K Number
K121444

Validate with FDA (Live)

Date Cleared
2012-06-07

(23 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The flat panel detector CSX-20 is designed to provide fluoroscopic and spot radiographic images of human anatomy during diagnostic, surgical and interventional procedures. Examples of clinical application may include angiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical-care and emergency room procedures or other-imaging applications at the physician's discretion. The device is intended to replace the spot-film devices. The device is also intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.

Device Description

The Canon CSX-20 is a digital radiography flat panel detector that can take fluoroscopic and spot radiographic images of any part of the body. It directly converts the X-ray images captured by the sensor into high-resolution digital images. The instrument is a component of an x-ray system and as such cannot be used outside of such a system. This unit converts the X-rays into digital signals. Not intended for mammography applications.

AI/ML Overview

The Canon CSX-20 is a digital radiography flat panel detector for fluoroscopic and spot radiographic images. The acceptance criteria and the study that proves the device meets the acceptance criteria are described below:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Improved spatial resolution via 16-bit A/D conversionAchieved with 16-bit A/D conversion from 14-bit
Enhanced signal-to-noise ratio (S/N)Achieved with 16-bit A/D conversion from 14-bit
Increased tonal precision (grayscale gradations)Grayscale gradations quadrupled from 4096 (CSX-10) to 16384 (CSX-20)
Improved maximum frame rate for fluoroscopyMaximum frame rate of 240 fps achieved
Safety and effectivenessDevice demonstrated to be safe and effective through performance testing, software validation, electrical safety, and electromagnetic compatibility testing. Complies with US Performance Standard for radiographic equipment.
Substantial equivalence to predicate device (K111824 Canon CSX-10)Determined to be substantially equivalent based on similarities in intended use, principles of operation, functional design, and medical use.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a specific "test set" in terms of clinical images or patient data that would have a sample size. The testing appears to be primarily device performance testing and refers to "non-clinical/test data" and "performance testing and software validation." This suggests the test set would consist of technical measurements and functional evaluations of the device itself rather than patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Given the nature of the testing described (performance testing, software validation, electrical safety, EMC), there is no mention of a "ground truth" derived from human expert evaluation of images. The ground truth for this type of device would generally be established by engineering and technical standards.

4. Adjudication Method for the Test Set

Not applicable, as the testing described focuses on technical and performance metrics of the device rather than clinical interpretation by multiple experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study was performed or is mentioned. The submission focuses on demonstrating substantial equivalence to a predicate device through technical improvements and performance testing of the device itself. There is no comparison of human reader performance with or without AI assistance.

6. Standalone Performance

The information provided describes the performance of the device (CSX-20 Flat Panel Detector) itself, rather than an algorithm. The device "directly converts the X-ray images captured by the sensor into high-resolution digital images." Its performance metrics (A/D conversion, grayscale, frame rate) relate to its standalone function as an imaging component.

7. Type of Ground Truth Used

The ground truth used for this submission is based on:

  • Engineering specifications and standards: For A/D conversion, S/N, grayscale gradations, and frame rate.
  • Regulatory standards: For electrical safety, electromagnetic compatibility, and compliance with the US Performance Standard for radiographic equipment.
  • Functional equivalence: Comparison to the predicate device (CSX-10) based on intended use, principles of operation, and functional design.

8. Sample Size for the Training Set

Not applicable. This device is a hardware component (digital X-ray detector). The concept of a "training set" typically applies to machine learning algorithms, which are not described as part of this device. The improvements are described as hardware modifications (A/D conversion).

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" or machine learning algorithm described. The improvements are based on hardware modifications and engineering design.

{0}------------------------------------------------

Canon

5. 510(k) SUMMARY

JUN - 7 2012

K1214444

.

