EZ-ASSIST-FILL
K121661 · Assist Fill, LLC · PCM · Dec 27, 2012 · Physical Medicine
Device Facts
| Record ID | K121661 |
| Device Name | EZ-ASSIST-FILL |
| Applicant | Assist Fill, LLC |
| Product Code | PCM · Physical Medicine |
| Decision Date | Dec 27, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 890.5050 |
| Device Class | Class 1 |
Intended Use
EZ-ASSIST-FILL is intended to assist the user with refilling a portable liquid oxygen tank. The device is for use in homes or clinics. The EZ-ASSIST-FILL is neither a life-sustaining, nor life-supporting device. The device is intended for Over-the-Counter use.
Device Story
EZ-ASSIST-FILL is a mechanical assist device for refilling portable liquid oxygen tanks. It consists of a telescoping spring arm assembly mounted on a fiberglass base plate. The device uses a pivot mechanism allowing the arm to rotate 90 degrees to position over the portable unit. The weight of the reservoir tank holds the assembly in place. The device replaces the physical strength required by the user to apply downward pressure during the oxygen transfer connection. It is intended for use in homes or clinics by patients or caregivers. By holding the lever in place during the filling process, it eliminates the need for manual intervention, facilitating the refill process for users with limited physical strength.
Clinical Evidence
No clinical data. Bench testing only, specifically the "EZ-ASSIST-FILL Production Unit Repeatability Test," confirmed the device holds the portable oxygen tank in place and allows successful filling without manual intervention.
Technological Characteristics
Mechanical assist device; fiberglass base plate; telescoping spring arm assembly; pivot mechanism. No energy source. Standalone device. No software.
Indications for Use
Indicated for users requiring assistance refilling portable liquid oxygen tanks in home or clinical settings. Not life-sustaining or life-supporting. Intended for Over-the-Counter use.
Regulatory Classification
Identification
A daily activity assist device is a modified adaptor or utensil (e.g., a dressing, grooming, recreational activity, transfer, eating, or homemaking aid) that is intended for medical purposes to assist a patient to perform a specific function.
Predicate Devices
- Puritan-Bennett Helios Universal (K993220)
- Fred Sammons Inc. Jar Holder & Opener (K813631)
Related Devices
- K050430 — MERITS HEALTH PRODUCTS OXYGEN CONCENTRATOR · Merits Health Products Co., Ltd. · Aug 4, 2005
- K041122 — PRECISION LIQUID OXYGEN SYSTEM · Precision Medical, Inc. · Oct 4, 2004
- K103324 — PRECISION LIQUID OXYGEN SYSTEM · Precision Medical, Inc. · Apr 21, 2011
- K993220 — PURITAN-BENNETT HELIOS PORTABLE LIQUID OXYGEN SYSTEM · Puritan Bennett Corp. · Oct 22, 1999
- K141967 — PUREFILL OXYGEN COMPRESSOR · Inovo, Inc. · Nov 7, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: k121661_.
DEC 2 7 2012
### Submitter's Identification: 1.
Mr. Peter Glassberg Assist Fill, LLC 7143 NW 67 Way Parkland, FL 33067 Tel: 954-232-8676
Date Summary Prepared: December 17, 2012
### Name of the Device: EZ-ASSIST-FILL 2.
- Common or Usual Name: Daily Activity Assist Device 3. Classification: 21 CFR 890.5050 product code: PCM
### Predicate Device Information: · 4.
Puritan-Bennett Helios Universal, cleared under K993220 Fred Sammons Inc. Jar Holder & Opener cleared under K813631
### ડ. Device Description:
EZ-ASSIST-FILL is designed to support any combination of portable and reservoir units that use downward pressure to effect the transfer connection. EZ-ASSIST-FILL consists of a telescoping spring arm assembly mounted on a fiberglass base plate using pivot which allows the spring arm to rotate through 90 degrees to swing over the portable unit when in refill position. The entire unit is held in place by the weight of the reservoir, which rests entirely on the base plate. There is no need to remove the reservoir from EZ-ASSIST-FILL between uses.
#### Intended Use: Q.
EZ-ASSIST-FILL is intended to assist the user with refilling a portable liquid oxygen tank. The device is for use in homes or clinics.
l
The EZ-ASSIST-FILL is neither a life-sustaining, nor life-supporting device. The device is intended for Over-the-Counter use.
{1}------------------------------------------------
### Technological Comparison to Predicate Devices: 7.
- The device performs the same function as a person applying pressure during filling. The only difference is that the assist device arm needs to be moved into position to hold the oxygen unit in place during filling.
The EZ-ASSIST-FILL is designed with similar intent as Fred Sammons Inc. Jar Holder & Opener that being to hold an object in place replacing the required user physical strength to achieve the same end results without using the assist device.
The predicate is mounted to a flat, stationary workspace in order to provide assistance holding the jar in place while the person opens a jar, while the subject device is mounted on to the reservoir tank to provide assistance holding the lever in place while the user fills the tank.
### Discussion of Non-Clinical Tests Performed for Determination of Substantial 8. Equivalence are as follows:
Bench testing was performed to assess functionality of the EZ-ASSIST-FILL. The EZ-ASSIST-FILL devices were tested according to the "EZ-ASSIST-FILL Production Unit Repeatability Test". The devices met the requirements of the predetermined acceptance criteria, which confirmed that the EZ-ASSIST-FILL held the portable oxygen tank in place without any manual intervention and the portable tank was successfully filled when using the EZ-ASSIST-FILL. All devices passed the test.
### Discussion of Clinical Tests Performed: 9.
Clinical testing was not performed.
### 10. Conclusions:
Based on the information provided in this submission we conclude that the EZ-ASSIST-FILL is substantially equivalent to the predicates and is safe and effective for its intended use.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure, with three wavy lines forming the body and head.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
# December 27, 2012
Assist Fill, Limited Liability Company C/O Ms. Maria F. Griffin MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 GREAT NECK NY 11021
Re: K121661
Trade/Device Name: EZ-Assist-Fill Regulation Number: 21 CFR 890.5050 Regulation Name: Daily Activity Assist Device Regulatory Class: I Product Code: PCM Dated: December 17, 2012 Received: December 20, 2012
Dear Ms. Griffin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
·http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Attachment 1
## Indications for Use
k121661
of 1 Page _
510(k) Number (if known):
Device Name
EZ-ASSIST-FILL
Indications For Use:
EZ-ASSIST-FILL is intended to assist the user with refilling the portable liquid oxygen tank. The device is for use in homes or clinics.
The EZ-ASSIST-FILL is neither a life-sustaining, nor life-supporting device. The device is intended for Over-the-Counter use.
Prescription Use (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nayan J. Patel -S 2012.12.27 12:10:33,-05'00
For Lex Schultheis
(Division Sign-Off) (Division Sigir-On)
Division of Anesthesiology, General Hospital Division Control, Dental Devices
510(k) Number:_K121661