(87 days)
GE Vivid E9 ultrasound system is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of Fetal: Abdominal (including renal and GYN); Pediatric: Small Organ (breast, testes, thyroid); Neonatal Cephalic: Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate), Transesophageal; Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).
GE Vivid E9 is a Track 3 diagnostic ultrasound system, which is primarily intended for cardiac imaging and analysis, but which also includes vascular and general radiology applications. The Vivid E9 incorporates a variety of electronic array transducers operating in linear, curved linear, sector/phased array or matrix array format, including two dedicated CW transducers and several real time 3D transducers. It consists of a mobile console with keyboard control panel; color LCD/TFT touch panel, LCD color video display and optional image storage and printing devices. It provides high performance ultrasound imaging and analysis and has comprehensive networking and DICOM capability.
The GE Healthcare Vivid E9 Diagnostic Ultrasound System is a general-purpose ultrasound system specialized for cardiac imaging. The submission describes the device as substantially equivalent to previously cleared predicate devices (Vivid E9 K101149, Vivid i / Vivid q K102388, and 6Tc Transducer K081921).
Acceptance Criteria and Device Performance (Summary)
The documentation provided does not include specific quantitative analytical or clinical performance acceptance criteria for the device's diagnostic performance (e.g., sensitivity, specificity, accuracy). Instead, the submission focuses on demonstrating substantial equivalence to predicate devices and adherence to medical device safety standards. The "device performance" is implicitly considered to be equivalent to the predicate devices and compliant with the listed voluntary standards.
The table below summarizes the types of tests mentioned as being conducted to ensure device safety and effectiveness.
| Acceptance Criteria Category | Reported Device Performance Assessment |
|---|---|
| Acoustic Output Safety | Evaluated and found to conform to applicable medical device safety standards (e.g., NEMA UD 3). |
| Measurement Accuracy | Evaluated and found to conform to applicable medical device safety standards. |
| Biocompatibility | Evaluated and found to conform to ISO 10993-1. |
| Cleaning and Disinfection Effectiveness | Evaluated and found to conform to applicable medical device safety standards. |
| Thermal Safety | Evaluated and found to conform to IEC60601-1, UL60601-1, IEC60601-1-2, IEC60601-2-37. |
| Electrical Safety | Evaluated and found to conform to IEC60601-1, UL60601-1, IEC60601-1-2, IEC60601-2-37. |
| Electromagnetic Safety | Evaluated and found to conform to IEC60601-1-2. |
| Mechanical Safety | Evaluated and found to conform to applicable medical device safety standards. |
| Software Quality Assurance | Risk Analysis, Usability Analysis, Requirement Reviews, Design Reviews, Unit Testing, Integration Testing, Final Acceptance Testing, Performance Testing, Safety Testing. |
| DICOM Conformance | Complies with NEMA PS 3.1 - 3.20. |
Study Information
-
Sample Size used for the test set and the data provenance:
The submission explicitly states that "The subject of this premarket submission, Vivid E9, did not require clinical studies to support substantial equivalence." Therefore, there is no mention of a defined test set of patient data or its provenance for diagnostic performance evaluation, as clinical studies were not conducted for this submission. The product's compliance was primarily based on non-clinical tests and comparison to predicate devices. -
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable, as clinical studies involving test sets and ground truth from experts were not conducted for this premarket submission. -
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable, as clinical studies involving test sets were not conducted for this premarket submission. -
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This submission is for an ultrasound system, not an AI-assisted diagnostic tool. -
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This submission is for an ultrasound system and its software features, not a standalone algorithm. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable, as clinical studies requiring ground truth were not conducted. The "ground truth" for the non-clinical tests mentioned would be established by validated test methods and reference standards for the specific safety and performance metrics (e.g., accuracy of measurements against known physical standards, successful completion of software tests). -
The sample size for the training set:
Not applicable. The document does not describe a machine learning model requiring a training set. The device is an ultrasound system with enhanced software features, and its development followed a standard quality assurance process. -
How the ground truth for the training set was established:
Not applicable, for the same reasons as above.
