K Number
K121534

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-06-20

(27 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CONE-TROL Hemoglobin A1c Control Set is intended for use as quality control material to monitor the performance and precision of Hemoglobin A 1 c determination methods.

Device Description

The CONE-TROL Hemoglobin A 1c Control Set is prepared from human blood to which chemicals, preservatives, and stabilizers are added. The control is provided in liquid form for user convenience.

AI/ML Overview

The provided documentation describes the acceptance criteria and study for the CONE-TROL Hemoglobin A1c Control Set, a quality control material, primarily focusing on its stability. It is not an AI/ML device, so many of the requested fields (e.g., MRMC studies, human reader improvement, multi-reader consensus) are not applicable.

Here's the breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criterion for the CONE-TROL Hemoglobin A1c Control Set is that the analyte activity (concentration) must fall within ±10% of the initial value at various storage conditions and time points.

AnalyteAcceptance Criteria (Concentration Range Accepted by Company)Initial ValueReported Performance (Concentration/Percent Change) at different conditions
Accelerated Closed Vial Stability (37°C, 30 hours)
HbA1c-Level 14.2-6.4% (implies ±10% around 5.3%)5.3%5.4% (1.89% change)
HbA1c-Level 28.6-13.0% (implies ±10% around 10.8%)10.8%10.8% (0% change)
Real Time Closed Vial Stability (-20°C, 210 days)
HbA1c-Level 14.2-6.4% (implies ±10% around 5.3%)5.3%5.4% (1.89% change)
HbA1c-Level 29.5-14.3% (implies ±10% around 11.9%)11.9%12.1% (1.68% change)
Open Vial Stability (2-8°C, 210 days)
HbA1c-Level 14.2-6.4% (implies ±10% around 5.3%)5.3%5.0% (5.7% change)
HbA1c-Level 29.5-14.3% (implies ±10% around 11.9%)11.9%11.5% (3.4% change)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • For stability studies, the "sample" typically refers to the number of vials or batches tested. The document mentions two vials were assayed in triplicate for value assignment, and data points were collected over several days (minimum 6 data points per level for value assignment). For stability, specific numbers of individual vials tested at each time point are not explicitly stated, but it follows a standard protocol of testing representative aliquots from the prepared control sets.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the studies were conducted by the manufacturer, Cone Bioproducts, located in Seguin, TX, USA, for regulatory submission to the FDA. The stability studies are prospective, as they track the product over time.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

Not applicable. This device is a quality control material whose "truth" is its chemical concentration, determined by analytical instrumentation. It does not involve human interpretation or expert opinions to establish ground truth in the way an AI diagnostic device would.

4. Adjudication Method for the Test Set

Not applicable. Ground truth for this chemical control is established through objective analytical measurements, not through adjudication of expert opinions.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This is not an AI/ML device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in a sense, the performance of the control material is assessed in a "standalone" manner. Its stability and value assignment are determined purely through analytical measurements on laboratory instruments (Tosoh G7 HPLC Analyzer). There is no "human-in-the-loop" component in determining the performance of the control material itself; it's an objective measurement of its chemical properties over time.

7. The Type of Ground Truth Used

The ground truth used is the measured concentration of HbA1c in the control material, determined by objective analytical methods using the Tosoh G7 HPLC Analyzer. This is a form of measurement data/analytical truth.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. As above, there is no training set for this type of device. The "ground truth" for the control material's initial value (value assignment) is established by assaying Levels 1 and 2 in triplicate on the applicable instrument system, with a minimum of 6 data points for each level. The mean, standard deviation, and coefficient of variance are then determined.

{0}------------------------------------------------

JUN 2 0 2012

SECTION V

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number: k

K121534

Submitter:

Cone Bioproducts 1012 N. Austin St. Seguin, TX 78155

Telephone: (830)379-0197 Facsimile: (830)379-0471

Contact Person:

Gail Schievelbein . Project Manager Telephone: (830)379-197 ext. 206 Facsimile: (839)379-0471

Preparation Date:

June 13, 2012

Device Information:

Device Classification Name: Single Analyte Controls

Proprietary Name: Regulation Number: Product Code: Regulatory Class:

(Assayed and Unassyed) CONE-TROL Hemoglobin A lc Control Set 21 CFR8862.1660 אוו Class I

Predicate Devices:

The CONE-TROL Hemoglobin Alc Control Set is substantially equivalent to Cone Bioproducts' HbA1c Linearity Set (K052010) for its stated intended use.

{1}------------------------------------------------

Device Description:

The CONE-TROL Hemoglobin A 1c Control Set is prepared from human blood to which chemicals, preservatives, and stabilizers are added. The control is provided in liquid form for user convenience.

Intended Use:

CONE-TROL Hemoglobin A1c Control Set is intended for use as quality control material to monitor the performance and precision of Hemoglobin A 1c determination methods.

Comparison to Predicate Device(s):

CONE-TROL Hemoglobin A 1c Control Set is substantially equivalent to Cone Bioproducts' HbA lc Linearity Set, (K052010) for its stated intended use.

