K Number
K120474
Manufacturer
Date Cleared
2012-07-10

(145 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Gyrus ACMI Telescope Storage-Sterilization Tray (TEL-GT) is intended to be used to enclose and protect Gyrus ACMI telescopes during Steam Sterilization using the following parameters:

STEAM Sterilization Parameters:

PreVacuum
Temp:275°F132°C
Pressure26psigs
Exposure3 minutes
Dry Time20 minutes

The trays are to be double wrapped with an FDA cleared sterilization wrap. The trays are optional accessories to the Gyrus ACMI telescopes. Telescopes without lumens up to 31 cm in length may be sterilized in the tray with a diameter between 2.7mm to 4mm. Two telescopes can be sterilized at one time.

Maintenance of Sterility is dependent on the sterilization wrap shelf life and not the tray. User must follow the instructions for use supplied with the approved sterilization wrap and follow recommended storage conditions and timeframes for shelf life of the sterilization wrap.

Device Description

The Gyrus ACMI Telescope Storage-Sterilization Trav is a two-piece container made of a plastic lid containing 14 holes, and a plastic trav containing 18 holes that permit ready ingress of Steam. The tray provides protection to the telescopes during sterilization and storage.

AI/ML Overview

The provided document, K120474, describes the Gyrus ACMI Telescope Storage-Sterilization Tray. This is a medical device, and the information presented is for a 510(k) premarket notification. The document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a study report with performance metrics in the way one would for a diagnostic or AI-driven decision support system.

Therefore, many of the requested categories (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set details) are not applicable to this type of device submission. The "study" described is a validation of the sterilization process through parameters, not a clinical performance study.

Here's an attempt to populate the table and answer the questions based on the available information:


Acceptance Criteria and Device Performance for Gyrus ACMI Telescope Storage-Sterilization Tray (K120474)

The Gyrus ACMI Telescope Storage-Sterilization Tray is intended for use in steam sterilization. The "acceptance criteria" for this device are fundamentally linked to its ability to facilitate effective steam sterilization and protect the enclosed telescopes. The "performance" is demonstrated by its compatibility with specific sterilization parameters and its design features that allow for proper steam ingress.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance / Compliance
Sterilization EffectivenessThe tray must permit steam sterilization of Gyrus ACMI telescopes using defined PreVacuum steam parameters (Temp: 275°F/132°C, Pressure: 26psigs, Exposure: 3 minutes, Dry Time: 20 minutes) when double-wrapped with FDA cleared sterilization wrap.Validation performed. The device was validated to sterilize Gyrus ACMI telescopes using Steam with the specified parameters. The plastic lid and tray contain 14 and 18 holes, respectively, "that permit ready ingress of Steam." The conclusion states "performance data" supports substantial equivalence.
Device ProtectionThe tray must enclose and protect Gyrus ACMI telescopes (without lumens, up to 31 cm length, 2.7mm to 4mm diameter) during sterilization and storage.The tray provides "protection to the telescopes during sterilization and storage."
BiocompatibilityMaterials must be compatible with testing identified in ISO10993-1."Materials are compatible with testing identified in ISO10993-1." The DEVICE does not come into contact with patients.
Material EquivalenceMaterials used must be the same as the predicate devices, or other legally marketed devices by Gyrus ACMI."The Gyrus ACMI Telescope Storage-Sterilization Tray is composed of the same materials as the predicates..." as well as other legally marketed devices.
Maintenance of SterilityMaintenance of sterility is dependent on the sterilization wrap shelf life, not the tray. Users must follow IFU for wrap.Explicitly stated: "Maintenance of Sterility is dependent on the sterilization wrap shelf life and not the tray. User must follow the instructions for use supplied with the approved sterilization wrap..."
Compliance to StandardsDesign must comply with relevant voluntary standards."The design of the DEVICE complies with the following standards: ISO10993-1, AAMI/ANSI ST77, AAMI/ANSI ST79."
Substantial EquivalenceThe device must be substantially equivalent to legally marketed predicate devices."In summary, based on intended use, technological characteristics and performance data, the Gyrus ACMI Telescope Storage-Sterilization Tray is substantially equivalent to previously cleared Gyrus ACMI storage-sterilization trays."

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of statistical analysis of a diagnostic or AI device. The validation described is for the sterilization process parameters using the tray. While specific numbers of sterilization cycles or tests are implied by "validation," the sample size and data provenance (country of origin, retrospective/prospective) are not detailed in this 510(k) summary. This is typical for a device like a sterilization tray where the focus is on process validation rather than clinical outcome studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. There is no mention of experts establishing a "ground truth" for a test set in the context of this device. The "ground truth" for sterilization effectiveness would be defined by standard microbiological testing and physical indicators during the validation process, which is established by established protocols and regulatory standards, not expert consensus on individual cases.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. There is no test set requiring expert adjudication for this type of device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not a diagnostic device or an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithmic device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For the sterilization effectiveness, the "ground truth" is established by adherence to validated sterilization parameters, successful inactivation of biological indicators if used, and sterility assurance levels as defined by relevant standards (e.g., AAMI/ANSI ST77, ST79). For biocompatibility, the ground truth is established by material testing against ISO10993-1. These are physical/microbiological truths, not expert consensus or pathology.

8. The sample size for the training set

Not applicable. This is not an AI/ML device requiring a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device requiring a training set.

