K Number
K121154

Validate with FDA (Live)

Date Cleared
2012-08-20

(126 days)

Product Code
Regulation Number
868.5630
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ultrasonic mesh nebulizer model NE403 is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize medications for inhalation by the patient. The device may be used by adult patients at home, hospital & sub-acute care settings.

Device Description

Nebulizer NE403 is a mesh screen ultrasonic nebulizer which was used ultrasonic Vibrating Mesh Technology ( VMT). The device creates aerosols of liquid medication by ejection of droplets from a mesh vibrated at ultrasonic frequencies. As with other mesh screen ultrasonic nebulizers, the mesh is caused to vibrate by fluidic contact with a ultrasonic piezoelectric member that is energized by application of a high frequency alternating voltage. The VMT nebulizer is portable and quiet. This nebulizer is powered by four AA alkaline batteries or a cable connecting to an AC adapter. Nebulizer NE403 consists of several function units: liquid medication containment, liquid nebulization and ejection, nebulized medicine (brume) channel (breathing parts), power supply, operating and display unit. liquid medication is poured into medicinde container, transformed into something like brume, which is inhaled by patient via breathing parts, thus, the intended use of device is achieved.

AI/ML Overview

The provided text describes a 510(k) submission for a medical device, the Nebulizer NE403. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing extensive clinical study data with specific acceptance criteria in the format you requested.

Therefore, many of the requested details about acceptance criteria derived from clinical performance studies, such as effect size of AI assistance or ground truth establishment for a training set, are not applicable to this document. The submission explicitly states that no clinical tests were conducted for substantial equivalence.

However, I can extract information related to non-clinical testing and the comparison to the predicate device.

Here's a breakdown of the available information based on your request:

1. A table of acceptance criteria and the reported device performance:

Since no clinical acceptance criteria are provided, and the submission focuses on non-clinical testing and substantial equivalence, the "acceptance criteria" can be inferred as meeting the requirements of the listed voluntary design standards and demonstrating comparable performance to the predicate device in specific non-clinical tests.

Acceptance Criteria (Inferred from Non-Clinical Tests & Standards)Reported Device Performance (Non-Clinical)
Conformance to IEC 60601-1 (General safety)Demonstrated conformance
Conformance to IEC 60601-1-2 (Safety - Collateral Standard)Demonstrated conformance
Conformance to EN 13544-1:2007 (Respiratory therapy equipment)Demonstrated conformance
Aerosol Performance Comparison (vs. predicate device)Comparable to predicate device (NE-U22)
Intra-Sample Dose and Particle Size VariabilityTesting conducted and demonstrated
Inter-Sample Dose and Particle Size VariabilityTesting conducted and demonstrated
Biocompatibility (Masks, Mouthpiece, gas path components)Certification conducted according to ISO 10993 series
Same intended use as predicate deviceYes, determined to have the same intended use
Same fundamental technology as predicate deviceYes, determined to have the same fundamental technology
Same energy type as predicate deviceYes
Same nebulizing method as predicate deviceYes
Comparable particle size to predicate deviceYes
Nebulizing rate within range of predicate deviceYes

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not applicable in the context of a clinical test set. Non-clinical tests were performed on the device itself and its components. The document doesn't specify the "sample size" of devices or components tested for each non-clinical criterion.
  • Data Provenance: The non-clinical testing was conducted by the manufacturer, Foshan Gaunying Electronics Co.,Ltd., or outsourced as part of their design validation. The context is a regulatory submission for a device manufactured in China. This is retrospective data from internal testing and compliance work.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. No clinical test set requiring expert-established ground truth was part of this 510(k) submission. Non-clinical tests are validated against engineering specifications, standards, and comparison to a predicate device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. There was no clinical test set or subjective assessment requiring an adjudication method.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This device is a nebulizer, a physical medical device for drug delivery, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not Applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used:

  • For Non-Clinical Tests: The "ground truth" for the non-clinical tests was based on engineering specifications, recognized international and national standards (IEC 60601-1, IEC 60601-1-2, EN 13544-1), and the performance characteristics of the legally marketed predicate device (Omron Micro Air Vibrating Mesh Nebulizer, K062263). For biocompatibility, it was adherence to ISO 10993 standards.

8. The sample size for the training set:

  • Not Applicable. There was no "training set" in the context of machine learning or an AI algorithm.

9. How the ground truth for the training set was established:

  • Not Applicable. As above, no training set was involved.

Summary from the document:

The Nebulizer NE403 demonstrates substantial equivalence through non-clinical testing by:

  • Meeting design requirements and conforming to voluntary design standards.
  • Showing comparable aerosol performance, dose variability, and particle size variability to the predicate device (Omron Micro Air Vibrating Mesh Nebulizer, K062263).
  • Demonstrating biocompatibility according to ISO standards for patient-contacting parts.
  • Having the same intended use and fundamental technological characteristics as the predicate device.

The study presented here is a 510(k) premarket notification which relies on demonstrating substantial equivalence to a predicate device, not on extensive clinical trials with detailed acceptance criteria for diagnostic performance as would be seen with AI or imaging devices.

{0}------------------------------------------------

ATTACHMENT 5

Image /page/0/Picture/1 description: The image contains a company logo and name. The logo is on the left side of the image and consists of three curved lines. The company name is "QUATEK" and is written in a bold, sans-serif font. Below the company name is the Chinese translation of the company name, followed by the English translation: "FOSHAN GAUNYING ELECTRONICS CO.,LTD".

