K Number
K120702

Validate with FDA (Live)

Date Cleared
2012-06-20

(104 days)

Product Code
Regulation Number
874.4710
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

E.G. Scan™ II Esophagoscope System is intended for use in endoscopic access and examination of the larynx, esophagus and gastroesophageal junction.

Device Description

E.G. Scan™ II Esophagoscope System and its accessories are used for diagnosis of patients. E.G. Scan™ II Probe takes pictures of the esophagus of human and sends image data to E.G. Scan™ II Controller. E.G. Scan™ II Controller processes and converts image data and upload. E.G. View™ image displaying software displays the image for diagnosis. The E.G Scan™ II Esophagoscope Probe is disposable. E.G. Scan™ II Esophagoscope system is a transnasal esophagoscope designed to capture images of the esophagus. Captured images are viewed via the E.G. View™ Software for diagnosis of diseases related to the esophagus.

AI/ML Overview

This 510(k) submission does not contain information about a study that proves the device meets specific acceptance criteria in the way described in the prompt. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and results.

Here's a breakdown of the requested information based on the provided text, highlighting what is not available:

1. Table of acceptance criteria and the reported device performance:

Acceptance CriteriaReported Device Performance
Not explicitly stated in terms of quantitative performance metrics. The submission states the device "has the same device characteristics as the predicate device" and "conforms to IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2" (safety standards)."Substantially equivalent" to the predicate device in areas of design, indication for use, technological characteristics, function, application, safety, and effectiveness. Conformity to: Biocompatibility (ISO 10993-1)Medical electrical equipment safety (IEC 60601-1)Specific safety for endoscopic equipment (IEC 60601-2-18)Electromagnetic compatibility (IEC 60601-1-2)

2. Sample size used for the test set and the data provenance:

  • Not applicable. No specific performance study with a test set is described in relation to clinical or diagnostic performance. The testing mentioned (biocompatibility, safety, EMC) would involve material samples or electrical measurements, not a patient test set for diagnostic accuracy.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. No such clinical or diagnostic performance study is described.

4. Adjudication method for the test set:

  • Not applicable. No such clinical or diagnostic performance study is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. The device is an esophagoscope system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a medical imaging device, not an algorithm being evaluated in isolation for diagnostic performance. Its function is to capture images for human interpretation.

7. The type of ground truth used:

  • Not applicable. Since no diagnostic performance study is presented, there is no mention of ground truth established through expert consensus, pathology, or outcomes data. The "ground truth" for the device's claims are its stated technical characteristics and adherence to safety and performance standards, proven through physical testing rather than clinical study.

8. The sample size for the training set:

  • Not applicable. This submission is for an imaging system, not a machine learning or AI algorithm that would require a training set.

9. How the ground truth for the training set was established:

  • Not applicable. As above, no training set for an AI/ML algorithm is involved.

Summary of what the document does provide:

The document focuses on establishing substantial equivalence to an existing legally marketed device (the E.G. Scan™ Esophagoscope System, K111030, and Vision-Science, Inc.'s Modified Flexible Trans-Nasal Esophagoscope, K072088). This is a common pathway for 510(k) clearances, where extensive new clinical performance data is often not required if the new device is technologically similar and has the same indications for use as a predicate.

The main evidence provided for safety and effectiveness is:

  • Comparison to predicate device: Stating "The E.G. Scan™ II Esophagoscope System and predicate device are substantially equivalent in the areas of design, indication for use, technological characteristics, function, application and safety and effectiveness."
  • Compliance with recognized standards:
    • Biocompatibility (ISO 10993-1)
    • Medical electric equipment safety (IEC 60601-1)
    • Specific safety for endoscopic equipment (IEC 60601-2-18)
    • Electromagnetic compatibility (IEC 60601-1-2)

Therefore, the "acceptance criteria" here are primarily framed as compliance with these standards and demonstrated substantial equivalence, rather than specific quantitative performance metrics from a clinical diagnostic study.

{0}------------------------------------------------

120702

JUN 2 0 2012

510/lx) Submission -- E.G. Scan™ II Esophagoscope System

5′10(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: Dec. 06. 2011

  1. Company and Correspondent making the submission:

Name - IntroMedic Co., Ltd.

