K Number
K120397

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-09-19

(224 days)

Product Code
Regulation Number
882.1870
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nihon Kohden MEB-2300A Neuropack Evoked Potential and EMG Measuring System is intended to monitor, record and display the bioelectric signals produced by muscles, to stimulate peripheral nerves, and to monitor, record and display the electrical activity produced by nerves to aid the clinician in the diagnosis and prognosis of neuromuscular disease (EMG). The device is also intended to measure and display nerve conduction time by applying a stimulus to a patient's nerve (NCV). The device may use electrical stimulus, visual stimulus, or sound stimulus for use in evoked response measurements (EP). The device may be used to determine autonomic responses as psychological indicators by measuring the electrical resistance of the skin and the tissue path between two electrodes applied to the skin. The device may also measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head (EEG).

The device is intended for use by medical personnel within a hospital, laboratory, clinic or nursing home setting or outside of a medical facility under direct supervision of a medical professional. The device may also be placed in the intensive care unit or operating room for continuous recording.

The device is available for use on any patient as determined by the medical professional including adults and children.

Device Description

The MEB-2200A Neuropack Evoked Potential and EMG Measuring System consists of a main acquisition unit, operation control panel, foot switch, electrode junction box (head box), electrical stimulation box and commercially available computer, monitor, printer, keyboard, and mouse. The device also includes optional visual, and auditory stimulators. The monitor, keyboard, and mouse connect to the computer. The computer, electrode junction box, electrical stimulation box, and control panel connect to the main unit. All components requiring AC power plug into the main unit's built-in, isolated power supply. The main unit plugs into a hospital grade AC power source. All components fit onto a portable cart. A stand is also available for the electrode junction box and electrical stimulation box. Patient data is stored to a encrypted database on the local hard drive or a remote server. The data can also be archived to other commercially available storage medias.

AI/ML Overview

This document is a 510(k) summary for the Nihon Kohden MEB-2300A Neuropack Evoked Potential and EMG Measuring System. It describes the device, its intended use, technical characteristics, and a comparison to a predicate device (MEB-2200A). The submission aims to demonstrate substantial equivalence to the predicate device.

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not present a formal table of acceptance criteria with corresponding performance results in the classical sense of an AI/ML device showing specific metrics like accuracy, sensitivity, or specificity against predefined thresholds for a task like disease detection.

Instead, the acceptance criteria are implicitly tied to demonstrating substantial equivalence to a predicate device (MEB-2200A Neuropack). The key performance aspects are the technical specifications and the device's ability to perform the same medical functions as the predicate.

The following table summarizes the comparison between the new device (MEB-2300A) and the predicate (MEB-2200A) at a technical specification level, which serves as the "reported device performance" in this context.

