K Number
K113800

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-01-20

(28 days)

Product Code
Regulation Number
878.5010
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Quill™ Quadrahelix Knotless Tissue-Closure Device is indicated for soft tissue approximation excluding closure of the epidermis

Device Description

The Quill™ Quadrahelix Knotless Tissue Closure Device is a non-absorbable isotactic polymer comprised of polypropylene per 21 CFR 878.5010. It is available sterile, undyed (beige) in suture and needle configurations in USP Size 3-0. Each suture has bi-directional quadrahelix barbs along the long axis of the suture monofilament. The Quill™ Quadrahelix Knotless Tissue Closure Device approximates tissue by using the opposing barbs on the suture surface to imbed in the tissues after the surgeon precisely places the suture within the tissues.

AI/ML Overview

Based on the provided text, here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Tensile StrengthConforms to defined product requirements
Wound Security StrengthConforms to defined product requirements

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data). It only mentions "Non-clinical laboratory performance testing was conducted."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. The testing described is "Non-clinical laboratory performance testing," which implies physical tests rather than expert-driven evaluation of clinical data.

4. Adjudication Method for the Test Set

This information is not applicable and not provided. The study is non-clinical laboratory testing, not a human-reader-based assessment that would require adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The study was non-clinical laboratory testing for physical properties of the suture.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This concept is not applicable to the device described. The device is a physical surgical suture, not a software algorithm. The "standalone" testing refers to the device's physical properties without human intervention during the test itself (e.g., pulling strength tests).

7. Type of Ground Truth Used

The ground truth used for this device would be engineering specifications and established standards for tensile strength and wound security for surgical sutures. For example, specific forces (e.g., in Newtons) that a suture of a certain size must withstand without breaking or losing its wound-holding capacity.

8. Sample Size for the Training Set

This information is not applicable and not provided. The device is a physical product, not a machine learning algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable to this device.


Summary of the Study:

The study was a non-clinical laboratory performance testing focused on the physical properties of the Quill™ Quadrahelix Knotless Tissue-Closure Device. The key performance metrics assessed were tensile strength and wound security strength. The study concluded that the device conforms to defined product requirements for these metrics and is substantially equivalent to the predicate devices. The details regarding the specific methodologies, sample sizes, and quantitative results of these non-clinical tests are not included in the provided 510(k) summary. The "ground truth" was established by pre-defined engineering and manufacturing specifications for surgical sutures.

{0}------------------------------------------------

K113800 page 1/2

Section 6 – 510(k) Summary

JAN 2 0 2012

.

.

· .

、「

: . 、

.

.

・.

. 19

Date Prepared:December 22, 2011
Company:Angiotech
100 Dennis Dr.
Reading, PA 19606
Contact:Kirsten Stowell
Regulatory Affairs Manager
Phone:610-404-3367
Fax:610-404-3924
Email:kstowell@angio.com
Device trade name:Quill™ Quadrahelix Knotless Tissue-Closure Device
Device Common Name:Polypropylene Non-Absorbable Surgical Suture
Device classification:Non-Absorbable Polypropylene Surgical Suture
Product code, GAW
21 CFR 878.5010
Class II
Legally marketed devices to which the device is substantially equivalent:K052373: Quill™ Nonabsorbable Polypropylene Barbed Suture
Description of the device:The Quill™ Quadrahelix Knotless Tissue Closure Device is a non-absorbable isotactic polymer comprised of polypropylene per 21 CFR 878.5010. It is available sterile, undyed (beige) in suture and needle configurations in USP Size 3-0. Each suture has bi-directional quadrahelix barbs along the long axis of the suture monofilament. The Quill™ Quadrahelix Knotless Tissue Closure Device approximates tissue by using the opposing barbs on the suture surface to imbed in the tissues after the surgeon precisely places the suture within the tissues.
Indications for Use:Quill™ Quadrahelix Knotless Tissue Closure Device is indicated for soft tissue approximation excluding closure of the epidermis.

control control control controllers.

:

CONFIDENTIAL

{1}------------------------------------------------

K113800 Page 2/2

Section 6 - 510(k) Summary

Substantial Equivalence: The Quill™ Quadrahelix Knotless Tissue-Closure Device has the same intended use, fundamental scientific technological characteristics, material and size as the predicate device.

Performance tests:

Non-clinical laboratory performance testing was conducted to confirm that the Quill™ Quadrahelix Knotless Tissue-Closure Device conforms to the defined product requirements for tensile strength and wound security strength. The results of this testing demonstrates that the Quill™ Quadrahelix Knotless Tissue-Closure Device is substantially equivalent to the predicate devices.

CONFIDENTIAL

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Angiotech % Ms. Kirsten Stowell Regulatory Affairs Manager 100 Dennis Drive Reading, Pennsylvania 19606

JAN 2 0 2012

Re: K113800

Trade/Device Name: Quill™ Quadrahelix Knotless Tissue Closure Device Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable propylene surgical suture Regulatory Class: Class II Product Code: GAW Dated: December 22, 2011 Received: January 09, 2012

Dear Stowell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Kirsten Stowell

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

KI13800 510(k) Number (if known):

Device Name: Quill™ Quadrahelix Knotless Tissue Closure Device

Indications for Use:

Quill™ Quadrahelix Knotless Tissue-Closure Device is indicated for soft tissue approximation excluding closure of the epidermis

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Keene for NXM

Page 1 of

(Division Sign-(Division Sign-On)
Division of Surgical. (Orthopedic, ovision of Storative Devices

510(k) Number K113500

§ 878.5010 Nonabsorbable polypropylene surgical suture.

(a)
Identification. Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.