K Number
K120563

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-03-30

(35 days)

Product Code
Regulation Number
866.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KeyPath™ MRSA/MSSA Blood Culture Test – BT is a qualitative in vitro diagnostic test for the timely identification of Staphylococcus aureus (S. aureus) and determination of methicillin susceptibility (MSSA) or methicillin resistance (MRSA) directly from positive blood cultures.

The Test uses bacteriophage amplification to identify the presence of S. aureus and assess the phenotypic response of the target organism to cefoxitin, an indicator of oxacillin (a methicillin analog) resistance.

The assay is performed directly on positive blood culture specimens that are determined as Gram Positive Cocci in singles (GPC) or as Gram Positive Cocci in Clusters (GPCC) by Gram stain.

The KeyPath™ MRSA/MSSA Blood Culture Test - BT is performed directly on positive blood culture specimens from BD BACTEC™ blood culture bottles (Plus Aerobic/F, Plus Anaerobic/F, Standard/10 Aerobic/F and Standard Anaerobic/F).

The Test is indicated for use in conjunction with other laboratory and clinical data available to the physician as an aid in the detection of MRSA/MSSA from positive blood cultures.

Subculturing of positive blood cultures is necessary for additional susceptibility test determinations, differentiation of mixed growth and for epidemiological typing.

Device Description

The KevPath™ MRSA/MSSA Blood Culture Test - BT (KevPath™ Test) is a bacteriophage amplification-enabled immunoasay to identify Staphylococcus aureus and determine its resistance or susceptibility to cefoxitin from positive blood cultures, which have been determined to have Gram Positive Cocci in singles (GPC) or Gram Positive Cocci in Clusters (GPCC) by Gram stain. This method utilizes the specificity of bacteriophage/bacteria interactions and their natural amplification processes to produce a surrogate signal. In the presence of S. aureus, bacteriophage will replicate, increasing to a detectable concentration. In the absence of S. aureus or the presence of bacteria other than S. aureus, the KevPath™ Test bacteriophage do not replicate and remain undetectable. In addition, susceptible strains of S. aureus do not grow in the presence of cefoxitin and therefore do not support bacteriophage amplification, while resistant strains will grow and support bacteriophage amplification.

To perform the KeyPath™ Test, a sample of the positive blood culture is added to each of the two provided reaction tubes, each comprised of KeyPath™ Test bacteriophage and proprietary reagents that enhance the growth of S. aureus and suppress other organisms. One Reaction Tube (Blue) is used for S. aureus identification. The second Reaction Tube (Red) is used for resistance/susceptibility testing. Following incubation, a small amount of the sample from each Tube is pipetted onto corresponding sample wells on the Detector. If the specimen contains S. aureus a pink-to-dark red/ purple Test Line (T) will appear in the Blue ID Window of the Detector.

If the sample is positive for S. aureus, Resistance/Susceptibility is then determined by reading the Red RS Window. Resistance (MRSA) is determined by the development of a visible pink-to-dark red/purple line at the Test Line (T) in the Red RS Window, while Susceptibility (MSSA) is determined by the absence of a visible colored line at the Test Line (T) in the Red RS Window.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the KeyPath™ MRSA/MSSA Blood Culture Test – BT, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Stated Goal)Reported Device Performance (SA Bottles - S. aureus ID)Reported Device Performance (SA Bottles - Resistance)Reported Device Performance (SN Bottles - S. aureus ID)Reported Device Performance (SN Bottles - Resistance)
Positive Agreement (S. aureus ID)≥ 95%100% (95% CI: 92.1-100)N/A97.8% (95% CI: 88.2-100)N/A
Negative Agreement (S. aureus ID)≥ 95%100% (95% CI: 90.0-100)N/A100% (95% CI: 88.4-100)N/A
Positive Agreement (Methicillin Resistance)≥ 95%N/A95.5% (95% CI: 77.2-99.9)N/A95.2% (95% CI: 76.2-99.9)
Negative Agreement (Methicillin Resistance)≥ 95%N/A95.7% (95% CI: 78.1-99.9)N/A95.7% (95% CI: 78.1-99.9)

