K Number
K120306

Validate with FDA (Live)

Device Name
RAPICIDE OPA-28
Manufacturer
Date Cleared
2012-09-27

(239 days)

Product Code
Regulation Number
880.6885
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Rapicide OPA-28 is a high level disinfectant solution for reprocessing of heat sensitive semicritical medical devices for which sterilization is not suitable. Rapicide OPA-28 may be used at or above its minimum recommended concentration (MRC) of 0.35% OPA as determined by Rapicide OPA-28 test strips in manual device reprocessing with an immersion time of at least 10 minutes at a minimum temperature of 20℃ for a reuse period not to exceed 28 days. Rapicide OPA-28 may also be used in compatible legally marketed automatic endoscope reprocessors at or above its MRC as determined by Rapicide OPA-28 test strips with an immersion time of at least 5 minutes at a minimum temperature of 25℃ for a reuse period not to exceed 28 days.

Device Description

Rapicide OPA-28 is a reusable liquid ortho-Phthalaldehyde (OPA) based high level disinfectant solution intended for reprocessing of heat sensitive semi-critical medical devices for which sterilization is not suitable. Rapicide OPA-28 may be used in manual reprocessing or in compatible legally marketed automatic endoscope reprocessors at the appropriate labeled use conditions. Rapicide OPA-28 is a clear liquid chemical germicide solution with a pH between 7.5 and 8.0. The active microbicidal ingredient ortho-Phthalaldehyde is formulated at a nominal concentration of 0.575% together with a solvent, buffers, surfactants, an antifoaming agent and water to create Rapicide OPA-28 solution. Rapicide OPA-28 is labeled with an unopened bottle shelf-life of 24 months, an open bottle use-period of 75 days and a reuse period not to exceed 28 days. The germicide must be used at or above its MRC, as determined by Rapicide OPA-28 test strips, with the indicated immersion time, use temperature and reuse period. Rapicide OPA-28 is supplied in cases containing high density polyethylene plastic bottles. Individual bottles are labeled with all information necessary to use the device safely.

AI/ML Overview

The provided text describes the regulatory clearance of a medical device, Rapicide OPA-28, a high-level disinfectant, rather than an AI/ML powered device. As such, many of the requested criteria for AI/ML device studies are not applicable.

However, I can extract information related to the device's performance, indications for use, and the types of studies performed as outlined in the submission.

Here's a breakdown of the relevant information:

Device: Rapicide OPA-28 High Level Disinfectant

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Performance ClaimReported Device Performance
High-level Disinfection Efficacy:
Sporicidal EfficacyDemonstrated
Tuberculocidal EfficacyDemonstrated
Fungicidal EfficacyDemonstrated
Virucidal EfficacyDemonstrated
Bactericidal EfficacyDemonstrated
Usage Conditions (Manual):
Minimum Recommended Concentration (MRC)0.35% OPA
Immersion TimeAt least 10 minutes
Minimum Temperature20°C
Reuse PeriodNot to exceed 28 days
Usage Conditions (Automatic Endoscope Reprocessors):
Minimum Recommended Concentration (MRC)0.35% OPA
Immersion TimeAt least 5 minutes
Minimum Temperature25°C
Reuse PeriodNot to exceed 28 days
Physical/Chemical Properties:pH between 7.5 and 8.0
0.575% nominal OPA concentration
Shelf-life (unopened)24 months
Use-period (open bottle)75 days
Other Performance Characteristics:
Simulated-use testsDemonstrated
In-use testsDemonstrated
Rinse ResidualDemonstrated
BiocompatibilityDemonstrated
Toxicity evaluationDemonstrated
Material compatibilityDemonstrated
StabilityDemonstrated
Test strip performanceDemonstrated

Study Proving Acceptance Criteria:

The studies performed were intended to demonstrate substantial equivalence to legally marketed predicate devices, as per the FDA's Guidance for Content and Format of Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/High Level Disinfectants (Jan 2000).

