K Number
K113867

Validate with FDA (Live)

Date Cleared
2012-10-19

(294 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LacriCATH Lacrimal Duct Balloon Catheter is a sterile, prescription device intended for single use during dilatation of an obstructed nasolacrimal duct system to correct epiphora.

Indications for Use Statement:

The LacriCATH Lacrimal Duct Balloon Catheter is a sterile, prescription device intended for single use during dilatation of an obstructed nasolacrimal duct system to correct epiphora as a result of the following.

5 mm - LacriCATH Adult Catheter: Functional or complete nasolacrimal duct obstruction Dacryocystitis Failed DCR (Dacryocystorhinostomy)

2mm/3mm LacriCATH Pediatric catheter (intended for patients 12 months of age and older):

Congenital nasolacrimal duct obstruction Dacryocystocele

Device Description

The Quest Medical, Inc. current legally marketed LacriCATH lacrimal duct catheters are sterile, hand held, ophthalmic surgical devices. The LacriCATH devices are designed to facilitate nasolacrimal duct surgeries by using balloon dilation to open occluded lacrimal passages. Each LacriCATH catheter has a stainless steel tube with positioning markers and an inflatable balloon. A luer connector on the proximal end of the catheter is attached to a separate inflation device or a syringe for inflation and deflation of the balloon. The balloon is designed to inflate to a known diameter and length at the specified pressure. The LacriCATH lacrimal duct catheters are sterile, single-use, biocompatible medical devices that are sterilized by ethylene oxide and labeled with a 3 year shelf life. The LacriCATH catheters are packaged with 1 device per pouch, and 2 pouches per carton. They are available by prescription only.

AI/ML Overview

The provided 510(k) summary for the K113867 device (LacriCATH® Lacrimal Duct Balloon Catheter) indicates that no clinical studies were performed or required to prove the device met acceptance criteria.

Instead, the submission relied on non-clinical performance data (bench testing) to demonstrate substantial equivalence to a predicate device.

Therefore, many of the requested details concerning acceptance criteria for clinical performance and clinical study specifics are not applicable in this case.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria: The document states that "Verification and Validation tests were performed to ensure the modified LacriCATH catheters continued to meet or exceed specification in accordance to the Design Requirements. Those specifications are the same as for the predicate device." While specific numerical acceptance criteria are not listed, the overarching criterion was to perform "as well or better than the predicate device" in non-clinical testing.

Reported Device Performance: The bench testing "concluded that the new modified LacriCATH catheters performed as well or better than the predicate device."

CriterionAcceptance Criteria (Implicit)Reported Device Performance
Mechanical/Functional PerformanceMeet or exceed predicate device specificationsPerformed as well or better than predicate device
BiocompatibilityMeet requirementsAchieved requirements
Physicochemical TestingMeet requirementsAchieved requirements
Shelf LifeMaintain integrity and function for 3 yearsDemonstrated 3-year shelf life
Sterilization ResidualsMeet safety limitsAchieved safety limits

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not applicable. No clinical test set was used.
  • Data Provenance: Not applicable. Data is from non-clinical (bench) testing.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Not applicable. Ground truth for clinical performance was not established as no clinical studies were performed. Ground truth for bench testing would typically be established by engineering specifications and standards.

4. Adjudication Method for the Test Set

  • Not applicable. No clinical test set was adjudicated.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Not applicable. This device is not an AI-assisted diagnostic tool, and no MRMC study was performed.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop)

  • Not applicable. This device is a physical medical device, not an algorithm.

7. Type of Ground Truth Used

  • For Non-Clinical Performance: Engineering specifications, industry standards, and internal design requirements. The "ground truth" for the device's technical performance was its ability to meet these established specifications, which were equivalent to or better than the predicate device.

8. Sample Size for the Training Set

  • Not applicable. There is no "training set" in the context of this device's non-clinical evaluation.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable.

In summary, the K113867 submission relied solely on non-clinical (bench) testing to demonstrate substantial equivalence to a predicate device. The acceptance criteria were based on meeting or exceeding the performance specifications of the predicate device. No clinical studies, human readers, or AI algorithms were part of this submission.

{0}------------------------------------------------

K113867
Page 1 of 3

510(k) Summary

Pursuant to 21CFR807.92(c), Quest Medical Inc. provides the following Summary for this Premarket Notification.

OCT 19 2012

A. Submitter Information

Quest Medical, Inc. One Allentown Parkway Allen, TX 75002 USA 972-390-9800 / 800-627-0226; Fax 972-390-2881 Amy Clendening-Wheeler October 18, 2012

B. Device Information

Trade/Device name: LacriCATH® Lacrimal Duct Balloon Catheter Product Code: OKS Device Classification Name: Lacrimal Stents and Intubation Stents Regulatory Class: Unclassified

C. Legally Marketed Predicate Device

lolab Lacrimal Duct Catheter, K935233

D. Device Description

The Quest Medical, Inc. current legally marketed LacriCATH lacrimal duct catheters are sterile, hand held, ophthalmic surgical devices. The LacriCATH devices are designed to facilitate nasolacrimal duct surgeries by using balloon dilation to open occluded lacrimal passages. Each LacriCATH catheter has a stainless steel tube with positioning markers and an inflatable balloon. A luer connector on the proximal end of the catheter is attached to a separate inflation device or a syringe for inflation and deflation of the balloon. The balloon is designed to inflate to a known diameter and length at the specified pressure. The LacriCATH lacrimal duct catheters are sterile, single-use, biocompatible medical devices that are sterilized by ethylene oxide and labeled with a 3 year shelf life. The LacriCATH catheters are packaged with 1 device per pouch, and 2 pouches per carton. They are available by prescription only.