Note: The acceptance criteria were stated as "≥ 95% positive and negative agreement for S. aureus identification and determination of methicillin resistance." The tables above show the results from the performance study, which met these criteria with high agreement percentages and confidence intervals.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • Performance Study: ≥ 20 MRSA, ≥ 20 MSSA, and ≥ 20 NSA clinical blood culture isolates from unique patients for each bottle type (Standard/10 Aerobic/F (SA) and Standard Anaerobic/F (SN)).
      • Specifically, for SA bottles: 45 positive controls, 35 negative controls for S. aureus ID. For resistance, the 2x2 table included 21 positive and 22 negative for resistance (total 43 S. aureus isolates).
      • Specifically, for SN bottles: 44 positive controls, 30 negative controls for S. aureus ID. For resistance, the 2x2 table included 20 positive and 22 negative for resistance (total 42 S. aureus isolates).
    • Media Interference Study: 16 strains (4 MRSA, 4 MSSA, and 8 NSA) with 5 replicates each, for both SA and SN bottles.
  • Data Provenance: Clinical blood culture isolates from unique patients. The blood for charge bottles was from healthy volunteers. The study was conducted per IRB-approved protocol MP2007A. The country of origin is not explicitly stated but implied to be the US given the submission to the FDA. The study is prospective in nature as bottles were inoculated and grown for testing, mimicking a real-world scenario.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • The document does not specify the number or qualifications of experts used to establish the ground truth. It refers to a "Reference" method for comparison but doesn't detail how this reference standard was itself established or by whom.

4. Adjudication Method for the Test Set

  • The document does not explicitly describe an adjudication method for the test set. It uses the "Reference" method as the comparative standard.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The device is an in vitro diagnostic test meant for laboratory use, not a medical imaging device or interpretative aid that would typically involve human readers.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, the performance study effectively evaluates the device in a standalone manner. The KeyPath™ test provides a qualitative result directly (pink-to-dark red/purple line or absence thereof) without active human interpretation or real-time human-in-the-loop decision making, beyond observing the test results per the package insert. The "Reference" method serves as the ground truth against which the device's accuracy is measured.

7. Type of Ground Truth Used

  • The type of ground truth used is a reference method, which is compared to the KeyPath™ Test results. The exact nature of this "Reference" method (e.g., conventional microbiological culture and susceptibility testing, molecular methods) is not explicitly detailed in the provided text, but it is implied to be the established gold standard for S. aureus identification and methicillin resistance determination.

8. Sample Size for the Training Set

  • The document does not directly describe a separate "training set" for the device, as it is a bacteriophage amplification-enabled immunoassay rather than a machine learning algorithm that typically requires a distinct training phase. The study described focuses on validation of the device's performance against a reference standard using clinical isolates.

9. How the Ground Truth for the Training Set Was Established

  • As mentioned above, there is no explicit "training set" described in the context of machine learning. The device's underlying biological mechanism (bacteriophage amplification) does not involve a training process in the conventional sense. The "ground truth" for evaluating its performance (as described in point 7) is based on established microbiology methods.

{0}------------------------------------------------

K120563

MAR 3 0 2012

510(k) SUMMARY

MICROPHAGE's KeyPath™ MRSA/MSSA Blood Culture Test – BT

MicroPhage, Inc. 2400 Trade Centre Ave. Longmont, CO 80503

Phone: 303-652-5049 Facsimile: 303-652-5080

Contact Person: Drew Smith, PhD.

Date Prepared: February 24, 2012

Name of Device and Name/Address of Sponsor

KeyPath™ MRSA/MSSA Blood Culture Test - BT

MicroPhage, Inc. 2400 Trade Centre Ave. Longmont, CO 80503

Common or Usual Name

Methicillin-resistant Staphylococcus aureus (MRSA) and Methicillin-susceptible Staphylococcus aureus (MSSA) from positive blood culture bottles test

Classification Name

Staphylococcal typing bacteriophage

Regulation

21 C.F.R. § 866.2050

Product Code

ous

Classification Panel

Microbiology

Predicate Devices

KeyPath™ MRSA/MSSA Blood Culture Test - BT (K102342)

{1}------------------------------------------------

Intended Use / Indications for Use

The KeyPath™ MRSA/MSSA Blood Culture Test – BT is a qualitative in vitro diagnostic test for the timely identification of Staphylococcus aureus (S. aureus) and determination of methicillin susceptibility (MSSA) or methicillin resistance (MRSA) directly from positive blood cultures.

