(74 days)
Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
Multi-Modality Image Fusion (MMIF) is software, which comprises of two software components, which is referred to as offline and online. The offline component perfains to the preparation of gland and suspected lesion boundaries on a DICOM image file days or hours prior to the biopsy procedure. The online component fuses the DICOM image files, which were, prepared on the offline component, with a snap shot incoming TRUS image. Each of the two software components can work together or independently.
The provided 510(k) summary for the Multi-Modality Image Fusion device does not contain a specific table of acceptance criteria or detailed results of a study proving the device meets those criteria. However, it does outline the types of tests performed and provides a general statement of success.
Based on the available information, here is a description of the acceptance criteria (inferred from the testing) and the study conducted:
1. Acceptance Criteria and Reported Device Performance
Since explicit acceptance criteria are not provided, they are inferred from the types of tests performed. The document states that "All product and engineering specifications were verified and validated" and that "All the above mentioned tests passed." This implies that the device successfully met the intended performance parameters for each test.
| Acceptance Criteria (Inferred from Tests) | Reported Device Performance |
|---|---|
| Functional/Hardware Verification | Tests Passed |
| - Hardware Verification | Successfully verified the MMIF hardware. |
| - Online Software Application Verification | Successfully verified the online software application. |
| - Offline Software Application Verification | Successfully verified the offline software application. |
| Risk Mitigation Verification | Tests Passed |
| - Risk Mitigation Verification | Successfully verified risk mitigation strategies. |
| Accuracy (Benchmarking) | Tests Passed |
| - Registration Accuracy for Clinical Data | Demonstrated accuracy in registration for clinical data. |
| - Phantom Volume Measurements | Demonstrated accurate volume measurements using phantoms. |
| - Phantom Registration Accuracy | Demonstrated accurate registration using phantoms. |
| System Performance (Static & Dynamic) | Tests Passed |
| - Overall System Performance | Confirmed static and dynamic performance of the complete system (both online and offline software on specified hardware) and compliance to specifications in a simulated real environment. |
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "clinical data collected from the hospital setting" for the "Registration Accuracy for Clinical Data (bench test)." However, the sample size for this test set (number of cases/patients) is not specified. The data provenance is generally stated as "clinical data collected from the hospital setting," indicating it is likely retrospective clinical data used for a bench test.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts used or their qualifications for establishing the ground truth of the "clinical data" used in testing. It only mentions that the "offline software would be typically executed by Radiologists."
4. Adjudication Method for the Test Set
The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for establishing ground truth on the test set.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not mentioned or described in the provided summary. The focus of the performance testing appears to be on the device's standalone accuracy and functionality.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, a standalone performance evaluation of the algorithm (referred to as "the new MMIF software") was performed. The "Testing and Performance Data" section describes "bench testing" activities conducted "to ensure the performance of the new MMIF software by verifying the accuracy of the specifications and simulating real customer data collected from the hospital setting." The registration accuracy and volume measurements were assessed for the software itself.
7. The Type of Ground Truth Used
The ground truth for the bench tests involved:
- Simulated prostates in test phantoms: Used for "Phantom Volume Measurements" and "Phantom Registration Accuracy." This represents a controlled and precisely known ground truth.
- "Clinical data collected from the hospital setting": Used for "Registration Accuracy for Clinical Data." The method for establishing ground truth for this clinical data (e.g., expert consensus, pathology, surgical outcomes) is not explicitly stated.
8. The Sample Size for the Training Set
The document does not specify a training set size. The summary focuses on device verification and validation, implying that the device was perhaps developed or configured based on various data, but the specific characteristics of a "training set" (in the machine learning sense) are not outlined.
9. How the Ground Truth for the Training Set Was Established
Since a training set is not explicitly mentioned, the method for establishing its ground truth is not provided. The device seems to be a software application for image processing and fusion, rather than a machine learning model that requires explicit training data in the context of this 510(k) summary. For software applications of this nature, ground truth is often implicitly established by the underlying physical principles of image registration algorithms and validated against physical models or existing clinical standards.
