K Number
K121246

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-06-15

(51 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LT Cage System is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone.

Device Description

The Medyssey LT Cage System consists of cages of various widths and heights, which can be inserted between two lumbar or lumbrosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The Medyssey LT Cage System is to be used with supplemental fixation.

The device is offered in 64 sizes, with heights ranging from 8 - 15mm, lengths from 25 - 32mm and a width of 11mm. They are also available with 0°, 4° and 8° lordosis. The devices are manufactured from titanium alloy (Ti-6Al-4V) meeting ASTM F136 and are supplied non-sterile.

The purpose of this submission is to provide a device that is manufactured from titanium alloy, which is similar to the PEEK device currently available from Medyssey Co. Ltd.

AI/ML Overview

This 510(k) summary describes a medical device, the Medyssey LT Cage System, which is an intervertebral body fusion device. The submission focuses on demonstrating substantial equivalence to a predicate device, the Medyssey LP Cage.

Here's an analysis of the provided information against the requested criteria:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Material Strength (Titanium vs. PEEK)The titanium alloy (Ti-6Al-4V) used in the Medyssey LT Cage is explicitly stated to be "stronger than the currently available PEEK material."
BiocompatibilityTitanium is reported to have "no biocompatibility issues," being "used in both spinal and general orthopedic implants."
Mechanical Performance (Static Compression, Torsion, Compressive Shear, Dynamic Compression, Subsidence)The Medyssey LT Cage System is considered to be of identical design and fundamental scientific technology to the predicate Medyssey LP Cage. The predicate device was subjected to and presumably met the following ASTM standards: - Static Compression ASTM F2077-03 - Static Torsion ASTM F2077-03 - Static Compressive Shear F2077-03 - Dynamic Compression F2077-03 - Subsidence ASTM F2267-04 The submission states, "No additional testing is required to support this submission" for the subject device due to the change only being material and the new material being stronger. This implies that the LT Cage is expected to meet or exceed these performance metrics.
Substantial Equivalence to PredicateThe overarching acceptance criterion for this 510(k) is substantial equivalence to the Medyssey LP Cage. This was met, and the FDA provided a clearance letter.

2. Sample Sizes Used for the Test Set and Data Provenance

This submission is for a material change to an existing device (the Medyssey LP Cage). No new clinical test set or data was used for the Medyssey LT Cage System to prove its performance directly. The basis for acceptance is derived from the established performance of the predicate device and the recognized properties of the new material. Therefore, information regarding sample size for a test set and data provenance (country, retrospective/prospective) is not applicable or provided in this context.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. As this is not a study involving a clinical test set requiring expert ground-truthing, this information is not provided.

4. Adjudication Method for the Test Set

Not applicable. There was no test set requiring an adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This submission is for a medical device (intervertebral cage) and not an AI or imaging diagnostic tool. Therefore, a MRMC comparative effectiveness study, human readers, or AI assistance is not relevant or described.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No. This is a physical medical implant device, not an algorithm or software. Therefore, a standalone algorithm performance study is not applicable.

7. Type of Ground Truth Used

The "ground truth" in this context is the established mechanical performance and biocompatibility of the titanium alloy (Ti-6Al-4V) and the mechanical performance data of the predicate Medyssey LP Cage, which was tested against ASTM standards.

8. Sample Size for the Training Set

Not applicable. There is no "training set" in the context of this device submission, as it does not involve machine learning or an algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set, the establishment of its ground truth is not relevant.

In summary, this 510(k) submission successfully argued for substantial equivalence based on:

  • Identical design and fundamental scientific technology between the subject and predicate devices.
  • A change only in material (from PEEK to titanium alloy).
  • The new material (titanium alloy) being stronger and having established biocompatibility for orthopedic implants.
  • Reliance on existing non-clinical test data for the predicate device, which demonstrated compliance with relevant ASTM standards for mechanical performance.

The FDA's clearance (K121246) indicates agreement with the argument that "No additional testing is required to support this submission" for the subject device.

