes in the clinical studies. It broadly states "clinical evaluations" and "clinical data collected confirms."

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, this is not applicable. The ShangRing™ is a physical surgical device (circumcision clamp), not an AI-assisted diagnostic or therapeutic tool for which MRMC comparative effectiveness studies (involving human readers improving with AI) would be relevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • No, this is not applicable. The ShangRing™ is a physical surgical device, not an algorithm. Therefore, "standalone" performance in the context of an algorithm does not apply. The device's performance is inherently tied to its use by a human healthcare practitioner.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The ground truth for the clinical studies appears to be based on clinical outcomes data (e.g., successful circumcision, absence/presence and severity of adverse events, complication rates, ease of application, duration of procedure, requirement for sutures). These outcomes would have been observed and recorded by the treating healthcare practitioners.

8. The sample size for the training set

  • The concept of a "training set" typically applies to machine learning algorithms. Since the ShangRing™ is a physical medical device, there isn't a "training set" in that sense. The device's design and mechanical properties are refined through engineering and non-clinical testing, while clinical "training" for the device's efficacy and safety for human use relies on the reported clinical evaluations.
  • If "training set" is interpreted as initial smaller clinical trials or studies that informed the final design and understanding before the reported "extensive clinical evaluations," the document does not provide details on such distinct phases or their sample sizes. The "approximately 3000 cases in these two countries" represents the cumulative clinical experience reported to demonstrate performance.

9. How the ground truth for the training set was established

  • As above, the concept of a "training set" in the context of an AI/algorithm is not applicable here. For the clinical evaluations (which serve as the basis for understanding the device's performance and implicit "truth"), the ground truth was established through direct observation of clinical outcomes by medical professionals in China and Kenya. These outcomes included the successful completion of circumcision, the occurrence and nature of adverse events, and complication rates, as documented during standard clinical practice and research protocols.

{0}------------------------------------------------

3 2012 AUG

<121789
page 1 of 4

510(k) Summary for the Wuhu Snnda Medical Treatment Appliance Technology Co., Ltd. ShangRing™

(per 21CFR 807.92 and lttp://www.fda.gov/MedicalDeviceRegulationandGuidance/default.htm)

1. SUBMITTER/510(K) HOLDER

Wuhu Snnda Medical Treatment Appliance Technology Co., Ltd. 4F Overseas Student Pioneer Park Science Innovation Center Economic & Technology Zone North Yinhu Road Wuhu, Anhui, China 241001

Contact Person:Mr. Jian Zhong Shang
Telephone:+(86) 553-5844860 or +(86) 553-5963971
Fax:+(86) 553-5844880

Date Prepared: June 15, 2012

2. DEVICE NAME

Trade Name:ShangRing™
Common Name:Circumcision clamp
Classification Name:Obstetric-Gynecologic Specialized Manual Instrument, Clamp,Circumcision, 21 CFR 884.4530
Device Class:Class II
Classification Panel:Obstetrical and Gynecological Panel
Product Code:HFX

3. PREDICATE DEVICES

  • · Plastibell (Hollister, pre-amendment device)
  • · Gomco Circumcision Clamp (Thomas Medical Inc., K926353) Type (K926535)
  • · Smart Klamp (Emergo Group, Inc., K032091)

4. DEVICE DESCRIPTION

The ShangRing is a sterile, single use, disposable circumcision device consisting of two concentric rings made from polycarbonate. The outer ring is formed from two semi-circular halves that are joined together on one end by a hinge, and by interlocking ratchets on the opposite end. The inner ring is lined with a soft silicone band.

{1}------------------------------------------------

K121789
page 2 of 4

The ShangRing is available in 14 sizes with inner ring diameters ranging from 26mm to 40mm. A disposable sizing guide is provided to assist in selecting the appropriate size ShangRing.

