IND DRUG HOME MULTI-PANEL TEST (2-5) IND MORPHINE HOME TEST

K121231 · Ind Ndiagnostic, Inc. · NGT · Jul 30, 2012 · General Hospital

Device Facts

Record IDK121231
Device NameIND DRUG HOME MULTI-PANEL TEST (2-5) IND MORPHINE HOME TEST
ApplicantInd Ndiagnostic, Inc.
Product CodeNGT · General Hospital
Decision DateJul 30, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5200
Device ClassClass 2

Intended Use

The IND Drug Home Test is a rapid chromatographic immunoassay multi-panel or single drug test for the qualitative detection of one or more of the following drugs: Amphetamine, Methamphetamine, Cocaine, Marijuana, and Morphine in human urine. It is intended for overthe-counter use. The test is intended for over-the counter (OTC) use as the first step in a two step process to provide consumers with information concerning the presence of the above stated drugs or metabolites in a urine sample. The test provides only preliminary test results. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. GC/MS (Gas Chromatography / Mass Spectrometry) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when the preliminary result is positive.

Device Story

Rapid chromatographic immunoassay for qualitative detection of drugs of abuse in human urine; includes multi-panel or single-drug cassette and strip formats. Input: human urine sample. Principle: immunoassay-based detection of specific drug metabolites at defined cut-off concentrations (AMP/MET 1000 ng/mL; COC/MOR 300 ng/mL; THC 50 ng/mL). Output: visual qualitative result (presence/absence of drug). Used in home setting by consumers. Provides preliminary results only; requires follow-up with GC/MS for confirmation. Benefits patient by providing initial information regarding potential drug presence.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Rapid chromatographic immunoassay; lateral flow cassette or strip format. Qualitative detection based on competitive binding. No electronic components, software, or external energy sources.

Indications for Use

Indicated for OTC qualitative detection of Amphetamine, Methamphetamine, Cocaine, Marijuana, and Morphine in human urine. Intended as a preliminary screening step; requires GC/MS confirmatory testing for positive results.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration Image /page/0/Picture/2 description: The image contains a black horizontal line. The line is straight and appears to be solid. The background is white. The line is centered vertically in the image. 10903 New Hampshire Avenue Silver Spring, MD 20993 IND Diagnostic, Inc c/o Kai Lou C/o Rai Exe Regulatory Affairs Department 1629 Fosters Way 1629 Fosters Way Delta, British Columbia, Canada V3M 6S7 JUL 30 2012 Re: k 121231 Trade Name: IND Drug Home Multi-Panel Test (2-5) – Cassette and Strip Formats IND Drug Home Multi-Pance Test – Cassette and Strip Formats IND Amphetamine Home Test – Cassette and Strip Formats IND Amphetamine Home Test – Cassette and Strip Formats IND Methamphetamine Home Test – Cassette and Strip Formats IND Methamplietannine Tions sees IND Cocaine Home Test – Cassette and Strip Formats IND Cocaine Home Test - Cassette and Strip Formats IND Morphine Home Test - Cassette and Strip Formats IND Morphine Home Test – Cassette and Strip Formats IND Marijuana Home Test – Cassette and Strip Formats 100 Marijuana 10010 – 100 Regulation Number: 21 CFR §862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Regulatory Class: Class Class Crass Crass NFW, NGI Dated: July 5, 2012 Received: July 10, 2012 Dear Kai Lou, Dear III. We have reviewed your Section 510(k) premarket notification of intent to market the device is a We have reviewed your Section 210(K) premaince is substantially equivalent (for the referenced above and have determined the device Is succedicate devices marketed in referenced above and have decembre) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate commerce prior to May 28, 1976, the macment with the provisions of Amendments, or to devices that have been reclassified in accordance with the provisions of each mercedes that have been reclassified in accondice with accordance while in the premaixel the Federal Food, Drug, and Cosmetic Act (Act ) that do not require approval of Andreal Food, Drug, and Cosmetic Act (Act) that of not require, subject to the general approval application (PMA). You may, therefore, market the device, subject to the gen approval application (PMA). You may, therefore, manse the Act include ontrols provisions of the Act. The general collible provisions of the more requirements for annual registration, listing of devices, good manufacturing practice, requirement requirements for annual registration, noting on the one of adulteration. Iabeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III ns affectives and the the same additional controls. Existing major regulations If your device is classified (see above) into either class II (Special Conneller) (PMA), it may be subject to such additional controls. Existing major regulations (CFR), Par (PMA), it may be subject to such additional controls. Exismly institutions of the Solot of Sept. your device can be found lick forther angouncements concerning your device i (111), which and of the 21, Code of Federal Regulation (Crity), Particle (Cricy, Particle) In addition, FDA may publish further announcements concerning your device in the Ee Register. Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA's issuance of a substantial equires with other requirements mean that FDA has made a determination that your device with other requires. Press of the Act of Acternination that your device compins with the research agencies. of the Act or any Federal stations and regulations administered by registration modified by any Federal statutes and regulations administered by only in the Act of eggistration You must comply Federal statises and regulations, but not limited to: regist You must comply with all the Act's requirements, then are and any medical device and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device r seats of the Part 807), labeling (21 CFR Parts 801 mill oos); and good and listing (21 CFR Part 807); labeling (21 CFR 803); and good reporting (reporting of medical device and in B (reporting of medical device-related adverse events) (21 of 100 )) degreens manufacturing practice requirements as set forth in the quality systems (QS) regulation ( CFR Part 820). {1}------------------------------------------------ ## Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you abon't office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 01160 of Dar reliestions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance... You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm Sincerely yours, S. H. Liu, Ph.D. Countney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health ## Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: IND Drug Home Multi-Panel Test (2-5) - Cassette and Strip Formats IND Amphetamine Home Test - Cassette and Strip Formats IND Methamphetamine Home Test - Cassette and Strip Formats IND Cocaine Home Test - Cassette and Strip Formats IND Morphine Home Test - Cassette and Strip Formats IND Marijuana Home Test - Cassette and Strip Formats ## Indications for Use: The IND Drug Home Test is a rapid chromatographic immunoassay multi-panel or single drug test for the qualitative detection of one or more of the following drugs: Amphetamine, Methamphetamine, Cocaine, Marijuana, and Morphine in human urine. It is intended for overthe-counter use. Drug Name (Code) Amphetamine (AMP) Methamphetamine (MET) Cocaine (COC) Marijuana (THC) Morphine (MOR) Cut-off 1000 ng/mL 1000 ng/mL 300 ng/mL 50 ng/mL 300 ng/mL The test is intended for over-the counter (OTC) use as the first step in a two step process to provide consumers with information concerning the presence of the above stated drugs or metabolites in a urine sample. The test provides only preliminary test results. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. GC/MS (Gas Chromatography / Mass Spectrometry) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when the preliminary result is positive. Information regarding confirmatory testing, the second step in the process, is provided in the package labeling. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR - Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (01VD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) k121231