{0}------------------------------------------------

K121246

510(k) Summary

  1. Company:

Medyssey Co. Ltd. 722-3, 4F, Science Tower 4F Dongducheon City Gyeonggido, Korea Tel: 82-31-879-0414 FAX: 82-31-879-0415

  1. Contact:

Michael Kvitnitskv Chief Operating Officer Medyssey Co. Ltd. 8001 N. Lincoln Ave. Suite 401 Skokie, IL 60077 Tel: 847-982-0100 FAX: 888-518-9070

  1. Proprietary Name:

.

Medyssey LT Cage System

  1. Classification Name: Intervertebral Body Fusion Device (21 CFR888.3080); Class II, Product Code MAX

Product Description: 5.

The Medyssey LT Cage System consists of cages of various widths and heights, which can be inserted between two lumbar or lumbrosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The Medyssey LT Cage System is to be used with supplemental fixation.

The device is offered in 64 sizes, with heights ranging from 8 - 15mm, lengths from 25 - 32mm and a width of 11mm. They are also available with 0°, 4° and 8° lordosis. The devices are manufactured from titanium alloy (Ti-6Al-4V) meeting ASTM F136 and are supplied non-sterile.

The purpose of this submission is to provide a device that is manufactured from titanium alloy, which is similar to the PEEK device currently available from Medyssey Co. Ltd.

6. Indications for Use:

The LT Cage System is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone.

{1}------------------------------------------------

7. Summary of Technological Characteristics

The purpose of this special 510(k) is to provide a cage with the same dimension design and implant technique as the Medyssey LP Cage, but manufactured from titanium alloy meeting ASTM F136. The design and fundamental scientific technology of the subject Medyssey LT Cage are identical to the predicate device. The only difference is the material.

8. Identification of Legally Marketed Predicate Device

Documentation was provided, which demonstrates that the subject Medyssey LT Cage is substantially equivalent to the predicate Medyssey LP Cage cleared on April 8, 2011 under K110067. The only difference is the material of manufacture and several lumbar cage devices are manufactured from titanium alloy including the Medtronic PERIMETER Interbody Fusion Device (K111525, Aug. 25, 2011) and the Ray Cage (P950019).

9. Brief Discussion of Non-Clinical Tests Submitted

The subject and predicate Medyssey Cage devices are identical in indications for use, intended use, performance specifications, and fundamental technological characteristics. Assessment of the design change (alternate material) was completed in accordance with Medyssey design control procedures. The titanium material is stronger than the currently available PEEK material. In addition titanium is used in both spinal and general orthopedic implants and has no biocompatibility issues. No additional testing is required to support this submission. The predicate LP Cage was subjected to the following:

  • Static Compression ASTM F2077-03 .
  • Static Torsion ASTM F2077-03 .
  • Static Compressive Shear F207703 .
  • Dynamic Compression F2077-03 .
  • Subsidence ASTM F2267-04 .

10. Conclusions from Non-Clinical Tests

A risk analysis was completed. Based on the risk analysis and additional supporting documentation provided in this premarket notification submission, Medyssey believes that the subject Medyssey LT Cage is substantially equivalent to the Medyssey LP Cage.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The text is black and the background is white.

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN 1 5 2012

Medyssy Co., Ltd. % Mr. Michael Kvitnitsky Chief Operating Officer 8001 North Lincoln Avenue Skokie, Illinois 60077

Re: K121246

Trade/Device Name: Medyssey LT Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: May 29, 2012 Received: May 30, 2012

Dear Mr. Kvitnitsky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 -- Mr. Michael Kvitnitsky

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson

Mark N. Melkerson // U Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Form

K121246 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Medyssey LT Interbody Fusion Cage System

Indications for Use:

The LT Cage System is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation: These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone.

Prescription Use × OR

Over-The-Counter Use

(Per 21 CFR 801.109)

(Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

K121246 510(k) Number_

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.