(335 days)
The Communify PACS Viewer 1.5 is a Picture Archiving and Communications System (PACS) viewer designed to be used to view Digital Imaging and Communications in Media (DICOM), DICOM-RT and non-DICOM information and data. The Communify PACS Viewer 1.5 is a software application that runs on standard "off-the-shell" personal computers, business computers, and servers running standard operating systems. Communify PACS Viewer is an image and display software that accepts DICOM data from any OEM modality which supports DICOM standard imaging data; the system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulations, create graphical representations of anatomical areas, and perform quantitative measurements.
Communify PACS Viewer 1.5 is designed to support the use and viewing of DICOM-RT (radiotherapy extensions), such as Structure Set, Contour, Dose, Dose Value Histogram and Plan information.
In Communify PACS Viewer 1.5, mammographic images should only be viewed on mammography monitors that are cleared or approved by FDA. Users should not apply lossy compression to mammographic images and only DICOM "for presentation" mammography images may be displayed. Communify PACS Viewer 1.5 does not manipulate or edit any mammography image acquisition system manufacturer's proprietary image processing methods.
Please note that Communify PACS Viewer 1.5 does not run on mobile smartphones and tablets.
Communify PACS Viewer (version 1.5) is a Picture Archiving and Communications System (PACS) viewer application which allows users to view, manipulate, annocate, transmit to other facilities, print, and animate all manners of DICOM and DICOM-RT images and modalities include, but are not limited to, CR, CT, DX, MR, NM, PT, RF, US, and XA.
Communify PACS Viewer 1.5 contains common image manipulation functions (such as zoom, pan, triangulation, and window/level) and common image labeling tools (including measurements tools, drawing tools, and annotation overlays). The annotation overlay displays all the important metadata (as configured by the user) for each displayed series study. Although annotation fields depend on the modality and the patient study, the Annotation Overlay Template Wizard provides a full list of annotation fields the user can assign into the image display. For RT studies, Communify PACS Viewer 1.5 allows the visualization of the RT Structure Set, Dose, Dose Value Histogram, and Plan data.
There are shortcut keys, toolbars, and right-click menus for easy access to tools and features. The Hanging Protocol editor of the Communify PACS Viewer configures the presentation layout of images on the screen when a study is loaded. This allows commonly used display formats and presets to be saved and easily accessed to allow for faster case study reviews.
The provided text does not contain detailed acceptance criteria or a specific study proving the device meets acceptance criteria in the format typically used for performance studies of medical AI devices. Instead, it describes a PACS viewer (Communify PACS Viewer 1.5) and its substantial equivalence to predicate devices (Communify PACS Viewer 1.0 and MIMviewer 4.1).
The "Performance Data" section mentions "Software verification and validation was performed following FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." and states that "Tests were conducted successfully to address issues in the hazard analysis, confirm the functionality in the verification and validation section and to provide comparisons of samples with 510(k)-cleared DICOM-RT readers." This indicates a software validation process, but not a clinical study with specific performance metrics such as sensitivity, specificity, or reader agreement that would typically be associated with an AI device's acceptance criteria.
Based on the provided information, I can extrapolate the following:
1. A table of acceptance criteria and the reported device performance
Since this is a PACS viewer and not an AI-powered diagnostic device, the "acceptance criteria" appear to be related to its functionality, display capabilities, and adherence to DICOM standards, rather than diagnostic accuracy metrics. The document implies that the device successfully met these functional and comparative criteria.