Feature / SpecificationMEB-2300A New Device Reported PerformanceMEB-2200A Predicate Device Performance
AMPLIFIERS:
Number of Channels6, 12 channels4, 8, 16 channels
Active Electrode jack4 portsNone
Input Impedance200 MΩ ±20% (Differential Mode), 1000 MΩ or greater (Common Mode)Same
Noise0.6 µVrmsSame
CMRR106 dB or greater (Differential mode), 112 dB or greater (Isolation mode)Same
Sensitivity1 to 500 µV/div and 1 to 10 mV/divSame
Low-cut Filter0.01 Hz to 3 KHz at 6 dB/octaveSame
High-cut Filter10 Hz to 20 KHz at 12 dB/octaveSame
Electrode Impedance Check2, 5, 10, 20, 50kΩ indicationSame
Amplitude Calibration1, 10, 100 μV, 1, 10mVSame
AC Filter50 or 60 HzSame
AVERAGERS:
A/D Converter18 bits16 bits
Conversion Speed5 µs/channel max.10 µs/channel max.
Monitor Time Base5 ms/div to 1 s/divSame
Analysis Time Base0.1 ms/div to 1 s/divSame
Time Base ModesIndividually selected for each channelSame
Delay Time-10 to +10 div in 1 div stepsSame
Trigger ModesRecurrent, Random, Foot Switch, Signal, EXT1/2/3/4/5/6, Somato1, Somato3Recurrent, Random, Foot Switch, Signal, EXT1/2/3/4 Somato1, Somato3, Trig1/2/3/4End
Number of Averages1 to 9999Same
Artifact Reject Inhibit Range±1 to ±5 div or OFFSame
DISPLAY:
DisplayColorSame
Number of Waveform Traces312 Maximum (26/channel x 12)416 Maximum (26/channel x 16)
Waveform Display ModesMonitor, Sweep, AnalysisSame
Cursors2 horizontal, 2 verticalSame
Scale5, 10, 15, 20 divSame
Grid (graticule) DisplayLine, Dot, OFFSame
STIMULATOR COMMON FUNCTIONS:
Stimulus ModesSingle, Double, Train, Train series (Multi train, Number of train: 1 to 10)Single, Double, Train
Stimulus Rates0.1 Hz to 100 HzSame
Delay Time0 to 10 secondsSame
ELECTRIC STIMULATORS:
Number of Outputs4Same
Stimulus Intensity0 to 200 mA0 to 100 mA
Stimulus Pulse Duration0.01 to 1 msSame
AUDITORY STIMULATORS:
Output TypeHeadphoneSame
Stimulus ModesClick, Tone BurstSame
Stimulus PhaseCondensation, Rarefaction, AlternatingSame
Stimulus Intensity0 to 135 dB SPLSame
Contralateral White Noise Masking0 to -40 dB, or OFFSame
Click Pulse Duration0.1 to 1 msSame
Tone Burst Frequency50 Hz to 10 KHzSame
Plateau Time0 to 1000 msSame
Rise/Fall Time0.1 to 10 msSame
VISUAL STIMULATORS:
Stimulus ModesPattern Reversal, Flash (with LED goggles), External Visual StimulationSame
PatternsCheckerboard, Horizontal Bars, Vertical BarsSame
Number of Horizontal Divisions4, 8, 16, 32, 64, 128Same
RECORDER:
Recording ModeHard Copy, Review, ReportSame
DIMENSIONS / WEIGHT:
Main Unit390(W) x 55(H) x 304(D) mm, 3.2 kg670(W) x 1320(H) x 800(D) mm, 14.3 kg
Electrode Junction Box180(W) x 56.5(H) x 234.5(D) mm, 1.5 kg(4/8 channels: 210x180x75mm, 1.8kg), (16 channels: 257x182x68mm, 2.4kg)
Power UnitMain Unit includes Power UnitSame
POWER REQUIREMENTS:
Line Voltage100-120 V AC, 220-240V AC117 V AC
Line Frequency50/60 HzSame
Power InputLess than 75 VALess than 450 VA
ENVIRONMENT:
Operating Temperature10 to 35 °CSame
Storage Temperature-20 to +65 °C-20 to +60 °C
Operating Humidity30 to 80%20 to 80%
Storage Humidity10 to 95%20 to 80%
Operating Atmospheric Pressure700 to 1060 hPaSame
Storage Atmospheric Pressure700 to 1060 hPaSame
TEST PROTOCOLS AVAILABLE:
Somatosensory evoked potentialSEP, SSEP, ECG-SSEP, ESCP, ELECTRICSame (only SEP listed explicitly)
Auditory evoked potentialABR, MLR, SVR, ECOCHG, AUDITORYSame (only ABR listed explicitly)
Visual evoked potentialPR-VEP, LED-VEP, EXT-VEP, ERG, EOG, VISUALSame (only PR-VEP listed explicitly)
EMG (electromyogram)EMG, EMG 2, QEMG, SF EMG, MACROEMG, QEMG, SF EMG, MACRO
Nerve ConductionMCS, SCS, NCS, REP.STIM, F-WAVE, H-REFLEX, BLINK, COLLISIONMCS, SCS, REP.STIM, F-WAVE, H-REFLEX, BLINK, COLLISION
Autonomic Nervous SystemMICRO-N, SSR, R-R IntervalSame (only MICRO-N listed explicitly)
Event related potentialsP-300, MRCP, CNVSame (only P-300 listed explicitly)
Trend MonitoringIOMSame

Key Differences and Rationale for Changes (Improved Performance/Functionality):

