K Number
K121422

Validate with FDA (Live)

Date Cleared
2012-08-16

(94 days)

Product Code
Regulation Number
892.1560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500, APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, and musculoskeletal (both conventional and superficial).

Device Description

The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500. APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 are mobile system. These systems are Track 3 devices that employ a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz.

AI/ML Overview

The provided text is a 510(k) summary for the Toshiba DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500, APLIO 400 MODEL TUS-A400, and APLIO 300 MODEL TUS-A300 Version 2.1. It primarily focuses on demonstrating substantial equivalence to a predicate device (Toshiba DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500 / APLIO 400 MODEL TUS-A400 / APLIO 300 MODEL TUS-A300 V2.0; 510(k) control number K110870) and outlining the intended uses for various transducers.

Based on the provided document, the information requested in your prompt about acceptance criteria and a study proving performance (specifically in the context of AI/ML or new clinical parameters) is not present. This document predates the widespread regulatory submissions for AI/ML devices in medical imaging, and its focus is on general ultrasound system functionality and traditional clinical applications.

Therefore, I cannot provide a table of acceptance criteria and reported device performance for AI features, nor can I provide details about sample sizes for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC comparative effectiveness studies, or standalone algorithm performance studies related to AI.

The document does include detailed tables for each transducer showing the "Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:", listing various clinical applications and modes of operation. For each application, it indicates whether it's a "P" (previously cleared by FDA) or "E" (added under this appendix) indication, or "N" (new indication) for a very few. This is essentially a declaration of the intended uses, but it does not specify performance acceptance criteria or provide study results to demonstrate performance for these applications.

The document mentions compliance with several standards, such as IEC 60601-1 and its parts, IEC 62304, and AIUM-NEMA UD2/UD3 standards. These are general safety and performance standards for medical electrical equipment and ultrasound output measurement/display. They are compliance standards, not specific acceptance criteria for AI or diagnostic performance in patient studies.

In summary, the provided content is a regulatory submission for device clearance based on substantial equivalence to a predicate device, focusing on intended uses and compliance with general safety and performance standards. It does not contain the kind of detailed information about acceptance criteria and performance studies you're asking for, particularly concerning AI or specific diagnostic efficacy metrics.

{0}------------------------------------------------

K121422

510(k) Summary of Safety and Effectiveness:

AUG 1 6 2012

Submitter's Name:Toshiba America Medical Systems, Inc.
Address:PO Box 2068,2441 Michelle Drive Tustin, CA 92781-2068
Contact:Charlemagne Chua, Manager Regulatory Affairs
Telephone No.:(714) 730-5000
Device Proprietary Name:DIAGNOSTIC ULTRASOUND SYSTEMAPLIO 500 MODEL TUS-A500 Version 2.1APLIO 400 MODEL TUS-A400 Version 2.1APLIO 300 MODEL TUS-A300 Version 2.1
Common Name:Diagnostic Ultrasound System
Classification:
Regulatory Class:II
Review Category:Tier II
Ultrasonic Pulsed Doppler Imaging System - Product Code: 90-IYN

[Fed.Reg.No.:892.1550] Ultrasonic Pulsed Echo Imaging System - Product Code: 90-IYO [Fed.Reg.No.:892.1560] Diagnostic Ultrasonic Transducer - Product Code: 90-ITX [Fed. Reg. No.: 892.1570]

Identification of Predicate Devices:

Toshiba America Medical Systems believes that this device is substantially equivalent to:

  • l) Toshiba DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500 / APLIO 400 MODEL TUS-A400 / APLIO 300 MODEL TUS-A300 V2.0 ; 510(k) control number K110870

Device Description:

The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500. APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 are mobile system. These systems are Track 3 devices that employ a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz.

Indications for Use:

The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500, APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, and musculoskeletal (both conventional and superficial).

{1}------------------------------------------------

Declaration of Conformity

p

This device is designed and manufactured in conjunction with the Quality System Regulation, IEC 60601-1 (applicable portions), IEC 60601-1-1 (applicable portion), IEC 60601-1-2 (applicable portion), IEC 60601-1-4 (applicable portion), IEC60601-2-37 (applicable portions), IEC 62304 (applicable portion) and the AIUM-NEMA UD2 Output Measurement Standard as applied to Track 3 Ultrasound systems and the AIUM-NEMA UD3 Output Display Standard.

