K Number
K121637

Validate with FDA (Live)

Date Cleared
2012-10-25

(143 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Auto-Disable Sterile Safety Syringe for Single Use ,5ml, is used to inject fluids into, or withdraw fluids from the body. In addition, the retractable needle of the syringe goes into the barrel after use. It is designed to aid in the prevention of needle stick injuries.

Device Description

The Auto-Disable Sterile Safety Syringe for Single Use is used to inject fluids into, or withdraw fluids from the body. In addition, the retractable needle of the syringe goes into the barrel after use. It is designed to aid in the prevention of needle stick injuries. The retractable type of piston syringe is a plastic disposable anti-needle stick syringe made of the below components: Cap, Needle, Needle house, Hub, Barrel, Gasket, Plunger.

AI/ML Overview

This looks like a 510K submission for a medical device. I will break down the information section by section as requested.

Acceptance Criteria and Device Performance Study for Auto-Disable Sterile Safety Syringe For Single Use, 5mL (K121637)

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state numerical acceptance criteria in a table format for a specific "device performance" study that would typically involve sensitivity, specificity, accuracy, etc., for an AI/diagnostic device. Instead, the performance testing focuses on compliance with various ISO and ASTM standards relevant to the physical and biological aspects of a sterile safety syringe.

However, the "Simulated Clinical Use Study" directly addresses the safety and usability of the device's main feature. We will infer the acceptance criteria from the study design and the conclusion of substantial equivalence.

Inferred Acceptance Criteria and Reported Device Performance for Simulated Clinical Use:

Acceptance Criteria CategorySpecific Criteria (Inferred from study design and intent)Reported Device Performance
Safety Feature ReliabilitySafety mechanism (retractable needle) functions as intended consistently, preventing needle stick injuries.Performed comparably to the predicate device, Invirosnap Safety Syringe. The study aimed to assess the "reliability and usability of the safety feature."
UsabilityThe device can be used effectively and safely by healthcare professionals (nurses). The user experience is comparable to predicate device.Nurses reported their experience with the device; implicit acceptance if similar to predicate.
Mechanical IntegrityDevice does not fail or malfunction during simulated use.No reported failures or malfunctions that would indicate non-substantial equivalence to the predicate.
General FunctionalityInjecting/withdrawing fluids is effective, and the retraction mechanism activates as designed.No issues reported, implying proper functionality for its intended use.

Other Performance Testing (Compliance with Standards):

The device underwent and passed a series of standard tests for biological evaluation, sterilization, packaging, and general syringe functionality to demonstrate substantial equivalence to the predicate device. These are not performance metrics in the sense of sensitivity/specificity but rather compliance with established safety and quality standards.

NoStandard/DescriptionReported Performance
1ISO10993-1:2009 (Biological evaluation)Passed
2ISO10993-5:2009 (In Vitro Cytotoxicity)Passed
3ISO 10993-10:2010 (Irritation and Delayed-Type Hypersensitivity)Passed
4ISO10993-11:2006 (Systemic toxicity - Acute)Passed
5ISO7886-1:1993/Corrigendum 1:1995 (Syringes for manual use)Passed
6ISO 7864 Third edition 1993-05-15 (Sterile hypodermic needles)Passed
7AAMI/ANSI/ISO 10993-7:2008 (Ethylene Oxide Sterilization Residuals)Passed
8AAMI/ANSI/ISO11607-1:2006/(R)2010 (Packaging)Passed (supports 3 years shelf life)
9ISO 11737-1:2006 (Microorganism population)Passed
10ISO 11737-2: 2009 (Sterility tests)Passed
11ISO/AAMI/ANSI11135-1 :2007 (EO Sterilization validation)Passed (all predetermined parameters met)
12ASTM F 1980-07 (Accelerated Aging Test)Passed (supports 3 years shelf life)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Simulated Clinical Use:
    • Test Device (Auto-Disable Syringe): 500 samples
    • Predicate Device (Invirosnap Safety Syringe): 100 samples
  • Data Provenance: The study was a "simulated clinical use study" conducted by Sunwell Biotech Co., Ltd. The document does not explicitly state the country of origin for the data or if it was retrospective or prospective. However, "simulated clinical use" implies a prospective collection of data in a controlled environment, likely in the country of the submitting company (China) or a contracted facility.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: 10 nurses.
  • Qualifications of Experts: They are identified as "nurses," implying healthcare professionals with practical experience in using syringes for injections and withdrawals. Specific years of experience or specialization are not provided.