The ShangRing is designed to perform a circumferential crushing of the foreskin (prepuce) between the inner ring and outer ring. The inner ring is placed over the penis, at the level of the coronal sulcus. After local anesthesia, the healthcare practitioner everts the patient's foreskin over the inner ring and the outer ring halves are locked together, sandwiching the foreskin between the two rings which effectively prevents bleeding. Skin forward of the crush line is trimmed away.

After the procedure, the ShangRing device remains on the patient for 7 days. After 7 days, the ShangRing is removed by the healthcare practitioner.

5. INDICATION FOR USE/INTENDED USE

!!

·

The ShangRing device is indicated for circumcision of adult males, defined as circumferential excision of the foreskin or prepuce at or near the level of the coronal sulcus.

6. TECHNOLOGICAL CHARACTERISTICS

The ShangRing and the predicate devices Plastibell, Gomco, and Smart Klamp all achieve circumcision by isolating the foreskin and applying pressure to form a crush line.

The ShangRing and the Smart Klamp are both constructed from plastic and are designed for the healthcare practitioner to position the foreskin over one component and engage a locking mechanism on a companion component. When the lock is engaged, the foreskin is crushed between plastic components.

The overall design of the predicate Plastibell and Gomco devices is similar. However, for both the Plastibell and Gomco, the foreskin is everted over a bell-shaped component (plastic for Plastibell, stainless steel for Gomco). With the Plastibell device, crushing is accomplished by a piece of string. With the Gomco device, a stainless steel clamp is applied.

7. NON-CLINICAL TESTING

Wuhu Snnda conducted mechanical testing of the ShangRing to (1) determine the force required to pull apart the ShangRing's locking mechanism and (2) measure the force required to pull skin out of the ShangRing when it is in the locked position.

{2}------------------------------------------------

K121789
page 3 of 4

These tests were conducted to supplement the clinical data collected to demonstrate that while the patient is wearing the ShangRing for the indicated period of time, when exposed to the forces anticipated during normal use conditions, the outer ring will not disengage and that the device will remain in place on the patient.

Finished, sterilized ShangRing devices were tested for biocompatibility according to ISO 10993-1 for a surface device in prolonged contact with intact skin. The results confirm that the ShangRing is biocompatible for its intended use.

8. CLINICAL TESTING

The ShangRing has been the subjective of extensive clinical evaluations, both in China and Kenya. Approximately 3000 cases in these two countries have been reported in clinical literature. These studies have demonstrated that the ShangRing is easily applied in a short surgical procedure with no suturing required. Adverse events were infrequent, and mild - requiring minimal management. Complication rates were lower than experienced with conventional methods. The clinical data collected confirms that the ShangRing is a safe and effective means of performing circumcision on adult males.

9. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS

The ShangRing has the same intended use and similar technological characteristics as the predicate devices. Both the ShangRing and predicate devices achieve circumcision by performing circumferential crushing of the foreskin. Non-clinical testing demonstrates that the ShangRing is biocompatible and possesses the required mechanical strength to prevent bleeding and remain attached to the foreskin for at least 7 days following closure of the locking ring. Clinical studies have demonstrated that the ShangRing is safe and effective for circumcision in the target population.

The data collected confirms that the differences in the design between the ShangRing and the predicate devices do not raise any new issues of safety and effectiveness. The information and data collected support a claim of substantial equivalence of the ShangRing to the specified predicate devices.

