AUTO-CHEK BLOOD GLUCOSE MONITORING SYSTEM, METER, TEST STRIP. AUTO-CHECK CONTROL SOLUTION

K121654 · I-Sens, Inc. · CGA · Sep 5, 2012 · Clinical Chemistry

Device Facts

Record IDK121654
Device NameAUTO-CHEK BLOOD GLUCOSE MONITORING SYSTEM, METER, TEST STRIP. AUTO-CHECK CONTROL SOLUTION
ApplicantI-Sens, Inc.
Product CodeCGA · Clinical Chemistry
Decision DateSep 5, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The AUTO-CHEK Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips and alternative sites such as the forearm, palm. thigh, and calf. Alternative site testing should be used only during steadystate blood glucose conditions. The AUTO-CHEK Blood Glucose Monitoring System is intended for self testing outside the body (in vitro) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes. The AUTO-CHEK Blood Glucose Test Strips are for use with the AUTO-CHEK Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips and alternative sites. The AUTO-CHEK Control Solutions are for use with the AUTO-CHEK Meter and AUTO-CHEK Test Strips to check that the meter and the test strips are working together properly and that the test is performing correctly.

Device Story

AUTO-CHEK Blood Glucose Monitoring System (BGMS) measures glucose levels in fresh capillary whole blood; utilizes amperometric method with glucose oxidase reagent on test strips. System generates small electrical current proportional to glucose concentration; meter converts current to numerical glucose value. Used by patients with diabetes at home for monitoring glycemic control. Components include meter, test strips, control solutions, lancing device, and lancets. Device is for single-patient use only. Healthcare providers use output to assess diabetes management effectiveness. Modifications from predicate include meter housing shape and test strip electrical connector pattern.

Clinical Evidence

Bench testing only. Validation focused on verifying that the modified electric connector pattern did not affect device safety or performance. Disinfection studies using CLOROX GERMICIDAL Wipes demonstrated complete inactivation of Hepatitis B virus and confirmed material robustness over 260 cleaning cycles, simulating 5 years of use.

Technological Characteristics

Amperometric glucose oxidase (Aspergillus sp.) biosensor. System includes meter, test strips, and control solutions. Modifications: updated meter housing and test strip electrical connector pattern. Connectivity: None specified. Sterilization: Not specified.