{0}------------------------------------------------
510(k) Summary
APR - 6 2012
510(k) Summary
Submitter's Name: Submitter's Address: Submitter's Telephone: Contact Name:
Date Summary was Prepared: Trade or Proprietary Name: Common or Usual Name: Classification Name:
Predicate Devices:
IGT, LLC dba Eigen 13366 Grass Valley Avenue, Grass Valley, CA 95945 530-274-1240 . William Mandel 21 Oct 2011
Multi-Modality Image Fusion Multi-Modality Image Fusion
System, Image Processing, Radiological, LLZ
Picture Archiving and Communications, 21CFR 892.2050
| Device Name | 510(k) Number |
|---|---|
| Syntegra | K041182 |
| UniSyn | K081987 |
| Aegis Navigation | K093672 |
| Abaris | K053610 |
Intended Use
Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
Description of the Device and Summary of the Technological Characteristics:
Multi-Modality Image Fusion (MMIF) is software, which comprises of two software components, which is referred to as offline and online. The offline component perfains to the preparation of gland and suspected lesion boundaries on a DICOM image file days or hours prior to the biopsy procedure. The online component fuses the DICOM image files, which were, prepared on the offline component, with a snap shot incoming TRUS image. Each of the two software components can work together or independently.
Offline Software
The offline software can delineate suspected areas of interest using visualization and save this data in a standard format, e.g. as a surface mesh in VTK or STL format. Furthermore, the
EIGEN
PREMARKET NOTIFICATION - MULTI-MODALITY IMAGE FUSION
{1}------------------------------------------------
offline software can ontionally delineate the gland boundary from a DICOM image, or other volumes where the boundary can be easily observed. This addition data is also saved in a standard format. e.g. surface mesh in VTK or STL or as a binary volume. This software would be typically executed by Radiologists.
Online Software
The online unit software is designed to fuse two images by entering a fusion mode and load the DICOM image and the boundary data provided by the offline unit. The online unit performs rigid alignment, either selecting globally align that orients the two DICOM volumes based on the known relative orientation difference between the two or specify landmarks to register images rigidly. The online unit then elastically warps the image to register on the incoming image from the offline unit.
The registered image volume is displayed on the incoming offline image frame of reference. Also overlaid on the incoming offline and registered image volumes annotating the suspected lesion boundaries. The two volumes are displayed side-by-side in a locked view. Rotation, slicing, panning, or zooming one of the volumes has the same effect on the other enabling visualization of similar features in both volumes simultaneously.
Plan targets on the volumes by selecting suitable locations within the suspected areas of interest boundary. When sufficient targets have been added, exiting fusion mode imports all targets planned in the offline frame of reference back to the original reconstructed ultrasound volume. This is done based on the deformation field that was estimated from registration (rigid followed by elastic registration).
Substantial Equivalence:
| Manufacturer | Eigen | ADACLaboratories | ConvergentImagingSolutions | Sentinelle MedicalInc. | TraxtalTechnologies lhc. |
|---|---|---|---|---|---|
| ProductName | Multi-ModalityImage Fusion | Syntegra | UniSyn | Aegis Navigation | Abaris |
| 510(k)number | pending | K041182 | K081987 | K093672 | K053610 |
| IntendedUse | Multi-ModalityImage Fusion is asoftwareapplication used by | Syntegra is asoftwareapplication formulti-modality | UniSyn is asoftwareapplication forimage | This deviceprovides two andthree-dimensionalimage review. | ABARIS is astereotaxicaccessory forComputed |
EIGEN
PREMARKET NOTIFICATION - MULTI-MODALITY IMAGE FUSION
{2}------------------------------------------------
| physicians in the | image registration | registration and | manipulation, and | Tomography | ||||
|---|---|---|---|---|---|---|---|---|
| clinic or hospital | and diagnostic | fusion display | analysis tools to | (CT), Magnetic | ||||
| for 2-D and 3-D | fusion. Images are | of scanned | assist users in | Resonance, (MR), | ||||
| visualization, | registered and | image data | screening, | Ultrasound (US), | ||||
| multi-modality | displayed in a | from CT, PET, | diagnosis, planning | Positron Emission | ||||
| image registration, | "fused" (overlaid | SPECT and | and performing | Tomography | ||||
