K Number
K120013

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-05-22

(140 days)

Product Code
Regulation Number
872.3660
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TempSpan Clear Matrix is indicated for:

  • Use as a clear template for intra oral polymerization of light activated materials to create temporary restorations.
  • Use as a general impression material for monophase impression techniques.
Device Description

TempSpan Clear Matrix Material is a clear preoperative impression material that will reproduce the finest detail resulting in extremely accurate provisional restorations. Use as the preliminary impression material for intraoral and extra-oral polymerization of dual-cure provisional materials. Light curing dual-cure provisional materials may eliminate or drastically reduce the oxygen-inhibited smear layer. TempSpan Clear Matrix Material is a medium viscosity vinyl polysiloxane material. It is packaged in a 50ml auto-mix cartridge, with an oral set time of 2 minutes. TempSpan Clear Matrix Material enables provisional restorations fabricated with dual cure materials to be light cured intra-orally and extra-orally. This process shortens the provisional procedure, minimizing chair time and patient discomfort. Additional features include: clear formula for intra-oral light curing, auto-mix delivery, and a quick set time with a working time of 1 minute and an oral set time of 2 minutes.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (TempSpan Clear Matrix)Predicate Device (Memosil 2) Performance (for comparison)Predicate Device (Elite Glass) Performance (for comparison)Predicate Device (Correct VPS) Performance (for comparison)
Biocompatibility (Cytotoxicity)Non-cytotoxicNon-cytotoxic (from ISO L929 MEM Elution Test)Not explicitly stated but implied to be non-cytotoxic as a legally marketed deviceNot explicitly stated but implied to be non-cytotoxic as a legally marketed deviceNot explicitly stated but implied to be non-cytotoxic as a legally marketed device
Biocompatibility (Irritation)Negligible irritantNegligible irritant (from ISO Intramuscular Injection Test)Not explicitly stated but implied to be negligible irritant as a legally marketed deviceNot explicitly stated but implied to be negligible irritant as a legally marketed deviceNot explicitly stated but implied to be negligible irritant as a legally marketed device
Work TimeNot explicitly stated, but expected to be comparable to predicate devicesNot explicitly stated, but evaluated and found comparableNot explicitly statedNot explicitly statedNot explicitly stated
Oral Set TimeNot explicitly stated, but expected to be comparable to predicate devices2 minutes (also evaluated and found comparable)Not explicitly statedNot explicitly statedNot explicitly stated
Shore A HardnessNot explicitly stated, but expected to be comparable to predicate devicesNot explicitly stated, but evaluated and found comparableNot explicitly statedNot explicitly statedNot explicitly stated

Note: The document states that the performance characteristics (Work Time, Oral Set Time, and Shore A Hardness) were "evaluated" and provided data for comparison to predicate devices, but the specific numerical results for TempSpan Clear Matrix or the predicate devices, and thus explicit acceptance criteria values, are not included in the provided text. The conclusion of substantial equivalence implies that these characteristics were within acceptable ranges compared to the predicates.

2. Sample Size for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not explicitly stated. The document refers to "bench testing" without providing specific sample sizes for each test.
  • Data Provenance: The studies were non-clinical (bench testing and biocompatibility). The country of origin of the data is not specified but is assumed to be from the manufacturer or accredited testing facilities. The data is retrospective in the sense that it supports a premarket notification for a new device, comparing it to existing ones.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • Number of Experts: Not applicable. This was a non-clinical bench testing study. The "ground truth" was established by objective physical and chemical testing methods (e.g., ISO standards for biocompatibility and laboratory measurements for physical properties).
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not applicable. This was a non-clinical bench testing study. The results of the tests (biocompatibility, work time, set time, hardness) would be objectively measured and compared, not adjudicated.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

  • MRMC Study: No. The document explicitly states: "Clinical testing has not been conducted on this product."

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was Done:

  • Standalone Study: Yes, in the sense that the device's physical and biocompatibility properties were tested independently. This device is a dental impression material, not an algorithm, so the concept of "standalone performance" applies to its inherent material properties rather than an AI algorithm.

7. The Type of Ground Truth Used:

  • Type of Ground Truth:
    • For biocompatibility: Adherence to ISO standards (ISO Intramuscular Injection Test and ISO L929 MEM Elution Test) results (negligible irritant, non-cytotoxic).
    • For performance characteristics (Work Time, Oral Set Time, Shore A Hardness): Objective physical measurements compared against the properties of legally marketed predicate devices. The "ground truth" here is the established performance of the predicate devices.

8. The Sample Size for the Training Set:

  • Sample Size for Training Set: Not applicable. This is a physical medical device, not an AI/machine learning algorithm. Therefore, there is no "training set."

9. How the Ground Truth for the Training Set was Established:

  • How Ground Truth for Training Set was Established: Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a swooping line above them. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are printed in a simple, sans-serif font.