  • Number of Channels: Increased (e.g., 4 to 6, 8 to 12). Rationale: Increased channel capacity.
  • Active Electrode jack: 4 ports added. Rationale: Noise is highly reduced by active electrode system.
  • A/D Converter: Improved from 16 bits to 18 bits. Rationale: Higher resolution.
  • Conversion Speed: Improved from 10 µs/channel max. to 5 µs/channel max. Rationale: Higher sampling speed.
  • Stimulus Modes: "Train series" added. Rationale: New functional mode.
  • Trigger Modes: External trigger input increased (EXT1/2/3/4/5/6) for complex waveform acquisition. Rationale: Increased connectivity and functionality.
  • Stimulus Intensity (Electric): Increased from 0 to 100 mA to 0 to 200 mA. Rationale: More stimulus intensity, comparable to other devices (e.g., Medtronic NIM-Spine).
  • EMG: "EMG2" added (Measuring mode with up to 4 measurement settings). Rationale: More flexible measurement settings.
  • Nerve Conduction: MCS, SCS, and T-WAVE integrated into one program (NCS) for efficient routine examination. Rationale: Improved workflow efficiency.

2. Sample size used for the test set and the data provenance

The document explicitly states: "No clinical testing performed for this submission."

Therefore, there is no "test set" in the context of clinical data for performance evaluation against human experts or ground truth. The evaluation is primarily based on non-clinical engineering and bench testing to demonstrate compliance with standards and equivalence in specifications to the predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

As no clinical testing was performed and no test set was used to evaluate a clinical performance claim (e.g., diagnostic accuracy), there were no experts used to establish ground truth in this context. The "ground truth" for non-clinical testing typically refers to engineering specifications and measurements.

4. Adjudication method for the test set

Not applicable, as no clinical test set was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an EMG/EP measuring system, not an AI-powered diagnostic algorithm that assists human readers in interpreting cases. No clinical MRMC study or AI assistance evaluation was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a medical device for acquiring and displaying bioelectric signals, not an algorithm with standalone diagnostic performance. The device provides data for clinicians to interpret.

7. The type of ground truth used

For the non-clinical testing, the "ground truth" implicitly refers to the device's technical specifications as measured on a bench, and verification against established engineering and safety standards (e.g., IEC 60601 series, CAN/CSA-C22.2 series). The design control process itself dictates verification against requirements and specifications.

8. The sample size for the training set

Not applicable. This device is not an AI/ML algorithm that requires a training set. It is a hardware and software system for physiological signal acquisition.

9. How the ground truth for the training set was established

Not applicable, as no training set was used.

{0}------------------------------------------------

510(k) Summary

Special 510(k) Notification MEB-2300A Neuropack

K120397

Company Name:Nihon Kohden Corporation90 Icon StreetFoothill Ranch, CA 92610
Device Name:Nihon Kohden MEB-2300A Neuropack Evoked Potentialand EMG Measuring System
510(k) Sponsor, Contact:Nihon Kohden America, Inc.90 Icon StreetFoothill Ranch, CA 92610Steve GeerdesDirector Quality Assurance and Regulatory AffairsPhone: (949) 580-1555 Ext. 3325Fax: (949) 580-1550
Revised Summary Date:09/17/2012

Electroencephalograph amplifier (EEG Amplifier) Common Name:

Classification Names:

Stimulator, Electrical, Evoked Response882.1870GWF
--------------------------------------------------------

Predicate Device(s):

MEB-2200A Neuropack Evoked Potential andEMG Measuring SystemK991899
---------------------------------------------------------------------------

Description of Device 1.0

The MEB-2200A Neuropack Evoked Potential and EMG Measuring System consists of a main acquisition unit, operation control panel, foot switch, electrode junction box (head box), electrical stimulation box and commercially available computer, monitor, printer, keyboard, and mouse. The device also includes optional visual, and auditory stimulators. The monitor, keyboard, and mouse connect to the computer. The computer, electrode junction box, electrical stimulation box, and control panel connect to the main unit. All

{1}------------------------------------------------

components requiring AC power plug into the main unit's built-in, isolated power supply. The main unit plugs into a hospital grade AC power source. All components fit onto a portable cart. A stand is also available for the electrode junction box and electrical stimulation box. Patient data is stored to a encrypted database on the local hard drive or a remote server. The data can also be archived to other commercially available storage medias .

2.0 Intended Use of Device

The Nihon Kohden MEB-2300A Neuropack Evoked Potential and EMG Measuring System is intended to monitor, record and display the bioelectric signals produced by muscles, to stimulate peripheral nerves, and to monitor, record and display the electrical activity produced by nerves to aid the clinician in the diagnosis and prognosis of neuromuscular disease (EMG). The device is also intended to measure and display nerve conduction time by applying a stimulus to a patient's nerve (NCV). The device may use electrical stimulus, visual stimulus, or sound stimulus for use in evoked response measurements (EP). The device may be used to determine autonomic responses as psychological indicators by measuring the electrical resistance of the skin and the tissue path between two electrodes applied to the skin. The device may also measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head (EEG).

