K Number
K113672

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-03-27

(105 days)

Product Code
Regulation Number
872.1800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PCH-2500 is digital extra oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, x-ray technicians and other professionals licensed by the law of the state in which the device is used.

Device Description

PCH-2500 is a dental digital radiographic imaging system which is available in two different image acquisition modes. Specifically designed for dental radiography of the teeth or jaws, PCH-2500 can be equipped with three dedicated sensors, each for two X-ray modalities : panoramic (Xmarul 501CF), cephalometric scan type (Xmaru2301CF) and oneshot ceph sensor type (1210SGA). PCH-2500 offers the digital panoramic and cephalometric X-ray modality for dental radiographs. The multi platforms of PCH-2500 imaging mode provides a wide range of imaging option based on the customer's diagnostic needs.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the PCH-2500 device, based only on the provided text:

Important Note: The provided text is a 510(k) summary for a medical device. This type of document is designed to demonstrate substantial equivalence to an already legally marketed device, not necessarily to provide a comprehensive clinical study report proving efficacy against a specific set of clinical acceptance criteria. Therefore, much of the requested information (like specific performance metrics for acceptance criteria, sample sizes for test/training sets, expert qualifications, and MRMC study details) is not present in the given text.


1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary for PCH-2500 primarily focuses on demonstrating substantial equivalence to the predicate device, PaX-Flex3D, by comparing technical characteristics and performance specifications. There are no explicit "acceptance criteria" for diagnostic accuracy or clinical outcomes stated. Instead, the performance is presented as meeting certain technical specifications.

CharacteristicAcceptance Criteria (from predicate/technical standards)Reported Device Performance (PCH-2500)
Indications for UsePCH-2500 is a digital extra-oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. Operated by licensed professionals.Same as Acceptance Criteria
Performance SpecificationPanoramic and cephalometric imaging capability (similar to predicate, excluding CT)Panoramic and cephalometric imaging capability
Tube Voltage50-90 kV (predicate)50-90 kV
Tube Current4-10 mA (predicate)4-10 mA
Focal Spot Size0.5 mm (predicate)0.5 mm
Exposure Time9-24 s (predicate)Max 20.2 s
Total Filtration2.8 mmAl (predicate)2.8 mmAl
Pixel Resolution (Panoramic)5 lp/mm (predicate)5 lp/mm
Pixel Resolution (Cephalometric - scan type)5 lp/mm (predicate)5 lp/mm
Pixel Resolution (Cephalometric - one shot type)N/A (predicate lacked this specific mode/sensor)3.9 lp/mm
Pixel Size (Panoramic)100 x 100 µm (predicate)100 x 100 µm
Pixel Size (Cephalometric - scan type)100 x 100 µm (predicate)100 x 100 µm
Pixel Size (Cephalometric - one shot type)N/A (predicate lacked this specific mode/sensor)127 x 127 µm
Safety, EMC, Performance StandardsCompliance with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, IEC 60601-1-2, NEMA PS 3.1-3.18, DICOM Set, IEC 61223-3-4, IEC 61223-3-5.All test results were satisfactory.

2. Sample Size Used for the Test Set and Data Provenance

The provided text does not specify a sample size for a test set in the context of clinical performance evaluation (e.g., number of patients/images used to assess diagnostic accuracy). The non-clinical performance and clinical consideration report for the SSXI detector (1210SGA) is "provided separately in this submission," but details are not given here.

The data provenance for the device performance testing mentioned (e.g., electrical, mechanical, EMC, acceptance tests according to IEC standards) would be from in-house engineering and testing as per standard regulatory requirements, likely conducted in Korea Republic (manufacturer's country). No clinical trial data is detailed. The expert review of image comparisons is mentioned, but without specifics of the dataset.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The text mentions an "expert review of image comparisons" for both the PCH-2500 and the predicate device. However, it does not specify the number of experts, their qualifications, or how they established a "ground truth" for a test set. This review seems to be part of the substantial equivalence argument, likely comparing image quality or diagnostic utility rather than rigorously establishing a ground truth for a clinical condition.