MAY 2 2 2012

510(k) Summary

Submitter:

Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, California 92867 (714) 516-7602 - Phone (714) 516-7488 - Facsimile Wendy Garman - Contact Person

April 2012 Date Summary Prepared:

  • Trade name TempSpan Clear Matrix .
  • Common name Impression Material .
  • Classification name ~ Material, Impression (21 CFR 872.3660, ELW) .

Devices for Which Substantial Equivalence is Claimed:

  • Memosil 2, Class II, K913537 , Product Code ELW, Heraeus Kulzer .
  • Elite Glass, Class II, K073005, ELW, Zhermack S.P.A. .
  • Correct VPS, Class II, K914178, Product Code ELW, Pentron Clinical .

Summary

Device Description

TempSpan Clear Matrix Material is a clear preoperative impression material that will reproduce the finest detail resulting in extremely accurate provisional restorations. Use as the preliminary impression material for intraoral and extra-oral polymerization of dual-cure provisional materials. Light curing dual-cure provisional materials may eliminate or drastically reduce the oxygen-inhibited smear layer.

TempSpan Clear Matrix Material is a medium viscosity vinyl polysiloxane material. It is packaged in a 50ml auto-mix cartridge, with an oral set time of 2 minutes. TempSpan Clear Matrix Material enables provisional restorations fabricated with dual cure materials to be light cured intra-orally and extra-orally. This process shortens the provisional procedure, minimizing chair time and patient discomfort. Additional

{1}------------------------------------------------

features include: clear formula for intra-oral light curing, auto-mix delivery, and a quick set time with a working time of 1 minute and an oral set time of 2 minutes.

When used in a clear impression tray, TempSpan Clear Matrix can be used to create light cured temporary restorations inside a patient's mouth. To begin the process, the clinician will select a rigid, clear plastic tray and use TempSpan Clear Matrix to take an impression of the patient's current state. The impression is set aside while the clinician prepares the tooth for a crown, bridge, etc. When it is time to make the temporary restoration, the impression made using TempSpan Clear Matrix is used to form the temporary restoration. This is completed through the following procedure:

    1. The temporary restoration material is applied directly into the TempSpan Clear Matrix impression.
    1. The impression and temporary restoration material are placed in the patient's mouth and the material is light-cured through the clear impression tray.
    1. Once the clear tray is removed, the temporary restoration remains attached to the patient's prepared tooth.

The clinician also has the option of using TempSpan Clear Matrix with the more traditional method of creating and cementing a temporary restoration. In this option, the clear impression is removed and the selected temporary restorative material is left to fully cure outside of the mouth. The clinician would then remove the temporary restoration from the clear impression and cement it onto the prepared tooth using a temporary cement.

Intended use of the Device

TempSpan Clear Matrix is indicated for:

  • Use as a clear template for intra oral polymerization of light activated materials to create temporary restorations.
  • · Use as a general impression material for monophase impression techniques.

Technological Characteristics Compared to Predicate

TempSpan Clear Matrix is substantially equivalent to other legally marketed devices in the United States. TempSpan Clear Matrix functions in a manner similar to Memosil 2, which is currently marketed by Heraeus Kulzer, Elite Glass which is currently marketed by Zhermack S.P.A., and Correct VPS which is currently marketed by Pentron Clinical.

{2}------------------------------------------------

Non-Clinical Performance Data

Biocompatibility studies were completed, which demonstrate that the material is safe for its intended use. TempSpan Clear Matrix was tested through the ISO Intramuscular Injection Test and the ISO L929 MEM Elution Test, and was found to be a negligible irritant and non-cytotoxic.

The 510(k) submission also includes data from bench testing used to evaluate performance characteristics of TempSpan Clear Matrix compared to the predicate devices, Memosil 2 currently marketed by Heraeus Kulzer, Elite Glass currently marketed by Zhermack S.P.A., and Correct VPS currently marketed by Pentron Clinical. The characteristics evaluated include Work Time, Oral Set Time and Shore A Hardness.

Clinical Testing

Clinical testing has not been conducted on this product.

Conclusion

Based upon the biocompatibility tests and bench testing, the clinical performance of TempSpan Clear Matrix is substantially equivalent to the predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Pentron Clinical Ms. Wendy Garman Director, Regulatory Affairs Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, California 92867

Re: K120013

Trade/Device Name: TempSpan Clear Matrix Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: May 2, 2012 Received: May 3, 2012

Dear Ms. Garman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

MAY 2 2 2012

{4}------------------------------------------------

Page 2 - Ms. Garman

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

K120013

Device Name: TempSpan Clear Matrix

Indications For Use:

TempSpan Clear Matrix is indicated for:

  • . Use as a clear template for intra oral polymerization of light activated materials to create
  • temporary restorations.
  • Use as a general impression material for monophase impression techniques.

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sueen Perry
(Division Sign-Off)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K120013

§ 872.3660 Impression material.

(a)
Identification. Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.(b)
Classification. Class II (Special Controls).