The types of data/information provided included:

  • Detailed description of physical and chemical properties.
  • Proposed labeling.
  • Sporicidal, tuberculocidal, fungicidal, virucidal, and bactericidal efficacy.
  • Simulated-use and in-use tests.
  • Rinse Residual and biocompatibility.
  • Toxicity evaluation.
  • Material compatibility.
  • Stability.
  • Test strip performance.

Non-Applicable Information for this type of Medical Device (Chemical Disinfectant):

The following requests are specific to AI/ML powered devices and are not applicable to this 510(k) submission for a liquid chemical germicide:

  1. Sample sizes used for the test set and data provenance (e.g., country of origin, retrospective/prospective).
  2. Number of experts used to establish ground truth for the test set and their qualifications.
  3. Adjudication method for the test set.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of human readers improving with AI vs. without AI assistance.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for test set.
  7. The sample size for the training set.
  8. How the ground truth for the training set was established.

This 510(k) submission focuses on the chemical efficacy and safety of the disinfectant itself through laboratory and simulated-use testing, rather than the performance of an interpretive or diagnostic algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Minntech, a Cantel Medical Company. The logo consists of a circular graphic on the left, with several horizontal lines inside. To the right of the graphic is the word "MINNTECH" in large, bold, sans-serif font, with the registered trademark symbol. Below the word "MINNTECH" is the text "A Cantel Medical Company" in a smaller font.

Premarket Notification 510(k) Rapicide OPA-28

K120306

Rapicide OPA-28 High Level Disinfectant

510(k) Summary

SEP 27 2012

Manufacturer: Minntech Corporation, A Cantel Medical Company

14605 28th Avenue North Address: Minneapolis, MN 55447 USA (763) 533-3300. 763-551-2653 fax

Official Contact:

Brent Geiger, MS, RAC Senior RA Specialist

Rapicide OPA-28 Trade Name: Liquid Chemical Germicide Common Name: Classification Name: Sterilant, Medical Devices Product Code: MED Device Class: II Classification Reg: 880.6885

January 20, 2012 Date Prepared:

Minntech Corporation has supplied the following information to the United States Food and Drug Administration to support substantial equivalence of Rapicide OPA-28 to other similar liquid chemical germicide solutions currently cleared for sale in the United States.

1. Device Description

Rapicide OPA-28 is a reusable liquid ortho-Phthalaldehyde (OPA) based high level disinfectant solution intended for reprocessing of heat sensitive semi-critical medical devices for which sterilization is not suitable. Rapicide OPA-28 may be used in manual reprocessing or in compatible legally marketed automatic endoscope reprocessors at the appropriate labeled use conditions.

Rapicide OPA-28 is a clear liquid chemical germicide solution with a pH between 7.5 and 8.0. The active microbicidal ingredient ortho-Phthalaldehyde is formulated at a nominal concentration of 0.575% together with a solvent, buffers, surfactants, an antifoaming agent and water to create Rapicide OPA-28 solution. Rapicide OPA-28 is labeled with an unopened bottle shelf-life of 24 months, an open bottle use-period of 75 days and a reuse period not to exceed 28 days. The germicide must be used at or above its MRC, as determined by Rapicide OPA-28 test strips, with the indicated immersion time, use temperature and reuse period. Rapicide OPA-28 is supplied in cases containing high density polyethylene plastic bottles. Individual bottles are labeled with all information necessary to use the device safely.

Minntech Corporation Attachment 4 - Page 1 of 2

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for MINNTECH, a Cantel Medical Company. The logo consists of a circular graphic on the left, followed by the word "MINNTECH" in bold, black letters. Below the word "MINNTECH" is the text "A Cantel Medical Company" in a smaller font. The "R" in a circle is a registered trademark symbol.