{0}------------------------------------------------

K120474

Image /page/0/Picture/1 description: The image shows the word "OLYMPUS" in a bold, sans-serif font. The letters are large and black, and they are spaced closely together. There is a registered trademark symbol to the right of the letter "S". The word is likely a logo or branding for the company Olympus.

Your Vision Our Fut

6) 510(k) Summary

JUL 1 0 2012

510(k) Summary for Gyrus ACMI Telescope Storage-Sterilization Tray

General Information

Manufacturer:

Gyrus ACMI Inc. 136 Turnpike Rd. Southborough, MA 01772-2104

Establishment Registration Number:

Contact Person:

Device Description

Classification Name: Product Code Trade Name:

Generic/Common Name:

Predicate Device

3003790304

Graham Baillie Manager, Regulatory Affairs

Sterilization Wrap КСТ Gyrus ACMI Telescope Storage-Sterilization Tray

Sterilization wrap containers, trays, cassettes & other accessories

K092682

Intended Use

The Gyrus ACMI Telescope Storage-Sterilization Tray (TEL-GT) is intended to be used to enclose and protect Gyrus ACMI telescopes during Steam Sterilization using the following parameters:

STEAM Sterilization Parameters:

PreVacuum
Temp:275°F132°C
Pressure26psigs
Exposure3 minutes
Dry Time20 minutes

The trays are to be double wrapped with an FDA cleared sterilization wrap. The trays are optional accessories to the Gyrus ACMI telescopes. Telescopes without lumens up to 31 cm in length may be sterilized in the tray with a diameter between 2.7mm to 4mm. Two telescopes can be sterilized at one time.

Gyrus ACMI, Inc 136 Turnpike Road Southborough, MA 01772

office 508.804.2600

508.804.2624 lax www.gyrusacmi.com

Image /page/0/Picture/28 description: The image shows the logo for "GYRUS ACMI". Above the text is a graphic of a triangle made of several lines. Below the text is the phrase "Power through integration".

  • Olympus

{1}------------------------------------------------

Maintenance of Sterility is dependent on the sterilization wrap shelf life and not the tray. User must follow the instructions for use supplied with the approved sterilization wrap and follow recommended storage conditions and timeframes for shelf life of the sterilization wrap.

Product Description

The Gyrus ACMI Telescope Storage-Sterilization Trav is a two-piece container made of a plastic lid containing 14 holes, and a plastic trav containing 18 holes that permit ready ingress of Steam. The tray provides protection to the telescopes during sterilization and storage.

Design

Compliance to Voluntary Standards

The design of the DEVICE complies with the following standards: ISO10993-1, AAMI/ANSI ST77, AAMI/ANSI ST79

Material

The DEVICE does not come into contact with patients. The materials used are the same as the materials in the predicate devices, as well as other legally marketed devices manufactured by Gyrus ACMI. Materials are compatible with testing identified in ISO10993-1.

Summary of Sterilization and Shelf Life Discussion

The DEVICE was validated to sterilize Gyrus ACMI telescopes using Steam with the following parameters:

Temp:Pre Vacuum
275°F
132°C
Pressure26 psigs
Exposure3 minutes
Dry Time20 minutes

Maintenance of Sterility is dependent on the sterilization wrap shelf life and not the tray. User must follow the instructions for use supplied with the sterilization wrap and follow manufacturers recommended storage conditions and timeframes for shelf life of the sterilization wrap.

Technological Characteristics and Substantial Equivalence

The Gyrus ACMI Telescope Storage-Sterilization Tray is composed of the same materials as the predicates and utilizes the same features as the predicates. Validation testing was performed to ensure sterility of devices recommended for use with the proposed device.

{2}------------------------------------------------

Conclusion:

In summary, based on intended use, technological characteristics and performance data, the Gyrus ACMI Telescope Storage-Sterilization Tray is substantially equivalent to previously cleared Gyrus ACMI storage-sterilization trays.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Graham Baillie Manager, Regulatory Affairs Gyrus ACMI, Incorporated 136 Turnpike Road Southborough, Massachusetts 01772

JUL 10 2012

Re: K120474

Trade/Device Name: Gyrus ACMI Telescope Storage-Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: June 14, 2012 Received: June 15, 2012

Dear Mr. Baillie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Baillie

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), . please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For

Anthony D. Watson, B.S., M.S., M.B.A. · Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{5}------------------------------------------------

K120474

5) Indications for Use Statement

Statement of Intended Use

510(k) Number (if Known):

Gyrus ACMI Telescope Storage-Sterilization Tray Device Name:

Indications For Use:

The Gyrus ACMI Telescope Storage-Sterilization Tray (TEL-GT) is intended to be used to enclose and protect Gyrus ACMI telescopes during Steam Sterilization using the following parameters:

STEAM Sterilization Parameters:

PreVacuum
Temp:275°F132°C
Pressure26psigs
Exposure3 minutes
Dry Time20 minutes

The trays are to be double wrapped with an FDA cleared sterilization wrap. The trays are optional accessories to the Gyrus ACMI telescopes. Telescopes without lumens up to 31 cm in length may be sterilized in the tray with a diameter between 2.7mm to 4mm. Two telescopes can be sterilized at one time.

Maintenance of Sterility is dependent on the sterilization wrap shelf life and not the tray. User must follow the instructions for use supplied with the sterilization wrap and follow manufacturers recommended storage conditions and timeframes for shelf life .

Prescription Use And/Or Over the Counter Use X

(21 CFR Part 801 Subpart D)

(21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation, ODE

Eglit D-Claudio Will

Page of

(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K120474

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).