K Number
K121706

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-01-14

(220 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FORA V10 No Code Blood Glucose Monitoring System, model TD-4244, is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. The meter contains some speaking functions but has not been validated for use by visually impaired users. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

This system is intended to be used by a single patient and should not be shared.

FORA V10 test strips are for use with FORA V10 No Code Blood Glucose meter model TD-4244 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, forearm, upper arm calf and thigh.

The FORA control solutions can only be used with FORA V10 No Code Blood Glucose meter, model TD-4244 and the FORA V10 test strips to check that meter and test strips are working together properly.

The alternative site testing in FORA V10 No Code Blood Glucose Monitoring systems model TD-4244 can be used only during steady-state blood glucose conditions.

Device Description

The system consists of three main products: the meter, test strips, and control solutions. These products have been designed, tested, and proven to work together as a system to produce accurate blood glucose test results. Use only FORA V10 test strips and FORA control solutions with the FORA V10 No Code Blood Glucose Monitoring System, model TD-4244.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the FORA V10 No Code Blood Glucose Monitoring System:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state specific acceptance criteria in terms of numerical thresholds (e.g., accuracy percentages, precision values) that the device must meet. Instead, the submission for the FORA V10 No Code Blood Glucose Monitoring System, model TD-4244, relies on demonstrating substantial equivalence to a predicate device (FORA V10 Blood Glucose Monitoring System, K093035).

The core of the "performance" claim is that it has the "same performance characteristics as the predicate device" and that "Software verification and validation testing confirmed that the performance, safety and effectiveness of the FORA V10 No Code Blood Glucose Monitoring System are equivalent to the predicate device."

Therefore, the implied acceptance criterion is that its performance is equivalent to the predicate. The "reported device performance" is essentially that it is equivalent to the predicate device, which would have had its performance demonstrated in its own 510(k) submission.

Table of (Implied) Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Implied)Reported Device Performance
Performance of the FORA V10 No Code Blood Glucose Monitoring System (model TD-4244) is equivalent to the predicate device (FORA V10 Blood Glucose Monitoring System, K093035)."The FORA V10 No Code Blood Glucose Monitoring System has the same performance characteristics as the predicate device." "Software verification and validation testing confirmed that the performance, safety and effectiveness of the FORA V10 No Code Blood Glucose Monitoring System are equivalent to the predicate device." "Based on the information provided in this submission, the FORA V10 No Code Blood Glucose Monitoring System is substantially equivalent to the predicate FORA V10 Blood Glucose Monitoring System."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the sample size used for any test set (accuracy, precision, etc.). It only mentions that the system was "compared to the YSI 2300 Glucose Analyzer in the Accuracy studies."
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish Ground Truth and Their Qualifications

  • Number of Experts: This information is not provided in the document.
  • Qualifications of Experts: This information is not provided in the document.

4. Adjudication Method for the Test Set

  • Adjudication Method: This information is not provided in the document.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study Done? No. This device is a blood glucose meter, not an imaging or diagnostic device typically requiring MRMC studies for human reader performance. The "performance characteristics" typically refer to analytical accuracy and precision in comparison to a reference method, rather than human interpretation of results.

6. Standalone (Algorithm Only) Performance

  • Standalone Performance Done? Yes, indirectly. The core of the performance evaluation is the device's ability to quantitatively measure glucose, which is its standalone analytical performance. The document states: "The FORA V10 No Code Blood Glucose Monitoring System is compared to the YSI 2300 Glucose Analyzer in the Accuracy studies." This confirms a standalone performance evaluation against a reference standard.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for the accuracy studies was established by the YSI 2300 Glucose Analyzer. This is a laboratory reference method widely accepted for determining glucose concentrations. The YSI itself is stated to be "calibrated with NIST (SRM) 917A reference material," confirming a traceable and high-quality reference standard.

8. Sample Size for the Training Set

  • Sample Size: The document does not provide information on a training set sample size. This is common for devices like blood glucose meters, as their underlying electrochemical biosensor technology isn't "trained" in the same way a machine learning algorithm is. The "development" and "testing" would involve product development, calibration, and verification/validation, rather than a distinct "training set."

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth Establishment for Training Set: As there's no mention of a "training set" in the context of an algorithm that learns, this question is not directly applicable. The device's calibration and design would rely on established laboratory methods and reference materials (like the NIST (SRM) 917A) during its development and manufacturing, but not in the sense of a machine learning training process with ground truth labels.