The device is intended for use by medical personnel within a hospital, laboratory, clinic or nursing home setting or outside of a medical facility under direct supervision of a medical professional. The device may also be placed in the intensive care unit or operating room for continuous recording.

The device is available for use on any patient as determined by the medical professional including adults and children.

3.0 Technical Characteristics

Setup for the device is performed by a medical professional in the graphical user interface environment of Windows XP or Windows 7. The operator enters patient identification data and specifies the testing protocol. The testing protocol specifies amplifier, acquisition, stimulation, and special parameters for recording data. Recording electrodes and/or sensors are attached to the patient and connected to the corresponding jacks on the junction box. Stimulation electrodes and/or accessories may also be connected. Skin-electrode impedance check is performed from the electrode junction box or the computer software to verify electrode connection and confirm impedance levels for non-invasive recording electrodes. Stimulation is activated to the patient if specified in the protocol.

The device provides 6 or 12 differential channels. Each channel has a pre-amplifier, post-amplifier, low pass filter, and high pass filter. Analog signals from the electrodes and/or sensors are transmitted to a preamplifier for conditioning. The signal output is then fed to the analog to digital converter (ADC).

{2}------------------------------------------------

The digitized electrophysiologic signals may be displayed on the monitor, printed, and/or stored to hard disk drives or other storage media. After completion of the study, the patient is disconnected from the device and medical professional prints and/or stores the data. The data can then be reviewed by a medical professional on the MEB-2300A/K system or another computer containing MEB-2300A/K review software.

4.0 Data Summary

  • . Non-clinical Testing Data Summary
    Testing of the Nihon Kohden MEB-2300A Neuropack Evoked Potential and EMG Measuring System was performed in compliance with Nihon Kohden Corporation design control process. Testing included:

Software and hardware verification and validation, and the device is in compliance with the following voluntary industrial standards:

IEC 60601-1Part1: General requirements for safety 1998-12
IEC 60601-1,Part 1: General Requirements for safety, Amendment 1,
Amendment 11991-11
IEC 60601-1,Part 1: General Requirements for safety, Amendment 2,
Amendment 21995-03
IEC 60601-1-12nd editionPart 1-1: General requirements for safety - Collateralstandard. Safety requirements for medical electricalsystems, 2000-12
IEC 60601-1-2Part 1-2: General requirements for safety - Collateral
2nd editionstandard. Electromagnetic compatibility, 2001-09
IEC 60601-1-2Part 1-2: General requirements for safety - Collateral
2nd edition,standard. Electromagnetic compatibility.
Amendment 1Amendment 1, 2004
IEC 60601-2-40Part 2-40: Particular Requirements for the safety ofelectromyographs and evoked response equipment, 1998-02
CAN/CSA-C22.2Medical electrical equipment, Part 1: General requirements
No. 601.1-M90for safety.
CAN/CSA-C22.2No. 601.1S1-94Supplement No. 1-94 to CAN/CSA-C22.2 No. 601-1-M90Medical Equipment- Part 1:General requirements forsafety.

{3}------------------------------------------------

.

CAN/CSA-C22.2No. 601.1B-90Amendment 2 to CAN/CSA-C22.2 No. 601.1-M90Medical equipment Part 1: General requirements forsafety: 2002
CAN/CSA-C22.2No. 60601-1-1-02Medical electrical equipment. Part 1-1: Generalrequirements for safety- Collateral: Safety requirements formedical electrical systems, 2006
CAN/CSA-C22.2No.60601-1-2-03Medical Electrical Equipment . Part 1-2: GeneralRequirements for Safety . Collateral Standard:Electromagnetic Compatibility . Requirements and Tests(Adopted IEC 60601-1-2:2001, second edition, 2001-09)
CAN/CSA-C22.2No. 60601-2-40-01Medical electrical equipment, Part 2-40: Particularrequirements for safety of electromyographs and evokedresponse equipment (adopted 60601-2-40: 1998)

Clinical Testing Data Summary .

No clinical testing performed for this submission.

