K Number
K120377

Validate with FDA (Live)

Date Cleared
2012-07-25

(170 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Device is intended for the incision, excision, vaporization, ablation and coagulation of soft tissue in oral and dentistry and for the ablation and vaporization of hard tissue in dentistry.

Hard Tissue
General Indications*
· Class I, II, III, IV and V cavity preparation
· Caries removal
· Hard tissue surface roughening or etching
· Enameloplasty, excavation of pits and fissures for placement of sealants

  • For use on adult and pediatric patients

Root Canal Hard Tissue Indications
· Tooth preparation to obtain access to root canal
· Root canal preparation including enlargement
· Root canal debridement and cleaning

Bone Surgical Indications
· Cutting, shaving, contouring and resection of oral osseous tissues (bone)
· Osteotomy

Endodontic Surgery (Root Amputation) Indications
· Flap preparation - incision of soft tissue to prepare a flap and expose the bone.
· Cutting bone to prepare a window access to the apex (apices) of the root(s).
· Apicoectomy - amputation of the root end.
· Root end preparation for retrofill amalgam or composite.
· Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex.
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.

Laser Periodontal Procedures
· Full thickness flap
· Partial thickness flap
· Split thickness flap
· Laser soft tissue curettage
· Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
· Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium
· Removal of granulation tissue from bony defects
· Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
· Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
· Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
· Osseous crown lengthening
· Removal of subgingival calculus

Soft Tissue Indications including Pulpal Tissues*
Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including:
· Excisional and incisional biopsies
· Exposure of unerupted teeth
· Fibroma removal
· Flap preparation incision of soft tissue to prepare a flap and expose the bone.
· Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions)
· Frenectomy and frenotomy
· Gingival troughing for crown impressions
· Gingivectomy
· Gingivoplasty
· Gingival incision and excision
· Hemostasis and coagulation
· Implant recovery
· Incision and drainage of abscesses
· Incision and drainage of periapical abscesses
· Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery
· Leukoplakia
· Operculectomy
· Oral papillectomies
· Pulpotomy
· Pulp extirpation
· Pulpotomy as an adjunct to root canal therapy
· Reduction of gingival hypertrophy
· Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and hyperplastic tissues (i.e. granulation tissue) from around the apex .
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
· Root canal debridement and cleaning
· Soft tissue crown lengthening
· Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
· Vestibuloplasty

  • For use on adult and pediatric patients
Device Description

Equipment is mobile Er: YAG Medical Treatment and Coagulation Laser of model MEY-1-A intended to be used for dental surgery in hospital environment. Er.YAG Laser emits an infrared beam with a wavelength 2.94 um which is readily absorbed by water contained by both hard and soft tissues of the human body. As a result, a energy of the laser beam instantly vaporized the water molecules in hard tissues of the tooth causing the tissues to crumble away or resection of the soft tissues of gingival.

AI/ML Overview

This document is a 510(k) summary for the AdvErL EVO Er: YAG Laser for Dentistry (Model: MEY-1-A). It focuses on establishing substantial equivalence to predicate devices rather than presenting a performance study with acceptance criteria in the typical sense for diagnostic accuracy.

The performance of the MEY-1-A is examined by using international standards for medical electrical equipment and laser products to ensure safety and essential performance. The document does not describe a clinical study comparing the device's efficacy against specific acceptance criteria for a diagnostic or therapeutic outcome. Instead, it relies on demonstrating that the device has the "substantially same equivalent technical specifications" and "intended use and Indications of use" as its predicate devices, implying that if the predicates met their performance criteria, this device would as well.

Here's an breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Safety Standards:
IEC 60601-1:2005 (General requirements for basic safety and essential performance)The device's performance was examined using this standard. (Implied compliance for substantial equivalence)
IEC 60601-2-22 (Particular requirements for safety of diagnostic and therapeutic laser equipment)The device's performance was examined using this standard. (Implied compliance for substantial equivalence)
IEC 60825-1 (Safety of laser products-Part 1: Equipment classification and requirements)The device's performance was examined using this standard. (Implied compliance for substantial equivalence)
IEC 60601-1-2:2007 (Electromagnetic compatibility - equipments and tests Edition 3)The device's performance was examined using this standard. (Implied compliance for substantial equivalence)
Intended Use & Indications:
Identical intended use and indications of use as predicate device I ("VERSAWAVE DENTAL ER: YAG LASER SYSTEM (K#041710)")The MEY-1-A has the same "intended use and Indications of use" as predicate device I, except for "Removal of subgingival calculus."
Substantially similar intended use and indications of use as predicate device II ("WATERLASE MD TURBO PLUS MODEL 7200XXX (K#101658)")The MEY-1-A has substantially similar "intended use and Indications of use" as predicate device II, but lacks the indication of "Root canal indications / Laser root canal disinfection after endodontic."
Technical Specifications:
Substantially same "equivalent technical specifications" as predicate device I and IIThe MEY-1-A has "substantially same 'equivalent technical specifications'" compared with both predicate devices. Minor differences in display, power calibration, and software are considered not to cause substantial impacts to effectiveness or safety.