{0}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

The Assigned 510(k) number is: K121706

    1. Submitter Information:
      Application Correspondence: Contact Person: Pinjung Chen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Address: 6F, No. 127, Wugong 2nd Rd., Wugu District, New Taipei City, 24888, Taiwan Phone: +886-2-6625-8188 #1176 -Fax: +886-2-6625-0868 Email: pinjung.chen@taidoc.com.tw

Date of submission: May 18th, 2012

Applicant: .. . Company Name: FORA Care Inc. Contact Person: Sophia Wu ' Address: 810 Lawrence Drive, Suite104, Newbury Park, CA 91320 Phone: (805) 498-8188 Fax: (805) 498-7188 E-mail: sophiawu@foracare.com

    1. Device name:
      Proprietary name: FORA V10 No Code Blood Glucose Monitoring System, model TD-4244

Regulatory information:

Regulation section: 21 CFR 862.1345 Glucose Test System, A. 21 CFR 862.1660 Single Analyte Control

B.Classification:Class II, Class I
C.Product Code:NBW, System, Test, Blood Glucose, Over The Counter

{1}------------------------------------------------

CGA. Glucose Oxidase, Glucose, JJX, Single Analyte Control

D. Panel:

Chemistry (75)

3. Intended Use:

FORA V10 No Code Blood Glucose Monitoring System, model TD-4244, is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. The meter contains some speaking functions but has not been validated for use by visually impaired users. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

This system is intended to be used by a single patient and should not be shared.

:"

ं "

FORA V10 test strips are for use with FORA V10 No Code Blood Glucose Monitoring System, model TD-4244, to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, forearm, upper arm calf and thigh. ------:一

The FORA control solutions can only be used with FORA V10 No Code Blood Glucose Monitoring System, model TD-4244, and the FORA V10 test strips to check that meter and test strips are working together properly.

The alternative site testing in FORA V10 No Code Blood Glucose Monitoring System, model TD-4244, can be used only during steady-state blood glucose conditions.

    1. The Indications for Use is identical to K093035. The Indications for Use can be found in Section 8.

5. Device Description:

The system consists of three main products: the meter, test strips, and control solutions. These products have been designed, tested, and proven to work together as a system to produce accurate blood glucose test results. Use only FORA V10 test strips and FORA control solutions with the FORA V10 No Code Blood Glucose

B25- Page 2 of 4

{2}------------------------------------------------

Monitoring System, model TD-4244.

    1. Substantial Equivalence Information:
    • A. Predicate device name: FORA V10 Blood Glucose Monitoring System
    • B. Predicate K number: K093035
    • C. Comparison with predicate:

The modified FORA V10 No Code Blood Glucose Monitoring System has the following similarities to the predicate device:

  • Same intended use.
  • Same operating principle.

. Same functions and physical appearance.

  • . Same fundamental scientific technology.
  • . Incorporate the same basic circuit design.
  • Incorporate the same materials.
  • . Same shelf life.
  • 1 Same user interface and software.
  • 1 Packaged using the same materials.
  • Manufactured by the same process. .

The modifications:

  • 510(k) applicant for the device is FORA Care Inc. ■
  • Labeling changes due to the modification.
    1. Test Principle:

The detection and measurement of glucose in blood is by an electrochemical biosensor technology using glucose oxidase.

8. Performance Characteristics:

The FORA V10 No Code Blood Glucose Monitoring System has the same performance characteristics as the predicate device.

{3}------------------------------------------------

Software verification and validation testing confirmed that the performance, safety and effectiveness of the FORA V10 No Code Blood Glucose Monitoring System are equivalent to the predicate device.

9. Traceability:

FORA V10 No Code Blood Glucose Monitoring System is compared to the YSI 2300 Glucose Analyzer in the Accuracy studies. The YSI is calibrated with NIST (SRM) 917A reference material. Test strips have been cleared under predicate K093035. Control Solutions were previously cleared under K093724.

10. Conclusion:

Based on the information provided in this submission, the FORA V10 No Code Blood Glucose Monitoring System is substantially equivalent to the predicate FORA V10 Blood Glucose Monitoring System.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

January 14, 2013

FORA Care, Inc. c/o Pinjung Chen 6F, NO.127, Wugong 2nd Rd., Wugu District New Taipei City, China (Taiwan) 24888

Re: K121706

Trade/Device Name: FORA V10 No Code Blood Glucose Monitoring System, model TD-4244 Regulation Number: 21 CFR $862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II. Product Code: CGA, NBW, JJX Dated: January 4, 2013 Received: January 7, 2013

Dear Pinjung Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 – Pinjung Chen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Ruth A. Chesler

for Courtney H. Lias, Ph.D Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K121706

Device Name: FORA V10 No Code Blood Glucose Monitoring System, model TD-4244

Indications for Use:

FORA V10 No Code Blood Glucose Monitoring System, model TD-4244, is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. The meter contains some speaking functions but has not been validated for use by visually impaired users. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

This system is intended to be used by a single patient and should not be shared.

FORA V10 test strips are for use with FORA V10 No Code Blood Glucose meter model TD-4244 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, forearm, upper arm calf and thigh.

The FORA control solutions can only be used with FORA V10 No Code Blood Glucose meter, model TD-4244 and the FORA V10 test strips to check that meter and test strips are working together properly.

The alternative site testing in FORA V10 No Code Blood Glucose Monitoring systems model TD-4244 can be used only during steady-state blood glucose conditions.

Prescription Use (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Ruth A. Chesler

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k) K121706

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.