K Number
K120375

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-06-28

(143 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MED-RX® Patient Delivery Set is to be used in the injection of contrast agents and flushing solutions for the purpose of enhancing diagnostic images.

Device Description

The MED-RX® Patient Delivery Sets each consist of a male luer lock, two female luer locks, 3 tubing pieces, and either a dual check valve, a single check valve, or a Y-connector. The tube is made of polyvinyl chloride (PVC). The total length of the sets varies from 18″ to 86″ and is also available in a coiled configuration. The MED-RX® Patient Delivery Sets are provided sterile and are not to be resterilized.

AI/ML Overview

This document describes the design and testing of a medical device, the MED-RX® Patient Delivery Set, but does not describe an AI/ML device. The information requested, particularly regarding acceptance criteria and studies proving the device meets those criteria, is typically provided for AI/ML devices to demonstrate their performance against defined metrics and ground truth.

Therefore, I cannot provide the requested information in the format of acceptance criteria and a study that proves the device meets those criteria from the provided text, as it pertains to a physical medical device and not an AI/ML system.

However, I can extract the non-clinical test summary which details performance tests:

1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical Test Summary)

TestStandardAcceptance Criteria (Implied by standard)Reported Device Performance
Tensile StrengthISO 8536-4:2010Withstand 15 N for 15 SecondsWithstand 15 N for 15 Seconds, Pass
Resistance to LeakageAs per Benlan internal requirements(Not specified, but implied "Pass")Pass
Particulate ContaminationISO 8536-4:2010Meet contamination index limitSamples met contamination index limit
Chemical RequirementsISO 8536-4:2007 PER Clause 5 & 7(Not specified, but implied "Pass")Pass
Natural Rubber Latex ContentModified Lowry MethodNot contain natural rubber latexDevice does not contain natural rubber latex
Method of Validation (Sterilization)ANSI/AAMI/ISO 11135-1: 2007Achieve a Sterility Assurance Level of 1 x 10^-6Validated to a Sterility Assurance Level of 1 x 10^-6
EO Sterilization ResidualsISO 10993-7: 2008(Not specified, but implied "Pass")Pass
Bacterial EndotoxinsCurrent edition of ANSI/AAMI ST72(Not specified, but implied "Pass")Pass
Hemolysis AssayASTM F-756-00Non-hemolyticProduct code 10-1254T is considered non-hemolytic and passes the test.
Acute Systemic Injection TestISO 10993-11Meet ISO Acute Systemic Injection Test requirementsThe findings indicate that the requirements of the ISO Acute Systemic Injection Test have been met.
Rabbit Pyrogen TestUSP 32:2009 <151>Non-pyrogenicProduct code 10-1254T was considered non-pyrogenic.
Intracutaneous Reactivity TestISO 10993-10:2002Meet ISO Intracutaneous Reactivity Test requirementsThe requirements of the ISO Intracutaneous Reactivity Test have been met.
Guinea Pig Maximization Sensitization TestISO 10993-10:2002No sensitization responseProduct code 10-1254T did not elicit a sensitization response.
CytotoxicityISO 10993-5: 2009 ISO MEM Elution with L-929 Mouse Fibroblast CellsNon-cytotoxicProduct code 10-1254T is considered non-cytotoxic.

Regarding the specific questions about AI/ML device studies:

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • This is not an AI/ML device; therefore, no test set of data (e.g., images, patient records) is described. The "samples" referred to in the non-clinical tests are physical units of the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not applicable as this is not an AI/ML device.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable as this is not an AI/ML device.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable as this is not an AI/ML device.

6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

  • Not applicable as this is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the physical device, the "ground truth" is established by compliance with the specified international and internal standards (e.g., ISO, ASTM, USP) for physical and biological characteristics.

8. The sample size for the training set

  • Not applicable as this is not an AI/ML device.

9. How the ground truth for the training set was established

  • Not applicable as this is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Benlan MED-RX Patient Delivery Sets. The logo features a stylized drop shape to the left of the word "Benlan" in bold, followed by "MED-RX Patient Delivery Sets" in a smaller font size. The text is black against a white background.

SECTION 5 - 510(K) SUMMARY

Date of Summary: February 3, 2012

JUN 2 8 2012

akville. ON LGH !

