K Number
K120544

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-03-19

(25 days)

Product Code
Regulation Number
870.1220
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

St Jude Medical (SJM) Electrophysiology Catheters can be used in evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

Device Description

The Response™ Electrophysiology Catheter with Lumen is a sterile, single use, electrophysiological diagnostic catheter with multiple electrodes and a fixed distal curve for intracardiac electrogram sensing (IECG). The body is a continuous tube with a central lumen for fluid infusion through a 3-way valve.

AI/ML Overview

This document is a 510(k) summary for the St. Jude Medical Response™ Electrophysiology Catheter with Lumen. It is a submission for a modified device, establishing substantial equivalence to a predicate device (K914278) which was also a Response™ Electrophysiology Catheter with Lumen.

The submission is primarily focused on demonstrating substantial equivalence for a modified version of an existing device, rather than proving a new device meets specific performance criteria through a clinical study. Therefore, the information typically requested about acceptance criteria, detailed device performance, sample sizes for test/training sets, expert involvement, and ground truth establishment, which are common for AI/ML device submissions, is not present in this document.

This 510(k) focuses on showing that changes (specifically, a resin modification) to an already cleared device do not adversely affect its safety and effectiveness.

Here's a breakdown of what can be extracted based on the provided text, and what cannot:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated as numerical performance metrics in this document. The overarching "acceptance criterion" is proving substantial equivalence to the predicate device. For the bench testing, the acceptance criterion was that the modified design meets "specifications" and that "biocompatibility was confirmed in accordance with ISO 10993-1." Specific numerical targets for performance parameters are not detailed.
  • Reported Device Performance: The document states that "Results of testing demonstrate that the modified design meets specifications as biocompatibility was confirmed in accordance with ISO 10993-1." However, no specific performance values are provided. The key claim is that the device "has the same intended use and fundamental scientific technology" as the predicate, and that the "resin modification does not adversely affect the safety and effectiveness."
Acceptance Criteria (Inferred)Reported Device Performance
Modified design meets specifications."Results of testing demonstrate that the modified design meets specifications."
Biocompatibility confirmed in accordance with ISO 10993-1."Biocompatibility was confirmed in accordance with ISO 10993-1."
Resin modification does not adversely affect safety and effectiveness."It was demonstrated that the resin modification does not adversely affect the safety and effectiveness."
Same intended use and fundamental scientific technology as predicate device."The modified Response™ Electrophysiology Catheter with Lumen has the same intended use and fundamental scientific technology as the predicate device."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • This information is not provided in the document. The submission mentions "bench testing" and "biocompatibility testing." These generally refer to laboratory (in vitro) and, potentially, animal (in vivo for biocompatibility) studies, not human clinical trials with "test sets" in the context of device performance in patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • This information is not provided as this type of expert review to establish ground truth is typical for diagnostic algorithms (e.g., AI/ML in imaging), not for a modified physical medical device like an electrophysiology catheter where performance is assessed via engineering and biocompatibility tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This information is not provided and is not applicable to the type of testing described in this 510(k).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This information is not provided and is not applicable. This is a physical medical device, not an AI/ML diagnostic or assistive device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This information is not provided and is not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the "bench testing," the "ground truth" would be the device's technical specifications and performance standards established by the manufacturer, often based on engineering principles and regulatory guidance.
  • For "biocompatibility testing," the "ground truth" is compliance with the ISO 10993-1 standard.

8. The sample size for the training set

  • This information is not provided and is not applicable. There is no concept of a "training set" for a physical medical device where manufacturing and materials are modified.

9. How the ground truth for the training set was established

  • This information is not provided and is not applicable.

{0}------------------------------------------------

12

MAR 1 9 2012

510(k) Summary
510(k) Number
Date Prepared:February 21, 2012
Submitter Information:
SubmitterName/Address:St. Jude Medical - Atrial Fibrillation Division14901 DeVeau PlaceMinnetonka, MN 55345Establishment Registration Number: 3005188751
Contact Person:Marlene PetersonPrincipal Regulatory Affairs SpecialistPhone (651) 756-3268Fax (952) 930-9481MPeterson07@sjm.com
Device Information:
Trade Name:Response™ Electrophysiology Catheter with Lumen
Common Name:Electrode Recording Catheter
Classification Name:Electrode Recording Catheter
Classification:Class II, 21 CFR 870.1220, Product Code DRF.
Predicate Device(s):Response™ Electrophysiology Catheter with Lumen(K914278).
Device Description:The Response™ Electrophysiology Catheter with Lumen is asterile, single use, electrophysiological diagnostic catheter withmultiple electrodes and a fixed distal curve for intracardiacelectrogram sensing (IECG). The body is a continuous tube with acentral lumen for fluid infusion through a 3-way valve.
Intended Use:(Indications for Use)St Jude Medical (SJM) Electrophysiology Catheters can be used inevaluation of a variety of cardiac arrhythmias from endocardial andintravascular sites.
Comparison toPredicate DevicesThe modified Response™ Electrophysiology Catheter with Lumenhas the same intended use and fundamental scientific technology asthe predicate device. The technological characteristics of themodified Response™ Electrophysiology Catheter with Lumen aresubstantially equivalent to the predicate devices includingpackaging, biocompatibility, sterilization, and labeling. Through
biocompatibility testing it was demonstrated that the resinmodification does not adversely affect the safety and effectiveness.
Summary on Non-Clinical TestingBench testing was performed to verify the device modifications.Results of testing demonstrate that the modified design meetsspecifications as biocompatibility was confirmed in accordancewith ISO 10993-1.
Statement ofEquivalenceThe modified St. Jude Medical Response™ ElectrophysiologyCatheter with Lumen has the same indications for use andtechnological characteristics as the predicate device. Based on thisand the data provided in this pre-market notification, St. JudeMedical's modified Response™ Electrophysiology Catheter withLumen has been shown to be substantially equivalent to thepredicate.

Response Electrophysiology Catheter
With Lumen
Special 510(k)

Confidential

St. Jude Medical

·

{1}------------------------------------------------

Note; K914278 was originally submitted under the Daig Corporation which was subsequently obtained by St. Jude Medical. The predicate to K914278 was K894500 for Electrophysiology Catheters

Response Electrophysiology Catheter With Lumen Special 510(k)

Confidential

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAR 1 9 2012

St. Jude Medical c/o Ms. Marlene Peterson Sr. Regulatory Affairs Specialist 14901 DeVeau Place Minnetonka, MN 55345

Re: K120544

Trade/Device Name: St Jude Medical Response Electrophysiology Catheter with Lumen Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode recording catheter or electrode recording probe. Regulatory Class: Class II (two) Product Code: DRF Dated: February 21, 2012 Received: February 23, 2012

Dear Ms. Peterson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Marlene Peterson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kmphe

Bram Ø. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

12 12 05 44 510(K) Number (if known):

Device Name: Response™ Electrophysiology Catheter with Lumen

Indications for Use:

St Jude Medical (SJM) Electrophysiology Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) .

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

11

(Division Si Division of C 510(k) Number

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).