K Number
K120592

Validate with FDA (Live)

Date Cleared
2013-02-27

(365 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dana Reusable Bowie Dick Type Test Pack, when used with a Dana Bowie Dick indicator, is designed for testing the air removal efficiency of pre-vacuum steam sterilizers. The indicator will demonstrate complete dye migration when proper sterilization conditions are met and proper vacuum is achieved. If there is inadequate vacuum such that would result in a 2°C (+1°-0°C) temperature difference between the center of the ANSVAAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle, the indicator will demonstrate incomplete dye migration (incomplete color change in the indicator window). The test parameters used for validating this device were 134℃ for 3.5 minutes with a 2 minute dry time.

Device Description

Dana Reusable Bowie Dick Type Pack and Dana Bowie Dick Indicators

AI/ML Overview

The provided document is a 510(k) premarket notification letter from the FDA for a device called "Dana Reusable Bowie Dick Type Pack and Dana Bowie Dick Indicators". This document typically outlines the FDA's decision on substantial equivalence based on data submitted by the manufacturer. However, it does not contain a detailed study report with specific acceptance criteria, sample sizes, ground truth establishment, or multi-reader multi-case study results.

The document states:

  • Device Name: Dana Reusable Bowie Dick Type Pack and Dana Bowie Dick Indicators
  • Intended Use: For testing the air removal efficiency of pre-vacuum steam sterilizers.
  • Performance Claim: The indicator will demonstrate complete dye migration when proper sterilization conditions are met and proper vacuum is achieved. If there is inadequate vacuum such that would result in a 2°C (+1°-0°C) temperature difference between the center of the ANSVAAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle, the indicator will demonstrate incomplete dye migration (incomplete color change in the indicator window).
  • Test Parameters for Validation: 134℃ for 3.5 minutes with a 2-minute dry time.

Based on the information provided in the given document, I cannot provide a complete answer to your request. The document is a regulatory approval letter, not a detailed scientific study report. It states that the device is "substantially equivalent" to predicate devices, implying that testing was conducted, but it does not include the specifics of that testing.

Here's what can be extracted and what is missing:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance (from document)
Complete Dye Migration (Positive Result): Demonstrates proper sterilization conditions and proper vacuum when used in a pre-vacuum steam sterilizer.The indicator will demonstrate complete dye migration when proper sterilization conditions are met and proper vacuum is achieved.
Incomplete Dye Migration (Negative Result): Demonstrates inadequate vacuum resulting in a 2°C (+1°-0°C) temperature difference between the center of the ANSVAAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle.If there is inadequate vacuum such that would result in a 2°C (+1°-0°C) temperature difference between the center of the ANSVAAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle, the indicator will demonstrate incomplete dye migration (incomplete color change in the indicator window).
Validation Test Parameters: 134℃ for 3.5 minutes with a 2-minute dry time.The test parameters used for validating this device were 134℃ for 3.5 minutes with a 2 minute dry time.

Missing Information (Not available in the provided document):

  • 2. Sample size used for the test set and the data provenance: Not specified.
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/specified. The ground truth here is a physical condition (temperature difference, vacuum presence/absence) measured by instruments, not expert interpretation of an image or signal.
  • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable (ground truth is instrumental/physical).
  • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable. This is a chemical indicator, not an AI or imaging device requiring human interpretation for its primary function.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device is a standalone indicator. Its "performance" is whether the dye changes correctly. There's no AI algorithm or human-in-the-loop for its direct function.
  • 7. The type of ground truth used: Based on the description, the ground truth is physical measurement of sterilization conditions, specifically the temperature difference in a standard test pack and the presence/absence of proper vacuum, conforming to ANSVAAMI/ISO 11140-5:2007 Standard. This is not expert consensus, pathology, or outcomes data in the traditional sense.
  • 8. The sample size for the training set: Not applicable for this type of device. There is no "training set" in the context of an AI/machine learning model. The device's performance is based on its chemical and physical properties.
  • 9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract emblem that resembles an eagle or bird-like figure with stylized wings and tail feathers.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 27, 2013

Mr. Harry Bala President Dana Products, Incorporated 7 Corey Drive SOUTH BARRINGTON IL 60010

Re: K120592

Trade/Device Name: Dana Reusable Bowie Dick Type Pack and Dana Bowie Dick Indicators Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: February 15, 2013 Received: February 19, 2013

Dear Mr. Bala:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Bala

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part. 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Image /page/1/Picture/6 description: The image contains a signature or handwritten text that is difficult to decipher due to its stylized and cursive nature. The text appears to be overlaid on top of some graphic elements, including what looks like a stylized letter or symbol on the left and a triangular shape on the right. The overall impression is that of a personalized autograph or a logo with a handwritten component.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

K120592

510(k) Number (if known):

Device Name: Dana Reusable Bowie Dick Type Test Pack and Dana Bowie Dick Indicators

Indications For Use:

The Dana Reusable Bowie Dick Type Test Pack, when used with a Dana Bowie Dick indicator, is designed for testing the air removal efficiency of pre-vacuum steam sterilizers. The indicator will demonstrate complete dye migration when proper sterilization conditions are met and proper vacuum is achieved. If there is inadequate vacuum such that would result in a 2°C (+1°-0°C) temperature difference between the center of the ANSVAAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle, the indicator will demonstrate incomplete dye migration (incomplete color change in the indicator window). The test parameters used for validating this device were 134℃ for 3.5 minutes with a 2 minute dry time.

Prescription Use AND/OR (Part 21 CFR 801 Subpart D) ·

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Claverie2013.02.25 17:26:13 -05'00'Page 1 of 1

sthesiology, General Hospital

510(k) Number:_**_

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).