Tel (905) 829-5004 Fax (905) 829-5006

Official Contact:Cheryl Brown - QA / RA Manager
Proprietary Name:Common Name:MED-RX® Patient Delivery SetIntravascular Administration Set
Classification Name:Intravascular Administration Set, 880.5440 Intravascular AdministrationSet.
Class:Class II
Product Code:FPA
Predicate Device:Liebel-Flarsheim Optistar MR Injector System→ Y-Tubing Set (K984088)

Device Description

The MED-RX® Patient Delivery Sets each consist of a male luer lock, two female luer locks, 3 tubing pieces, and either a dual check valve, a single check valve, or a Y-connector. The tube is made of polyvinyl chloride (PVC). The total length of the sets varies from 18″ to 86″ and is also available in a coiled configuration. The MED-RX® Patient Delivery Sets are provided sterile and are not to be resterilized.

Indications for Use

The MED-RX® Patient Delivery Set is to be used in the injection of contrast agents and flushing solutions for the purpose of enhancing diagnostic images.

Substantial Equivalence

The information provided in the premarket notification demonstrates that the proposed device is substantially equivalent to legally marketed devices. The proposed MED-RX® Patient Delivery Sets are substantially equivalent to the Liebel-Flarsheim Optistar Y-Tubing Set (K984088). Both devices have the same intended use for injection of contrast agents and flushing solutions for the purpose of enhancing diagnostic images. The product configurations are identical between the proposed device and the predicate.

NOTE: The predicate device upon which to determine substantial equivalence is the Ytubing set included as a sterile disposable in K984088.

A comparison of features and principles of operation between the proposed device and predicate device is provided in Table 1 below.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Benlan MED-RX Patient Delivery Sets. The logo features a stylized droplet shape on the left, followed by the word "Benlan" in bold, sans-serif font. Below "Benlan" is the text "MED-RX Patient Delivery Sets", with "MED-RX" in a slightly larger font size than "Patient Delivery Sets".

Table 1: Comparison between MED-RX® Patient Delivery Set and Liebel-Flarsheim Optistar Y-Tubing Set (K984088

ATTRIBUTEPROPOSED DEVICE - MED-RX®Patient Delivery SetPREDICATE DEVICE - Liebel- Flarsheim Optistar Y-Tubing Set (K984088)
General Indications
Indications for UseThe MED-RX® Patient Delivery Setis to be used in the injection ofcontrast agents and flushingsolutions for the purpose ofenhancing diagnostic images.The Optistar MR is designed toinject MR contrast agents andflushing solutions for the purposeof enhancing diagnostic images ofhumans.
DisposableYes .Yes
PrescriptionYesYes
Non-PyrogenicYesYes
Intended Environment of UseHospitalSame
Material Composition
TubingPolyvinyl Chloride - DEHP freeMedical grade tubing
ComponentsMedical grade plastics (PVC, LDPE,HDPE, Polycarbonate)Medical grade plastics
Packaging and Sterilization
SterileYesYes
Sterilization MethodEthylene Oxide (EO).Same
Packaging ConfigurationMedical grade paper/film pouchSame
Physical Specifications
Tubing outer diameter (OD)0.156"0.130"
Tubing inner diameter (ID)0.060"0.060"
Total set length18" - 86"66"
Design Features
Distal ConfigurationMale Luer LockSame
Proximal ConfigurationTwo Female Luer LocksSame
Y-SiteDual Check Valve OR Single CheckValve OR Y-ConnectorDual Check Valve OR Single CheckValve OR Y-Connector
Y-Leg Tube ConfigurationStraightStraight
Long Arm Tube ConfigurationStraight OR CoiledCoiled
Maximum Pressure Rating400 PSISame
CapsLuer capsSame

Summary of Differences

There are no significant differences between the proposed MED-RX® Patient Delivery Sets and the predicate device, Liebel-Flarsheim Optistar Y-Tubing Set (K984088). Similarities between the proposed device and the predicate device include indications for use and design configuration. The MED-RX® Patient Delivery Sets and the Liebel-Flarsheim Optistar Y-Tubing Set are sterile, disposable devices, packaged in paper/film pouches and sterilized using ethylene oxide.

Both the proposed MED-RX® Patient Delivery Sets and the predicate device feature two female luer locks on the proximal end of the set, followed by two short Y-leg tubes that meet at either a

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Benlan MED-RX Patient Delivery Sets. The logo features a stylized drop-like shape to the left of the word "Benlan" in a bold, sans-serif font. Above the "lan" portion of the name is a small "TM" symbol, indicating a trademark. Below the name is the text "MED-RX Patient Delivery Sets" in a smaller, sans-serif font.

dual check valve, a single check valve, or a Y-connector. The Y-site is followed by one long arm tube that terminates with a male luer lock. Protective luer caps are provided for all luer locks. All proximal and distal components are compatible with standard luer devices, similar to the predicate device. The predicate Liebel-Flarsheim Optistar Y-Tubing Set is available in identical configurations to the predicate device.