| Acceptance Criteria (Implied) | Reported Device Performance (Implied) |
|---|---|
| Functional Equivalence to Predicate Devices | The device performs all capabilities of Communify PACS Viewer 1.0 and supports DICOM-RT data like MIMviewer 4.1. |
| DICOM Compliance and Modality Support | Processes and displays medical images from DICOM compliant modalities (CR, CT, DX, MR, NM, PT, RF, US, XA, and others). |
| Image Manipulation and Annotation Features | Supports common image manipulation (zoom, pan, triangulation, window/level) and labeling tools (measurements, drawing, annotation overlays). |
| DICOM-RT Data Visualization | Allows visualization of RT Structure Set, Dose, Dose Value Histogram, and Plan data. |
| Maintenance of Source Data Integrity | Does not alter the source data and original image. |
| Software Verification and Validation | Tests were conducted successfully to address issues in hazard analysis and confirm functionality per FDA guidance. |
| Comparison to 510(k)-cleared DICOM-RT Readers | Comparisons of samples were positive, indicating comparable performance. |
| Mammographic Image Handling | Viewed on FDA-cleared/approved monitors; no lossy compression; only DICOM "for presentation" mammography images displayed. |
| Operating Environment | Runs on standard "off-the-shelf" personal computers, business computers, and servers; does not run on mobile smartphones/tablets. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "comparisons of samples with 510(k)-cleared DICOM-RT readers." However, it does not specify the sample size, type of data (e.g., patient cases), country of origin, or whether the tests were retrospective or prospective. It refers to general software verification and validation, which typically involves testing with a representative set of data but not necessarily a clinical "test set" in the context of diagnostic performance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided. The study described is a software verification and validation study, not a clinical study involving human expert review for establishing ground truth on diagnostic performance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as this was not a clinical study involving expert adjudication of cases.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was performed. The device is a PACS viewer, not an AI diagnostic tool designed to assist human readers in making diagnoses or interpretations that would be measured by an MRMC study. The comparison was for functional equivalency, not diagnostic improvement.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is a viewer, and its performance is inherently human-in-the-loop (the user viewing the images). It does not have a "standalone algorithm" performance in the traditional sense of an AI diagnostic device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Given that the study was software verification and validation and comparison to predicate devices for functional equivalence, the "ground truth" would likely be adherence to DICOM standards, correct display of image data, accurate execution of manipulation functions, and consistency with data rendered by cleared predicate devices. It would not typically involve expert consensus, pathology, or outcomes data for diagnostic accuracy.
8. The sample size for the training set
Not applicable. The device is a software application (PACS viewer), not an AI model that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device.
{0}------------------------------------------------
- 510(k) Summary (21 CFR 807.92)
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
12085 "The assigned 510(k) number is:
| Submitter's name: | Christopher Relyea |
|---|---|
| Submitter's address: | 160 West 71st Street, 18th Floor, New York, NY 10023, USA |
| Phone number: | 646-526-7111 |
| Fax number: | 203-621-3198 |
| Name of contact person: | Christopher Relyea |
| Date the summary was prepared: | 10/18/2012 |
| Device Name/Trade Name: | Communify PACS Viewer 1.5 |
| Common Name: | Medical Image Workstation |
| Classification Name: | 21 CFR 892.2050 |
| Product Code | LLZ |
Substantially equivalent to the devices are the following:
| Manufacturer | Device | 510(k) |
|---|---|---|
| Flying Squirrel Storage | Communify PACS Viewer 1.0 | K110797 |
| MIM Software Inc. (formerly MIMvista Corp.) | MIMviewer 4.1 | K071964 |
{1}------------------------------------------------
Comparison to Predicate Devices
Communify PACS Viewer 1.5 is an extension of the Communify PACS Viewer 1.0 that includes all of the 1.0 version's capabilities, such as the ability to process and display medical images from DICOM compliant modalities such as CR, CT, DX, MR, NM, PT, RF, US, XA, and others. MIM is also a PACS viewer. Like MIM, Communify PACS Viewer 1.5 supports the ability to read and visualize DICOM-RT data files, such as Structure Set, Dose Value Histogram and Plan (Beam) data.
Conclusion
Similar to the predicate devices, the Communify PACS Viewer 1.5 system allows the display and measurement of medical images, but does not alter the source data and original image. All devices require that use is restricted to suitably licensed healthcare professionals. It is our conclusion that Communify PACS Viewer 1.5 is significantly comparable to the Communify PACS Viewer 1.0 and MIM devices.
Device Description
Communify PACS Viewer (version 1.5) is a Picture Archiving and Communications System (PACS) viewer application which allows users to view, manipulate, annocate, transmit to other facilities, print, and animate all manners of DICOM and DICOM-RT images and modalities include, but are not limited to, CR, CT, DX, MR, NM, PT, RF, US, and XA.