Standards form:

Please see the attached standard form of the IEC 60601-1-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37 and IEC 62304.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized emblem. The emblem consists of a white eagle with its wings spread, superimposed over a blue background. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the emblem.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Toshiba Medical Systems Corporation, Japan % Mr. Charlemagne Chua Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780-2068

AUG 16 2012

Re: K121422

Trade/Device Name: Aplio™ 500/400/300 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: July 13, 2012 Received: July 16, 2012

Dear Mr. Chua:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Aplio™ 500/400/300 Diagnostic Ultrasound System, as described in your premarket notification:

Transducer Model Number

PST-25BTPVT-661VTPVT-770RTPLT-1204BT
PST-30BTPVT-674BTPLT-604ATPLT-1204BX
PST-50BTPVT-675MVPLT-704ATPLT-1204MV
PST-65ATPVT-681MVPLT-704SBTPET-508MA
PVT-375BTPVT-712BTPLT-705BTFPET-510MB
PVT-375MVPVT-745BTFPVT-705BTHPET-512MC
PVT-382BTPVT-745BTHPLT-805ATPC-20M
PVT-382MVPVT-745BTVPLT-1202SPC-50M

{3}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Joshua Nipper at 1 (301) 796-6524.

Sincerely Yours,

Michael D O'Shea for

Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Aplio™ 500 /400/300 Diagnostic Ultrasound System

Indications for Use:

The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500, APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, and musculoskeletal (both conventional and superficial).

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)


(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) Number K124222

{5}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: __________________________________________________________________________________________________________________________________________________________________ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
OphthalmicP2PPPP5,7
FetalPPPPP2PPPP5,7
AbdominalPPPPP2,3PPPP5,7,12
Intra-operative (Abdominal)PPPPP2PPP4,5
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP2,3PPPP5,7,12
Small Organ (Note 1)PPPPP2PPP4,5,6,7,11
Neonatal CephalicPPPPP3PPP
Adult CephalicPPPPP3PPP
Trans-rectalPPPPP2PPPP4,5,7,11
Trans-vaginalPPPPP2PPPP4,5,7
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPP2PPP4,5,6,7,11
Musculo-skeletal (Superficial)PPPPP2PPP4,5,6,7,11
Intravascular
Other (Specify)
Cardiac AdultPPPPP3PPPP4
Cardiac PediatricPPPPP3PPPP4
Intravascular (Cardiac)
Trans-esoph. (Cardiac)PPPPP3P4
Intra-cardiac
Other (Specify)
Peripheral vesselPPPPP2PPP4,5,6,7,11
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI

  • Note 5 ApliPure
    Note 13 2D WMT

Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion


(Division Sign-Off)

Division of Radiological Devices Office of In Vitro ation and Safety

510K K12.1422

Prescription Use Only (Per 21 CRF801.109)

{6}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: ______________________________________________________________________________________________________________________________________________________________________ Transducer: PST-25BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
AbdominalPPPPP3PPP11
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP3PPP
Small Organ (Specify) (1)
Neonatal CephalicPPPPP3PPP
Adult CephalicPPPPP3PPP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial).
Intravascular
Other (Specify)
Cardiac AdultPPPPP3PPPP4,13
Cardiac PediatricPPPPP3PPPP4,13
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
(Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

516

Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Off)

{7}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PST-30BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
AbdominalPPPPP3PPP11
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP3PPP
Small Organ (Specify) (1)
Neonatal CephalicPPPPP3PPP
Adult CephalicPPPPP3PPP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultPPPPP3PPPP4,13
Cardiac PediatricPPPPP3PPPP4,13
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle.