Section 05-510 (k) Summary Version: 1.2

AUG

510(k) SUMMARY

  1. Sponsor:

Date of Submission Preparaton: 02/20/2012

510(k) S ubmitter's Na me: Foshan Gaunying Electronics Co.,Ltd.

Address: 4F, #4 IndustryCountry, ChengNanPark of Foshan Hi-tech Industrial Development Zone

  1. 510(k) Correspondent contact:

510(k) Correspondent Shenzhen ZYTC consulting Co., Ltd.

Address:05C.fuhai kangle building Nanshan district Shenzhen, CHINA 518000

Telephone: +86-755-8606 9197

Fax: +86-755-8606 9197

E-mail: cefda13485 @163.com

Contact person: Field.Fu

3.Proposed Device:

Trade Name: Nebulizer NE403 Classification Name: Nebulizer (Direct Patient Interface) Regulation Number: 868.5630 Product Code: CAF Device Class: II

  1. Predicate Device:

Predicate Device: Micro Air Vibrating Mesh Nebulizer 510(k) Number: K062263 Manufacturer: Omron Healthcare Inc.

  1. Description of Proposed Device:

Nebulizer NE403 is a mesh screen ultrasonic nebulizer which was used ultrasonic Vibrating Mesh Technology ( VMT). The device creates aerosols of liquid medication by ejection of droplets from a mesh vibrated at ultrasonic frequencies. As with other mesh screen ultrasonic nebulizers, the mesh is caused to vibrate by fluidic contact with a ultrasonic piezoelectric member that is energized by application of a high frequency alternating

Section 05 Page 1 of 3

{1}------------------------------------------------

魔雷子有限公司

Section 05-510 (k) Summary Version: 1.2

UNVING ELECTRONICS CO.,LTD

voltage. The VMT nebulizer is portable and quiet. This nebulizer is powered by four AA alkaline batteries or a cable connecting to an AC adapter.

Nebulizer NE403 consists of several function units: liquid medication containment, liquid nebulization and ejection, nebulized medicine (brume) channel (breathing parts), power supply, operating and display unit.

liquid medication is poured into medicinde container, transformed into something like brume, which is inhaled by patient via breathing parts, thus, the intended use of device is achieved.

6. Statement of indications for Use:

The ultrasonic mesh nebulizer model NE403 is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize medications for inhalation by the patient.

The device may be used by adult patients at home, hospital & sub-acute care settings.

7. Biocompatibility Certification:

Biocompatibility Testing conducted on the Masks, Mouthpiece and the components which contact the gas path of the patients according to ISO10993-1, ISO10993-3, ISO10993-5, ISO10993-6, ISO10993-10.

8. Comparison to predicate device

Both the nebulizer and the Predicate device have the same intended use and fundamental technology. They have the same energy type, nebulizing method, particle size, etc. although differences between them in the respect of the target population, nebulizing rate, those items of nebulizer NE403 fall into the range of predicate device. A side-by-side comparison of the NE403 and the cited predicate devices is included in the 510(k) submission (Section 09). The NE403 is substantially equivalent to the technological features as the predicate devices.

9. Discussion of Non -Clinical Tests Perf ormed for Determi nation of Substantial Equivalence are as follows:

The NE403 did not conduct, nor rely upon, clinical tests to determine substantial equivalence. Non-clinical testing was performed in order to validate the design according to the company's specified design requirements, and to assure conformance with the following voluntary design standards:

IEC 60601-1 "Medical electrical equipment - Part 1: General requirements for safety". IEC 60601-1-2 "Medical electrical equipment - Part 1-2: General requirements for safety -Collateral Standard"

EN 13544-1:2007 "Respiratory therapy equipment - Part 1: Nebulizing systems and their components".

In addition, according to FDA guidance (reviewer guidance for nebulizers,metered dose inhalers, spacers andactuators), the Aerosol Performance Comparison Testing of the proposed device(NE403) and predicate device(NE-U22) , as well as the Intra-Sample and Inter-Sample Dose and Particle Size Variability Testing of NE403 were conducted.

{2}------------------------------------------------

AN GAUNYING ELECTRONICS CO.,LTD

Section 05-510 (k) Summary Version: 1.2

10. Conclusions:

The NE403 has the same intended use and technological characteristics as the predicate device. Moreover, bench testing and safety report documentation supplied in this submission demonstrates that the difference in the submitted models could maintain the same safety and effectiveness as that of predicate device. In the other words, those engineering difference do not affect the intended use or alter the fundamental scientific technology of the device. Thus, the NE403 is substantially equivalent to the predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three tail feathers, representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Foshan Gaunying Electronics Company, Limited C/O Mr. Guenter Ginsberg Media Trade Corporation 11820 Red Hibiscus Drive Bonita Springs, Florida 34135

AUG 20 2012

Re: K121154

Trade/Device Name: Nebulizer NE403 Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: July 7, 2012 Received: July 10. 2012

Dear Mr. Ginsberg:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Ginsberg

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

th fa.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

ATTACHMENT 4

Section 04- Indications for use Version: 1.1

Indications for Use

510(k) Number (

):

Device Name:

Nebulizer NE403

Indications for Use:

The ultrasonic mesh nebulizer model NE403 is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize medications for inhalation by the patient.

The device may be used by adult patients at home, hospital & sub-acute care setting.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schult

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

Section 04 -Page 1 of 1 ·

§ 868.5630 Nebulizer.

(a)
Identification. A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.(b)
Classification. Class II (performance standards).