The Test uses bacteriophage amplification to identify the presence of S. aureus and assess the phenotypic response of the target organism to cefoxitin, an indicator of oxacillin (a methicillin analog) resistance.

The assay is performed directly on positive blood culture specimens that are determined as Gram Positive Cocci in singles (GPC) or as Gram Positive Cocci in Clusters (GPCC) by Gram stain.

The KeyPath™ MRSA/MSSA Blood Culture Test - BT is performed directly on positive blood culture specimens from BD BACTEC™ blood culture bottles (Plus Aerobic/F, Plus Anaerobic/F, Standard/10 Aerobic/F and Standard Anaerobic/F).

The Test is indicated for use in conjunction with other laboratory and clinical data available to the physician as an aid in the detection of MRSA/MSSA from positive blood cultures.

Subculturing of positive blood cultures is necessary for additional susceptibility test determinations, differentiation of mixed growth and for epidemiological typing.

Technological Characteristics

The KevPath™ MRSA/MSSA Blood Culture Test - BT (KevPath™ Test) is a bacteriophage amplification-enabled immunoasay to identify Staphylococcus aureus and determine its resistance or susceptibility to cefoxitin from positive blood cultures, which have been determined to have Gram Positive Cocci in singles (GPC) or Gram Positive Cocci in Clusters (GPCC) by Gram stain. This method utilizes the specificity of bacteriophage/bacteria interactions and their natural amplification processes to produce a surrogate signal. In the presence of S. aureus, bacteriophage will replicate, increasing to a detectable concentration. In the absence of S. aureus or the presence of bacteria other than S. aureus, the KevPath™ Test bacteriophage do not replicate and remain undetectable. In addition, susceptible strains of S. aureus do not grow in the presence of cefoxitin and therefore do not support bacteriophage amplification, while resistant strains will grow and support bacteriophage amplification.

To perform the KeyPath™ Test, a sample of the positive blood culture is added to each of the two provided reaction tubes, each comprised of KeyPath™ Test bacteriophage and proprietary reagents that enhance the growth of S. aureus and suppress other organisms. One Reaction Tube (Blue) is used for S. aureus identification. The second Reaction Tube (Red) is used for resistance/susceptibility testing. Following incubation, a small amount of the sample from each Tube is pipetted onto corresponding sample wells on the Detector. If the specimen contains S. aureus a pink-to-dark red/ purple Test Line (T) will appear in the Blue ID Window of the Detector.

If the sample is positive for S. aureus, Resistance/Susceptibility is then determined by reading the Red RS Window. Resistance (MRSA) is determined by the development of a visible pink-to-dark red/purple line at the Test Line (T) in the Red RS Window, while Susceptibility (MSSA) is determined by the absence of a visible colored line at the Test Line (T) in the Red RS Window.

{2}------------------------------------------------

Performance Data

Study Design

The purpose of this submission is to establish equivalent performance of the KeyPath™ MRSA/MSSA Blood Culture Test - BT in additional bottle types, as compared to cleared bottle types. Previously cleared bottle types are BACTEC™ Plus Aerobic/F and Plus Anaerobic/F. The additional bottle types for which clearance is being sought are BACTEC™ Standard /10 Aerobic/F and Standard Anaerobic/F.

A Media Interference Study was performed to demonstrate that blood culture matrix does not act as an interferent in the KeyPath™ Test; and a performance study was performed to demonstrate that the KevPath™ Test shows ≥ 95% positive and negative agreement for S. aureus identification and determination of methicillin resistance.

Media Interference Study

Study description and methods

Performance of the KevPath™ MRSA/MSSA Blood Culture Test ~ BT was analyzed in presence of 10 µL (nominal) and 20µL (twice the nominal volume) of blood culture matrix from appropriate bottles. Five replicates of 16 strains (4 MRSA, 4 MSSA and 8 NSA) were tested. For test matrix and positive controls, blood culture bottles containing blood from volunteers were inoculated with approximately 100 CFU of bacteria and grown to alarm in the BACTEC™ 9050. Negative controls were comprised of un-inoculated blood culture mix containing blood. No interference was demonstrated if the correct results for each strain type were observed.