Note 1 Small organ includes BM; B/PWD; BDF/PWD; BDF/PWD; BDF/A/DF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion
    Note 13 2D WMT Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Off)

of Radiological Devices

510

{8}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:__ PST-50BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther(Note)
Ophthalmic
Fetal
AbdominalEEEEE3EEE11
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricEEEEE3EEE
Small Organ (Specify) (1)
Neonatal CephalicEEEEE3EEE
Adult CephalicEEEEE3EEE
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultEEEEE3EEEE4,13
Cardiac PediatricEEEEE3EEEE4,13
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K090158

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 2 Combined mode includes BAN, BR WD, BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI
  • Note 5 ApliPure
  • Note 6 MicroPure
  • Note 7 Precision Imaging
  • Note 8 STIC
  • Note 9 3D Color (Volume Color) Note 10 STIC Color
  • Note 11 Elastography
  • Note 12 Fusion
  • Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Off)
Division of Radiological Devices
510k K121492

MAY 2012

{9}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PST-65AT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of OperationBMPWDCWDColorDopplerCombined(Specify)*TIIIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
AbdominalEEEEE3EEE11
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricEEEEE3EEE
Small Organ (Specify) (1)
Neonatal CephalicEEEEE3EEE
Adult CephalicEEEEE3EEE
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultEEEEE3EEEE4,13
Cardiac PediatricEEEEE3EEEE4,13
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645

Note I Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color
    Note 11 Elastography · Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

BealEBS

(Division Sign-Off)

on of Radiological

ﻟﻘﺎﺩ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ

MAY 2012

{10}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __Aplio 300, Aplio 400, Aplio 500 v2,1 Transducer: _ PVT-375BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWD CWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
FetalPPPP2PPP5,7
AbdominalPPPP2PPP5, 7,11,1
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPP5, 7,12
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note I Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
    Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Off)

510(k): APLIO 500/400/300, v2.1 Page 194 of 3006

{11}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PVT-375MV

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
FetalPPPPP2PPPN5,7,8,9,10
AbdominalPPPPP2PPPN5,7,8,9,10
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP2PPPN5,7,8,9,10
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645

Note 1 Small organ includes thyroid, breast and testicle. Note 1 Small organ alstades arfrenz, e. BRWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

A-7

{12}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PVT-382BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of OperationBMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
FetalPPPPP2PPP5,7
AbdominalPPPPP2PPP5,7
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP2PPP5,7
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N= new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography
    Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

Ruch D. 82

{13}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PVT-382MV

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of OperationBMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
FetalPPPPP2PPPP5, 7,9
AbdominalPPPPP2PPPP5, 7,9
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP2PPPP5,7,9
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix ·Previous 510(k) of the transducer: K103645

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color
    Note 11 Elastography Note 12 Fusion

Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Off)
Division of Radiological Devices
and

A-9 -

{14}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: ______________________________________________________________________________________________________________________________________________________________________ Transducer:_ PVT-661VT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:


Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPP2PPP4,5,7,11
Trans-vaginalPPPPP2PPP4,5,7,11
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle. .

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT .
    Prescription Use Only (Per 21 CRF801.109)

vision Sion-Off

ological Devices

{15}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PVT-674BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of Operation
BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
FetalPPPP2PPP5,7
AbdominalPPPP2PPP5,7
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPP5,7
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes BM; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF 801.109)
(Division Sign-Off)

{16}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PVT-675MV ___

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
FetalPPPP2PPPP5,7,8,9,10
AbdominalPPPP2PPPP5,7,8,9,10
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPPP5,7,8,9,10
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K 110870

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Division of Radiological Devices

510k
MAY 2012

{17}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1)TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:___ PVT-681MV

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPP2PPPP4,5,7,9,1
Trans-vaginalPPPP2PPPP4,5,7,9,1
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI
  • Note 5 ApliPure
  • Note 6 MicroPure

Note 7 Precision Imaging

  • Note 8 STIC Note 9 3D Color (Volume Color)
    Note 10 STIC Color

  • Note 11 Elastography
    Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

Michael D. O'Keefe

(Division Sign-Off)
Division of Radiological Devices

2012

{18}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1)TUS-A500 Ultrasound System

System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PVT-712BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
AbdominalPPPP2PPPP5,7
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPP5,7
Small Organ (Specify) (1)
Neonatal CephalicPPPP2PPP5,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication: P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CFR801.109)

(Division Sign-Off Division of Radiological Devices

STOR
MAY 20

{19}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 PVT-745BTF Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
AbdominalEEEEE2EEE5,7
Intra-operative (Abdominal)EEEEE2EEE5,7
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)EEEEE2EEE5,7
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: None

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Notc 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per

(Division Sign-Off)
Division of Radiological Devices
OIVD

{20}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PVT-745BTH

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
AbdominalEEEE2EEE5,7
Intra-operative (Abdominal)EEEE2EEE5,7
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)EEEE2EEE5,7
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: None

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

A-16

100 rescription Use On.

