K Number
K120356

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-12-11

(309 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NCS Fish-Fit MD System is intended for the attachment of soft tissue to bone during arthroscopic or open treatment of rotator cuff tendons lesions with the purpose of repairing the rotator cuff (supraspinatus, infraspinatus and subscapularis).

Device Description

NCS Fish-Fit MD System is an implantable anchor made from Titanium grade 4. Sutures are not pre-assembled with bone anchor. The system includes an instrumentation package, Compass MD. The instrument package is intended to facilitate implantation of the anchor.

AI/ML Overview

The provided 510(k) summary for the NCS Fish-Fit MD System is for a medical device (an implantable suture anchor) and not for a software or AI/ML-based device. Therefore, the questions related to clinical studies, ground truth, expert adjudication, MRMC studies, and training sets are not applicable to this submission.

The acceptance criteria and supporting study described are for the mechanical performance of the device, comparing it to predicate devices.

Here's the information that can be extracted from the provided text, adapted for the nature of this device:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Adequate fixation strength (comparable to predicate devices)Non-clinical testing performed to verify adequate fixation strength. Data provided to show that NCS Fish-Fit MD System and Bone Anchor perform as well as the predicate device with respect to fixation strength.
Adequate pull-out strength (comparable to predicate devices)Non-clinical testing performed to verify adequate pull-out performances. Data provided to show that NCS Fish-Fit MD System and Bone Anchor perform as well as the predicate device with respect to pull-out strength.

2. Sample size used for the test set and the data provenance

  • Test Set: The document refers to "non-clinical testing" and "data have been provided," but does not explicitly state the sample size (e.g., number of anchors tested) or the provenance (e.g., in vitro, ex vivo). It implies laboratory-based mechanical testing.
  • Data Provenance: Non-clinical (likely in-vitro mechanical testing).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. This is a mechanical device, and the "ground truth" is measured physical properties like strength, not a diagnostic interpretation. The testing likely conformed to engineering standards.

4. Adjudication method for the test set

  • Not applicable. This is a non-clinical, mechanical testing study, not a study involving human interpretation or adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI/ML device or a diagnostic imaging device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a hardware medical device.

7. The type of ground truth used

  • The "ground truth" for this device's performance measurement would be quantitative mechanical measurements (e.g., force in Newtons, displacement in millimeters) obtained from standardized engineering tests (e.g., universal testing machine results).

8. The sample size for the training set

  • Not applicable. This is a hardware medical device; there is no "training set."

9. How the ground truth for the training set was established

  • Not applicable. There is no training set for this device.

Summary of the Study Proving Acceptance Criteria:

The study was a non-clinical testing program conducted to assess the mechanical properties of the NCS Fish-Fit MD System, specifically its fixation strength and pull-out strength. The testing was performed in accordance with the FDA guidance document "Guidance Document for Testing Bone Anchor Devices" dated April 20, 1996. The purpose was to demonstrate that the device performs as well as (i.e., is substantially equivalent to) the identified predicate devices in these critical mechanical parameters. While specific numerical results or sample sizes are not provided in this summary, the conclusion states that the testing verified adequate performance, demonstrating substantial equivalence.

{0}------------------------------------------------

510(k) Summary

NCS Fish-Fit MD System

DEC 1 1 2012

000078

K120356

Submitted by:

NCS Lab Srl Via Pola Esterna 4/12 41012 Carpi (MO) Italy

Contact Person:

Matteo Mantovani Technical Director Telephone: +39 (059) 669813 Fax: +39 (059) 669813

Preparation:

November 27, 2012

  1. Device Name and Classification Information:

Trade/Proprietary Name:

NCS Fish-Fit MD System, which consists of the following:

    1. the Fish-Fit MD, a sterile, single use, implantable suture anchor; and,

{1}------------------------------------------------

  1. the Compass MD, a series of reusable surgical

instruments used to implant the Fish-Fit MD.

