K Number
K113597

Validate with FDA (Live)

Date Cleared
2012-06-03

(181 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.

Device Description

The MiniLap Bipolar Electrocautery devices are a family of disposable bipolar electrosurgical graspers that directly penetrate soft tissue to access certain areas of the human anatomy without the need for a traditional insertion conduit. These disposable instruments are used for electrosurgical cauterization during general laparoscopic procedures. Fabricated from a stainless steel alloy, MiniLap Bipolar instruments consist of an integrated 2.4mm (+/- .1mm) insulated needle/cannula shaft that houses a retractable grasper instrument. The shaft can be introduced percutaneously to the surgical site, after which the working portion of the instrument can be deployed to approximate and cauterize soft tissue. Each BiPolar instrument has a male connection provided as an integral part of the instrument that may be utilized for the bipolar connection when attached to standard two plug bipolar cables and their generators. The bipolar connection is compatible with manufacturers of standard dual female bipolar plug cables. The instruments come in multiple lengths from 150mm to 300mm.

AI/ML Overview

The provided 510(k) summary describes a medical device called "MiniLap Bipolar Electrocautery Devices." However, it does not contain any information about acceptance criteria or a study designed to prove the device meets specific performance metrics in the way a diagnostic AI/ML device would.

This document is a submission for a traditional medical device (electrosurgical grasper), not an AI/ML-driven diagnostic tool. The performance claim for such a device focuses on mechanical, electrical, and biocompatibility aspects, and its "acceptance" is typically based on demonstrating substantial equivalence to a predicate device.

Therefore, many of the requested sections about studies, sample sizes, ground truth, and expert-read performance are not applicable to this type of device submission and are not found in the provided text.

Here's how to address the request based on the provided text, highlighting the absence of AI/ML specific information:

Acceptance Criteria and Device Performance

The provided document doesn't explicitly state "acceptance criteria" in the format of performance metrics for an AI/ML device (e.g., sensitivity, specificity, AUC). Instead, the acceptance is based on demonstrating substantial equivalence to predicate devices through a comparison of technological characteristics and compliance with general safety and performance standards for electrosurgical instruments.

The "performance data" section states: "The subject device has been subjected to and passed a variety of mechanical and electrical tests and evaluations. Additionally, the device is composed of biocompatible materials with a history of usage in the medical device industry." This broadly indicates that the device met internal performance benchmarks related to its function as a surgical tool.

Table of Acceptance Criteria and Reported Device Performance (as inferred for a traditional medical device):

Acceptance Criteria Category (Inferrred from 510(k) structure)Reported Device Performance (from text)
Mechanical Performance"Subjected to and passed a variety of mechanical tests and evaluations."
Electrical Performance"Subjected to and passed a variety of electrical tests and evaluations."
Biocompatibility"Composed of biocompatible materials with a history of usage in the medical device industry."
Technological Characteristics Equivalence (to Predicate A - Richard Wolf Medical Instruments Corp.)- Max Power: Device: 70W max, Predicate: 50W max (Device has higher max) - Normal Operating Range: Device: 20-35W, Predicate: 20-35W (Equivalent) - Safety: Device: Safety Interlock, Insulated shaft/jaws, IEC compliant. Predicate: No safety interlock, Insulated jaws, IEC compliant. (Device has more safety features) - Sterilization: Device: Sterile, single use. Predicate: Non sterile, reusable. (Difference noted) - Needle Diameter: Device: 2.4mm ± .1mm. Predicate: 3.5mm, 5mm. (Device has smaller options) - Length: Device: 150-300mm. Predicate: 320mm, 450mm. (Device has shorter options)
Technological Characteristics Equivalence (to Predicate B - MINI LAP Instruments K070686)- Needle Diameter: Device: 2.4mm ± .1mm, Predicate: 2.4mm (Equivalent) - Device Length: Device: 150-300mm, Predicate: 150-300mm (Equivalent) - Material Composition: Device: Medical Grade Stainless Steel 17-7 SS, 300 SS. Predicate: Medical Grade Stainless Steel 470 SS, 300 SS. (Similar) - Sterilization: Device: Sterile, single use. Predicate: Sterile, single use. (Equivalent) - Rotation: Device: 175 degree shaft rotation each direction. Predicate: 360 degree shaft rotation. (Difference noted)
Intended Use Equivalence"The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue." (Compared to predicate indications for open and laparoscopic surgery for grasping, coagulating, and transecting tissue or penetrating soft tissue to access areas and grasp/manipulate tissues).

1. Sample sizes used for the test set and the data provenance:

  • Not Applicable. This is a physical medical device (surgical instrument), not an AI/ML diagnostic system. There is no "test set" in the context of image data or clinical scenarios for an algorithm. Testing involved mechanical, electrical, and biocompatibility evaluations of the device itself. The provenance of such "data" would be internal lab testing, not patient data.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. As above, there is no "test set" in the AI/ML sense, and thus no ground truth established by experts for algorithmic performance. The "acceptance" of the device is based on engineers and regulatory experts reviewing its design, testing results, and comparison to predicate devices.

3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not Applicable. No adjudiciation method for a clinical test set is relevant here.

4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is a physical surgical instrument, not an AI-assisted diagnostic tool.

5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. There is no algorithm to evaluate.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not Applicable. The "ground truth" for a physical device's performance is typically defined by engineering standards, material specifications, and operational parameters (e.g., electrical impedance, material strength, insulation effectiveness).

7. The sample size for the training set:

  • Not Applicable. There is no AI/ML model that requires a training set.