510

{21}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

Toshiba America Medical Systems, Inc.

System: Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:__ PVT-745BTV

Intended Use; Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
AbdominalPPPPP2PPP5,7
Intra-operative (Abdominal)PPPPP2PPP5,7
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

cal Devices

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M: B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per 21 CRF80

Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Off)

MAY 2012

{22}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PVT-770RT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectalEEEE2EEE4,5,7,1
Trans-vaginalEEEE2EEE4,5,7,1
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 2 Combined mode includes BMF, BF WD, BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
    Prescription Use Only (Per 21 CRF801.109)

Mulin D'then
(Division Sign-Off)

of Radiological Devices

{23}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

Toshiba America Medical Systems, Inc.

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PLT-604AT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)EEPWDE2EEEE5,7
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)EEEE2EEEE5,7
Musculo-skeletal (Superficial)EEEE2EEEE5,7
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselEEEE2EEEE5,7
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645

Note 1 Small organ includes thyroid, breast and testicle. Note 1 Shartorgan motiades arrivate, 6 R/F/FWD; BDF/PWD; BDF/A/DF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per 21 CRF801.109)

Prescription Use Only (Per 21 CFR 801.109)

Division

MAY 2012

{24}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 PLT-704AT Transducer:_

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of OperationBMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)EEEE2EEE5,7
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)EEEE2EEE5,7
Musculo-skeletal (Superficial)EEEE2EEE5,7
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselEEEE2EEE5,7
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645

Note 1 Small organ includes thyroid, breast and testicle. Note 1 Sman organ includes BlM; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per 21 CRF

Prescription Use Only (Per 21 CFR 801.109)

(Division Sign-Off)

Sign-Off)
Radiological Devices
OUD
422
A-20

510(k): APLIO 500/400/300, v2.1 Page 208 of 3006

{25}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

Toshiba America Medical Systems, Inc.

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PLT-704SBT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)PPPPP2PPP5,7
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPP2PPP5,7
Musculo-skeletal (Superficial)PPPPP2PPP5,7
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselPPPPP2PPP5,7
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only

(Division Sign-Off)
Division of Radiological Devices

{26}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 PLT-70SBTF Transducer:_

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of OperationBMPWD CWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
AbdominalEEEE2EEE5,7
Intra-operative (Abdominal)EEEE2EEE5,7
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

al Devices

{27}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 PVT-705BTH Transducer:_

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWD CWD]ColorDopplerCombined(Specify)*THI DynamicFlowPowerCHI2D4DOtherNote
Ophthalmic
Fetal
AbdominalEEEE2E। ਜE5,7
Intra-operative (Abdominal)EEEE2EEE5,7
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
(Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

lological Devices

Previous 510(k) of the transducer: K103645

:

MAY 201

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion ` Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Off)

{28}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PLT-805AT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)PPPPP2PPP5,6,7,11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPPP2PPP5,6,7,11
Musculo-skeletal (Superficial)PPPPP2PPP5,6,7,11
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselPPPPP2PPP5,6,7,11
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870 i

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M: B/PWD; BDF/PWD; BDF/MDF; 8DF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
    Prescription Use Only (Per 21 CRF801.109)

(Division Sign-Off)

ion of Radiological Devices


MAY 2012

510k

{29}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

·

System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PLT-1202S

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of Operation
BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)PPPP2PP4,5,11
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)PPPP2PP4,5,11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPP2PP4,5,11
Musculo-skeletal (Superficial)PPPP2PP4,5,11
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselNNNN2NN4,5,11
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI
    Note 5 ApliPure

  • Note 6 MicroPure Note 7 Precision Imaging
    Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color

Note 11 Elastography

Note 12 Fusion Note 13 2D WMT

MAY 2012

Prescription Use Only (Per 21 CRF801.109)