Results and conclusions

Negative controls - For both SA and SN bottles the KeyPath™ Test returned negative results (NSA) in all replicates, as shown in Tables 6.1 and 6.2.

Positive controls - For both SA and SN bottles, the KeyPath™ Test returned the expected results in all replicates, as shown in Tables 6.1 and 6.2.

Test matrix - For both SA and SN bottles, the KeyPath™ Test returned the expected results in all replicates, as shown in Tables 6.1 and 6.2.

{3}------------------------------------------------

KP Test Result
ConditionTypeMRSAMSSANSA
Negative ControlNone0020
ControlMRSA2000
MSSA0200
NSA0040
TestMRSA2000
MSSA0200
NSA0040

Table 6.1 – Summary Data for SA Bottle Interference Summary data for SA bottle interference

Table 6.2 – Summary Data for SN Bottle Interference Summary data for SN bottle interference

KP Test Result
ConditionTypeMRSAMSSANSA
Negative ControlNone0020
ControlMRSA2000
MSSA0200
NSA0040
TestMRSA2000
MSSA0200
NSA0040

All test results are as expected. We conclude that blood culture matrix from SA and SN bottles does not interfere with the KeyPath™ MRSA/MSSA Blood Culture Test - BT.

{4}------------------------------------------------

Performance Study

Study description and methods

A panel of ≥ 20 MRSA, ≥ 20 MSSA and ≥ 20 NSA clinical blood culture isolates from unique patients was tested in the KeyPath™ MRSA/MSSA Blood Culture Test - BT.

BACTEC™ Standard/10 Aerobic/F (SA) and Standard Anaerobic/F (SN) bottles were charged with blood from healthy volunteers per IRB-approved study protocol MP2007A. Bottles were inoculated with approximately 100 CFU from fresh overnight cultures.

Testing and external daily controls were performed per Package Insert instructions (181-00005 Rev. A) within 24 hours of alarm. Bottles that did not alarm within 3 days were discarded.

Results and conclusions

Tables 6.3 and 6.4 summarize the SA bottle data for S. aureus identification, and determination of resistance and susceptibility within S. aureus, respectively. Tables 6.5 and 6.6 summarize the SN bottle data for S. aureus identification, and determination of resistance and susceptibility within S. aureus, respectively.

. Table 6.3 - S. aureus identification summary, SA bottles

Analysis: S. oureus Identification

Image /page/4/Figure/9 description: The image shows a table of data related to SA bottles and KeyPath. The table includes a reference section with values 45 and 35, and a statistic section with positive and negative agreement percentages. Both positive and negative agreement are at 100%, with 95% confidence intervals ranging from 92.1 to 100 and 90.0 to 100, respectively.

Table 6.4 - Resistance summary, SA bottles

Analysis: Methicillin resistance

SA bottles

Image /page/4/Figure/13 description: The image shows a 2x2 contingency table comparing two methods, KeyPath and Reference. The table displays the counts of agreements and disagreements between the two methods. The top left cell shows 21, indicating the number of instances where both KeyPath and Reference are positive. The bottom right cell shows 22, indicating the number of instances where both KeyPath and Reference are negative.

StatisticPercent95 % CI
LowerUpper
Positive Agreement95.577.299.9
Negative Agreement95.778.199.9

{5}------------------------------------------------

Image /page/5/Figure/0 description: The image shows an analysis of S. aureus Identification using SN bottles. A 2x2 table compares KeyPath to a reference, with values 44 and 0 in the first row, and 1 and 30 in the second row. The totals are 45 and 30. A table shows the positive agreement is 97.8% with a 95% confidence interval of 88.2 to 100, and the negative agreement is 100% with a 95% confidence interval of 88.4 to 100.