4. Adjudication Method for the Test Set

Since the text does not detail a clinical test set with specific ground truth establishment, no adjudication method (e.g., 2+1, 3+1) is described.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No MRMC comparative effectiveness study is mentioned or implied. The PCH-2500 is a digital X-ray imaging system, not an AI-powered diagnostic tool, and the submission predates widespread clinical AI in radiology.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

Not applicable. The PCH-2500 is a hardware imaging device, not a standalone algorithm.

7. The Type of Ground Truth Used

For the technical performance aspects, the "ground truth" is adherence to established technical standards (e.g., IEC, NEMA) and the performance of the predicate device. For potential image quality assessments by experts (implied by "expert review of image comparisons"), the ground truth would inherently be expert opinion/consensus on image quality and diagnostic features. Pathology or outcomes data are not mentioned as being used for ground truth.

8. The Sample Size for the Training Set

Not applicable/Not provided. As this is an X-ray imaging system, there isn't a "training set" in the machine learning sense. Any calibration or factory testing would involve proprietary methods not detailed here.

9. How the Ground Truth for the Training Set Was Established

Not applicable/Not provided. See point 8.

{0}------------------------------------------------

K113672
Page 1 of 5

MAR 2 7 2012

510(k) Submission - PCH-2500

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 872.1800.

Date

レ 12/8/2011

Manufacturer

Vatech Co., Ltd.

23-4, Seogu-Dong, Hwaseong-Si, Gyeonggi-Do, 445-170, Korea Republic

Tel: +82-31-379-9585 Fax: +82-31-379-9638 Contact person: Mr. Sonny Park

Official Correspondent (U.S. Designated agent)

Mtech Group 12946 Kimberley Ln, Houston, TX 77079 Tel: +713-467-2607 Fax: +713-464-8880 Contact person: Mr. Dave Kim (davekim@mtech-inc.net)

Trade/Proprietary Name:

PCH-2500

Common Name:

. Digital X-ray Imaging System

Classification Name:

System, X-Ray, Extra oral Source, Digital (21CFR 872.1800, Product code MUH, class2)

{1}------------------------------------------------

510(k) Submission - PCH-2500

Description:

PCH-2500 is a dental digital radiographic imaging system which is available in two different image acquisition modes. Specifically designed for dental radiography of the teeth or jaws, PCH-2500 can be equipped with three dedicated sensors, each for two X-ray modalities : panoramic (Xmarul 501CF), cephalometric scan type (Xmaru2301CF) and oneshot ceph sensor type (1210SGA).

PCH-2500 offers the digital panoramic and cephalometric X-ray modality for dental radiographs. The multi platforms of PCH-2500 imaging mode provides a wide range of imaging option based on the customer's diagnostic needs.

Indication for use:

PCH-2500 is digital extra oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, x-ray technicians and other professionals licensed by the law of the state in which the device is used.

· Predicate Device:

Manufacturer: Vatech Co., Ltd
Device: PaX-Flex3D
510(k) Number: K 102259 (Decision Date - 2/18/2011)

Substantial Equivalence:

PCH-2500 described in this 510(k) has the similar intended use and technical characteristics as PaX-Flex3D of Vatech Co., Ltd.