Manfred D. Krahn

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Division C Divisi un of Radiolo

A-25

510(k): APLIO 500/400/300, v2.1 Page 213 of 3006

{30}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _ Aplio 300, Aplio 400, Aplio 500 v2, 1 Transducer:___PLT-1204BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOtherNote
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)PPPPP2PPP4,5,6,7,1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPP2PPP4,5,6,7,1
Musculo-skeletal (Superficial)PPPPP2PPP4,5,6,7,1
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselNNNNN2NNN4,5,6,7,1
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI
    Note 5 ApliPure

  • Note 6 MicroPure
    Note 7 Precision Imaging

  • Note 8 STIC
    Note 9 3D Color (Volume Color)

  • Note 10 STIC Color
    Note 11 Elastography

Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

(Division Sign

of Radiolo

  • 2313 401 401 10.

{31}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

Aplio 500 v2.1 System: Transducer:_ PLT-1204BX

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation Clinical Application Specific PWD CWD Color Combined THI Dynamic Power CHI 4D Other B M Doppler (Specify) Flow 2D [Note] (Tracks 3) Ophthalmic Fetal Abdominal Intra-operative (Abdominal) Intra-operative (Neuro) Laparoscopic Pediatric P P P P 5,7 Small Organ (Specify) (1) P P P 2 Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skeletal P P P P 5,7 b P P 2 (Conventional) Musculo-skeletal (Superficial) P P P P 2 P P P 5,7 Intravascular Other (Specify) Cardiac Adult Cardiac Pediatric Intravascular (Cardiac) Trans-esoph. (Cardiac) Intra-cardiac Other (Specify) Peripheral vessel P P P P P P P 5,7 2 Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note I Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M: B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109)

Division G

{32}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

A M SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE FELLER

System: _Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PLT-1204MV

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)PPPPP2PPPP5,7,8,9,1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPP2PPPP5,7,8,9,1
Musculo-skeletal (Superficial)PPPPP2PPPP5,7,8,9,1
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselPPPPP2PPPP5,7,8,9,1
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI
    Note 5 ApliPure

Note 6 MicroPure

Note 7 Precision Imaging

Note 8 STIC Note 9 3D Color (Volume Color)

Note 10 STIC Color

Note 11 Elastography Note 12 Fusion

Note 13 2D WMT

Prescription Use Only (Per 21 CFR 801.109)

{33}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: __ Aplio 300, Aplio 400. Aplio 500 v2.1 Transducer:_ PET-508MA


Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)EEEEE3E4,13
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: None

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

ical Devices

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI
    Note 5 ApliPure

  • Note 6 MicroPure
    Note 7 Precision Imaging

  • Note 8 STIC
    Note 9 3D Color (Volume Color)

  • Note 10 STIC Color
    Note 11 Elastography

Note 12 Fusion

Note 13 2D WMT

MAY

Prescription Use Only (Per 21 CRF801.109)

A-29

510(k): APLIO 500/400/300, v2.1 Page 217 of 3006

{34}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PET-510MB

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)PPPPP3P4,13
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: KI 10870

Note 1 Small organ includes thyroid, breast and testicle.

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

  • Note 4 TDI
    Note 5 ApliPure

Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color

Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRF801.109) (Division Sign-Off)

vision of Radiological Devices

510k MAY 2012

{35}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PET-512MC

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)EEEEE3E4,13
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103629

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21 CRE801 109 Division Sign-Off

vision of Radiological Devices

MAY 2012

A-3 I

{36}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

System: Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther(Note)
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultP
Cardiac PediatricP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselP
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per 21 CRF801.109)

Division Sign-Off

ogical Devices

.

{37}------------------------------------------------

510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System

Aplio 300, Aplio 400, Aplio 500 v2.1 System: Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColorDopplerCombined(Specify)*THIDynamicFlowPowerCHI2D4DOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify) (1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultP
Cardiac PediatricP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselP
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870

Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT

Prescription Use Only (Per 21, GRF80 109 (Division Sign-Off

vision of Radiological Devices

  • 4 - 8 - 4 - 4 - 4 - 4 - 4 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -

N/A