(455 days)
The Gmate® VOICE Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, hand, upper arm, forearm, calf or thigh as an aid in monitoring the effectiveness of diabetes management in the home by individuals with diabetes. The Gmate® VOICE Blood Glucose Monitoring System is intended to be used by a single user and should not be shared with any other person.
The Gmate® VOICE Blood Glucose Monitoring System is for self-testing outside the body (in vitro diagnostic use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Gmate® VOICE Blood Glucose Monitoring System should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
The Gmate® VOICE Blood Glucose Monitoring System includes a speaking feature that provides audible test results for diabetic users.
The Gmate® Blood Glucose Test Strips are for use with the Gmate® VOICE Meter for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, hand, upper arm, forearm, calf or thigh.
The Gmate® Control Solution is for use with the Gmate® VOICE Blood Glucose Monitoring System and is intended as a quality control measure to verify the accuracy of your blood glucose test results and to ensure that the Gmate® VOICE meter and Gmate® Test Strips are working properly. The Gmate® Control Solution is intended for use by people with diabetes at home.
The Gmate® VOICE Blood Glucose Monitoring System is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood, which is used with the Gmate® Blood Glucose Test Strips.
The test principle is:
This device is an in vitro diagnostic product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the glucose oxidase that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. This reaction is measured by the Meter and displayed as your blood glucose result.
This document is a 510(k) summary for the Gmate® VOICE Blood Glucose Monitoring System, which primarily focuses on establishing substantial equivalence to a predicate device rather than detailing specific clinical study results against acceptance criteria. Therefore, much of the requested information regarding acceptance criteria and study particulars for a medical AI device is not present. However, I can extract the available relevant information and highlight the missing details based on the context of a 510(k) submission.
Missing Information:
This document is a 510(k) summary for a blood glucose monitoring system, not specifically an AI-powered diagnostic device in the modern sense. Consequently, several of the requested categories are not applicable or not detailed in this type of regulatory submission. Specifically, there is no mention of "AI" or "machine learning," and therefore no information on:
- Multi-reader multi-case (MRMC) comparative effectiveness study
- Effect size of human readers with vs. without AI assistance
- Standalone algorithm performance
- Training set sample size
- How ground truth for the training set was established
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in the format typically seen for an AI diagnostic device. However, regulatory submissions like this often implicitly adhere to performance benchmarks for blood glucose meters. The study described is an "accuracy study" that aims to compare the device's performance against a reference method and implicitly against established standards for blood glucose meters.
| Acceptance Criteria (Implied for Blood Glucose Meters) | Reported Device Performance (from Accuracy Studies as implied by 510(k)) |
|---|---|
| (Not explicitly stated in this 510(k) summary) | Not detailed in this 510(k) summary. The document states "The Gmate® VOICE Blood Glucose Monitoring System is substantially equivalent to the following predicate device: OneTouch® ULTRA® System Manufactured by LifeScan, Inc., K002134." This implies that the device's performance met equivalence standards based on studies comparing it to the predicate. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not specified in this 510(k) summary.
- Data Provenance: Not specified in this 510(k) summary. The company has offices in New York and Seoul, but the origin of the study data is not stated. The studies would typically be prospective to evaluate a new device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This is not applicable as the device measures blood glucose which is typically compared against a laboratory reference method, not expert-established ground truth.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable for a blood glucose monitoring system that compares readings to a reference lab method.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a blood glucose meter, not an AI-assisted diagnostic tool requiring human reader interpretation. No AI component is mentioned.
6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done
This device operates as a standalone blood glucose meter. The "algorithm" is the electrochemical detection and glucose concentration calculation. The performance described in the underlying studies (not detailed in this document) would be of the device in its entirety without human-in-the-loop performance modifications, other than the user performing the test.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For blood glucose monitoring systems, the ground truth is typically established by laboratory reference methods (e.g., YSI glucose analyzer) which are considered highly accurate and precise.
8. The sample size for the training set
Not applicable or not specified. This is a traditional medical device, not an AI model requiring a distinct "training set" in the machine learning sense. The device is calibrated and validated through internal testing before market submission.
9. How the ground truth for the training set was established
Not applicable or not specified, for the same reasons as #8. Ground truth for calibration and validation would be established via highly accurate laboratory reference methods.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Philosys, Inc. The logo consists of a stylized graphic to the left of the company name. The company name is written in a bold, sans-serif font.