2. Sample size used for the test set and the data provenance

The document does not describe a clinical performance study with a test set of data in the context of diagnostic or therapeutic accuracy, but rather compliance with safety and electrical standards. Therefore, information on sample size and data provenance for a test set is not applicable here. The "test set" in this context refers to the device itself undergoing engineering and safety testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is not a study requiring expert-established ground truth for a diagnostic or imaging performance metric.

4. Adjudication method for the test set

Not applicable. No clinical adjudication method is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a laser device, not an AI-assisted diagnostic or imaging device, and no MRMC study is mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm, but a physical medical device (laser).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this submission is based on adherence to recognized international safety and performance standards (IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2) and a demonstration of substantial equivalence in intended use and technical specifications to previously cleared predicate devices.

8. The sample size for the training set

Not applicable. This device does not use an algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

K120377
pg 1 of 4

JUL 2 5 2012

510(k) summary

Company name and address :

Date of preparation: Contact:

J.MORITA MFG. CORP. 680 Higashihama Minami cho Fushimi-ku, Kyoto, Japan Telephone: +81-75-611-2141 Facsimile: +81-75-605-2354 April 10, 2012 Yoshihide Okagami · General Manager of Technical Administration

and for the ablation and vaporization of hard tissue in

1 . NAME OF DEVICE and Intended Use/Indications of Use

Trade or Proprietary Name:AdvErL EVO Er: YAG Laser for Dentistry
Model: MEY-1-A
Note: The name is referred to as " MEY-1-A" hereafter.
Common Name:Dental medical laser equipment
Usual Name:Medical laser system
Classification Name:Laser surgical instrument for use in general andplastic surgery and in dermatology(21CFR878.4810)
Product Code :GEX
Intended Use :Device is intended for the incision, excision, vaporization,ablation and coagulation of soft tissue in oral and dentistry

Predicate devices and Indications of use:

The MEY-1-A is substantially equivalent to both predicate devices of "VERSAWAVE DENTAL ER: YAG LASER SYSTEM ( K#041710) " made by HOYA CONBIO and WATERLASE MD TURBO PLUS MODEL 7200XXX( K#101658)" made by BIOLASE TECHNOLOGY, INC.

dentistry.

The indications of use of the MEY-1-A covers all identical indications of " VERSAWAVE DENTAL ER:YAG LASER SYSTEM ( K#041710 ) " , however, lacks the indications of "Root canal indications / Laser root canal disinfection after endodontic" which is included in the indications of "WATERLASE MD TURBO PLUS MODEL 7200XXX( K#101658)

As for details of indications refer to FOOT NOTE of this summary ..

ESTABLISHMENT REGISTRATION NUMBER 2.

Registration No. 2081055Registration No. 3002807636
Initial Distributor:Manufacturer:
J. MORITA USA, INC.J. MORITA MFG. CORP.
9 Mason680 Higashihama Minami-cho
Irvine, CA 92618Fushimi-ku, Kyoto
USAJapan 612-8533
Telephone: 949-581-9600+81-75-611-2141
Facsimile: 949-581-9688+81-75-605-2354

283

{1}------------------------------------------------

K120377
pg 2 of 4

  1. DEVICE CLASSIFICATION / CLASSIFICATION PANEL Device classification: Class 2 medical devise 21CFR878,4810 Laser surgical instrument for use in general and plastic surgery and in dermatology Laser classification: Class 4 medical laser product Per 21CFR 1040.10 and per 21CFR 1040.11 Device classification panel: 76 Dental

4. PERFORMANCE STANDARDS

We examine the performance of the MEY-1-A by using the international standards:

IEC 60601-1:2005Medical electrical equipment Part 1: General requirements for basicsafety and essential performance
IEC 60601-2-22Medical electrical equipment Part 2: Particular requirements for safety of diagnostic and therapeutic laser equipment
IEC 60825-1Safety of laser products-Part 1: Equipment classification and requirements
IEC 60601-1-2:2007Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility -equipments and tests Edition 3)

DEVICE DESCRIPTION and SUBSTANTIAL EQUIVALENCE 5.