Address ~ Suite 1104, E&C Venture Dream Tower 6-Cha, 197-28 Guro-Dong, Guro-Gu,

Seoul, 152-719 Korea

Telephone - +82-2-801-9300

Fax – +82-2-801-9330

Contact - JinYoung, Lee

Internet – http://www.intromedic.com

    1. Device :
Proprietary name: E.G. Scan™ II Esophagoscope System
Common Name: Esophagoscope System
Classification Name: Esophagoscope System
  1. Predicate Device :
Manufacturer :IntroMedic Co., Ltd,
Device :E.G. Scan™ Esophagoscope System
510(k) Number :K111030
Manufacturer: Vision-Science, Inc.
Device: Modified Flexible Trans-Nasal Esophagoscope withDigital Video Processor and Disposable EndoSheath®System
510(k) Number: K072088
  1. Classifications Names & Citations :

. 21CFR874.4710, EOX, Esophagoscope System, Class2

IntroMedic Co., Ltd.

{1}------------------------------------------------

510(k) Submission - E.G. Scan™ II Esophagoscope System

  1. Description :

5.1 Introduction

E.G. Scan™ II Esophagoscope System and its accessories are used for diagnosis of patients. E.G. Scan™ II Probe takes pictures of the esophagus of human and sends image data to E.G. Scan™ II Controller. E.G. Scan™ II Controller processes and converts image data and upload. E.G. View™ image displaying software displays the image for diagnosis.

The E.G Scan™ II Esophagoscope Probe is disposable.

5.2 General Technology

E.G. Scan™ II Esophagoscope system is a transnasal esophagoscope designed to capture images of the esophagus. Captured images are viewed via the E.G. View™ Software for diagnosis of diseases related to the esophagus.

    1. Indication for use :
      The E.G. Scan™ II Esophagoscope System is intended for use in endoscopic access and examination of the larynx, esophagus and gastroesophageal junction.
    1. Comparison with predicate device :
      The E.G. Scan™ II Esophagoscope System and predicate device are substantially equivalent in the areas of design, indication for use, technological characteristics, function, application and safety and effectiveness. This was determined by reviewing the information provided in the 510(k) in comparison to the content specified in the FDA guidance documents.

8. Safety, EMC and Performance Data :

The E.G. Scan™ II Esophagoscope System has the same device characteristics as the predicate device, E.G. Scan™ Esophagoscope System of IntroMedic Co., Ltd.; intended use, material, design and use concept are similar. The biocompatibility of the patient contact parts has been demonstrated through the cytotoxicity, sensitization and irritation testing by ISO 10993-1 Biological evaluation of medical devices. The E.G. Scan™ II Esophagoscope System conforms to IEC 60601-1 Medical electric equipment, Part 1: General requirements for safety, IEC 60601-2-18

IntroMedic Co., Ltd.

{2}------------------------------------------------

K120702

510(k) Submission ~ E.G. Scan™ II Esophagoscope System

Medical electrical equipment-Part 2: Particular requirements for the safety of endoscopic equipment and IEC 60601-1-2 Medical electric equipment, General requirements for safety collateral standard electromagnetic compatibility.

9. Conclusions :

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification IntroMedic Co., Ltd. concludes that The E.G. Scan™ II Esophagoscope System is safe and effective and substantially equivalent to predicate devices as described herein.

IntroMedic Co., Ltd.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

IntroMedic Company, Limited % Mr. Marc M. Mouser Underwriters Laboratories, Inc. 2600 NW Lake Road Camas, WA 98607

JUN 2 0 2012

Re: K120702

Trade/Device Name: E.G. Scan™ II Esophagoscope System Regulation Number: 21 CFR 874.4710 Regulation Name: Esophagoscope (flexible or ridgid) and accessories Regulatory Class: II Product Code: EOX Dated: May 25, 2012 Received: June 1. 2012

Dear Mr. Mouser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 – Mr. Marc M. Mouser

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

E. E. Ivanov

Malvina B. Eydelman M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K120702

Indications for Use

510(k) Number(if known):

Device Name: E.G. Scan™ II Esophagoscope System

Indications for Use:

E.G. Scan™ II Esophagoscope System is intended for use in endoscopic access and examination of the larynx, esophagus and gastroesophageal junction.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation(ODE)

Page 1 of 1

S.M. Mofizur

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number_

× Prescription Use (Per 21 CFR 801.109)

§ 874.4710 Esophagoscope (flexible or rigid) and accessories.

(a)
Identification. An esophagoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the esophagoscope and is intended to examine or treat esophageal malfunction symptoms, esophageal or mediastinal disease, or to remove foreign bodies from the esophagus. When inserted, the device extends from the area of the hypopharynx to the stomach. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.