K Number
K121356

Validate with FDA (Live)

Date Cleared
2012-11-09

(189 days)

Product Code
Regulation Number
888.3040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Micro/Mini N-PK Bone Anchor is intended for the reattachment of soft tissue to bone for the following indications: Elbow, Hand and Wrist - Biceps tendon reattachment - Ulnar or lateral collateral ligament reconstruction . - ◆ Scapholunate ligament reconstruction - . Lateral epicondylitis repair

Device Description

The Micro/Mini N-PK(H) Bone Anchor consists of an adjustable loop of non-absorbable suture (Sizes 2, 0 and 2- 0) with one (1) or two (2) attached anchors. The bone anchor is a two part system, one part nickel-titanium (nitinol) and the second part polyetheretherketone (PEEK). The PEEK part resides within the nitinol sleeve and retains the suture. The nitinol part expands upon delivery providing resistance to pullout. The construct is provided sterile and preloaded on a disposable delivery instrument.

AI/ML Overview

This 510(k) summary describes the Micro/Mini N-PK(H) Bone Anchor, a device intended for the reattachment of soft tissue to bone. The study focuses on demonstrating substantial equivalence to predicate devices through functional and safety testing.

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria (Test)Reported Device Performance
Functional and Safety Testing
BiocompatibilityConducted in accordance with ISO 10993-1 standards.
Corrosion testingConducted in accordance with ASTM F2129.
Tensile testingPerformed to verify compliance with USP suture requirements.
Fixation/Static disassembly strengthComparison testing performed.
Cyclic fatigue performanceComparison testing performed.

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample sizes used for each specific test (biocompatibility, corrosion, tensile, fixation/static disassembly, cyclic fatigue). The data provenance is not specified either (e.g., country of origin, retrospective or prospective). These tests are typically performed in a laboratory setting.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The study involves laboratory-based functional and safety testing of a medical device, not a diagnostic or prognostic assessment requiring expert interpretation of results. Therefore, there is no "ground truth" in the clinical sense established by experts.

4. Adjudication method for the test set

Not applicable, as this is laboratory testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a 510(k) submission for a bone anchor, not an AI-based diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used

For the functional and safety testing, the "ground truth" is defined by the established industry standards and requirements against which the device's performance is measured:

  • Biocompatibility: ISO 10993-1 standards.
  • Corrosion: ASTM F2129 standards.
  • Tensile testing: USP suture requirements.
  • Fixation/Static disassembly strength and Cyclic fatigue performance: Comparison testing to predicate devices, implying that the performance of the predicate devices serves as the benchmark or "ground truth" for acceptable performance.

8. The sample size for the training set

Not applicable. There is no concept of a "training set" for this type of medical device testing. The device's design and materials are tested against established performance standards and predicate device performance.

9. How the ground truth for the training set was established

Not applicable, as there is no training set involved. The ground truth for the device's performance is established by recognized international and industry standards (ISO, ASTM, USP) and the performance characteristics of legally marketed predicate devices.

{0}------------------------------------------------

510(k) Summary
K121356

Page 1/2

.

As Required by 21 CFR 807.92

NOV 9 2012
Anulex Technologies, Inc.
Submitter:5600 Rowland Road, Suite 280
Minnetonka, MN 55343
Contact Person:Rachel Kennedy
Director, Regulatory Affairs and Quality Systems
Anulex Technologies, Inc.
Direct: 952.224.4034Fax: 952.224.5417E-Mail: rkennedy@anulex.com
Date Prepared:October 25, 2012
Trade Name:Micro/Mini N-PK(H) Bone Anchor
Classification:II
Product Code:MBI
21 CFR 888.3040
Predicate Device(s):The subject device is substantially equivalent to thefollowing predicate devices:
Smith & Nephew, Inc., Bioraptor PK Suture Anchor(K071586 cleared August 17, 2007) DePuy Mitek Mini QuickAnchor Plus (K071257cleared June 29, 2007)
Device Description:The Micro/Mini N-PK(H) Bone Anchor consists of anadjustable loop of non-absorbable suture (Sizes 2, 0 and 2-0) with one (1) or two (2) attached anchors. The boneanchor is a two part system, one part nickel-titanium(nitinol) and the second part polyetheretherketone (PEEK).The PEEK part resides within the nitinol sleeve and retainsthe suture. The nitinol part expands upon deliveryproviding resistance to pullout.The construct is provided sterile and preloaded on adisposable delivery instrument.
Indications for Use:The Micro/Mini N-PK Bone Anchor is intended for thereattachment of soft tissue to bone for the followingindications:Elbow, Hand and Wrist Biceps tendon reattachment Ulnar or lateral collateral ligament

Anulex Technologies, Inc.

.

K121356A2

{1}------------------------------------------------

Page 2/2

reconstruction

  • Scapholunate ligament reconstruction
  • Lateral epicondylitis repair ●

Functional and Safety Testing:

Comparison to Predicate:

Conclusion:

Biocompatibility testing was conducted in accordance with ISO 10993-1 standards. Corrosion testing was conducted in accordance with ASTM F2129. Tensile testing was performed to verify compliance with USP suture requirements and comparison testing to support the safety and effectiveness of the Micro/Mini N-PK(H) Bone Anchor included evaluation of fixation/static disassembly strength and cyclic fatigue performance.

The Micro/Mini N-PK(H) Bone Anchor that is the subject of this submission has the same materials and similar technological characteristics in comparison to the predicate suture anchors. The intended use is the same as the predicates, fixation of soft tissue to bone for a variety of orthopedic procedures in the elbow, hand and wrist while the predicate anchors include procedures in additional anatomical locations.

Substantial equivalence is demonstrated through the detailed device description, performance testing and conformance with voluntary performance standards.

Anulex Technologies, Inc.

K121356A2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Letter Dated: November 9, 2012

Anulex Technologies, Incorporated % Ms. Rachel Kennedy Director, Regulatory Affairs and Quality Assurance 5600 Rowland Road, Suite 280 Minnetonka, Minnesota 55343

Re: K121356

Trade/Device Name: Micro/Mini N-(PK)(H) Bone Anchor Regulation Number: 21 CFR 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener. Regulatory Class: Class II Product Code: MBI Dated: October 26, 2012 Received: October 31, 2012

Dear Ms. Kennedy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Rachel Kennedy

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address · http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Appendix 4: Indications for Use Form

KI2i356

Device Name: Micro/Mini N-PK(H) Bone Anchor

Indications for Use:

The Micro/Mini N-PK Bone Anchor is intended for the reattachment of soft tissue to bone for the following indications:

Elbow, Hand and Wrist

  • Biceps tendon reattachment
  • Ulnar or lateral collateral ligament reconstruction .
  • ◆ Scapholunate ligament reconstruction
  • . Lateral epicondylitis repair

Prescription Use _________X (Part 21 CFR 801 Subpart D AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ASh

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K171356

Micro/Mini N-PK(H) Bone Anchor Anulex Technologies, Inc. Confidential

Traditional 510(k) Premarket Notification

Page 52

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.