Substantial equivalence comparison 5.0

5.1 Intended Use

MEB-2300A NEW DEVICEMEB-2200A PREDICATE (K991899)
1.Monitor, record and display bioelectric signals produced by muscles and nerves (EMG).1.Same
2.Measure, record and display nerve conduction time (NCV).2.Same
3.Determine autonomic responses by measuring electrical resistance of the skin and tissue path3.Same
4.Measure evoked response using electrical, visual or auditory stimulus (EP).4.Same
5.Measure, record and display the electrical activity of the patient's brain (EEG).5.Same

5.2 Physical Characteristics

:

MEB-2300A NEW DEVICEMEB-2200A PREDICATE (K991899)

.

{4}------------------------------------------------

.

1.Bio-potential signals are amplified, filtered and digitized in the input junction box and sent to the main unit.1.Same
2.The main unit controls stimulation triggers.2.Same
3.Waveforms are displayed on monitor and/or printed.3.Same
4.Waveforms may be permanently stored to storage media for future retrieval.4.Same

Target Population and Environment 5.3

.

MEB-2300A NEW DEVICE
1.Adults and children.
2.The device is intended for use by medical personnel within a hospital, laboratory, clinic or nursing home setting or outside of a medical facility under direct supervision of a medical professional. The device may also be placed in the intensive care unit or operating room for continuous recording.
MEB-2200A PREDICATE (K991899)
1.Same
2.Same

