(168 days)
The Treasure Bipolar Head is intended for use as follows:
- A. Fractures of the proximal femur;
- B. Nonunions of proximal femoral neck fractures;
- C. Aseptic necrosis of the femoral head;
- D. Osteo-rheumatoid and post-traumatic arthritis of the hip with minimal distortion of the acetabulum;
- E. Salvage of failed total hip arthroplasty.
The Treasure Bipolar Head is intended for cementless use. The Treasure Bipolar Head is intended for use with the Helicon Hip System.
The TGM Medical, Inc. Treasure Bipolar Head consists of a bipolar femoral head component, locking ring, and a bipolar insert component. The Treasure Bipolar Head is available either fully preassembled, or with the bipolar femoral head and locking ring preassembled and the insert separate. The bipolar femoral head component is manufactured from cobalt chrome alloy (CoCrMo, ASTM F75, ASTM F799, or ASTM F1537). The bipolar head has a highly polished spherical outer surface with a cylindrical bored internal diameter which accepts the polyethylene bipolar insert. The bipolar insert component and locking ring are manufactured from ultra-high molecular weight polyethylene (UHMWPE, ASTM F648). The bipolar insert and locking ring are designed for use with the appropriate size bipolar head component.
The provided text describes a 510(k) premarket notification for the "Treasure Bipolar Head." This is a medical device submission, and the focus of such submissions is typically demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a performance study with specific acceptance criteria in the way a clinical trial might.
Based on the provided text, here is the information regarding acceptance criteria and the "study" (which in this context is the substantial equivalence claim):
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Substantial Equivalence to Predicate Device: The Treasure Bipolar Head must be substantially equivalent in materials, design, packaging, sterilization, and indications for use to legally marketed predicate devices. | The components of the new device (CoCr bipolar head, UHMWPE locking ring, and insert) employ identical materials, design features, packaging, and sterilization to the respective COI components (K922560, K102399, K111472) and have similar indications. |
| No New Performance Data Required: No new non-clinical or clinical performance data should be required to demonstrate substantial equivalence, implying that the device performs equivalently to the predicate(s). | "No new performance data was required for the Treasure Bipolar head." |
2. Sample size used for the test set and the data provenance
The document explicitly states, "No new performance data was required for the Treasure Bipolar head." Therefore, there was no "test set" in the sense of a new study involving human subjects or laboratory testing specifically for this 510(k) application. The substantial equivalence claim relies on the existing clearances and data for the predicate devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. No new test set requiring expert ground truth establishment was conducted for this 510(k) submission. The ground truth for the safety and effectiveness of the predicate devices would have been established during their original clearance processes.
4. Adjudication method for the test set
Not applicable. No new test set requiring adjudication was conducted for this 510(k) submission.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a hip prosthesis, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a hip prosthesis, not an algorithm.
7. The type of ground truth used
The "ground truth" for the substantial equivalence claim is the pre-existing FDA clearance of the predicate devices (Consensus Bipolar System K922560, Consensus TaperSet Hip System K102399, and Helicon Hip System K111472), implying that these devices were deemed safe and effective based on their original data and regulatory review. The current submission's argument is that the Treasure Bipolar Head is sufficiently similar to these predicate devices not to raise new questions of safety or effectiveness.
8. The sample size for the training set
Not applicable. There was no "training set" in the context of this 510(k) submission, as no new performance data or clinical study was conducted.
9. How the ground truth for the training set was established
Not applicable, as no training set was used.