5. 510(k) SUMMARYJUN - 7 20
Submitter:Canon, Inc. - Medical Equipment Group30-2 Shimomaruko, 3-chromeOhta-ku, Tokyo 146-8501 Japan
Contact Person:Mr. Naoyasu AsakaStaff ManagerTEL: 81-3-3758-2111FAX: 81-3-5482-3960asaka.naoyasu@canon.co.jp
Date Prepared:April 24, 2012 (Revised June 7, 2012)
Trade Name:CSX-20 Flat Panel Detector
Common Name:Solid State X-Ray Imager (Flat Panel/Digital Imager)
Classification Name:OWB, JAA 892.1650 Image-intensified fluoroscopic x-ray system
Predicate Device:K111824 Canon CSX-10 (MQB 892.1650)
Device Description:The Canon CSX-20 is a digital radiography flat panel detector that can takefluoroscopic and spot radiographic images of any part of the body. It directlyconverts the X-ray images captured by the sensor into high-resolution digitalimages. The instrument is a component of an x-ray system and as such cannot beused outside of such a system. This unit converts the X-rays into digital signals. Notintended for mammography applications.
Statement ofIntended Use:The flat panel detector CSX-20 is designed to provide fluoroscopic and spotradiographic images of human anatomy during diagnostic, surgical andinterventional procedures. Examples of clinical application may includeangiography, endoscopy, urologic, orthopedic, neurologic; vascular, cardiac,critical-care and emergency room procedures or other imaging applications at thephysician's discretion. The device is intended to replace spot-film devices. Thedevice is also intended to replace fluoroscopic images obtained through imageintensifier technology. Not intended for mammography applications.
Summary ofTechnologicalCharacteristics:Comparison with the predicate shows the technological characteristics of theCSX-20 are substantially equivalent to the predicate device. The flat panel detectorunits are functionally identical.The A/D Conversion was modified from 14-bit precision to 16-bit precision toimprove the resolution and S/N. This modification allows for improved tonalprecision.Grayscale is a function of the A/D conversion. As a result of the A/D conversionmodification, the grayscale gradations have quadrupled from the 4096 of the CSX-10 to the 16384 gradations now available with the CSX-20.By modifying the A/D conversion to 16-bits, a maximum frame rate of the 240 fpshas been improved.The modifications described above are all related to the change in A/D conversionfrom 14-bit to 16-bit precision. These modifications have been incorporated into theCSX-20 in an effort to improve the product performance.

. . . . .

{1}------------------------------------------------

Canon

Summary of Non-Clinical /Test Data:

Tests were performed on the device which demonstrated that the device is safe and effective, performs comparably to and is substantially equivalent to the predicate device. Tests include: Performance testing and Software Validation. Electrical safety and Electromagnetic Compatibility testing has been performed. The unit complies with the US Performance Standard for radiographic equipment. .

Conclusion:

Canon, Inc. - Medical Equipment Group considers the CSX-20 to be substantially equivalent to the predicate device listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use for the flat panel detector.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

JUN - 7 2012

Canon, Inc. - Medical Equipment Group % Ms. Diane Rutherford Regulatory Engineer Ken Block Consulting 1201 Richardson Drive, Suite 280 RICHARDSON TX 75080

Re: K121444

Trade/Device Name: CSX-20 Flat Panel Detector Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB and JAA Dated: April 24, 2012 Received: May 15, 2012

Dear Ms. Rutherford:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA): You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number:

Device Name: CSX-20 Flat Panel Detector

Indications for Use:

The flat panel detector CSX-20 is designed to provide fluoroscopic and spot radiographic images of human anatomy during diagnostic, surgical and interventional procedures. Examples of clinical application may include angiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical-care and emergency room procedures or other-imaging applications at the physician's discretion. The device is intended to replace the spot-film devices. The device is also intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.

Prescription Use _ X (21 CFR 801 Subpart D)

Over-the-Counter Use AND/OR (21. CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDHR, Office of Device Evaluation (ODE)

Jahmal DD
(Division Sign-Off)

(Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510KLA21444

Page 1 of 1

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.