K Number
K120602

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-03-23

(24 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Apaceram™ Bone Graft Substitute is a synthetic hydroxyapatite provided in several particulate and shaped sizes. It is intended for use as a bone void filler for bony voids, gaps, or defects that are not intrinsic to the stability of the bony structure. Apaceram™ is intended to be placed into bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, or pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. It also can be used with autograft as a bone graft extender. Apaceram™ is resorbed and replaced with bone during the healing process.

Device Description

Apaceram™ is a hydroxyapatite osteoconductive bone void filler. It is available in four types: AX, B, G, and R, which vary in porosity, shape and sizes. Apaceram™ is provided sterile for single patient use. Apaceram™ is composed of calcium salts, is osteoconductive, and provides an interconnected, highly porous scaffold environment for new bone ingrowth.

AI/ML Overview

The provided text is a 510(k) summary for the Apaceram™ Bone Graft Substitute. This document primarily focuses on demonstrating substantial equivalence to a predicate device and does not contain information about acceptance criteria or a study proving the device meets specific performance metrics in the way a clinical trial or algorithm validation study would.

Here’s why it’s not possible to provide the requested information based on the given text:

  • Type of Submission: This is a "Special 510(k)" primarily for a "change in ownership" from Pentax Corporation to HOYA Corporation. Special 510(k)s are typically used for well-understood, minor modifications to existing devices that do not affect the device's fundamental safety or effectiveness, often relying heavily on the equivalence to the predicate device.
  • Focus on Equivalence: The core argument in this submission is that "The Apaceram™ Bone Graft Substitute is identical to the predicate device." It explicitly states, "The safety, performance and biocompatibility testing were submitted in the original Apaceram™ 510(k)." This means any detailed performance data, acceptance criteria, and specific studies would have been included in the original 510(k) submission (K071912), not in this update.
  • Device Type: This is a bone graft substitute, a physical material. The questions you've asked (e.g., sample size for test set, experts for ground truth, adjudication method, MRMC studies, standalone algorithm performance, training set sample size) are highly specific to the validation of AI/ML-based diagnostic or prognostic devices. They are not applicable to the regulatory approval process for a bone graft material like Apaceram™.

Therefore, I cannot provide the requested table and study details because the provided document does not contain this type of information. It falls outside the scope of this specific 510(k) submission.

{0}------------------------------------------------

KI20602

MAR 2 3 2012

Submitter: HOYA Corporatio

Apaceram™ Special 510(k)

, Haga-

HOYA Corporation PENTAX New Ceramics Division
510(k) SUMMARYApaceram™ Bone Void Filler
Submitter Name:HOYA Corporation PENTAX New Ceramics Division
Submitter Address:2-7-5 Naka-Ochiai, Shinjuku-ku, Tokyo 161-8525 JAPAN
Contact Person:Manufacturing Site:PENTAX Mashiko Factory: 858 Hanawa, Mashiko-machi, Hgun, Tochigi 321-4292 JapanNobuyuki AsaokaInternational Sales GroupNew Ceramics Division
Phone Number:813-5840-6141
Fax Number:813-5840-6143
Date Prepared:February 27, 2012
Device Trade Name:Apaceram™ Bone Graft Substitute
Device Common Name:Synthetic, porous hydroxylapatite
Classification Number:21 CFR 888.3045
Classification Name:Resorbable calcium salt bone void filler
Product Code:MQV
Predicate Devices:K071912 Apaceram™ Bone Graft Substitute, Pentax Corp.

Apaceram™ Bone Graft Substitute is a synthetic hydroxyapatite provided in several particulate and shaped sizes. It is intended for use as a bone void filler for bony voids, gaps, or defects that are not intrinsic to the stability of the bony structure. Apaceram™ is intended to be placed into bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, or pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. It also can be used with autograft as a bone graft extender. Apaceram™ is resorbed and replaced with bone during the healing process.

Apaceram™ is a hydroxyapatite osteoconductive bone void filler. It is available in four types: AX, B, G, and R, which vary in porosity, shape and sizes. Apaceram™ is provided sterile for single patient use.

Apaceram™ is composed of calcium salts, is osteoconductive, and provides an interconnected, highly porous scaffold environment for new bone ingrowth. The safety, performance and biocompatibility testing were submitted in the original Apaceram™ 510(k).

The Apaceram™ Bone Graft Substitute is identical to the predicate device. The purpose of this 510(k) is to reflect the change in ownership of Apaceram™ from Pentax Corporation to HOYA Corporation PENTAX New Ceramics Division.

Statement of Intended Use:

Device Description:

Technological Characteristics and Testing:

Substantial Equivalence:

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle or bird in flight, with three curved lines representing its wings and body. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Hova Corporation PENTAX New Ceramics Division % Trisler Consulting Patsy J. Trisler, JD, RAC 5600 Wisconsin Avenue, #509 Chevy Chase, Maryland 20815

MAR 2 3 2012

Re: K120602

Trade/Device Name: Apaceram™ Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV Dated: February 27, 2012 Received: February 28, 2012

Dear Ms. Trisler

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. `The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Patsy J. Trisler, JD, RAC

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you desire specific advice to: JDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

10 /1

Enclosure

{3}------------------------------------------------

Submitter: Hoya Corporation Pentax New Ceramics Division

Apaceram™ Special 510(k)

510(k) Number (if known):

K120602

Device Name:

Apaceram™ Bone Graft Substitute

Indications for Use:

Apaceram™ Bone Graft Substitute is a synthetic hydroxyapatite provided in several particulate and shaped sizes. It is intended for use as a bone void filler for bony voids, gaps, or defects that are not intrinsic to the stability of the bony structure. Apaceram™ is intended to be placed into bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, or pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. It also can be used with autograft as a bone graft . extender. Apaceram™ is resorbed and replaced with bone during the healing process.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Knutsen N. Loria

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K120602

Page 1 of 1

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.