K Number
K113576

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-05-18

(165 days)

Product Code
Regulation Number
868.5740
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ETView Double Lumen Video Tracheoscope (DLVT™) is intended to isolate the left or right lung of a patient for intensive care or surgery, one lung ventilation or one lung anesthesia. The DLVT™ System is indicated for verifying tube placement and repositioning.

Device Description

The ETView Double Lumen Video Tracheoscope (DLVT™) System functions as a standard 37 French size endobronchial tube that additionally has an embedded video imaging device in its tracheal lumen. The system provides a video image of the patient's bronchus, which is displayed on the monitor, for as long as the DLVT™ is inside the patient's bronchus. Visualization of the bronchus is used to verify placement and repositioning of the endobronchial tube during the intubation procedure or throughout surgery. The device is indicated for non-MRI environment only.

AI/ML Overview

The provided text describes the ETView Double Lumen Video Tracheoscope (DLVT™) System and its performance testing for 510(k) clearance, but it does not include acceptance criteria in a table format, nor does it detail a study that explicitly 'proves' the device meets acceptance criteria in the way a clinical trial might for AI/ML devices. Instead, the document outlines a series of performance tests to demonstrate substantial equivalence to predicate devices.

However, based on the provided text, I can infer the general approach and key information relevant to your request, adapting the structure to fit the available data.

Here's a breakdown of the requested information based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria in a quantitative table with corresponding device performance metrics for features like sensitivity, specificity, or accuracy, which are typical for AI/ML devices. Instead, the acceptance criteria are demonstrated through compliance with established medical device standards and functional testing for properties like flow, diameter, and material integrity. The "reported device performance" is essentially that the device "complies with the following standards" and "demonstrated that the Double Lumen Video Tracheoscope (DLVT™) is as safe and effective as the cleared predicate devices."

Acceptance Criteria (Inferred from Standards & Testing Goals)Reported Device Performance (Summary)
Material & Design Standards Compliance:
ISO 5361:1999 Anaesthetic and respiratory equipment -- Tracheal tubes and connectorsComplies with the standard.
ISO 16628:2008 Tracheobronchial Tubes Sizing and markingComplies with the standard.
Sterilization Standards Compliance:
ANSI/AAMI/ISO 11135-1:2007 Sterilization of health care products - Ethylene oxideComplies with the standard.
AAMI TIR28:2001 Product adoption and process equivalency for ethylene oxide sterilizationComplies with the standard.
Risk Management & Biocompatibility Standards Compliance:
ISO 14971-1:2007 Risk management for medical devicesComplies with the standard.
ISO 10993-1:2003(E), Biological evaluation of medical devices -- Part 1: Evaluation and testingComplies with the standard.
Functional Performance (bench & limited in-vivo):
Determination of Cuff Resting DiameterPerformance demonstrated to be safe and effective as predicate.
Cuff Leak Resistance IntegrityPerformance demonstrated to be safe and effective as predicate.
Resistance to Cuff HerniationPerformance demonstrated to be safe and effective as predicate.
Cuff SymmetryPerformance demonstrated to be safe and effective as predicate.
Resistance to tube collapsePerformance demonstrated to be safe and effective as predicate.
Air Flow ResistancePerformance demonstrated to be safe and effective as predicate.
Thermal SafetyPerformance demonstrated to be safe and effective as predicate.
Determination of effective inside diameterPerformance demonstrated to be safe and effective as predicate.
Tracheal SealPerformance demonstrated to be safe and effective as predicate.
Determination of the bronchial segmentPerformance demonstrated to be safe and effective as predicate.
Cadaver Test (to verify tube placement and repositioning by visualization)Performance demonstrated to be safe and effective as predicate.
MR Environment Non-Clinical Test (to confirm not for MRI environment)Performance demonstrated to be safe and effective as predicate (indicates non-MRI environment only).