{0}------------------------------------------------
K1202cl
Page 1 of 3
APR 1 9 2012
GE Healthcare 510(k) Premarket Notification Submission
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided: January 19, 2012 Date: GE Healthcare [GE Vingmed Ultrasound AS] Submitter: Strandpromenaden 45 N-3191, Horten, Norway Brvan Behn Primary Contact Person: Regulatory Affairs Manager GE Healthcare T:(414)721-4214 F:(414)918-8275 Jan Tore Thollefsen Secondary Contact Person: Regulatory Affairs Leader GE Vingmed Ultrasound AS T:(+47)3302-1269 F:(+47)3302-1357 Trade Name: Vivid E9 Diagnostic Ultrasound System Device: Vivid E9 Common/Usual Name: Classification Names: Class II Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-I YN Product Code: Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-1YO Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX K101149 Vivid E9 Diagnostic Ultrasound System Predicate Device(s): K102388 Vivid i / Vivid q Diagnostic Ultrasound System K081921 6Tc Transducer (Vivid E9) Device Description: GE Vivid E9 is a Track 3 diagnostic ultrasound system, which is primarily intended for cardiac imaging and analysis, but which also includes vascular and general radiology applications. The Vivid E9 incorporates a variety of electronic array transducers operating in linear, curved linear, sector/phased array or matrix array format, including two dedicated CW transducers and several real time 3D transducers. It consists of a mobile console with keyboard control panel; color LCD/TFT touch panel, LCD color video display and optional image storage and printing It provides high performance ultrasound imaging and devices. analysis and has comprehensive networking and DICOM
capability.
Image /page/0/Picture/5 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, creating a recognizable emblem for the company.
{1}------------------------------------------------
K 1202c/
Page 2 of 3
GE Healthcare 510(k) Premarket Notification Submission
The system level clinical applications, imaging modes and device description remain unchanged compared to the predicate device Vivid E9 K101149. This modification to the predicate device adds additional software features and applies Real time 3D imaging to the Transesophageal application and additionally adds the 6VT-D 4D Transesophageal transducer.
- Intended Use: GE Vivid E9 ultrasound system is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of Fetal: Abdominal (including renal and GYN); Pediatric: Small Organ (breast, testes, thyroid); Neonatal Cephalic: Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate), Transesophageal; Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).
The Vivid E9 employs the same fundamental scientific Technology: technology as its predicate devices.
Summary of Non-Clinical Tests: Determination of
Substantial Equivalence: The device has been evaluated for acoustic output safety, measurement accuracy, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. The Vivid E9 and its applications comply with voluntary standards:
- IEC60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety.
-
UL60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety.
-
IEC60601-1-2. Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests
-
IEC60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
-
NEMA UD 3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment:
-
ISO 10993-1, Biological evaluation of medical devices - Part 1:
{2}------------------------------------------------
K12C2CI
Page 3 of 3
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The circle has decorative swirls around its perimeter, giving it a classic and recognizable appearance.
510(k) Premarket Notification Submission
Evaluation and testing within a risk management process. (Biocompatibility)
- NEMA PS 3.1 - 3.20, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology)
The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- Usability Analysis .
- . Requirement Reviews
- Design Reviews
- . Testing on unit level (Module verification)
- . Integration testing (System verification)
- . Final Acceptance Testing (Validation)
- . Performance testing (Verification)
- . Safety testing (Verification)
Summary of Clinical Tests:
The subject of this premarket submission, Vivid E9, did not require clinical studies to support substantial equivalence.
GE Healthcare considers the Vivid E9 to be as safe, as effective, Conclusion: and performance is substantially equivalent to the predicate device(s).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
APR 1 9 2012
GE Vingmed Ultrasound AS % Mr. Bryan Behn Regulatory Affairs Manager GE Healthcare GE Medical Systems Ultrasound and Primary Care Diagnostic, LLC 9900 W. Innovation Drive, RP-2138 WAUWATOSA WI 53226
Re: K120201
Trade/Device Name: Vivid E9 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: January 19, 2012 Received: January 23, 2012
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Vivid E9 Diagnostic Ultrasound System, as described in your premarket notification:
| Transducer Model Number | ||
|---|---|---|
| ML6-15-D | 11L-D | 6VT-D |
| 12S-D | M4S-D | 6T or 6Tc |
| 4V-D | M5S-D | 9T |
| i13L | 6S-D | P2D |
| 4C-D | 3V-D | P6D |
| 9L-D | E8C-D |
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Joshua Nipper at (301) 796-6524.
Sincerely Yours,
Janine M./Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{5}------------------------------------------------
GE Healthcare Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. There are three small, stylized swirls or flourishes positioned around the circle's perimeter, adding a decorative element to the design.
510(k) Number (if known):
Device Name:
Vivid E9 Diagnostic Ultrasound System
Indications for Use:
GE Vivid E9 ultrasound system is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of Fetal; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate), Transesophageal; Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).
Prescription Use: X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use: N/A (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Devices (OIVD)
Thum M. Herro
(Division Sign-Off)
Division of Radiológical Devices Office of In Vitto Diagnostic Device Evaluation and Safety
510(k) Number K120201
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. The letters and the border are black, creating a strong contrast. The logo is simple and recognizable.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form GE Vivid E9 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | P | P | P | P | P | P | P | P | P | P | P |
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | P | P |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | P |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | P |
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | P | P | P | P | P | P | P | P | P | P | N |
| Transrectal | P | P | P | P | P | P | P | P | P | P | |
| Transvaginal | P | P | P | P | P | P | P | P | P | P | |
| Transuretheral | |||||||||||
| Intraoperative[5] | P | P | P | P | P | P | P | P | P | P | |
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] Other use includes Urology/Prostate
[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV),
[*] Combined modes are B/M, B/Color M, B/PWD.or CWD, B/Color/PWD or CWD, B/Power/PWD.