Specifications 5.4

MEB-2300A New DeviceMEB-2200A Predicate Device(K991899)
AMPLIFIERS:
Number of Channels6, 12 channels4, 8, 16 channels
Active Electrode jack4 portsNone
Input Impedance200 $MΩ$ ±20% (Differential Mode)1000 $MΩ$ or greater (CommonMode)Same
Noise0.6 $\mu$ VrmsSame
Common Mode RejectionRatio (CMRR)106 dB or greater(Differential mode)112 dB or greater (Isolation mode)Same
Sensitivity1 to 500 µV/div and 1 to 10 mV/divSame
Low-cut Filter0.01 Hz to 3 KHz at 6 dB/octaveSame
High-cut Filter10 Hz to 20 KHz at 12 dB/octaveSame
Electrode ImpedanceCheck2, 5, 10, 20, 50kΩ indicationSame
Amplitude Calibration1, 10, 100 μV, 1, 10mVSame
AC Filter50 or 60 HzSame
AVERAGERS:
A/D Converter18 bits16 bits
Conversion Speed5 µs/channel max.10 µs/channel max.
Monitor Time Base5 ms/div to 1 s/divSame
Analysis Time Base0.1 ms/div to 1 s/divSame
Time Base ModesIndividually selected for eachchannelSame
Delay Time-10 to +10 div in 1 div stepsSame
Trigger ModesRecurrent, Random, Foot Switch,Signal,EXT1/2/3/4/5/6, Somato1,Somato3Recurrent, Random, FootSwitch, Signal, EXT1/2/3/4Somato1,Somato3, Trig1/2/3/4End
Number of Averages1 to 9999Same
Artifact Reject InhibitRange±1 to ±5 div or OFFSame
DISPLAY:
DisplayColorSame
Number of WaveformTraces312 Maximum416 Maximum
Waveform Display ModesMonitor, Sweep, AnalysisSame
Cursors2 horizontal, 2 verticalSame
Scale5, 10, 15, 20 divSame
Grid (graticule) DisplayLine, Dot, OFFSame
STIMULATOR COMMONFUNCTIONS:
Stimulus ModesSingle, Double, Train, Train series(Multi train, Number of train: 1 to10)Single, Double, Train
Stimulus Rates0.1 Hz to 100 HzSame
Delay Time0 to 10 secondsSame
ELECTRICSTIMULATORS:
Number of Outputs4Same
Stimulus Intensity0 to 200 mA0 to 100 mA
Stimulus Pulse Duration0.01 to 1 msSame
AUDITORYSTIMULATORS:
Output TypeHeadphoneSame
Stimulus ModesClick, Tone BurstSame
Stimulus PhaseCondensation, Rarefaction,AlternatingSame
Stimulus Intensity0 to 135 dB SPLSame
Contralateral White NoiseMasking0 to -40 dB, or OFFSame
Click Pulse Duration0.1 to 1 msSame
Tone Burst Frequency50 Hz to 10 KHzSame
Plateau Time0 to 1000 msSame
Rise/Fall Time0.1 to 10 msSame
VISUAL STIMULATORS:
Stimulus ModesPattern Reversal, Flash (with LEDgoggles), External VisualStimulationSame
PatternsCheckerboard, Horizontal Bars,Vertical BarsSame
Number of HorizontalDivisions4, 8, 16, 32, 64, 128Same
RECORDER:
Recording ModeHard Copy, Review, ReportSame
DIMENSIONS / WEIGHT:
Main Unit390 (W) x 55 (H) x 304 (D) mm3.2 kg670 (W) x 1320 (H) x 800 (D) mm14.3 kg
Electrode Junction Box180 (W) x 56.5 (H) x 234.5 (D) mm1.5 kg4/8channels :210 (W) x 180 (H) x 75 (D) mm 1.8kg16 channels: 257(W) x 182(H) x 68D) mm 2.4kg
Power UnitThe Main Unit includes Power Unit.Same
POWERREQUIREMENTS:
Line Voltage100-120 V AC,220-240V AC117 V AC
Line Frequency50/60 HzSame
Power InputLess than 75 VALess than 450 VA
ENVIRONMENT:
Operating Temperature10 to 35 °CSame
Storage Temperature-20 to +65 °C-20 to +60 °C
Operating Humidity30 to 80%20 to 80%
Storage Humidity10 to 95%20 to 80%
Operating AtmosphericPressure700 to 1060 hPaSame
Storage AtmosphericPressure700 to 1060 hPaSame
TEST PROTOCOLSAVAILABLE:
Somatosensory evokedpotentialSEP (somatosensory evokedpotential)Same
SSEP (short-latencysmoatosensory evoked potential)
ECG-SSEP (ECG triggered SSEP)
ESCP (evoked spinal cordpotential)
ELECTRIC (customizable protocol)
Auditory evokedpotentialABR (auditory brainstem response)Same
MLR (middle latency response)
SVR (slow vertex response)
ECOCHG (eletrocochlegram)
AUDITORY (customizable protocol)
Visual evoked potentialPR-VEP (pattern reversal visualevoked potentialSame
LED-VEP (LED visual evokedpotential)
EXT-VEP (External stimulator visualevoked potential)
ERG (Electroretinogram)
EOG (Electroculogram)
VISUAL (customizable protocol)
EMG (electromyogram)EMG (electromyogram)EMG (electromyogram)
EMG 2(electromyogram2)QEMG (Quantitative EMG)
QEMG (Quantitative EMG)SF EMG (single fiber EMG)
SF EMG (single fiber EMG)MACRO (macro EMG)
Nerve Conduction
MCS (motor nerve conduction)MCS (motor nerve conduction)
SCS (sensory nerve conduction)SCS (sensory nerve conduction)
NCS (nerve conduction studies)REP.STIM (repetitivestimulation)
REP.STIM (repetitive stimulation)F-WAVE
F-WAVEH-REFLEX (monosynaptic reflexH-wave)
H-REFLEX (monosynaptic reflex H-wave)BLINK (blink reflex)
BLINK (blink reflex)COLLISION (collision method)
COLLISION (collision method)
Autonomic NervousSystemMICRO-N (microneurography ofsympathetic nerve activity)Same
SSR (Sympathetic skin response)R-R Interval
Event related potentialsP-300 (Long latency potential)Same
MRCP (Movement-related corticalpotential)
CNV (Contingent negative variation)
Trend MonitoringIOMSame

{5}------------------------------------------------

· ·

:

{6}------------------------------------------------

{7}------------------------------------------------

י :

{8}------------------------------------------------

.

.

.

: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

{9}------------------------------------------------

:

:

:

:

.

,

·

Differences 5.5

{10}------------------------------------------------

י

.