Any minor differences between the proposed device and the predicate have been evaluated to have no impact on safety or effectiveness of the MED-RX® Patient Delivery Sets. Therefore the proposed device can be considered substantially equivalent to legally market devices and raises no new issues of safety and effectiveness.

Non-Clinical Test Summary

The MED-RX® Patient Delivery Sets were subject to numerous performance tests including tensile strength, resistance to leakage, particulate contamination, and for chemical requirements. The MED-RX® Patient Delivery Sets have successfully completed all required performance testing following the applicable guidelines of ISO 8536-4: 2010. The MED-RX® Patient Delivery Sets were also tested for natural rubber latex content. Please refer to Table 2.

TestStandardResults
Tensile StrengthISO 8536-4:2010Withstand 15 N for 15 Seconds, Pass
Resistance to LeakageAs per Benlan internalrequirementsPass
Particulate ContaminationISO 8536-4:2010Samples met contamination indexlimit
Chemical RequirementsISO 8536-4:2007 PERClause 5 & 7Pass
Natural Rubber Latex ContentModified Lowry MethodDevice does not contain naturalrubber latex

Table 2: Non-Clinical Test Summary

Summary of Sterilization

Each MED-RX® Patient Delivery Set is individually packaged using a medical grade breathable coated paper heat sealed to a polypropylene medical grade film and sterilized using ethylene oxide. Please see Table 3 for a summary.

2760 Brighton Rd. Oakville ON L6H 5T4 www.benlan.com

{3}------------------------------------------------

Sets

Table 3: Sterilization Summary

Test DescriptionStandardResults
Method of ValidationANSI/AAMI/ISO 11135-1: 2007Validated to a SterilityAssurance Level of 1 x 10-6
EO Sterilization ResidualsISO 10993-7: 2008Pass
Bacterial EndotoxinsCurrent edition of ANSI/AAMI ST72Pass

Summary of Biocompatibility Tests

Biocompatibility testing was successfully completed on sterile finished devices. The MED-RX® Patient Delivery Sets are classified as external communicating devices with limited to prolonged use, indirect blood path contact. A summary of the testing completed and the relevant standards are listed in Table 4.

Table 4: Biocompatibility Test Summary

StandardTest DescriptionResults
ASTM F-756-00Hemolysis Assay - Extract MethodProduct code 10-1254T isconsidered non-hemolytic andpasses the test.
ISO 10993-11Acute Systemic Injection TestThe findings indicate that therequirements of the ISO AcuteSystemic Injection Test have beenmet.
USP 32:2009 <151>Materials Mediated Rabbit Pyrogen TestProduct code 10-1254T wasconsidered non-pyrogenic.
ISO 10993-10:2002Intracutaneous Reactivity TestThe requirements of the ISOIntracutaneous Reactivity Test havebeen met.
ISO 10993-10:2002Guinea Pig Maximization SensitizationTestProduct code 10-1254T did not elicita sensitization response.
ISO 10993-5: 2009ISO MEM Elution with L-929 MouseFibroblast Cells (Cytotoxicity)Product code 10-1254T isconsidered non-cytotoxic.

Benla

r 905 829 5004 2760 Brighton Rd. Ookville OR L6H 5T4 www.benlan.com

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Cheryl Brown Quality Assurance/Regulatory Affairs Manager Benlan, Incorporated 2760 Brighton Road Oakville, Canada L6H 5T4

JUN 2 8 2012

Re: K120375

Trade/Device Name: MED-RX® Patient Delivery Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: June 15, 2012 Received: June 20, 2012

Dear Ms. Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Ms. Brown

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Fd Shef
SyrohitnD

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Benlan MED-RX Patient Delivery Sets. The logo features a stylized letter "O" to the left of the word "Benlan" in a bold, sans-serif font. Below "Benlan" is the text "MED-RX Patient Delivery Sets," with "MED-RX" in a slightly larger font size and a trademark symbol next to it.

SECTION 4 - INDICATIONS FOR USE

510(K) Number (If Known): ヒ (2003 フ S

Device Name: MED-RX® Patient Delivery Set

Indications For Use:

The MED-RX® Patient Delivery Set is to be used in the injection of contrast agents and flushing solutions for the purpose of enhancing diagnostic images.

Prescription Use:
(Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use
(21 CFR 801 Subpart C)
N/A

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mli C. 6/28/12

Page 1 of 1

(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices

510(k) Number: K120373

4-1

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.