New York Office 304 Park Avenue South, Suite 218 New York, NY 10010 USA TEL: 855-GO-GMATE FAX: 800-939-8670
MAR
8 2013
Seoul Office 635-5 Yeoksam-dong, Kangnam-ku, Seoul, 135-908, Korea TEL: +82-2-553-1421 FAX: +82-2-553-1423
510(k) SUMMARY (As required by 21.CFR.807.92)
ﺮ"
| Introduction: | According to the requirements of 21 CFR.807.92, the following information provides sufficient data to understand the basis. for a determination of substantial equivalence. |
|---|---|
| Submitted By: | Philosys Co Ltd304 Park Avenue SouthSuite 218New York, NY 10010 |
| Contact Person: | Linda ChanPhone: 917-757-3793Fax: 800-931-9137 |
| Date Summary, Prepared: | December 8, 2011 |
| Device Name: | Propriety Name: Gmate® VOICE Blood Glucose Monitoring System and ChCommon Name: Blood Glucose Test SystemClassification Name: Class II, 862.1345 Blood Glucose Monitoring SystemProduct Code:CGA - Glucose Oxidase, GlucoseNBW - System, Test, Blood Glucose, Over the Counter |
| Propriety Name: Gmate® Control SolutionClassification Name: Class I, 21 CFR 862.1660, Quality Control Material (assayed and unassayed)Product Code:JJX - Single Analyte Controls (assayed and unassayed) | |
| Predicate Device: | We claim substantial equivalence to the OneTouch® ULTRA® System Manufactured by LifeScan, Inc., K002134. |
| Device Description: | The Gmate® VOICE Blood Glucose Monitoring System is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood, which is used with the Gmate® Blood Glucose Test Strips. |
The test principle is:
This device is an in vitro diagnostic product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the glucose oxidase that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. This reaction is measured by the Meter and displayed as your blood glucose result.
{1}------------------------------------------------
New York Office 304 Park Avenue South, Suite 218 New York, NY 10010 USA TEL: 855-GO-GMATE FAX: 800-939-8670
Seoul Office 635-5 Yeoksam-dong, Kangnam-ku, Seoul, 135-908, Korea TEL: +82-2-553-1421 FAX: +82-2-553-1423
510(k) SUMMARY (As required by 21.CFR.807.92)
Intended Use:
The Gmate® VOICE Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, hand, upper arm, forearm, calf or thigh as an aid in monitoring the effectiveness of diabetes management in the home by individuals with diabetes. The Gmate® VOICE Blood Glucose Monitoring
System is intended to be used by a single user and should not be shared with any other person.
Gmate® VOICE Blood Glucose Monitoring System is for self-testing outside the body (in vitro diagnostic use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Gmate® VOICE Blood Glucose Monitoring System should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
The Gmate® VOICE Blood Glucose Monitoring System includes a speaking feature that provides audible test results for diabetic users.
The Gmate® Blood Glucose Test Strips are for use with the Gmate® VOICE Meter for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, hand, upper arm, forearm, calf or thigh.
The Gmate® Control Solution is for use with the Gmate® VOICE Blood Glucose Monitoring System and is intended as a quality control measure to verify the accuracy of your blood glucose test results and to ensure that the Gmate® VOICE meter and Gmate® Test Strips are working properly. The Gmate® Control Solution is intended for use by people with diabetes at home.
| Comparison toPredicate Device: | The Gmate® VOICE Blood Glucose Monitoring System is substantiallyequivalent to the other products in commercial distribution intended forsimilar use. The most notable, it is substantially equivalent to the currentlymarketed item, OneTouch® ULTRA® System Manufactured by LifeScan, Inc.,K002134. |
|---|---|
| Conclusion: | The Gmate® VOICE Blood Glucose Monitoring System is substantiallyequivalent to the following predicate device:OneTouch® ULTRA® System Manufactured by LifeScan, Inc., K002134. |
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a human figure with outstretched arms, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2013
Philosys. Inc. C/O Linda Chan 304 Park Avenue South, Suite 218 New York, NY 10010
Re: K113636
Trade/Device Name: Gmate VOICE™ Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CGA, NBW, JJX Dated: February 14, 2013 Received: February 15, 2013
Dear Ms. Chan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for
{3}------------------------------------------------
Page 2-Ms. Chan
the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Carol G. Benson -S for
Courtney H. Lias, Ph.D.
Director
Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics
and Radiological Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K113636
Gmate® VOICE Blood Glucose Monitoring System Device Name:
Indications for Use:
The Gmate® VOICE Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingerting, hand, upper ann, forearm, calf or thigh as an aid in monitoring the effectiveness of diabetes magagement in the home by individuals with diabetes. The Gmate® VOICE Blood Glucose Monitoring System is intented to be used by a single user and should not be shared with any other person.
The Gmate® VOICE Blood Glucose Monitoring System is for self-testing outside the body (in vitro diagnostic use only) by people with diabetes at home as an aid to monitor the effectiveness of diabels control. The Gmate® VOICE Blood Glucose Monitoring System should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-ste times (when glucose is not changing rapidly).
The Cmate® VOICE Blood Glucose Monitoring System includes a speaking feature that provides audible test results for diabetic users.
The Gmate® Blood Glucose Test Strips are for use with the Gmate® VOICE Meter for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fine grannumment upper arm, forearm, calf or thigh.
The Gmate® Control Solution is for use with the Gmate® VOICE Blood Glucose Monitoring System and is intended as a quality control measure to verify the accuracy of your blood glucose test results and to ensure that the Gmate® VOICE meter and Gmate® Test Strips are working properly. The Gmate® and to Control Solution is intended for use by people with diabetes at home.
Prescription Use AND/OR Over-The-Counter Use__X (Part 21 CFR 801 Subpart D) (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Kathering Serrano
Division Sign-Off Office of In Vitro Diagnostic and Radiological Health
510(k) K113636
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.