MEB-2300ANew DeviceMEB-2200APredicate DeviceRationale for Change
Number ofChannels6, 12 channels4,8,16 channelsNumber of channels ofeach electrode junction boxincreases as follows,- 4 to 6- 8 to 12- 16 to (32 pending, it isnot. available yet.)
ActiveElectrodejack4 portsNoneNoise is highly reduced byactive electrode system.
A/DConverter18 bits16 bitsHigher resolution
ConversionSpeed5 µs/channel max.10 µs/channel max.Higher sampling speed
Number ofWaveformTraces312 Maximum(26/channel x 12)416 Maximum(26/channel x 16)Number of waveform tracesper channel remains thesame.
StimulusModesSingle, Double, Train,Train series (Multi train,Number of train: 2 to 10)Single, Double, TrainTrain series is added.
TriggerModesRecurrent, Random,Foot Switch,Signal,EXT1/2/3/4/5/6,Somato1, Somato3Recurrent, Random,Foot Switch, Signal,EXT1/2/3/4 Somato1,Somato3,Trig1/2/3/4EndExternal trigger inputincreases and connectableexternal device alsoincreases. As a result,complex waveform can beacquired. Train series wasrequired but not available incurrent device. Therefore,"trig 1/2/3/4End" was usedas substitute for trainseries.
StimulusIntensity0 to 200 mA0 to 100 mAMore stimulus intensityNIM-Spine by Medtronic(K#031510) also output theintensity up to 200mA.
MEB-2300ANew DeviceMEB-2200APredicate DeviceRationale for Change
EMG(electromyogram)EMG (electromyogram)EMG2(electromyogram2)QEMG (QuantitativeEMG)SF EMG (single fiberEMG)MACRO (macro EMG)EMG (electromyogram)QEMGSF EMG (single fiberEMG)MACRO (macro EMG)EMG2: Measuring modewith up to 4 measurementsettings availableEMG: Only onemeasurement setting
NerveConductionMCS (motor nerveconduction)SCS (sensory nerveconduction)NCS (nerveconduction studies)REP.STIM (repetitivestimulation)F-WAVEH-REFLEX(monosynaptic reflexH-wave)BLINK (blink reflex)COLLISION (collisionmethod)MCS (motor nerveconduction)SCS (sensory nerveconduction)REP.STIM (repetitivestimulation)F-WAVEH-REFLEX(monosynaptic reflexH-wave)BLINK (blink reflex)COLLISION (collisionmethod)MCS, SCS, and T-WAVEare integrated in oneprogram as NCS forefficient routineexamination.

·

{11}------------------------------------------------

6.0 Conclusions

The safety and effectiveness of the Nihon Kohden MEB-2300 was demonstrated by testing in compliance with the Design Control process. The intended use and technology of the Nihon Kohden MEB-2300A is equivalent to the predicate devices MEB-200A. No new questions of safety or effectiveness are raised.

{12}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/12/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Nihon Kohden America, Inc. c/o Mr. Steve Geerdes Director of Regulatory Affairs and Ouality Assurance 90 Icon Street Foothill Ranch, CA 92610

SEP 19 2012

Re: K120397

Trade/Device Name: MEB-2300A Neuropack Evoked Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked Response Electrical Stimulator Regulatory Class: Class II Product Code: GWF, GWE, GWJ, GWQ, GZO, IKN Dated: September 4, 2012 Received: September 5, 2012

Dear Mr. Steve Geerdes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indicationsfor use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{13}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

n

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kesia Alexander

Malvina B. Eydelman, M.D.. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

Indications for Use

510(k) Number (if known): K120397

Device Name: MEB-2300A Neuropack Evoked Potential and EMG Measuring System

Indications For Use:

The Nihon Kohden MEB-2300A Neuropack Evoked Potential and EMG Measuring System is intended to monitor, record and display the bioelectric signals produced by muscles, to stimulate peripheral nerves, and to monitor, record and display the maooloo, to earning produced by nerves to aid the clinician in the diagnosis and prognosis of neuromuscular disease (EMG). The device is also intended to measure and display nene conduction time by applying a stimulus to a patient's nerve (NCV). The device may use electrical stimulus, visual stimulus, or sound stimulus for use in evoked response measurements (EP). The device may be used to determine autonomic responses as psychological indicators by measuring the electrical resistance of the skin and the tissue path between two electrodes applied to the skin. The device may also and the tlood pain boom betwity of the patient's brain obtained by placing two or more electrodes on the head (EEG).

The device is intended for' use by medical personnel within a hospital, laboratory, clinic or nursing home setting or outside of a medical facility under direct supervision of a or naroing nome oot... The device may also be placed in the intensive care unit or operating room for continuous recording.

The device is available for use on any patient as determined by the medical professional including adults and children.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

John Grimes, Ph.D.

(Division Sign-Off) (Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K120397

Page 1 of 1

§ 882.1870 Evoked response electrical stimulator.

(a)
Identification. An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.(b)
Classification. Class II (performance standards).