| and fusion of | in the same spatial | MR scanners. | image-guided | (PET), Single | ||||
| medical images. | orientation) | UniSyn creates | interventional | Photon Emission | ||||
| Additional | format to provide | multi-planar | procedures. | Computed | ||||
| software featuresinclude database | combined | reformat and | Supported imaging | Tomography | ||||
| functional and | maximum | modalities include | (SPECT), | |||||
| management, | anatomical data | intensity | Magnetic | Fluoroscopy, | ||||
| communication. | providing | projection | Resonance (MR), | Endoscopy and | ||||
| surface rendering, | different angular | displays of the | Ultrasound (US), | other imaging | ||||
| segmentation, ROI | perspectives for | data and | Single Photon | systems. it | ||||
| delineation, | interpretation by | provides | Emission | displays the | ||||
| measurements, and | trained | measurements | Computed | simulated image | ||||
| reporting. | professionals | such as area, | Tomography | of a tracked | ||||
| volume and | (SPECT), | insertion tool | ||||||
| Standard | Computed | such as a biopsy | ||||||
| Uptake Values | Tomography (CT), | needle, guidewire | ||||||
| for user defined | Positron | or probe on a | ||||||
| regions on the | Emission | computer monitor | ||||||
| image | Tomography | screen that shows | ||||||
| (PET), | images of the | |||||||
| Fluoroscopy and | target organs and | |||||||
| Endoscopy. Images | the current and the | |||||||
| and data are | projected future | |||||||
| received from | path of the | |||||||
| various imaging | interventional | |||||||
| systems and other | instrument taking | |||||||
| sources such as | into account | |||||||
| calibrated spatial | movements of the | |||||||
| positioning | patient. This is | |||||||
| devices. | intended for | |||||||
| This device | treatment | |||||||
| provides the | planning and | |||||||
| capability to | intra-operative | |||||||
| overlay annotations | guidance for | |||||||
| on 2D or 3D | surgical | |||||||
| medical image | procedures. | |||||||
| displays. | The device also | |||||||
| These annotations | supports an | |||||||
| may represent the | image-free mode | |||||||
| position of | in which the | |||||||
| instruments | proximity of the | |||||||
| including but not | interventional | |||||||
| limited to biopsy | device is | |||||||
| needles, guidance | displayed relative | |||||||
| wires, imaging | to another device. | |||||||
| probes or other | The device is | |||||||
| tracked devices. | intended to be | |||||||
| This device is | used in clinical | |||||||
| intended to assist | interventions and | |||||||
| skilled medical | for anatomical | |||||||
| professionals in | structures where | |||||||
| skilled medicalprofessionals inclinical screeningand | clinical screening | for anatomicalstructures whereimaging iscurrently used for | imaging is |
.
510(k) Summary
.
EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
・
・
.
. . . . .
,
.
{3}------------------------------------------------
| 510(k) Summary | |||||
|---|---|---|---|---|---|
| interventions, foranatomicalstructures whereimaging iscurrently used forvisualizing suchstructures,including head andneck, breast,thoracic, andabdominalapplications(including | visualizing suchprocedures. Thedevice is alsointended for usein clinicalinterventions todetermine theproximity ofone devicerelative toanother. | ||||
| pelvis).Lossy compressedmammographicimages anddigitized filmscreen images mustnot be | |||||
| reviewed forprimary imageinterpretations.Mammographicimages may onlybe interpretedusing | |||||
| a FDA approvedmonitor that offersat least 5 MPixelresolution andmeets othertechnical | |||||
| specificationsreviewed andaccepted by theFDA. | |||||
| WhereUsed | Office settings inclinic or hospital | Office settings inclinic or hospital | Office settingsin clinic orhospital | Not specified | clinicalinterventions |
| SoftwareDevice | Yes | Yes | Yes | Yes | Yes |
| ImageRegistration | Multi-modalityimage registration | Multi-modalityimage registration | Multi-modalityimageregistration | Multi-modalityimage registration | Yes |
| ImageFusion/OverlayDisplay | Fused overlay ofimages frommultiple modalities | Fused overlay ofimages frommultiplemodalities | Fusion overlayof images frommultiplemodalities | overlay of imagesfrom multiplemodalities | Yes |
| OpacityControl | Yes | Yes | Yes | Not Specified | Not Specified |
| 3-DRendering | Yes | Yes | Yes | Yes | Yes |
| SurfaceRendering | Yes | Yes | Yes | Yes | Yes |
| Regions ofinterest | Yes | Yes | Yes | Yes | Yes |
| Configura | Yes | Yes | Yes | Yes | Yes |
EIGEN
.
