(107 days)
The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.
The Aspire HD Plus and Aspire HD-s are integrated FFDM systems combining an X-ray system made by Fujifilm with Fujifilm's a-Se detector and Acquisition Workstation (AWS). The Aspire HD Plus and Aspire HD-s create digital mammography images by direct capture of x-ray energy using the a-Se detector. The detector is a Fujifilm design utilizing an exclusive dual layer a-Se scintillator with Direct Optical switching circuitry to acquire image data and transfer images to the AWS for automated post processing, technologist preview and QC, and subsequent transmission to hard copy printers, diagnostic workstations and archiving systems. The Aspire HD Plus and Aspire HD-s provide automated compression and three AEC modes.
The Aspire HD Acquisition Workstation (FDR 2000AWS) includes an off the shelf personal computer, the application software, Windows 7 Operating System, a 5megapixel portrait type monitor, and a hub. The hub transmits signals between the personal computer and control cabinet, and between the personal computer and exposure stand.
The AWS display primarily consists of three windows:
. Patient Information Input window
Exposure Menu Selection window .
. Study window.
The user may switch between these windows depending on the operation being performed. The X-ray control panel, which controls and observes the exposure stand, is always displayed in the lower part of each window. This allows setting the exposure conditions and confirming the radiation conditions on a single view.
Here's a breakdown of the acceptance criteria and study information for the Fujifilm Aspire HD Plus and Aspire HD-s digital mammography systems, based on the provided 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document focuses on establishing substantial equivalence to a predicate device (FUJIFILM Aspire HD (FDR MS-1000) (K110729)) rather than presenting a formal table of specific, quantitative acceptance criteria with corresponding performance metrics for the proposed devices. Instead, it states that the devices meet the requirements through a comparison of imaging characteristics and a clinical image attribute review.
Here's an interpretation based on the information provided:
| Acceptance Criteria Category | Reported Device Performance (Comparison to Predicate) |
|---|---|
| Technological Characteristics | "The technological characteristics of the devices are similar as demonstrated by the comparison of imaging characteristics such as MTF, Noise Analysis, DQE, CNR, Phantom testing, etc." |
| Non-Clinical Testing | Compliance with FFDM 510(k) Guidance document for: Sensitometric Response, Spatial Resolution, Noise Analysis, Signal-to-Noise Ratio Transfer - DQE, Dynamic Range, Image Erasure and Fading, Repeated Exposure Test, AEC Performance, ACR MAP Phantom Testing, Contrast Detail Phantom Testing, Patient Radiation Dose Testing, and Breast Compression system Testing. |
| Clinical Image Quality | "The Aspire HD Plus provides sufficiently acceptable quality for mammographic use." (This is stated to apply to both proposed devices due to identical imaging geometry and components.) |
| Indications for Use (IFU) | Same IFU as the predicate device: screening and diagnosis of breast cancer, intended for use in the same clinical applications as traditional screen-film mammography systems. |
| Safety & Efficacy | "no new safety or efficacy issues other than those already identified with the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Six (6) image sets per exposure mode of screening and diagnostic cases.
- Data Provenance: Not explicitly stated, but clinical image attribute reviews typically involve retrospective analysis of images. The document does not specify the country of origin.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Not explicitly stated, but described as "independent mammographic radiologists."
- Qualifications of Experts: "independent mammographic radiologists." Specific years of experience are not mentioned.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly stated. The document only mentions a "clinical image attribute review" by independent radiologists. It does not detail if there was a consensus process, independent reads followed by adjudication, or similar methods.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described. The study mentioned is an image attribute review for image quality, not a comparative effectiveness study of human readers with vs. without AI assistance.
- Effect Size: Not applicable, as an MRMC study was not detailed.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done
- Standalone Study: No, this device is a complete Full-Field Digital Mammography X-ray System, not an AI algorithm. The performance evaluation is for the entire imaging system.
7. The Type of Ground Truth Used
- Type of Ground Truth: For the clinical image attribute review, the "ground truth" was the subjective assessment of "sufficiently acceptable quality for mammographic use" by independent mammographic radiologists. This is closer to expert consensus on image quality rather than pathology, outcomes data, or directly diagnosing disease. For the non-clinical tests (MTF, DQE, etc.), established physical phantom measurements and engineering standards were used.
8. The Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. These are hardware medical devices (digital mammography systems), not AI algorithms that are "trained" on a dataset in the typical machine learning sense. The "development" would involve engineering and design, but not a distinct "training set" like for AI.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as there isn't a training set in the context of an AI algorithm. The system's design and engineering parameters are based on established physics principles and clinical requirements for mammography.