A. DEVICE DESCRIPTION

General Product Information

Equipment is mobile Er: YAG Medical Treatment and Coagulation Laser of model MEY-1-A intended to be used for dental surgery in hospital environment.

Er.YAG Laser emits an infrared beam with a wavelength 2.94 um which is readily absorbed by water contained by both hard and soft tissues of the human body. As a result, a energy of the laser beam instantly vaporized the water molecules in hard tissues of the tooth causing the tissues to crumble away or resection of the soft tissues of gingival.

B. SUBSTANTIAL EQUIVALENCE

Company's proposal for establishing substantial equivalence of the MEY-1-A by demonstrating Comparison Table where the MEY-1-A is compared with two different predicate devices.

As a result of this comparison, the MEY-1-A has not only the same " intended use and Indications of use", but also has the substantially same " equivalent technical specifications" compared with the predicate device I ( Versa Wave Dental Er: YAG Laser System K#041710 ) and the predicate device II (WaterLase MD Turbo Plus MODEL 7200XXX : K#101658 }.

There are some differences in respects of display, power calibration and software, however, these are esteemed to be minor differences since we presume that they use similar existing technology with ours to realize their functions in each design stage, so that such slight difference would not cause substantial impacts to effectiveness or safety problems.

284

{2}------------------------------------------------

FOOT NOTE:

Indications for Use Statement

AdvErL EVO Er: YAG Laser for Dentistry Device Name: Model: MEY-1-A

Indications for Use:

Hard Tissue

General Indications*

  • · Class I, II, III, IV and V cavity preparation
  • · Caries removal
  • · Hard tissue surface roughening or etching
  • · Enameloplasty, excavation of pits and fissures for placement of sealants
    • For use on adult and pediatric patients

Root Canal Hard Tissue Indications

  • · Tooth preparation to obtain access to root canal
  • · Root canal preparation including enlargement
  • · Root canal debridement and cleaning

Bone Surgical Indications

  • · Cutting, shaving, contouring and resection of oral osseous tissues (bone)
  • · Osteotomy

Endodontic Surgery (Root Amputation) Indications

  • · Flap preparation incision of soft tissue to prepare a flap and expose the bone.
  • · Cutting bone to prepare a window access to the apex (apices) of the root(s).
  • · Apicoectomy amputation of the root end.
  • · Root end preparation for retrofill amalgam or composite.

· Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues(i.e., granulation tissue) from around the apex.

NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.

Laser Periodontal Procedures

  • · Full thickness flap
  • · Partial thickness flap
  • · Split thickness flap
  • · Laser soft tissue curettage

· Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket

· Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium

· Removal of granulation tissue from bony defects

· Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)

· Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)

· Ostectorny (resection of bone to restore bony architecture, resection of bone for grafting,

285

{3}------------------------------------------------

K120377
pg 4 of 4

etc.)

  • · Osseous crown lengthening
    · Removal of subgingival

Soft Tissue Indications including Pulpal Tissues*

Incision, excision, vaporization, ablation and coaqulation of oral soft tissues, including:

  • · Excisional and incisional biopsies
  • · Exposure of unerupted teeth
  • Fibroma removal

· Flap preparation - incision of soft tissue to prepare a flap and expose the bone.

  • · Flap preparation incision of soft tissue to prepare a fiap and expose unerupted teeth (hard and soft tissue impactions)
    · Frenectomy and frenotomy

  • · Gingival troughing for crown impressions

  • · Gingivectomv

  • · Gingivoplasty

  • · Gingival incision and excision

  • · Hemostasis and coagulation

  • · Implant recoverv

  • · Incision and drainage of abscesses

  • · Incision and drainage of periapical abscesses

  • · Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area

during apical surgerv

  • · Leukoplakia
  • · Operculectomy
  • Oral papillectomies
  • · Pulpotomy
  • · Pulp extirpation
  • · Pulpotomy as an adjunct to root canal therapy
  • · Reduction of gingival hypertrophy
  • · Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and hyperplastic tissues (i.e. granulation tissue) from around the apex .
  • NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
  • · Root canal debridement and cleaning
  • · Soft tissue crown lengthening
  • · Treatment of canker sores, herpe tic and aphthous ulcers of the oral mucosa
  • · Vestibuloplastv
    • For use on adult and pediatric patients

N.B.