Indications for Use

Indicated for quantitative blood glucose measurement in people with diabetes. For home use (in vitro) via capillary whole blood from fingertips or alternative sites (forearm, palm, thigh, calf). Not for neonates, diagnosis, or screening. Alternative site testing restricted to steady-state conditions.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KI21654 Image /page/0/Picture/1 description: The image shows the logo for i-sens. The logo features a lowercase letter 'i' inside of a circle. To the right of the circle is the word 'sens' in lowercase letters. The font is bolded and black. י א 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191 # 510(k) Summary SEP 5 2012 (As required by 21 CFR 807.92) | Type of 510(k): | Special 510(k) | |-------------------------|-------------------------------------------------------------| | Submitted By: | i-SENS, Inc. | | | 465-6, Wolgye-dong, Nowon-gu, Seoul 139-845, Korea | | | Tel.) +82-2-916-6191 | | | Fax) +82-2-942-2514 | | | www.i-sens.com | | Contact Person: | Dr. Hyun Joon Oh | | | Tel.) +82-33-903-0760 | | | Fax) +82-33-748-6191 | | | e-mail: hjoh@i-sens.com | | Prepared Date: | May 25, 2012 | | Device Name: | Trade name: AUTO-CHEK Blood Glucose Monitoring System | | | Common Name: Glucose Test System | | Regulatory Information: | 1) Regulation section: 21 CFR 862.1345 Glucose Test System, | | | 21 CFR 862.1660, Quality control material | | | 2) Classification: Class II, Class I | | | 3) Product Code: CGA - glucose oxidase, glucose | | | NBW - system, test, blood glucose, over the counter | | | JJX - Quality control material | | | 4) Panel: Clinical Chemistry (75) | . {1}------------------------------------------------ Intended Use: The AUTO-CHEK Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips and alternative sites such as the forearm, palm. thigh, and calf. Alternative site testing should be used only during steadystate blood glucose conditions. The AUTO-CHEK Blood Glucose Monitoring System is intended for self testing outside the body (in vitro) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes. > The AUTO-CHEK Blood Glucose Test Strips are for use with the AUTO-CHEK Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips and alternative sites. The AUTO-CHEK Control Solutions are for use with the AUTO-CHEK Meter and AUTO-CHEK Test Strips to check that the meter and the test strips are working together properly and that the test is performing correctly. Device Description: The AUTO-CHEK Blood Glucose Monitoring System (BGMS) measures the glucose level in whole blood samples using a small electrical current generated in the test strips. The system consists of the followings: the AUTO-CHEK Meter, AUTO-CHEK Test Strips, AUTO-CHEK Control Solutions with two different glucose concentrations ("Control A" and "Control B" ranges, sold separately), a Lancing Device, Lancets, a User's manual, and a Logbook. ## l ) Predicate Device Name: COOL Blood Glucose Monitoring System Substantial Equivalence Information: 2) Predicate 510(k) Number: k103396 2) Comparison with Predicate Device: The modified AUTO-CHEK BGMS has the following features that are identical to the predicate device: {2}------------------------------------------------ i-SENS, Inc. 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191 - Intended use - . Measurement principle - I Fundamental scientific technology - 1 Technical specifications - Operating ranges The modifications that was made on the AUTO-CHEK BGMS compared to COOL BGMS are: - . The shape of meter's housing - . The electric connector pattern of the test strip Type of Test: Quantitative, Amperometric method, Glucose oxidase (Aspergillus sp.) Test Principle: The reagent on the test strip produces a small electrical current using glucose as a substrate in the blood sample. The meter converts electrical current to glucose concentration. Technological The AUTO-CHEK BGMS has the same fundamental scientific technology as Characteristics: the predicate device. Assessment of Performance Characteristics: When compared with the predicate device, the basic features of the candidate device, the intended use, measurement principle, fundamental scientific technology, technical specifications, and operating ranges, are all the same, The only differences are the shape of meter's housing and electric connector pattern of the test strip. The validations were conducted in order to verify that the modified electric connector pattern has not caused any adverse effects on the safety and effectiveness of the candidate device. The test results showed that the candidate device operated effectively, accurately, and safely. Summary of Pre-cleaning and Disinfection: Disinfection studies were performed on the AUTO-CHEK meter and lancing device by an outside commercial testing service to evaluate effectiveness of disinfectant, CLOROX GERMICIDAL Wipes (EPA Reg. No: 67619-12), in preventing the spread of blood-borne pathogens, using hepatitis B virus {3}------------------------------------------------ (HBV). The results demonstrated complete inactivation of live virus inoculated on the materials of the meter and lancing device. We have also demonstrated that 260 each of pre-cleaning and disinfection cycles with the same disinfectant designed to simulate 5 years of use has not affected either the performance of the meter and the lancing device or the external materials of the meter and lancing device demonstrating the robustness of the meter and lancing device. #### Conclusion: Based on the validation results, the candidate device, AUTO-CHEK BGMS, is substantially equivalent to the predicate device. Further, the AUTO-CHEK BGMS has met the performance, safety, and effectiveness of the device for its intended use. 7 - 4 {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES 10903 New Hampshire Avenue Silver Spring, MD 20993 I-SENS, Inc. c/o Hyun Joon Oh Division Manager, Quality Assurance 465-6 Wolgye-Dong, Nowon-GU SEOUL, Republic of Korea 139-845 SEP 5 2012 k121654 Re: Trade Name: AUTO-CHEK Blood Glucose Monitoring System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Codes: CGA, NBW, JJX Dated: August 1, 2012 Received: August 6, 2012 Dear Dr. Hyun Joon Oh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). Image /page/4/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a human figure with outstretched arms, representing care and protection. {5}------------------------------------------------ ### Page 2 . If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your do need in our incomes on Safety at (301) 796-5450. Also, contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (3 contact the Office of in Virto Diagnound Boding by reference to premarket notification" (21) please note the regarding postmarket surveillance, please contact CDRH's CFK Farl 607.97). Por questions regulains postmarket Surveillance at (301). Office of Survelliance and Drometre s (OB a) of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical CFK Fall 805), picaso go to map default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance... You may obtain other general information on your responsibilities under the Act from the You may outain other general mormation on your consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address (800) 036-2041 of (594-2041 of (50-10-2017) ou/Industry/default.htm Sincerely yours, N Couriney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use Form 510(k) Number (if known): K121654 Device Name: AUTO-CHEK Blood Glucose Monitoring System. Indications for Use: The AUTO-CHEK Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips and alternative sites such as the forearm, palm, thigh, and calf. Alternative site testing should be used only during steady-state blood glucose conditions. The AUTO-CHEK Blood Glucose Monitoring System is intended for self testing outside the body (in vitro) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes. The AUTO-CHEK Blood Glucose Test Strips are for use with the AUTO-CHEK Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips and alternative sites. The AUTO-CHEK Control Solutions are for use with the AUTO-CHEK Meter and AUTO-CHEK Test Strips to check that the meter and the test strips are working together properly and that the test is performing correctly. Prescription Use _____________________________________________________________________________________________________________________________________________________________ ਮ (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Kathleen Lewis Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K121654 Page 1 of
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%