(28 days)
The Navigator™ Delivery System (Navigator DS) is intended for use in the delivery of medication and/or fluids in a controlled manner. The Navigator DS is indicated for use in the intermittent delivery of medications and other fluids in intra-articular applications.
The Navigator Delivery System (Navigator DS) consists of a fluid delivery module, a daily disposable cassette, a per-patient disposable handpiece and tubing set, and wired foot pedal. Image integration with qualified ultrasound units occurs by Ethernet cable connection, if desired, allowing simultaneous display of Navigator treatment information on the ultrasound screen and printing of ultrasound images on the patient treatment record.
This is a 510(k) summary for a minor modification to an existing device, the Navigator™ Delivery System (Navigator DS). The modification involves an "alarm modification" and does not alter the device design, intended use, or indication for use. Therefore, the information provided is limited, and many of the requested criteria for studies would not be applicable or available in this type of submission.
Here's an attempt to answer your questions based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly state numerical acceptance criteria in the format typically seen for diagnostic performance studies (e.g., sensitivity, specificity thresholds). Instead, the acceptance criterion for this modification was that the safety and performance of the device are not affected by the alarm modification.
| Acceptance Criteria | Reported Device Performance | Study Type |
|---|---|---|
| Safety and performance of the Navigator DS are not affected by the alarm modification. | All data demonstrated that the safety and performance of the Navigator DS is not affected by the alarm modification. | Performance Testing, Safety Case and Hazard Analysis |
| Acceptable risk profile. | A Safety Case and Hazard Analysis demonstrated an acceptable risk profile based on design-based risk mitigation and satisfactory performance testing. | Safety Case and Hazard Analysis |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not specified. The document states "Testing of the Navigator DS was carried out, including performance testing." It does not provide details on the number of units or test scenarios.
- Data Provenance: Not specified. Given the nature of a medical device modification, it is almost certainly prospective testing conducted by the manufacturer, rather than retrospective clinical data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- This information is not provided. Given that this is a performance test for a mechanical/software alarm modification, "ground truth" would likely be established by engineering specifications and direct observation/measurement of system behavior by qualified engineers/testers, rather than clinical experts.
4. Adjudication Method for the Test Set
- This information is not provided. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies for diagnostic accuracy, which is not the focus of this submission. The "adjudication" for performance testing of an alarm would be based on whether the alarm triggered correctly under expected conditions and did not trigger incorrectly.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic aids, assessing how AI impacts human interpretation. The Navigator DS is an infusion pump, and this submission concerns a minor alarm modification, not AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- This question is not applicable to this device. The Navigator DS is a physical medical device, an infusion pump, not an AI algorithm. "Standalone performance" would refer to the pump's ability to deliver fluids as intended, which was assured to be unaffected by the alarm modification.
7. The Type of Ground Truth Used
- The term "ground truth" as typically used in AI/diagnostic studies (e.g., pathology, outcomes data) is not directly applicable here. For performance testing of an alarm modification on an infusion pump, the "ground truth" would be the engineering specifications and design requirements for alarm behavior in various conditions (e.g., occlusion, low fluid), and the observed behavior of the device during testing.
8. The Sample Size for the Training Set
- Not applicable/Not provided. This is not an AI/machine learning device that requires a training set. The device's functionality is based on its physical and software design, not learned from data.
9. How the Ground Truth for the Training Set was Established
- Not applicable. As stated above, this device does not use a "training set" in the context of AI/machine learning.
{0}------------------------------------------------
.
APR 1 6 2012
:
510(k) Summary
SUBMITTER'S INFORMATION
| Owner: | Carticept Medical, Inc. |
|---|---|
| Address: | 6120 Windward Parkway, Suite 220, Alpharetta, GA 30005 |
| Phone: | 770-754-3800 |
| Fax Numbers: | 770-754-3808 |
| Contact Person: | Tanya Eberle, Director, Regulatory Affairs |
| Date Summary Prepared: | March 16, 2012 |
DEVICE INFORMATION
| Name of Device: | Navigator™ Delivery System (Navigator DS) |
|---|---|
| Common/Usual Name: | Infusion Pump, External |
| Classification Name: | Infusion Pump, Class II, 21 CFR 880.5725 (Product CodeFRN) |
| Predicate Device(s): | Navigator™ Delivery System (K112067) |
| Device Description: | The Navigator Delivery System (Navigator DS) consists of afluid delivery module, a daily disposable cassette, a per-patient disposable handpiece and tubing set, and wired footpedal. Image integration with qualified ultrasound unitsoccurs by Ethernet cable connection, if desired, allowingsimultaneous display of Navigator treatment information onthe ultrasound screen and printing of ultrasound images onthe patient treatment record. |
| Indication for Use: | The Navigator™ Delivery System (Navigator DS) is intendedfor use in the delivery of medication and/or fluids in acontrolled manner. The Navigator DS is indicated for use inthe intermittent delivery of medications and other fluids inintra-articular applications. |
| TechnologicalCharacteristics: | All technological and design aspects of the Navigator DSdevice are preserved. |
| Comparison to PredicateDevice: | The Navigator DS intended use and performancecharacteristics are not altered by this modification. |
{1}------------------------------------------------
Performance Data:
Testing of the Navigator DS was carried out, including performance testing. All data demonstrated that the safety and performance of the Navigator DS is not affected by the alarm modification.
A Clinical Evaluation was determined not to be required as the device design, intended use and indication for use are preserved.
A Safety Case and Hazard Analysis demonstrated an acceptable risk profile based on design-based risk mitigation and satisfactory performance testing.
Rationale for Substantial Equivalence:
This minor modification falls within the FDA regulations for 510(k) review. The indication for use, intended use, technological characteristics, principles of operation, and performance have not been altered. The removal of this feature does not raise any new questions of safety or effectiveness and the system has been demonstrated to provide the same level of performance as the predicate device. The Navigator DS with aligned alarm settings is substantially equivalent to the predicate device (Navigator DS K112067).
Conclusion:
The Navigator DS, as modified by this 510(k), does not raise any new issues regarding safety or effectiveness, and therefore is suitable for commercial sale.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Tanya Eberle Director. Regulatory Affairs Carticept Medical, Incorporated 6120 Windward Parkway, Suite 220 Alpharetta, Georgia 30005
APR 1 6 2012
Re: K120830
Trade/Device Name: Navigator Delivery System (Navigator DS) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: II Product Code: FRN Dated: March 16, 2012 Received: March 19, 2012
Dear Ms. Eberle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Ms. Eberle
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) No. (if known): K120830
Device Name:
Navigator™ Delivery System (Navigator DS)
Indications for Use:
The Navigator™ Delivery System (Navigator DS) is intended for use in the delivery of medication and/or fluids in a controlled manner. The Navigator DS is indicated for use in the intermittent delivery of medications and other fluids in intra-articular applications.
Prescription Use: × (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page of
Lijl Uli for RZC Apr 13, 2012
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: KJ20830
§ 880.5725 Infusion pump.
(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).