    1. The abovementioned indications of use of the MEY-1-A is same to all the terms of " VERSAWAVE DENTAL ER:YAG LASER SYSTEM (K#041710)" except " Removal of subgingival calculus".
    1. On the other hand, the indications of use of the MEY-1-A lacks the indication of "Root canal indications / Laser root canal disinfection after endodontic" which is included in the indications of "WATERLASE MD TURBO PLUS MODEL 7200XXX( K#101658)

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized eagle with its wings outstretched, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

J. Morita USA, Incorporated % Fish & Richardson P.C. Mr. Keith A. Barritt 1425 K. Street Northwest Suite 1100 Washington, District of Columbia 20005

JUL 25 2012

Re: K120377

Trade/Device Name: AdvErl EVO ER: YAG laser for Dentistry Model: MEY-1-A Regulation Number: 21 CFR 878.4810

· Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology

Regulatory Class: Class II

Product Code: GEX Dated: July 13, 2012 Received: July 16, 2012

Dear Mr. Barritt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{5}------------------------------------------------

Page 2- Mr. Keith A. Barritt

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Eunil Keith

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Device Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use statement

510(K) Number: K120377

AdvErL EVO Er: YAG Laser for Dentistry Device Name:

Model: MEY-1-A

Indications for Use:

Hard Tissue

General Indications*

· Class I, II, III; IV and V cavity preparation

· Caries removal

· Hard tissue surface roughening or etching

· Enameloplasty, excavation of pits and fissures for placement of sealants

  • For use on adult and pediatric patients

Root Canal Hard Tissue Indications

· Tooth preparation to obtain access to root canal

· Root canal preparation including enlargement

· Root canal debridement and cleaning

Bone Surgical Indications

· Cutting, shaving, contouring and resection of oral osseous tissues (bone)

· Osteotomy

Endodontic Surgery (Root Amputation) Indications

· Flap preparation - incision of soft tissue to prepare a flap and expose the bone.

· Cutting bone to prepare a window access to the apex (apices) of the root(s).

· Apicoectomy - amputation of the root end.

· Root end preparation for retrofill amalgam or composite.

· Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex.

NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation. (Division Sign-Off)

Laser Periodontal Procedures

· Full thickness flap

· Partial thickness flap

· Split thickness flap

· Laser soft tissue curettage

· Laser removal of diseased, inflamed and necrosed soft tissue within

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number

for
man

K12037

{7}------------------------------------------------

periodontal pocket

· Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium

· Removal of granulation tissue from bony defects

· Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)

· Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiological osseous contours)

· Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)

  • · Osseous crown lengthening
  • · Removal of subgingival calculus

Soft Tissue Indications including Pulpal Tissues*

Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including:

  • · Excisional and incisional biopsies
  • · Exposure of unerupted teeth
  • · Fibroma removal
  • · Flap preparation incision of soft tissue to prepare a flap and expose the bone.
  • · Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth
  • (hard and soft tissue impactions)
  • · Frenectomy and frenotomy
  • · Gingival troughing for crown impressions
  • · Gingivectomy
  • · Gingivoplasty
  • · Gingival incision and excision
  • Hemostasis and coagulation
  • · Implant recovery
  • · Incision and drainage of abscesses
  • · Incision and drainage of periapical abscesses
  • · Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery
  • · Leukoplakia
  • · Operculectomy
  • · Oral papillectomies
  • · Pulpotomy

Nailk Qal for mkn
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

290

510(k) Number K120377

{8}------------------------------------------------

  • · Pulp extirpation
    • Pulpotomy as an adjunct to root canal therapy

· Reduction of gingival hypertrophy

• Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and hyperplastic tissues (i.e. granulation tissue) from around the apex .

NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.

• Root canal debridement and cleaning

· Soft tissue crown lengthening

· Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa

· Vestibuloplasty

  • For use on adult and pediatric patients

Nekrp Ogden forman

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K120377

291

{9}------------------------------------------------

Prescription Use_X (Part21CFR801 Subpart D)

AND/OR

Over The Counter Use (Part21CFR807 Subpart C)

Concurrence of CDRH, Office of Device Evaluation(ODE)

(Division Sign-Off)

. .

Division of Surgical, Orthopedic, and Restorative Devices.

510(K) Number: K120377

Neil RPDgden forman

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K120377

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.