Common/Usual Name:

Anchor

Orthopedic Manual Surgical Instruments

Classification Information

NCS Fish-Fit MD System

Classification Name: Smooth or Threaded Metallic Bone Fixation Fastener

Classification Regulation: 21 CFR 888.3040

Regulatory Class: II

Product Code: MBI - Fastener, Fixation, Nondegradable, Soft Tissue

3. Predicate Devices

The NCS Fish-Fit MD System is substantially equivalent to the following predicate devices:

  • Smith and Nephew, TwinFix Ultra Ti Suture Anchor, . cleared by FDA on April 19, 2010 in K100159; and
  • ConMed Linvatex, Y-Knot All-Suture Anchor, cleared by . FDA on August 8, 2011 in K111779.

2

000079

{2}------------------------------------------------

Description of the Device 4.

NCS Fish-Fit MD System is an implantable anchor made from Titanium grade 4. Sutures are not pre-assembled with bone anchor. The system includes an instrumentation package, Compass MD. The instrument package is intended to facilitate implantation of the anchor.

5. Intended Use and Indications for Use

The NCS Fish-Fit MD System is intended for use for the reattachment of soft tissue (tendons) to bone in the shoulder for rotator cuff repairs. The device has the following indications for use:

The NCS Fish-Fit MD System is intended for the attachment of soft tissue to bone during arthroscopic or open treatment of rotator cuff tendons lesions with the purpose of repairing the rotator cuff (supraspinatus, infraspinatus and subscapularis).

ે. Comparison to the Predicate Devices

The NCS Fish-Fit MD System is substantially equivalent to the legally marketed predicate devices. The NCS Fish-Fit MD System

{3}------------------------------------------------

and its predicate devices are all smooth or threaded metallic bone fixation fastener as defined in 21 CFR 888.3040.

The NCS Fish-Fit MD System has the same intended use as both predicate devices. All three devices are intended for use during arthroscopic or open surgeries to repair rotator cuff injuries. All three devices serve the same primary function: namely, to secure the tendons to bone during healing. All three devices are used by the same health care providers (orthopedic surgeons) in the same target patient population: adults who require repair of rotator cuff injuries. All devices are implanted into a cavity drilled into the bone.

All three devices share similar design features, including materials, similar sizes and dimensions, and insertion into a pre-drilled cavity into the bone. Each device utilizes a slightly different design to fix the anchor to the bone. Data have been provided to show that the NCS Fish-Fit MD System and Bone Anchor perform as well as the predicate device with respect to fixation strength and pull-out strength. These data have been provided in accordance with the FDA guidance document entitled "Guidance Document for Testing Bone Anchor Devices" dated April 20, 1996 to show that the NCS Fish-Fit MD System.

The provided non-clinical testing has been performed to verify adequate fixation strength and pull-out performances of the device, demonstrating substantially equivalence to the predicate devices.

1

{4}------------------------------------------------

Conclusion 7.

:

  1. Sales of the same .

י

·

Based upon the similarities in performances, materials and intended uses, NCS Fish Fit MD System is substantially equivalent to the : predicate devices.

:

.

.

. .

000082

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, sans-serif font. The text is arranged on a single line and is centered horizontally. The words are all capitalized, and the overall appearance is clean and professional.

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-002

Letter dated: December 11, 2012

NCS Lab SRL % Mr. Matteo Mantovani Technical Director Via Pola Esterna 4/12 Carpi, Italy 41012

Re: K120356

Trade/Device Name: NCS Fish-Fit MD System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: November 27, 2012 Received: November 29, 2012

Dear Mr. Mantovani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{6}------------------------------------------------

Page 2 – Mr. Matteo Mantovani

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050:

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number:

K120356

Device Name:

NCS Fish-Fit MD System

Indications For Use:

The NCS Fish-Fit MD System is intended for the attachment of soft tissue to bone during arthroscopic or open treatment of rotator cuff tendons lesions with the purpose of repairing the rotator cuff (supraspinatus, infraspinatus and subscapularis).

Prescription Use _____________________________________________________________________________________________________________________________________________________________

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Casey Hanley

For Division of Orthopaedic Devices

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.