CharacteristicProposedVatech Co., Ltd.PCH-2500PredicateVatech Co., Ltd.PaX-Flex3D
510(k) number-K102259
Indicationsfor usePCH-2500 is digital extra oralsource x-ray system intendedPaX-Flex3D is a computedtomography x-ray system intended to
to take panoramic andcephalometric images of theoral and maxillofacial anatomyto provide diagnosticinformation for adult andpediatric patients. The deviceshould be operated and usedby dentists, x-ray techniciansand other professionalslicensed by the law of the statein which the device is used.produce panoramic, cephalometric orcross-sectional images of the oralanatomy on a real time basis bycomputer reconstruction of X-rayimage data from the same axial planetaken at different angles. Itprovides diagnostic details of theanatomic structures by acquiring 360rotational image sequences of oraland maxillofacial area for a precisetreatment planning in adult andpediatric dentistry. The device isoperated and used by physicians,dentists, and X-ray technicians.
PerformanceSpecificationPanoramic and cephalometricPanoramic and cephalometricand compute tomography
Input VoltageAC 100-120 / 200-240 VAC 110 / 230 V
Tube Voltage50-90 kV50-90 kV
Tube Current4-10 mA4-10 mA
Focal Spot Size0.5 mm0.5 mm
Exposure TimeMax 20.2 s9-24 s
Total Filtration2.8 mmAl2.8 mmAl
Pixel ResolutionPanoramic : 5 lp/mmPanoramic : 5 lp/mm
Cephalometric (scan type)5 lp/mm :Cephalometric (scan type)5 lp/mm :
Cephalometric (one shot type)3.9 lp/mm :
CT – N/ACT - 3.3 lp/mm
Pixel SizePanoramic : 100 x 100 µmPanoramic - 100 x 100 µm
Cephalometric (scan type):100 x 100 µm :Cephalometric (scan type):100 x 100 µm :
Cephalometric (one shot type)
127 x 127 µm :
CT - N/ACT : 150 x 150 µm/200 x 200 µm
Image ReceptorCMOS photodiode array –panoramic (Xmaru1501CF) &cephalometric(Xmaru2301CF)Amorphous silicon TFT withscintillator – Cephalometric(1210SGA)CMOS photodiode array –panoramic (Xmaru1501CF) &cephalometric (Xmaru2301CF)

{2}------------------------------------------------

510(k) Submission – PCH-2500

:

. : .

672

5

{3}------------------------------------------------

K 1111672
Page 4 of 5

510(k) Submission - PCH-2500

Indications for use, safety characteristics, and non-clinical performance for panoramic and cephalometric sensors (scan type) of PCH-2500 and PaX-Flex3D are similar. The primary differences are as follows: PCH-2500 lacks the cone beam CT mode. PCH-2500 offers optional solid state X-ray sensor for cephalometric mode. (The non-clinical performance and clinical consideration report for the SSXI detector, 1210SGA, are provided separately in this submission). Based on the non-clinical and clinical consideration and the outcome of an expert review of image comparisons for both devices, we can claim the substantial equivalence of PCH-2500 in comparison with PaX-Flex3D, the predicate device, in terms of safety and effectiveness.

Safety, EMC and Performance Data:

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1(A1+A2, 1995), IEC 60601-1-1 (2001), IEC 60601-1-3 (Ed. 1, 1994), IEC 60601-2-7 (1998), IEC 60601-2-28 (Ed. 1, 1993) and IEC 60601-2-32 (Ed. 1, 1994) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2.

PCH-2500 also meets the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.

{4}------------------------------------------------

K1111672
Page 5 of 5

510(k) Submission - PCH-2500

Clinical considerations according to FDA Guidance Non-clinical & & "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" were performed.

Acceptance test according to IEC 61223-3-4 and IEC 61223-3-5 was performed.

All test results were satisfactory.

Conclusion:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Vatech Co., Ltd. concludes that PCH-2500 is safe and effective and substantially equivalent to predicate device as described herein.

END

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three lines representing the eagle's feathers. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

MAR 2 7 2012

Vatech Co., Ltd. Mr. Dave Kim Official Correspondent Mtech Group 12946 Kimberly Lane HOUSTON TX 77079

Re: K113672

Trade/Device Name: PCH-2500 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH and MOB Dated: February 14, 2012 Received: February 16, 2012

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{6}------------------------------------------------

Page 2

medical device-related adverse events) (21 CFR 803); and good manufacturing practice mcuirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as set form in the quarty, I device as described in your Section 510(k) premarket will anow you to begin marketing your covivalence of your device to a legally marketed nontication. The I Dr I mailing of succion for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you decire operitio da research Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

KIN672

Indications for Use

510(k) Number(if known):

Device Name: PCH-2500

Classification: System, X-Ray, Extraoral Source, Digital (21 CFR 872.1800, Product code MUH, Class2)

Indications for Use:

PCH-2500 is digital extra oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, and x-ray technicians and other professionals licensed by the law of the state in which the device is used.

Prescription Use

(Part 21 CFR 801 Subpart D)

Over-The-Counter Use_ AND/OR (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)

(Division Sign-Off)
Division of Radiological Devices
Office of in Vitro Diagnostic Device Evaluation and Safety
510K. K113672

Concurrence of CDRH, Office of Device Evaluation(ODE)

Page 1 of 1

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.