K Number
K113627

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-02-17

(71 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Hemasorb Resorbable Hemostatic Bone Putty is intended for use in the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade. The material may be used during surgical procedures and in treating traumatic injuries.

Device Description

Orthocon Hemasorb Resorbable Hemostatic Bone Putty is a sterile, soft, moldable, biocompatible, absorbable material of putty-like consistency intended for use in the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade. The material is a mixture of calcium stearate, vitamin E acetate and a liquid surfactant. The material is virtually odorless, off-white in color and can be spread easily with minimal adhesion to surgical gloves. Hemasorb requires no kneading prior to application and does not soften appreciably at body temperature. The subject of this 510(k) is the addition of a spatula as an alternative method of Hemasorb application.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called Hemasorb® Resorbable Hemostatic Bone Putty. The submission is for a modification to an already cleared device, specifically the addition of a spatula for application.

Based on the information provided, the "study" conducted is a non-clinical performance testing to demonstrate the substantial equivalence of the modified device (with the spatula) to previously cleared predicate devices.

Here's the breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Implied)Reported Device Performance
Integrity & PerformanceAbility to maintain integrity under simulated use and shipping conditions, and to function as intended."The device was determined to... demonstrate adequate integrity and performance characteristics."
Resistance to cyclic compression and tensile failure (for the spatula/device as a system).Specific thresholds not given, but results confirmed "adequate integrity and performance characteristics."
Overall stability (including sterile package integrity)"stability (including sterile package integrity)" was tested and deemed adequate.
BiocompatibilityAbsence of adverse biological reactions (implied for the material itself)."The device was determined to be biocompatible."
Mechanical PropertiesMechanical performance relevant to handling and application (e.g., tensile failure, cyclic compression).Test results (though not quantitatively detailed) indicated adequate performance.
Material PropertiesConsistency, odor, color, spreadability, adhesion to gloves, softening at body temperature.Described as "soft, moldable," "virtually odorless, off-white," "spread easily with minimal adhesion to surgical gloves," and "does not soften appreciably at body temperature." (This is a description of the device, not necessarily a performance outcome of the study, but it speaks to its inherent characteristics).

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: The document does not specify a numerical sample size for the non-clinical tests. It refers to "simulated use and shipping" and "mechanical tests," implying a methodology but not concrete numbers of units tested.
  • Data Provenance: The study is described as "in vitro tests." No information is provided regarding country of origin of data, nor whether it's retrospective or prospective, as it's a non-clinical, laboratory-based evaluation of a product, not human subject data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This question is not applicable to this submission. The "ground truth" for non-clinical performance testing typically refers to engineering standards, validated test methods, and pre-defined specifications rather than expert human interpretation of results. The determination of "adequate integrity and performance characteristics" would have been made by the manufacturer's R&D and Quality teams against these internal specifications. No external experts are mentioned for establishing ground truth.

4. Adjudication Method for the Test Set

This question is not applicable. Adjudication methods (like 2+1, 3+1) are used in studies involving human interpretation (e.g., image reading) where there might be disagreement among readers on a diagnosis or finding. This was a non-clinical, objective performance test.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. This submission is for a medical device (bone putty and a spatula), not an AI-powered diagnostic or interpretive tool. Therefore, no MRMC comparative effectiveness study was performed, and there is no AI component.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this type of non-clinical performance testing is based on pre-defined engineering specifications, validated test methods, and industry standards for material performance, integrity, and biocompatibility. The goal was to show that the modified device (with spatula) met these performance characteristics and was equivalent to the predicate devices.

8. The Sample Size for the Training Set

This question is not applicable. There is no "training set" as this is not a machine learning or AI-based device.

9. How the Ground Truth for the Training Set was Established

This question is not applicable. As there is no training set, this information is not relevant.

{0}------------------------------------------------

K113627 68 1 64 5

FEB 1 7 2012

510(k) Summary

Contact:Rosemary HarryVice President, Regulatory and Quality AffairsOrthocon, Inc1 Bridge Street, Suite 121Irvington, NY 10533harry@orthocon.com914-357-2600
Date Prepared:January 18, 2012
Device Trade Name:Hemasorb® Resorbable Hemostatic Bone Putty
Manufacturer:Orthocon, Inc.1 Bridge StreetSuite 121Irvington, NY 10533
Common Name:Wax, bone
Classification:None applicable
Class:Unclassified
Product Code:MTI

Indications For Use:

Hemasorb Resorbable Hemostatic Bone Putty is intended for use in the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade. The material may be used during surgical procedures and in treating traumatic injuries.

Device Description:

Orthocon Hemasorb Resorbable Hemostatic Bone Putty is a sterile, soft, moldable, biocompatible, absorbable material of putty-like consistency intended for use in the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade. The material is a mixture of calcium stearate, vitamin E acetate and a liquid surfactant. The material is virtually odorless, off-white in color and can be spread easily with minimal adhesion to surgical gloves. Hemasorb requires no kneading prior to application and does not soften appreciably at body temperature. The subject of this 510(k) is the addition of a spatula as an alternative method of Hemasorb application.

Predicate Devices:

Substantial equivalence was shown with the previously cleared devices in K043260 and K102762.

{1}------------------------------------------------

Summary of Non-Clinical Performance Testing

The in vitro tests performed were: simulated use and shipping (including mechanical tests of cyclic compression and tensile failure testing), finite element analysis, and stability (including sterile package integrity).

The device was determined to be biocompatible and to demonstrate adequate integrity and performance characteristics.

Conclusion:

The modification subject to this 510(k) was shown to be substantially equivalent to previously cleared devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure, composed of three curved lines.

Public Health Service

FEB 1 7 2012

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Orthocon, Inc. % Ms. Rosemary Harry VP. RA/QA 1 Bridge Street, Suite 121 · Irvington, New York 10533

Re: K113627

Trade/Device Name: Hemasorb® Resorbable Hemostatic Bone Putty Regulatory Class: Unclassified · Product Code: MTJ Dated: January 18, 2012 Received: January 20, 2012

Dear Ms. Harry:

We have reviewed your Section 510(k) premarket notification of intent to market the device wo have a row a your and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Fror (. 10.) has the device, subject to the general controls provisions of the Act. The I ou may, thereforminens of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability addition. Trease now. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register_

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a and regulations administered by other Federal agencies. You must or uny 1 cithe Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CI K I all 607), labornig (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Rosemary Harry

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Mark A. Milikurn

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

4. Indications for Use

510(k) Number (if known): 长川3 627

Device Name: Hemasorb® Resorbable Hemostatic Bone Putty

Hemasorb Resorbable Hemostatic Bone Putty is intended for use in the control of Flemaboro Tesserouser Traged bone by acting as a mechanical barrier or tamponade. The material may be used during surgical procedures and in treating traumatic injuries.

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

K11362

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Moore for NXM
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K113627

N/A