E. Intended Use/Indications for Use

The intended use and the Indications for use for the new LacriCATH model numbers are the same as for the predicate devices. Compared to the labeling for the predicate device, the modified labeling contains the same information but organized to be clearer to the intended user. There is no impact to the use of the device or the safety and effectiveness of the device.

{1}------------------------------------------------

Intended Use Statement:

The LacriCATH Lacrimal Duct Balloon Catheter is a sterile, prescription device intended for single use during dilatation of an obstructed nasolacrimal duct system to correct epiphora.

Indications for Use Statement:

The LacriCATH Lacrimal Duct Balloon Catheter is a sterile, prescription device intended for single use during dilatation of an obstructed nasolacrimal duct system to correct epiphora as a result of the following.

5 mm - LacriCATH Adult Catheter: Functional or complete nasolacrimal duct obstruction Dacryocystitis Failed DCR (Dacryocystorhinostomy)

2mm/3mm LacriCATH Pediatric catheter (intended for patients 12 months of age and older):

Congenital nasolacrimal duct obstruction Dacryocystocele

ட் Technological Characteristics:

The new LacriCATH catheters have the same technological characteristics compared to the predicate device. They have the same design and mode of operation. Only the length of the catheter has been modified, and the materials for the balloon, strain relief, and adhesive were updated. The sterilization method has been updated and the shelf life extended to 3 years. Table 1 below provides a comparison of the new LacriCATH catheters to the predicate device. This table illustrates equivalency of the LacriCATH catheters with the predicate device.

Predicate DeviceModified Device
510(k)K935233Under Review
Brand NameLacrimal Duct CatheterLacriCATH® Lacrimal DuctCatheter
Model #LDC213, LDC315, LDC508LDC213T, LDC315T, LDC508T,
ManufacturerIolab originally, then Quest Medical, Inc.Quest Medical, Inc
Materials
Balloon, OutertubingLDPE - low density PolyethyleneNylon balloonNo outer tubing
Sheath/SleeveTeflonSAME
Catheter304 stainless SteelSAME
AdhesiveLoctite™ 4541 - medical grade cyanoacrylateDymax 203A-CTH-F medical gradeadhesive

TABLE 1

{2}------------------------------------------------

Predicate DeviceModified Device
Luer ConnectorPolycarbonateSAME
Strain ReliefPolyether polyamide polymerOnflex TPE
Material:MarkersMedical Grade Ink - BlackSAME
Dimensions
Balloon Size2 mm diameter / 13 mm length-3 mm diameter/ 15 mm length5 mm diameter/ 8 mm lengthSAME
Balloon DiameterDeflated: 0.8 mm Inflated: 2 mm0.9 mm 3 mm1.1 mm 5 mmSAME
Catheter Length245 mm200 mm
Sterilization/Pkg
MethodGamma radiationEthylene Oxide (EtO)
Minimum SAL$1 x 10^{-6}$SAME
PackagingTyvek polyethylene/Mylar pouch; heat-sealedSAME
Shelf LifeOriginal: Indefinite as long as package integrity is not compromisedQuest: 3 years3 years

G. Non-clinical Performance Data:

Preclinical (bench) testing completed for the modified LacriCATH catheters included mechanical/functional testing, biocompatibility, physicochemical testing, shelf life testing, and sterilization residuals testing. There was no additional testing required as a result of the Risk Assessments performed for the modification for the new LacriCATH catheters. Verification and Validation tests were performed to ensure the modified LacriCATH catheters continued to meet or exceed specification in accordance to the Design Requirements. Those specifications are the same as for the predicate device. There were no clinical tests performed. The bench-testing concluded that the new modified LacriCATH catheters performed as well or better than the predicate device. Testing for the new LacriCATH Model numbers demonstrates they are as safe and as effective as the predicate device.

Clinical Testing ાન.

None required.

ட் Conclusions

Based on the information presented in this 510(k) premarket notification, the Quest Medical LacriCATH lacrimal duct catheters is considered substantially equivalent to the currently marketed predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT 19 2012

Quest Medical, Inc. c/o Amy Clendening-Wheeler Sr. Regulatory Affairs Specialist One Allentown Parkway Allen, TX 75002

Re: K113867

Trade/Device Name: LacriCATH®Lacrimal Duct Balloon Catheter Regulation Number: None Regulation Name: Lacrimal Stents and Intubation Sets Regulatory Class: Unclassified Product Code: OKS Dated: October 11, 2012 Received: October 12, 2012

Dear Ms. Clendening-Wheeler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely vours.

Debrah Falls

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): KII 3867

Device Name:

Indications for Use: The LacriCATH Lacrimal Duct Balloon Catheter is a sterile, prescription device indicated for single-use during dilatation of an obstructed nasolacrimal duct to correct epiphora as a result of the following:

5mm LacriCATH Adult catheter:

functional or complete nasolacrimal duct obstruction dacrvocystitis failed DCR (dacryocystorhinostomy)

2mm/3mm LacriCATH Pediatric catheter (intended for patients 12 months of age and older): congenital nasolacrimal duct obstruction dacryocystocele

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ivision Sign-Off)

Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices

510(k) Number_K113867

N/A