4. Adjudication Method for the Test Set

The document states that the 10 nurses "reported their experience with each sample." This suggests a direct user feedback or observation method rather than a formal adjudication process where multiple experts collectively establish a single ground truth for each case. Given the nature of a safety syringe (does it retract or not, is it easy to use), direct user experience reports serve as the primary data. There is no mention of a 2+1, 3+1, or similar adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted in the context of an AI/diagnostic device improving human reader performance. This device is a physical medical instrument (syringe), not a diagnostic algorithm. The "simulated clinical use study" involved human users (nurses) testing the device, but it was not designed to measure an effect size of how much human readers improve with AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • No, this question is not applicable. The device is a physical auto-disable safety syringe, not an algorithm or AI system. Therefore, there is no "standalone" algorithm performance to evaluate.

7. The Type of Ground Truth Used

For the simulated clinical use study, the "ground truth" was based on user experience and observation of functional performance by the participating nurses, guided by the objective to assess the "reliability and usability of the safety feature." The success or failure of the needle retraction, ease of use, and overall experience reported by the nurses effectively constitute the ground truth for this type of device performance evaluation.

8. The Sample Size for the Training Set

  • This device is a physical safety syringe, not an AI/ML algorithm that requires a "training set" of data. Therefore, the concept of a training set sample size is not applicable to this submission.

9. How the Ground Truth for the Training Set Was Established

  • As the concept of a training set is not applicable to this physical device, there is no information on how ground truth for a training set was established.

{0}------------------------------------------------

K121637

510(k) Summary

OCT 25 2012

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: June 4, 2012

Company and Correspondent making the submission:

Name - Sunwell Biotech Co., Ltd.

Address - No.9,Yongxing Road, Gangzha Economic Development Zone, Nantong City, Jiangsu, China 226002

Telephone - 86-513-85308888

Fax - 86-513-85308899

Contact - Mr. Charlie Mack

Email - charliemack@irc-us.com

Device :

Trade/proprietary name: Auto-Disable Sterile Safety Syringe For Single Use, 5mL

Common Name

: Antistick Syringe

Classification Name : Piston, Syringe

Predicate Device:

ModelManufacturerK NumberSubmitted Device
InvirosnapSafetySyringe, 1ml2ml 3ml 5ml10ml and20mlInviro MedicalDevices, Inc.K092430Auto-Disable SterileSafety Syringe forSingle Use, 5mL

Classifications Names & Citations : 21CFR 880.5860, MEG, Anti-Stick Syringe, Class II

{1}------------------------------------------------

Description :

General

The Auto-Disable Sterile Safety Syringe for Single Use is used to inject fluids into, or withdraw fluids from the body. In addition, the retractable needle of the syringe goes into the barrel after use. It is designed to aid in the prevention of needle stick injuries.

The retractable type of piston syringe is a plastic disposable anti-needle stick syringe made of the below components:

  1. Cap - Covers the needle until the syringe is to be used

  2. Needle - The needle penetrates the patient's skin to inject/withdraw fluid from the body

  3. Needle house - a plastic part, as a assembly part with needle to fasten together with epoxy glues

  4. Hub - a plastic part for minimizes the risk of leakage around the needle assembly

  5. Barrel - The barrel has a scale showing the capacity of the syringe , in addition ,the top of the barrel has a airproof ring lock fitting for minimize risk of leakage around the Hub

  6. Gasket - A TPE (Thermoplastic Elastomer) molding part. The gasket is to be assembled with plunger for bring plus/minus pressure within barrel when the plunger to be fully depressed/retracted

  7. Plunger - Once the plunger is fully depressed, it engages the needle assembly, as the plunger is retracted ,the needle assembly is retracted into the barrel, once the plunger is fully retracted and locked in place, the plunger is snapped off leaving the needle in the barrel of the syringe

Indications For Use

The Auto-Disable Sterile Safety Syringe for Single Use ,5ml, is used to inject fluids into, or withdraw fluids from the body. In addition, the retractable needle of the syringe goes into the barrel after use. It is designed to aid in the prevention of needle stick injuries.