{0}------------------------------------------------
FUJKFILM
419 WEST AVENUE STAMFORD, CT 06902 203-324-2000 PHONE: TOLL-FREE: 800-431-1850 203-353-0926 FAX:
510(k) Summary
k 121674
Date Prepared: June 1, 2012
Submitter's Information
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (203) 602-3576 Facsimile: (203) 353-0296 Contact: Peter Altman
Device Name and Classification:
| Product Name: | Aspire HD Plus and Aspire HD-s |
|---|---|
| Model Number: | FDR MS-2500 and FDR MS-2000 |
| Classification Name: | Full-Field Digital Mammography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR 892.1715 |
| Device Class: | Class II |
| Product Code: | MUE |
Substantial Equivalence/Predicate Devices:
FUJIFILM Aspire HD (FDR MS-1000) (K110729)
The predicate and proposed devices have the same Indication For Use (IFU). All three devices generate digital mammographic images that are intended for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.
All systems employ the same digital amorphous selenium x-ray detector integrated into gantry (stand) based x-ray systems. The x-ray stand, tube and generator of the Aspire HD Plus and Aspire HD-s are essentially identical. Note that the Aspire HD's X-ray system was manufactured by Siemens, whereas the Aspire HD Plus and Aspire HD-s X-ray systems are manufactured by Fujifilm. The technological characteristics of the devices are similar as demonstrated by the comparison of imaging characteristics such as MTF, Noise Analysis, DQE, CNR, Phantom testing, etc. measured during non-clinical testing. Non-clinical testing was conducted in accordance with the FFDM 510(k) Guidance document covering Sensitometric Response, Spatial Resolution, Noise Analysis, Signal-to-Noise Ratio Transfer - DQE, Dynamic Range, Image Erasure and Fading, Repeated Exposure Test, AEC Performance, ACR MAP Phantom Testing, Contrast Detail Phantom Testing, Patient Radiation Dose Testing, and Breast Compression system Testing.
A clinical image attribute review was conducted by independent mammographic radiologists in accordance with the FFDM 510(k) Guidance document. The mammographic attributes of six (6) image sets per exposure mode of screening and diagnostic cases were reviewed concluding that the Aspire HD Plus provides sufficiently acceptable quality for mammographic use. Since the Aspire HD Plus and Aspire HD-s have exactly the same imaging geometry and components (detector, and X-ray system), the results of the image attribute review apply to both systems.
SEP
21 2012
{1}------------------------------------------------
FUJKFILM
419 WEST AVENUE STAMFORD, CT 06902 203-324-2000 PHONE: TOLL-FREE: 800-431-1850 203-353-0926 FAX:
Description of the Device:
The Aspire HD Plus and Aspire HD-s are integrated FFDM systems combining an X-ray system made by Fujifilm with Fujifilm's a-Se detector and Acquisition Workstation (AWS). The Aspire HD Plus and Aspire HD-s create digital mammography images by direct capture of x-ray energy using the a-Se detector. The detector is a Fujifilm design utilizing an exclusive dual layer a-Se scintillator with Direct Optical switching circuitry to acquire image data and transfer images to the AWS for automated post processing, technologist preview and QC, and subsequent transmission to hard copy printers, diagnostic workstations and archiving systems. The Aspire HD Plus and Aspire HD-s provide automated compression and three AEC modes.
The Aspire HD Acquisition Workstation (FDR 2000AWS) includes an off the shelf personal computer, the application software, Windows 7 Operating System, a 5megapixel portrait type monitor, and a hub. The hub transmits signals between the personal computer and control cabinet, and between the personal computer and exposure stand.
The AWS display primarily consists of three windows:
- . Patient Information Input window
- Exposure Menu Selection window .
- . Study window.
The user may switch between these windows depending on the operation being performed. The X-ray control panel, which controls and observes the exposure stand, is always displayed in the lower part of each window. This allows setting the exposure conditions and confirming the radiation conditions on a single view.
Intended Use:
The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.
Safety Information:
The Aspire HD Plus and Aspire HD-s introduce no new safety or efficacy issues other than those already identified with the predicate device. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the devices are of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and is consistent with the level of concern indicated in the "Class II Special Controls Guidance Document: Full-Field Digital Mammography System" document issued on: November 5, 2010.
Conclusion:
This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject devices to be as safe and effective as the predicate device based upon the clinical and non-clinical data summarized above.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is a stylized image of a bird or eagle with its wings spread. The bird is made up of three curved lines that form the wings and body. The seal is black and white.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
SEP 21 2012
Mr. Peter Altman Regulatory Consultant FUJIFILM Medical Systems, USA Inc. 419 West Avenue STAMFORD CT 06902
Re: K121674
Trade/Device Name: Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000) Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: September 21, 2012. Received: September 21, 2012
Dear Mr. Altman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely Yours,
Janine M. Morris
Director Bivision of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K121674
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Raciological D Offica of litro Diagnostic Devica Evaluation and Salety
510K 121674
Page 1 of 1
Page 22 of 2288
N/A