Table 6.5 - S. aureus Identification Summary, SN bottles

Table 6.6 - Methicillin Resistance Testing Summary, SN bottles

Analysis: Methicillin resistance

SN bottles

Reference
+-
KeyPath+201
-122
Total2123
StatisticPercent95 % CI
LowerUpper
Positive Agreement95.276.299.9
Negative Agreement95.778.199.9

For all evaluations of S. aureus identification and determination of resistance and susceptibility, the KeyPath™ MRSA/MSSA Blood Culture Test - BT functioned as intended and the levels of positive and negative agreement observed were as expected.

Substantial Equivalence

The KeyPath™ MRSA/MSSA Blood Culture Test - BT, when used on samples from BACTEC™ Standard/10 Aerobic/F or Standard Anaerobic/F bottles, is as safe and effective as the KeyPath™ MRSA/MSSA Blood Culture Test - BT when used on samples from cleared bottle types (BACTEC™ Plus Aerobic/F and Plus Anaerobic/F). The KeyPath™ MRSA/MSSA Blood Culture Test - BT technological characteristics and principles of operation are unchanged. Other than the addition of bottle types, the intended use is unchanged. The expanded intended use raises no new issues of safety or effectiveness. Performance data demonstrate that the KeyPath™ MRSA/MSSA Blood Culture Test – BT, when used on samples from BACTEC™ Standard/10 Aerobic/F or Standard Anaerobic/F bottles, is as safe and effective as the KeyPath™ MRSA/MSSA Blood Culture Test – BT when used on samples from cleared bottle types (BACTEC™ Plus Aerobic/F and Plus

{6}------------------------------------------------

Thus, the KeyPath™ MRSA/MSSA Blood Culture Test – BT is substantially Anaerobic/F). equivalent when used with additional BACTEC™ bottle types.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . .

:

.

.

.

:

. . . . . .

:

:

:

:

:

.

:

. .

·

. .

:

:

.

. 19

. . . . . . . .

:

1:00 pm 1 personal program and

.

.

:

. .

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around it. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of curved lines.

Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

MAR 3 0 2012

MicroPhage, Inc. c/o Drew Smith, Ph.D. Chief Science Officer 2400 Trade Centre Ave Longmont, CO 80503

Re: K120563

Trade/Device Name: KeyPath™ MRSA/MSSA Blood Culture Test-BT Regulation Number: 21 CFR § 866.2050 Regulation Name: Staphylococcal typing bacteriophage Regulatory Class: I Product Code: OUS Dated: February 24, 2012 Received: February 24, 2012

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice

{8}------------------------------------------------

Page 2 - Drew Smith Ph.D.

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus; permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Silla Adams

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use Statement

K120563 510(k) Number (if known):

Device Name: KeyPath™ MRSA/MSSA Blood Culture Test - BT

Indications for Use:

The KeyPath™ MRSA/MSSA Blood Culture Test – BT is a qualitative in vitro diagnostic test for the timely identification of Staphylococcus aureus (S. aureus) and determination of methicillin susceptibility (MSSA) or methicillin resistance (MRSA) directly from positive blood cultures.

The Test uses bacteriophage amplification to identify the presence of S. aureus and assess the phenotypic response of the target organism to cefoxitin, an indicator of oxacillin (a methiciliin analog) resistance.

The assay is performed directly on positive blood culture specimens that are determined as Gram Positive Cocci in singles (GPC) or as Gram Positive Cocci in Clusters (GPCC) by Gram stain.

The KeyPath™ MRSAMSSA Blood Culture Test - BT is performed directly on positive blood culture specimens from BD BACTEC™ blood culture bottles (Plus Aerobic/F, Plus Anaerobic/F, Standard/10 Aerobic/F and Standard Anaerobic/F).

The Test is indicated for use in conjunction with other laboratory and clinical data available to the physician as an aid in the detection of MRSA/MSSA from positive blood cultures.

Subculturing of positive blood cultures is necessary for additional susceptibility test determinations, differentiation of mixed growth and for epidemiological typing

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sudali lu-Pode
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K120563

16

§ 866.2050 Staphylococcal typing bacteriophage.

(a)
Identification. A staphylococcal typing bacteriophage is a device consisting of a bacterial virus intended for medical purposes to identify pathogenic staphylococcal bacteria through use of the bacteria's susceptibility to destruction by the virus. Test results are used principally for the collection of epidemiological information.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.