{2}------------------------------------------------

Sterilization:

Sunwell Biotech Co., Ltd., has successfully performed the EO validation according to ISO 11135-1:2007 for Auto-Disable Sterile Safety Syringe for Single Use. All predetermined parameters have been met.

Biocompatibility:

The biocompatibility study has been conducted for Auto-Disable Sterile Safety Syringe for Single Use 5ml according to the below applicable standards:

In Vitro Cytotoxicity (ISO 10993-5:2009) Delayed Contact Sensitization Study in the Guinea Pig (ISO10993-10:2010) Intracutaneous Test (ISO10993-10:2010) Systemic Toxicity (Acute) (ISO10993-11:2006) Hemolysis Test (ISO 10993-4-2002/Amd. 1:2006) Complement Activity Test (ISO 10993-4-2002/Amd. 1:2006) Thrombosis Test (ISO 10993-4-2002/Amd. 1:2006)

Package and Shelf Life:

Testing was performed to demonstrate package and shelf life verification. The study shows that the subject sterilization package complies with the applicable standards and support the 3 years shelf life claim. The tests noted below were performed:

Accelerated Aging of Sterile Barrier Systems (ASTM F1980-07) Sterilization of medical devices - Microbiological methods (AAMI/ANSI/ISO 11737-2:2009 - Part 2)

Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission (ASTM D3078-02)

Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration (ASTM F1929-98)

Test Method for Seal Strength of Flexible Barrier Materials (ASTM F88/F88M-09) for terminally sterilized medical devices - Part 1: (AAMI/ANSI/ISO 11607-1:2006/) Sterile hypodermic syringes for single use - Part 1: (ISO 7886-1:1993) Sterile hypodermic needles for single use (ISO 7864 Third edition 1993-05-15) Stainless steel needle tubing for the manufacture of medical devices (ISO 9626 First edition 1991-09-01, AMENDMENT 1 2001-06-01)

{3}------------------------------------------------

Simulated Clinical Use Study

Sunwell Biotech Co., Ltd conducted simulated clinical use study of Auto-Disable Sterile Safety Syringe for Single Use, 5ml, in accordance with the FDA Guidance for industry and FDA Staff, Medical Devices with Sharps Injury Prevention Features, August 9, 2005. The objective was to assess the reliability and usability of the safety feature of that device. Based on the simulated clinical use test recommended in that document, 10 nurses were recruited. They tested a total of 500 Auto-Disable Sterile Safety Syringe for Single Use and 100 samples of a similar products, the Invirosnap Safety Syringe( Which is already marketed in the United States with K092430), and reported their experience with each sample. The safety mechanisms and the user actions of the two devices are similar. Samples of the both syringes were provided by Sunwell Biotech Co., Ltd.

Product Picture

Image /page/3/Picture/3 description: The image shows a syringe with a needle. The syringe is made of clear plastic and has a black plunger. The syringe is marked with measurements from 1 to 5 ml. The brand name "Dr. Needl" is printed on the side of the syringe.