Communify PACS Viewer 1.5 contains common image manipulation functions (such as zoom, pan, triangulation, and window/level) and common image labeling tools (including measurements tools, drawing tools, and annotation overlays). The annotation overlay displays all the important metadata (as configured by the user) for each displayed series study. Although annotation fields depend on the modality and the patient study, the Annotation Overlay Template Wizard provides a full list of annotation fields the user can assign into the image display. For RT studies, Communify PACS Viewer 1.5 allows the visualization of the RT Structure Set, Dose, Dose Value Histogram, and Plan data.
There are shortcut keys, toolbars, and right-click menus for easy access to tools and features. The Hanging Protocol editor of the Communify PACS Viewer configures the presentation layout of images on the screen when a study is loaded. This allows commonly used display formats and presets to be saved and easily accessed to allow for faster case study reviews.
Performance Data
Software verification and validation was performed following FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Tests were conducted successfully to address issues in the hazard analysis, confirm the functionality in the verification and validation section and to provide comparisons of samples with 510(k)-cleared DICOM-RT readers.
Intended Use Statement
Communify PACS Viewer is a Picture Archiving and Communications System (PACS) viewer designed to be
{2}------------------------------------------------
used to view Digital Imaging and Communications in Media (DICOM-RT and non-DICOM information and data. Communify PACS Viewer is a software application that runs on standard "off-theshelf" personal computers, business computers, and servers running standard operating systems. Communify PACS Viewer is an image and display software that accepts DICOM data from any OEM modality which supports DICOM standard imaging data; the system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulations, create graphical representations of anatomical areas, and perform quantitative measurements.
Communify PACS Viewer is designed to support the use and viewing of DICOM-RT (radiotherapy extensions), such as Structure Set, Contour, Dose, Dose Value Histogram and Plan information.
In Communify PACS Viewer, mammographic images are viewed in mammography monitors that are cleared or approved by FDA. Users should not apply lossy compression to mammographic images and only DICOM "for presentation" mammography images may be displayed. Communify PACS Viewer does not manipulate or edit any mammography image acquisition system manufacturer's proprietary image processing methods.
Please note that Communify PACS Viewer does not run on mobile smartphones and tablets.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
February 19, 2013
Communify Health c/o Chris Relyea 160W 71ST 18 Floor New York, NY 10023
Re: K120851
Trade/Device Name: Communify PACS Viewer (version 1.5) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 30, 2012 Received: February 12, 2013
Dear Mr. Relyea:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Relyea
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Image /page/4/Figure/5 description: The image shows a logo with the letters FDA in a stylized, geometric design. The letters are interconnected and have a three-dimensional appearance due to the use of parallel lines. The logo is black and white and appears to be a graphic representation of the Food and Drug Administration.
for
Janine M. Morris Director
Division of Radiological Health
Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
· Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K120851 .
Device Name:
Communify PACS Viewer (version 1.5)
Indications for Use:
The Communify PACS Viewer 1.5 is a Picture Archiving and Communications System (PACS) viewer designed to be used to view Digital Imaging and Communications in Media (DICOM), DICOM-RT and non-DICOM information and data. The Communify PACS Viewer 1.5 is a software application that runs on standard "off-the-shell" personal computers, business computers, and servers running standard operating systems. Communify PACS Viewer is an image and display software that accepts DICOM data from any OEM modality which supports DICOM standard imaging data; the system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulations, create graphical representations of anatomical areas, and perform quantitative measurements.
Communify PACS Viewer 1.5 is designed to support the use and viewing of DICOM-RT (radiotherapy extensions), such as Structure Set, Contour, Dose, Dose Value Histogram and Plan information.
In Communify PACS Viewer 1.5, mammographic images should only be viewed on mammography monitors that are cleared or approved by FDA. Users should not apply lossy compression to mammographic images and only DICOM "for presentation" mammography images may be displayed. Communify PACS Viewer 1.5 does not manipulate or edit any mammography image acquisition system manufacturer's proprietary image processing methods.
Please note that Communify PACS Viewer 1.5 does not run on mobile smartphones and tablets.
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
FDA
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health 510(k) K120851