{3}------------------------------------------------

Regulatory StatusIntended UseShangRingHollisterPlastibellThomas Medical Inc.Gomco Circumcision ClampK926535Emergo Group, Inc.Smart Klamp®K032091
ProposedThe ShangRing™ device is indicated forcircumcision of adult males, defined ascircumferential excision of the foreskin orprepuce at or near the level of the coronalsulcus.Pre-amendment deviceThe plastic bell with handle and string isused to tie off the foreskin or prepuce tostop blood flow such that the excessforeskin forward of the string can be cut offwith minimal bleeding.This clamp is used to hold the foreskin ofthe penis in place while it is being cutduring a circumcision procedure.The device is indicated for circumcisionof newborns and older males, defined ascircumferential excision of the foreskin orprepuce at or near the level of the coronalsulcus, with minimal amount of preputialskin remaining.
Patient populationAdult malesNewborns to adultsNewborns to adultsNewborns to older males
Mechanism ofActionCircumferential crushing of the foreskin(i.c., crush the preputial skin between theouter plastic ring and inner plastic ring)Circumferential crushing of the foreskinCircumferential crushing of the foreskin(i.e., crush the preputial skin around therim of the Gomco bell and an appropriatesized hole in the base plate)Circumferential crushing of the foreskin(i.e., crush the foreskin between the innertube and outer clamp)
Sizes14 sizes (2.6, 2.8, 2.9, 3.0, 3.1, 3.2, 3.3,3.4, 3.5, 3.6, 3.7, 3.8, 3.9, 4.0 cm)6 sizes (1.1, 1.2, 1.3, 1.4, 1.5, 1.6, 1.7 cm)8 sizes (1.1, 1.3, 1.6, 2.1, 2.6, 2.9, 3.2, 3.5cm)4 sizes (1.0, 1.3, 1.6, 2.1 cm)
Material(s)Inner and outer rings: Lexan(polycarbonate)Soft lining surrounding inner ring:SiliconeBell: PlasticString: sutureStainless steelCylinder: transparent polycarbonateOuter clamp: nylon
Dorsal incisionYes, in cases when foreskin is too tight toevert. This may occur in patients withoutphimosis.YesYesNo
Need for sutureNoNoNoNo
AdditionalinstrumentsneededClamp, blunt end scissors, surgicalscissors, 1/8" or 3 mm flat bladedscrewdriverClamp, surgical scissorsClamp, Cold knifeClamp, scissors
Device remains onthe patient afterprocedureYes, removed after 7 daysYes, falls off in 3-7 days*NoYes, removed after 5 days**
Single UseYesYesNoYes
Provided SterileYesYesNoYes
BiocompatibilityEstablishedYesInformation not availableInformation not availableInformation not available
http://www.circlist.com/instrstechs/plastibell.htmlhttp://smartcircumcision.com/smartklamp_2.html

'able 5-1. Side-by-Side Comparison of the ShangRing with Predicate Devic

CONFIDENTIAL
Page 5-4

June 15, 2012

ShangRing™, Traditional 510(k)
Wuhu Snnda Medical Treatment Appliance Technology Co.

K121789
Page 4 of Y

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Wuhu Snnda Medical Treatment Appliance Technology Co., Ltd. % Cynthia J.M. Nolte, Ph.D. Principal Regulatory Consultant Aptiv Solutions 49 Plain Street NORTH ATTLEBORO MA 02760

Re: K121789

Trade/Device Name: ShangRingTM Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: HFX Dated: June 15, 2012 Received: June 19, 2012

Dear Dr. Nolte:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

AUG 3 2012

{5}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21.CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin K. Ticho

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

KI21789

Indications for Use

510(k) Number (if known): K121789

ShangRing™ Device Name:

Indications for Use:

The ShangRing™ device is indicated for circumcision of adult males, defined as circumferential excision of the foreskin or prepuce at or near the level of the coronal sulcus.

Prescription Use _x (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hubert Lemmer

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K121789

§ 884.4530 Obstetric-gynecologic specialized manual instrument.

(a)
Identification. An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes.
(2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant.
(3) An umbilical clamp is an instrument used to compress the umbilical cord.
(4) A uterine curette is an instrument used to scrape and remove material from the uterus.
(5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix.
(6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus.
(7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination.
(8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix.
(9) A gynecological cerclage needle is a looplike instrument used to suture the cervix.
(10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus.
(11) A gynecological fibroid screw is an instrument used to hold onto a fibroid.
(12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity.
(13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina.
(14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures.
(15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus.
(16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis.
(17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina.
(18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
(b)
Classification. (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.