Comparison with predicate device :- Please see the next page

{4}------------------------------------------------

SE Table

Element of comparisonSubject DeviceClaimed SE Device
CompanySunwell Biotech Co., Ltd.Inviro Medical Devices, Inc.
FDA510(K) NumberN/AK092430
Device NameAuto-Disable Sterile Safety Syringe for Single UseInvirosnap Safety Syringe
Model Number5ml1ml 2ml 3ml 5ml 10ml and 20ml
Intended UseThe Auto-Disable Sterile Safety Syringe for Single Useis used to inject fluids into, or withdraw fluids from thebody. In addition, the retractable needle of the syringegoes into the barrel after use. It is designed toaid in the prevention of needle stick injuriesThe InviroSnap Safety Syringe is used to inject fluids into, orwithdraw fluids from, the body. In addition, the InviroSnapSafety Syringe is designed to aid in the prevention of needlestick injuries.
Principle of operationIdenticalAfter use, the health care professional fully depresses theplunger to engage the luer assembly. Once the luerassembly is engaged, pulling back the plunger causes theAdapter and the attached needle to be withdrawn into thesafety of the barrel. In this position against the flange, lateralpressure on the plunger results in a controlled fracture ofthe plunger. Both the syringe and plunger are discarded in asharps container.
Safety featureIdenticalManually retractable safety syringe with permanent disable
Specific drug useIdenticalConventional drugs
SterilizationIdenticalSterilized by ethylene oxide gas
Needle lengthIdentical1 1/4 Inch
Needle gauge22G21/22/23G
Needle tip configurationIdenticalTri-Beveled Tip
Wall typeIdenticalRegular wall
Nozzle typeIdenticalNeedle and hub are integral to the syringe, not separable.
Barrel marking specsidenticalconforms to ISO 7886-1:1993/ Corrigendum 1:1995
Gradations legibilityidentical0.2ml
Needle cover coloridenticalclarity
Lubricant compositionidenticalPolydimethylsiloxane oil (PDMS)
Lubricant amount/cm²Identical<0.25mg/cm²
Barrel transparencyIdenticaltransparency
Delivery accuracyIdenticalconforms to ISO 7886-1-1993/Corrigendum 1:1995

:

.

21 - 21 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 21 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 -

.

. .

{5}------------------------------------------------

.

{6}------------------------------------------------

Performance Testing:

Biological, sterility and performance testing have been ac complished according to the table below to demonstrate this product is substantially equivalent to the predicate device. The submitted device passed all of these tests.

NoSerial Number/versionStandard and description
1ISO10993-1:2009Biological evaluation of medical devices
2ISO10993-5:2009Biological Evaluation of Medical Devices- Part 5 Tests for In Vitro Cytotoxicity.
3ISO 10993-10:2010Biological Evaluation of Medical Devices- Part 10: Tests for Irritation andDelayed-Type Hypersensitivity
4ISO10993-11:2006Test for systemic toxicity (Acute)
5ISO7886-1:1993/Corrigendum 1:1995Sterile hypodermic syringes for singleuse - Part 1: Syringes for manual use
6ISO 7864 Third edition 1993-05-15Sterile hypodermic needles for singleuse.
7AAMI/ANSI/ISO 10993-7:2008Biological Evaluation of Medical Devices -Part 7: Ethylene Oxide SterilizationResiduals
8AAMI/ANSI/ISO11607-1:2006/(R)2010Packaging for terminally sterilized medicaldevices - Part 1: Requirements formaterials, sterile barrier systems andpackaging systems
9ISO 11737-1:2006Sterilization of medical devices-Microbiological methods-Part 1:Determination of the population ofmicroorganisms on
10ISO 11737-2: 2009Sterilization of medical devicesMicrobiological methods -Part 2: Tests ofsterility performed in the definition,validation
11ISO/AAMI/ANSI11135-1 :2007Sterilization of health care products -Ethylene Oxide - Part 1: Requirementsfor development, validation and routinecontrolof a sterilization process for medicaldevices
12ASTM F 1980-07Accelerated Aging Test

{7}------------------------------------------------

Conclusions:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification The Sunwell Biotech Co., Ltd. Auto-Disable Sterile Safety Syringe for Single Use, 5mL is substantially equivalent to the predicate device as described herein.

END

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Sunwell Biotech Company, Limited
C/O IRC
Mr. Charlie Mack
Principal Engineer
77325 Joyce Way
Echo, Oregon 97826

Re: K121637

Trade/Device Name: Auto-Disable Sterile Safety Syringe For Single Use, 5ml Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: October 15, 2012 Received: October 19, 2012

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

OCT 25 2012

{9}------------------------------------------------

Page 2- Mr. Mack

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{10}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Auto-Disable Sterile Safety Syringe For Single Use, 5ml

Indications for Use:

The Auto-Disable Sterile Safety Syringe for Single Use ,5ml, is used to inject fluids into, or withdraw fluids from the body. In addition, the retractable needle of the syringe goes into the barrel after use. It is designed to aid in the prevention of needle stick injuries.

Prescription Use _ × AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

11

hijd Mhi for RZC at 22,2012

(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices

510(k) Number: K12)637

Page 1 of

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).