[*] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
System provides real-time 3D and 4D acquisition when used with special 4D probes.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Laure Re. Thomas
Prescription User (Per 21 CFR 801.109)
(Division Sign-O4) Division of Radiological Devices Office of In Waro Diagnostic Device Evaluation and Safety
510K K120201
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with ML6-15-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | |||||||||||
| Abdominal | |||||||||||
| Pediatric[2] | P | P | P | P | P | P | P | P | P | P | |
| Small Organ[1][2] | P | P | P | P | P | P | P | P | P | P | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | |
| Adult Cephalic | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Peripheral Vascular(2) | P | P | P | P | P | P | P | P | P | P | |
| Musculo-skeletal Conventional[2] | P | P | P | P | P | P | P | P | P | P | |
| Musculo-skeletal Superficial | |||||||||||
| Other | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transurethral | |||||||||||
| Intraoperative | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication, P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Small organ includes breast, testes, thyroid.
[2] Needle guidance imaging
of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first of th
510K
[1] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
( • ] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
| Prescription User (Per 21 CFR 801.109) | |
|---|---|
| (Division Sign-Off) | |
| Division of Radiological Devices | |
| Office of In Vitro Diagnostic Device Evaluation and Safety | |
| 12/20/21 |
16
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circle. The letters and the circle are black, creating a strong contrast. The logo is simple and recognizable, representing the company's brand identity.
GE Healthcare Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with 12S-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics | ||||||||||||
| Abdominal | ||||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | P | ||
| Small Organ | ||||||||||||
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | ||
| Adult Cephalic | ||||||||||||
| Cardiac(1) | P | P | P | P | P | P | P | P | P | P | ||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative (specify) | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: {1} Cardiac is Adult and Pediatric.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ •] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
turns (Division Sign-Off)
Division of Radiological Devices
510K
Office of In Vjtro Diagnostic Device Evaluation and Safaty
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are black, while the background is white. The logo has a vintage or slightly distressed appearance.
GE Healthcare Traditional 510(k) Premarket Notification
GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with 4V-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Cardiac[2] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[3] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Cardiac is Adult and Pediatric.
[3] Other use includes Urology/Prostate
- [*]. Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
- [ + ] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO
Concurren
(Division Sign-Off)
Division of Radiological Devices.
Office of In Vitro Diagnostic Device Evaluation and Safety
510K
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular border. The letters and the border are depicted in black, creating a strong contrast against the white background.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form GE Vivid E9 with i13L Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/ Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | |||||||||||
| Abdominal [1] | P | P | P | P | P | P | P | P | |||
| Pediatric | |||||||||||
| Small Organ (specify) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac[2] | P | P | P | P | P | P | P | P | |||
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other (specify) | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[3] | P | P | P | P | P | P | P | P | |||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Cardiac is Adult and Pediatric via Intraoperative;
[3] Intraoperative includes abdominal, thoracic, and vascular.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ • ] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
| (Division Sign-Off) | |
|---|---|
| Division of Radiological Devices | |
| Office of In Vitro Diagnostic Device Evaluation and Safety |
| 510K. | |
|---|---|
| ------- | -- |
| 19 | |
|---|---|
| -- | ---- |
19
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The letters and the interior of the circle are black, while the space around the letters is white.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form GE Vivid E9 with 4C-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | |||
| Abdominal | P | P | P | P | P | P | P | P | P | |||
| Pediatric | P | P | P | P | P | P | P | P | P | |||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac | ||||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | |||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | |||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative (specify) | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [2] Small organ includes breast, testes, thyroid.
[4] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ • ] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
1111 (Division Sign-Off) Division of Radioingical Devices
of In Vitro Diagnostic Device Evaluation and Safety
Office
510K
Prescription User (Per 21 CFR 801.109)
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are in a dark color, contrasting with the lighter background.
GE Healthcare Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form GE Vivid E9 with 9L-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | |||||||||||
| Abdominal | |||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | ||
| Small Organ[2] | P | P | P | P | P | P | P | P | P | ||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | ||
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | ||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [2] Small organ includes breast, testes, thyroid.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ + ] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K120201
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The letters and the border are in white, set against a black background, creating a high-contrast design. The logo is simple, recognizable, and represents the General Electric brand.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form GE Vivid E9 with 11L-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | |||||||||||
| Abdominal | |||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | ||
| Small Organ[2] | P | P | P | P | P | P | P | P | P | ||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | ||
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | ||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[5] | P | P | P | P | P | P | P | P | P | ||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| l anarosconic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [2] Small organ includes breast, testes, thyroid.