:
DIGE."
PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
·
{4}------------------------------------------------
510(k) Summary
| ble image layouts | |||||
|---|---|---|---|---|---|
| 3DContouring | Yes | Not Specified | Yes | Yes | Yes |
| Export of3D contoursforplanning | Yes | Not Specified | Yes | Not Specified | Yes |
| ImageStorageandCommunication | DICOM, JPEG | DICOM (check) | DICOM | DICOM | DICOM |
| Modalities | CT, MRI,Ultrasound | Positron EmissionTomography(PET) andX-ray ComputedTomography(CT). | CT, PET,SPECT andMR scanners | MRI, US, SPECT,CT, PET,Fluoroscopy,Endoscopy andothers | ComputedTomography(CT), MagneticResonance, (MR),Ultrasound (US),Positron EmissionTomography(PET), SinglePhoton EmissionComputedTomography(SPECT),Fluoroscopy,Endoscopy andother imaging |
The product's technical features are substantially equivalent to Syntegra (K041182). UniSyn (K081987), Aegis Navigation (K093672) and Abaris (K053610). The Multi-Modality Image Fusion is a software product that runs on PC-based workstations. Image data is input to the devices and used to generate 3-D views and perform image processing. Like the four predicate devices, the software has image measurement, multi-planar reformatting, segmentation and image registration abilities, fusion of images from different modalities, image storage and retrieval, as well as patient information management functions.
The Syntegra and UniSyn are software products that accept multiple image data types including magnetic resonance, and computed tomography. Aegis Navigation and Abaris are systems that accept multiple modalities that include magnetic resonance, computed tomography and ultrasound. The Multi-Modality Image Fusion has been designed for processing medical images in standard DICOM format. The system is composed of two modules: an offline and an online module for image fusion for planning a procedure.
EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
{5}------------------------------------------------
Testing and Performance Data:
All product and engineering specifications were verified and validated. Test phantoms incorporating simulated prostates were developed and were used to verify system performance.
The device has been designed and manufactured to conform to the following standards:
ACRINEMA PS3.1-3.18 Digital Imaging and Communications in Medicine (DICOM)
ISO 14971 Application of Risk Management to Medical Devices
ISO 13485 Medical Devices - Quality Management Systems - Requirements For Regulatory purposes
The following tests were performed:
MMIF Hardware verification Test Procedure
MMIF Online Verification Test Procedure
MMIF Offline Software Application Verification Test Procedure .
MMIF Risk Mitigation Verification Test Procedure
Registration Accuracy for Clinical Data (bench test)
Phantom Volume Measurements (bench test)
Phantom Registration Accuracy (bench test)
All the above mentioned tests passed. Theses testing activities were conducted to establish the performance, safety, effectiveness, functionality usability, and reliability characteristics of the new device.
The purpose of the performance testing (bench testing) was to ensure the performance of the new MMIF software by verifying the accuracy of the specifications and simulating real customer data collected from the hospital setting.
The conducted verification tests have confirmed static and dynamic performance of the complete system consisting of both the online and off line software installed on specified hardware. Furthermore these tests establish compliance to the specification under conditions simulation real environment of use.
Conclusion:
The results of comparing the intended use, function, technological characteristics, mode of operation and specifications of the Multi-Modality Image Fusion with those of the two predicate devices demonstrate that the Multi-Modality Image Fusion is substantially
- EIGEN
PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
{6}------------------------------------------------
equivalent to existing products on the market today.
. . . .
.
ﺎ ﺳﮯ ﺍ
1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -
·
The verification and validation testing included bench performance tests, functionality tests, and system test. All tests were successfully completed.
. '·
'
.
: :
EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
. .
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
APR - 6 2012
IGT LCC dba Eigen % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services 1394 25th Street NW BUFFALO MN. 55313
Re: K120187
Trade/Device Name: Multi-Modality Image Fusion Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications System Regulatory Class: II Product Code: LLZ Dated: March 1, 2012 Received: March 2 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls: Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny 1 vict all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{8}------------------------------------------------
Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Indications for Use Form
510(k) Number (if known):
Device Name: Multi-Mogality Image Fusion
Indications for Use: Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
|---|---|
| AND/OR |
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE OF ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K120187
1-2
N/A