DeviceCharacteristicsSubject DevicePredicate Device(s)K052010
Intended UseCONE-TROL Hemoglobin A1cControl Set is intended for use asquality control material to monitor theperformance and precision ofHemoglobin A1c determinationmethods.HbA1c Linearity Set is intended foruse as quality control material tomonitor the linearity throughout thereportable range of Hemoglobin A1cassays using protocols established inindividual laboratories.
AnalyteA1cA1c
MethodologyCompatible with Immunoassay andHPLC HbA1c test methods.Compatible with Immunoassay andHPLC HbA1c test methods.
MatrixHuman Blood, stabilizers, andpreservative.Human Blood, stabilizers, andpreservative.
Control FormLiquidLiquid
Levels1-21-4
Storage-20°C-20°C
StabilityUnopened vial stability (-20°C):2 YearsUnopened vial stability (2-8°C):180 daysOpened vial stability (2-8°C):180 daysUnopened vial stability (-20°C):2 YearsOpened vial stability (2-8°C):14 days

{2}------------------------------------------------

Summary:

Value Assignment:

Levels 1 and 2 are prepared so that Level 1 meets the clinically normal range of HbAIc, and Level 2 meets the clinically abnormal range of HbA lc on the manufacturer's instrument. Levels 1 and 2 are assayed in triplicate on the applicable instrument system. When the recovery falls within ±10% of the target concentration, the CONE-TROL Hemoglobin Alc Control Set is labeled and filled. QC testing for value assignment is performed for Levels 1 and 2 of control. Two vials are assayed in triplicate over a minimum of 2 days and maximum of 7 day period on the Immunoassay or HPLC test method for which the product is designed. A minimum of 6 data points for each level is required for value assignment. One instrument may be used if used on two different days or by two different operators. The mean, standard deviation and coefficient of variance is determined for each instrument for each level of control.

CLOSED VIAL STABILITY

The shelf life for CONE-TROL HbA ic Control Set has been determined based upon accelerated stability studies to be confirmed by real time stability studies. The expiration for CONE-TROL HbA1c Control Set is 2 years at -20°C and for 180 days at 2-8°C from the date of manufacture.

The product stability has been extrapolated from accelerated studies with elevated temperatures and will be confirmed by real time studies. Hemoglobin A 1c concentrations were analyzed using the Tosoh G7 HPLC Analyzer. Vials were stored at -20°C and 37°C and values were obtained between 0 and 30 hours.

The O Rule Table based on the Arrhenius Equation was used to estimate the shelf life during the accelerated stability study. The Q Rule states that a product's degradation rate decreases by a constant factor (Q=4) every 10°C decrease in temperature. For the test point to be valid the analyte activity must fall ± 10% from the initial value.

Table VII-1: Accelerated Closed Vial Stability of CONE-TROL Hemoglobin A1c
Control Set at 37°C
AnalyteInitial ValueConcentration at30 hoursPercentChangeConcentrationRange Acceptedby Company
HbA1c-Level 15.3%5.4%1.89%4.2-6.4%
HbA1c-Level 210.8%10.8%0%8.6-13.0%

3

{3}------------------------------------------------

To date, real time stability studies confirm the accelerated stability studies through 210 days. Closed Vial real time stability will be monitored throughout the life of the product. Closed Vial stability data will be collected at six month intervals. For the test point to be valid the analyte activity must fall ± 10% from the initial value.

AnalyteInitial ValueConcentration at210 daysPercentChangeConcentrationRange Acceptedby Company
HbA1c-Level 15.3%5.4%1.89%4.2-6.4%
HbA1c-Level 211.9%12.1%1.68%9.5-14.3%

Table VII-2: Real Time Closed Vial Stability at -20°C

Based on the data given in Table VII-1, CONE-TROL Hemoglobin A1c Control Set will be stable for 2 years when stored at -20°C as specified in the product insert.

OPEN VIAL STABILITY

The open vial, 2-8°C stability of CONE-TROL Hemoglobin A 1c Control Set has been determined from real time stability testing. HbA1c concentrations were determined using a Tosoh G7 HPLC Analyzer. CONE-TROL Hemoglobin A1c Control Set was stored as described in the package insert. Values for Hemoglobin A1c were collected through 210 days.

Table VII-3: Open Vial Stability of CONE-TROL Hemoglobin A1c Control Set at 2-8°C

AnalyteInitial ValueConcentration at210 daysPercentChangeConcentrationRange Acceptedby Company
HbA1c-Level 15.3%5.0%5.7%4.2-6.4%
HbA1c-Level 211.9%11.5%3.4%9.5-14.3%

Based on the data presented in Table VII-3, at the present time, CONE-TROL Hemoglobin A 1c Control Set will be given an expiration date of 180 days when stored at 2-8°C as described in the package insert.

Substantial Equivalence:

The information provided in this pre-market notification demonstrates CONE-TROL Hemoglobin A 1 c Control Set is substantially equivalent to Hemoglobin A 1 c Linearity Set (K052010). Substantial equivalence was demonstrated through comparison of intended use and physical properties to the commercially available and analytical predicate device. The information supplied in this pre-market notification provides reasonable assurance that the CONE-TROL Hemoglobin A 1c Control Set is safe and effective for its stated intended use.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/2 description: The image shows a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the left side of the seal. To the right of the text is a symbol that resembles an eagle or a bird in flight, with three stylized human profiles incorporated into the design. The overall impression is that of an official emblem or logo, likely representing a government agency related to health and human services in the United States.

10903 New Hampshire Avenue Silver Spring, MD 20993

Cone Bioproducts Ms. Gail Schievelbein Project Manager 1012 N. Austin St. Seguin, TX 78155

JUN 2 0 2012

Re: K121534

Trade/Device Name: CONE-TROL Hemoglobin Alc Control Set Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material Regulatory Class: Class I, reserved Product Code: JJX Dated: May 2, 2012 Received: May 24, 2012

Dear Ms. Schievelbein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours.

N

Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE STATEMENT

+121.534 510(k) Number (if known):

CONE-TROL Hemoglobin A1c Control Set Device name:

Indications for Use:

CONE-TROL Hemoglobin A1c Control Set is intended for use as quality control material to monitor the performance and precision of Hemoglobin A 1 c determination methods.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Russell Chun

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

K121534 510(k)

Page 1 of 1

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.