- [5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).
- [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
- RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
P
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
SIOK
Prescription User (Per 21 CFR 801.109)
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, creating a strong contrast. The logo is simple, recognizable, and represents the General Electric brand.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form GE Vivid E9 with M4S-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ◆] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
| Prescription User (Per 21 CFR 801.109) | |
|---|---|
| -- | ---------------------------------------- |
| (Division Sign-Off) | |
|---|---|
| Division of Radiological Devices | |
| Office of In Vitro Diagnostic Device Evaluation and Safety |
| 510K | K120201 |
|---|---|
| ------ | --------- |
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. The letters and the border are in black, creating a strong contrast. The logo is simple and recognizable.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form GE Vivid E9 with M5S-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ •] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sigh-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
રાજ્યન
Prescription User (Per 21 CFR 801.109)
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and the entire logo is presented in black against a white background.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form GE Vivid E9 with 6S-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Adult Cephalic | |||||||||||
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ • ] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign Off)
Division of Radiological Devices Office of In Vitro Diagnostic Devices
Office of In Vitro Diagnostic Device Evaluation and Safety 510K
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular shape. The letters and the circle are black, creating a strong contrast.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with 3V-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type. Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[*] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
| (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) |
|---|
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division $ign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K.
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The letters are in a stylized font, and the circle has small, leaf-like shapes around its perimeter.
GE Healthcare Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with E8C-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: .
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/ Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | |||
| Abdominal[1] | P | P | P | P | P | P | P | P | |||
| Pediatric | |||||||||||
| Small Organ (specify) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | |||
| Exam Type, Means of Access | - | ||||||||||
| Transesophageal | |||||||||||
| Transrectal | P | P | P | P | P | P | P | P | |||
| Transvaginal | P | P | P | P | P | P | P | P | |||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN/Pelvic;
[4] Other use includes Urology/Prostate;
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[*] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
510K.
Division of Rastialogical Disvices Office of In Vitro Diagnostic Device Evaluation and Safety
--
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, contrasting with the white background. The logo is a well-known symbol associated with the multinational conglomerate corporation.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with 6VT-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/ Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics | ||||||||||||
| Abdominal | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (specify) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac | N | N | N | N | N | N | N | N | N | N | N | |
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other (specify) | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | N | N | N | N | N | N | N | N | N | N | N | |
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative (specify) | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. [] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off)
Division of Radiological Devices
Office of in Vitro Diagnostic Devics Evaluation and Safety
510K
Prescription User (Per 21 CFR 801.109)
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and there are small, rounded shapes placed around the border, giving it a decorative appearance.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with 6T or 6Tc Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/ Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | |||||||||||
| Abdominal | |||||||||||
| Pediatric | |||||||||||
| Small Organ (specify) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other (specify) | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | P | P | P | P | P | P | P | P | P | P | |
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. [] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off)
SIOK
Prescription User (Per 21 CFR 801.109)
Division of Radiological Devices Office of In Vitro Diagnestic Device Evaluation and Safety
O
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "g" and "e" intertwined within a circular border. The logo is black and white.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with 9T Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/ Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | |||||||||||
| Abdominal | |||||||||||
| Pediatric | |||||||||||
| Small Organ (specify) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac [3] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other (specify) | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | P | P | P | P | P | P | P | P | P | P | P |
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. [3] Cardiac is Adult & Pediatric
[*] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Fanny Ross
(Division Sign-Off)
Prescription User (Per 21 CFR 801.109)
Division of Radiglogical Devices Office of In Vitro Diagnostic Davica Evaluation and Safe
510K K120201
{22}------------------------------------------------
Image /page/22/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the outline of the circle are white, while the background within the circle is black.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with P2D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/ Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | |||||||||||
| Abdominal | |||||||||||
| Pediatric | |||||||||||
| Small Organ (specify) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac[3] | P | P | |||||||||
| Peripheral Vascular | P | P | |||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other (specify) | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [3] Cardiac is Adult and Pediatric.
[*] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Batoty 510K
{23}------------------------------------------------
Image /page/23/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are black, creating a bold and recognizable symbol.
Traditional 510(k) Premarket Notification GE Vivid E9, January 19, 2012
Diagnostic Ultrasound Indications for Use Form
GE Vivid E9 with P6D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/ Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | RT3DMode* |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics | |||||||||||
| Abdominal | |||||||||||
| Pediatric | |||||||||||
| Small Organ (specify) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac[3] | P | P | |||||||||
| Peripheral Vascular | P | P | |||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other (specify) | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [3] Cardiac is Adult and Pediatric.
[*] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign Off)
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510K
N/A