8. How the ground truth for the training set was established:

  • Not Applicable. There is no AI/ML model that requires a training set or its associated ground truth.

{0}------------------------------------------------

510(k) Summary

JUN - 3 2012

Date Prepared [21 CFR 807.92(a)(1)] November 1, 2011, Updated May 31, 2012

Submitter's Information [21 CFR 807.92(a)(1)]

Regulatory Contact Allan Alward 88 Ashford Avenue Dobbs Ferry, NY 10522

Sponsor/Manufacturer Mini Lap Technologies, Inc. 88 Ashford Avenue Dobbs Ferry, NY 10522 Contact: Dr. S. Ravikumar Tel: 914 591 8400

FDA Establishment Registration Number is 3007123990

Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)|

Trade Names

  • MINI LAP Bipolar Electrocautery Devices ◆
    Device Common, Usual or Classification Names

Laparoscopic Instruments, Graspers, Cannula, Trocar, Manual Surgical Instruments, Bipolar Instruments

Classification Panel

Classification of this device would fall under the responsibility of the General & Plastic Surgery panel.

Class

Class 2 device under the following product codes/regulations:

  • KOG, 21 CFR 876.1500, Endoscope Accessories
  • 트 FBQ, 21 CFR 878.5090, Trocar
  • . GEI, 21 CFR 878.4400, Electrosurgical cutting and coagulation accessories

{1}------------------------------------------------

Predicate Device [21 CFR 807.92(a)(3)]

The following devices have been identified as predicate devices:

  • . Wolf Medical Instruments Bipolar Forceps - K023813
  • MINI LAP Instruments K070686 .

Description of the Device [21 CFR 807.92(a)(4)]

The MiniLap Bipolar Electrocautery devices are a family of disposable bipolar electrosurgical graspers that directly penetrate soft tissue to access certain areas of the human anatomy without the need for a traditional insertion conduit. These disposable instruments are used for electrosurgical cauterization during general laparoscopic procedures.

Fabricated from a stainless steel alloy, MiniLap Bipolar instruments consist of an integrated 2.4mm (+/- .1mm) insulated needle/cannula shaft that houses a retractable grasper instrument. The shaft can be introduced percutaneously to the surgical site, after which the working portion of the instrument can be deployed to approximate and cauterize soft tissue. Each BiPolar instrument has a male connection provided as an integral part of the instrument that may be utilized for the bipolar connection when attached to standard two plug bipolar cables and their generators. The bipolar connection is compatible with manufacturers of standard dual female bipolar plug cables.

The instruments come in multiple lengths from 150mm to 300mm.

Intended Use [21 CFR 807.92(a)(5)]

The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.

Technological Characteristics [21 CFR 807.92(a)(6)]

We believe the MINI LAP Bipolar Electrocautery Devices are substantially equivalent to the predicate devices. see chart

{2}------------------------------------------------

CharacteristicMini Lap TechnologiesRichard Wolf MedicalInstruments Corp.Mini Lap Technologies
BiPolar Electrocautery DevicesMINI LAP Instruments
Bipolar Forceps
510(k)K113597K023813K070686
Indications for UseThe MiniLap Technologies familyof BiPolar instruments withbipolar cautery have applicationsin a variety of general, thoracic,gynecologic (except for use infemale sterilization), urologic,laparoscopic and endoscopicprocedures for manipulation andcoagulation of tissue.For use in open and laparoscopicsurgery where grasping,coagulating, and transectingtissue is indicated.Penetrate soft tissue to accesscertain areas of the humananatomy. The devices are used tograsp, hold and manipulate othersoft internal tissues as well asitems such as hernia mesh.
Needle diameter2.4mm ± .1mm3.5 mm, 5 mm2.4 mm
Device length150-300 mm320 mm, 450 mm150-300 mm
Material CompositionMedical Grade Stainless Steel17-7 SS, 300 SSMedical Grade Stainless SteelMedical Grade Stainless Steel470 SS, 300 SS
SterilizationSterile, single useNon sterile, reusableSterile, single use
Power SourceBipolar Electrosurgical GeneratorBipolar Electrosurgical GeneratorN/A
Power Ranges70 watts maximum, 20-35 wattsnormal operating range50 watts maximum, 20-35 wattsnormal operating range
SafetySafety InterlockSafety Interlock is not applicableSafety Interlock
Insulated shaft and jawsInsulated jaws
Compliant to applicable IECelectrical safety standardsCompliant to applicable IECelectrical safety standards
Rotation175 degree shaft rotation, eachdirection360 degree shaft rotation

Technological Characteristics Comparison Chart

Performance Data [21 CFR 807.92(b)(1)]

The subject device has been subjected to and passed a variety of mechanical and electrical tests and evaluations. Additionally, the device is composed of biocompatible materials with a history of usage in the medical device industry.

Conclusion [21 CFR 807.92(b)(3)]

We believe the changes are minimal and conclude that the subject device is as safe and effective as the predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing-like shapes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MiniLap Technologies % Mr. Allan Alward Vice President Research and Development 88 Ashford Ave. Dobbs Ferry. NY 10522

JUN. - 3 2012

Re: K113597

Trade/Device Name: Mini-Lan Bipolar Electrocautery Devices Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: May 15, 2012 Received: May 16, 2012

Dear Mr. Alward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Mr. Allan Alward

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic & Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K113597

Device Name: Mini-Lap Bipolar Electrocautery Devices

Indications for Use:

The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sign-Off

(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic.
Division of Surgical, Orthopedic. Division of Surgices and Restorative Devices

K113597

510(k) Number

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.