{0}------------------------------------------------
K12150
NOV. 5 2012
2. 510(k) SUMMARY
Sponsor Name:
TGM Medical, Inc. 5145 Golden Hills Parkway, Suite 175 El Dorado Hills, CA 95762
510(k) Contact:
Prakash Pai
Phone: (774) 277-1312 prakash.pai63@yahoo.com
| Date Prepared: | Revised October 21, 2012 |
|---|---|
| Trade Name: | Treasure Bipolar Head |
| Common Name: | Bipolar head prosthesis for cementless use |
| Classification Name: | Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis (21 CFR 888.3390, Class II device, Product Code KWY). |
Device Description:
The TGM Medical, Inc. Treasure Bipolar Head consists of a bipolar femoral head component, locking ring, and a bipolar insert component. The Treasure Bipolar Head is available either fully preassembled, or with the bipolar femoral head and locking ring preassembled and the insert separate. The bipolar femoral head component is manufactured from cobalt chrome alloy (CoCrMo, ASTM F75, ASTM F799, or ASTM F1537). The bipolar head has a highly polished spherical outer surface with a cylindrical bored internal diameter which accepts the polyethylene bipolar insert. The bipolar insert component and locking ring are manufactured from ultra-high molecular weight polyethylene (UHMWPE, ASTM F648). The bipolar insert and locking ring are designed for use with the appropriate size bipolar head component.
Indications for Use:
The Treasure Bipolar Head is intended for use as follows:
- A. Fractures of the proximal femur;
- B. Nonunions of proximal femoral neck fractures;
- C. Aseptic necrosis of the femoral head;
- D. Osteo-rheumatoid and post-traumatic arthritis of the hip with minimal distortion of the acetabulum;
2
- E. Salvage of failed total hip arthroplasty.
The Treasure Bipolar Head is intended for cementless use. The Treasure Bipolar Head is intended for use with the Helicon Hip System.
Substantial Equivalence:
f
{1}------------------------------------------------
Technological Characteristics/Substantial Equivalence:
Consensus Orthopedics, Inc. (COI) licensed the previously cleared TaperSet femoral stem (K102399) to TGM Medical, Inc. which was cleared as the Helicon Hip System (K111472). COI has also licensed their previously cleared Bipolar Head (K922560) to TGM for use with Helicon. The components of the new device employ identical materials, design features, packaging and sterilization to the respective COI components and have similar indications. Therefore, the CoCr bipolar head, UHMWPE locking ring, and insert are substantially equivalent to the legally marketed COI predicate devices (Table 2.1).
| 510(k)Number | Trade Name | 510(k) holder | 510(k)Release Date |
|---|---|---|---|
| K922560 | Consensus Bipolar System | U.S. Medical Products, Inc. | 09/11/1992 |
| K102399 | Consensus TaperSet Hip System | Consensus Orthopedics, Inc. | 12/02/2010 |
| K111472 | Helicon Hip System | TGM Medical, Inc. | 9/06/2011 |
3
Table 2.1: Legally marketed devices to which substantial equivalence is claimed:
Non-Clinical Performance Data:
No new performance data was required for the Treasure Bipolar head.
Z of Z.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure with three curved lines representing its body and wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
TGM Medical, Inc. % Prakash Pai 5145 Golden Foothill Parkway Suite 175 & 180 El Dorado Hills, CA 95762 US
November 5, 2012
Re: K121509
Trade/Device Name: Treasure bipolar head, femoral head, treasure bipolar head, insert Regulation Number: 21 CFR 21 CFR 888.3390
Regulation Name: Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented Or Uncemented Prosthesis
Regulatory Class: Class II Product Code: KWY
Dated: September 21, 2012 Received: September 26, 2012
Dear Prakash Pai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Prakash Pai
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the qualify of ovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Ìnternet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Peter D. Rumm -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
1. INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: Treasure Bipolar Head
Indications for Use:
The Treasure Bipolar Head is intended for use as follows:
- A. Fractures of the proximal femur;
- B. Nonunions of proximal femoral neck fractures;
- C. Aseptic necrosis of the femoral head;
- D. Osteo-rheumatoid and post-traumatic arthritis of the hip with minimal distortion of the acetabulum:
- E. Salvage of failed total hip arthroplasty.
The Treasure Bipolar Head is intended for cementless use. The Treasure Bipolar Head is intended for use with the Helicon Hip System.
Over the Counter Use _ AND/OR Prescription Use X (21 CFR Part 801 Subpart C) (21 CFR Part 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ase
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121509
N/A