2. Indications for Use

Ravicide OPA-28 is a high level disinfectant solution for reprocessing of heat sensitive semicritical medical devices for which sterilization is not suitable. Rapicide OPA-28 may be used at or above its minimum recommended concentration (MRC) of 0.35% OPA as determined by Rapicide OPA-28 test strips in manual device reprocessing with an immersion time of at least 10 minutes at a minimum temperature of 20℃ for a reuse period not to exceed 28 days. Rapicide OPA-28 may also be used in compatible legally marketed automatic endoscope reprocessors at or above its MRC as determined by Rapicide OPA-28 test strips with an immersion time of at least 5 minutes at a minimum temperature of 25°C for a reuse period not to exceed 28 days.

3. Comparison to Other Devices in Commercial Distribution Within the United States

Rapicide OPA-28 is equivalent in performance and indications to predicate devices Metricide OPA Plus (K070627). Cidex OPA (K030004) and Rapicide (K993042). All of the products are liquid chemical high level disinfectant solutions with equivalent intended use and indications for use. The chemical formulation of Rapicide OPA-28 is similar to that of predicate devices Metricide OPA Plus and Cidex OPA Solution.

4. Summary of Non-Clinical Performance Data

Minntech Corporation has provided testing to show that Rapicide OPA-28 is safe and effective for its intended use based on the requirements listed in FDA's Guidance for Content and Format of Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/High Level Disinfectants (Jan 2000). The following types of data/information were provided to FDA in support of substantial equivalence to predicate devices and to demonstrate that Rapicide OPA-28 performs as intended.

  • Detailed description of physical and chemical properties .
  • Proposed labeling .
  • Sporicidal, tuberculocidal, fungicidal, virucidal and bactericidal efficacy ●
  • . Simulated-use and in-use tests
  • Rinse Residual and biocompatibility .
  • Toxicity evaluation .
  • Material compatibility ●
  • Stability .
  • Test strip performance .

5. Conclusion

Minntech Corporation has provided appropriate premarket notification information in the form of a 510(k) to support the substantial equivalence of Rapicide OPA-28 to legally marketed predicate devices. The information and performance data provided indicates that Rapicide OPA-28 is safe and effective for its intended use when used in accordance with the device labeling.

Minntech Corporation Attachment 4 - Page 2 of 2

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 27 2012

Mr. Brent Geiger, MS, RAC Senior Regulatory Affairs Specialist Minntech Corporation, A Cantel Medical Company 14605 280 Avenue North Minneapolis, Minnesota 55447 .

Re: K120306

Trade/Device Name: Rapicide OPA-28 Regulation Number: 21 CFR 880.6885 Regulation Name: Liquid Chemical Sterilants / High level Disinfectants Regulatory Class: II Product Code: MED Dated: September 11. 2012 Received: September 11, 2012

Dear Mr. Geiger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Geiger

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For Peter Smith

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Minnetech logo. The logo consists of a circular graphic on the left and the word "MINNTECH" in bold, uppercase letters on the right. Below the word "MINNTECH" is the phrase "A Cantel Medical Company" in a smaller font. The logo is black and white.

K120306

Premarket Notification 510(k)

Rapicide OPA-28

Indications for Use

510(k) number (if known):

Device Name: Rapicide OPA-28

Indications for Use:

Rapicide OPA-28 is a high level disinfectant solution for reprocessing of heat sensitive semicritical medical devices for which sterilization is not suitable. Rapicide OPA-28 may be used at or above its minimum recommended concentration (MRC) of 0.35% OPA as determined by Rapicide OPA-28 test strips in manual device reprocessing with an immersion time of at least 10 minutes at a minimum temperature of 20℃ for a reuse period not to exceed 28 days. Rapicide OPA-28 may also be used in compatible legally marketed automatic endoscope reprocessors at or above its MRC as determined by Rapicide OPA-28 test strips with an immersion time of at least 5 minutes at a minimum temperature of 25℃ for a reuse period not to exceed 28 days.

E. Lylita D. Lumini-Web

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K12 0306

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use__X (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Minntech Corporation Attachment 3 - Page 1 of 1

§ 880.6885 Liquid chemical sterilants/high level disinfectants.

(a)
Identification. A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.(b)
Classification. Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.