2. Sample Size Used for the Test Set and Data Provenance

The document mentions a "Cadaver Test" as part of performance testing. However, it does not specify the sample size (i.e., number of cadavers) used for this test. It also does not explicitly state the country of origin for the cadaver data or whether it was retrospective or prospective. Given it's a cadaver test, it would be considered a prospective test for the device's functional performance in a cadaveric model.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not specify the number of experts or their qualifications used to evaluate the Cadaver Test results. Given the nature of a 510(k) for a medical device (as opposed to an AI/ML diagnostic tool), such details are often less extensively documented in the summary unless it involves subjective interpretation of images that are integral to a diagnostic claim.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the Cadaver Test or any other performance testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This document describes a traditional medical device (a video tracheoscope) and not an AI/ML-driven diagnostic or assistive technology. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed/reported.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This device is not an algorithm, so standalone algorithm performance is not applicable. The DLVT™ System is a physical medical device that provides direct video visualization for a clinician.

7. The Type of Ground Truth Used

For the functional performance tests, particularly the "Cadaver Test" regarding verifying tube placement and repositioning, the "ground truth" would be established by direct observation and professional judgment of the clinicians/researchers performing the test. For the other tests (e.g., cuff integrity, airflow), the ground truth is based on objective measurements against engineering and physical standards. There is no mention of pathology, outcome data, or expert consensus in the context of a diagnostic ground truth.

8. The Sample Size for the Training Set

This device is a physical medical instrument, not an AI/ML algorithm that requires a training set. Therefore, a training set sample size is not applicable.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for an AI/ML algorithm, this question is not applicable.

{0}------------------------------------------------

(ETVIEW Airway Management _

MAY 1 8 2012

510(K) SUMMARY

Double Lumen Video Tracheoscope (DLVT™) System 510(k) Number K113576

Applicant's Name: ETView Ltd.

Misgav Business Park

M.P Misgav 20174

Israel

Tel: (972)72-260-7063

Fax: (972)72-260-7263

  • Contact Person: Yoram Levy, Qsite
    31 Haavoda St.

Binyamina, Israel 30500

Tel (972)4-638-8837

Fax (972)4-638-0510

Yoram@gsitemed.com

Double Lumen Video Tracheoscope (DLVT™) System Trade Name:

Endobronchial tube (EBT) Device Type:

Preparation Date: November 23, 2011

Regulatory Name: Tracheal/bronchial differential ventilation Classification:

tube

Product Code: CBI Regulation No: 21 CFR 868.5740 Class: II Classification Panel: Anesthesiology

Device Description:

The ETView Double Lumen Video Tracheoscope (DLVT™) System functions as a standard 37 French size endobronchial tube that additionally has an embedded video imaging device in its tracheal lumen. The system provides a video image of the patient's

İ

{1}------------------------------------------------

bronchus, which is displayed on the monitor, for as long as the DLVTTM is inside the patient's bronchus.

Visualization of the bronchus is used to verify placement and repositioning of the endobronchial tube during the intubation procedure or throughout surgery.

The device is indicated for non-MRI environment only.

Intended Use Statement:

The ETView Double Lumen Video Tracheoscope (DLVT™) is intended to isolate the left or right lung of a patient for intensive care or surgery, one lung ventilation or one lung anesthesia. The DLVT' " System is indicated for verifying tube placement and

repositioning.

Predicate Devices: claimed:

Substantial equivalence to the following predicate devices is

Device Name510k NoDate of Clearance
ETView Tracheoscopic VentilationTube (TVT™)K082420October 8, 2008
Well LEAD Endobronchial TubesK092886March 11, 2010

Performance Standards

Double Lumen Video Tracheoscope (DLVT™) was tested and

complies with the following standards:

  • ISO 5361:1999 Anaesthetic and respiratory equipment --. Tracheal tubes and connectors
  • ISO 16628:2008 Tracheobronchial Tubes Sizing and . marking
  • ANSI/AAMI/ISO 11135-1:2007 Sterilization of health care . products - Ethylene oxide
  • AAMI TIR28:2001 Product adoption and process equivalency . for ethylene oxide sterilization
  • ISO 14971-1:2007 Risk management for medical devices .

{2}------------------------------------------------

  • ISO 10993-1:2003(E), Biological evaluation of medical . devices -- Part 1: Evaluation and testing
    A detailed description appears in Section 14.

Performance Testing

Performance testing demonstrated that the Double Lumen Video

Tracheoscope (DLVT™). is as safe and effective as the cleared predicate devices.

The following performance tests were conducted:

  • Determination of Cuff Resting Diameter
  • Cuff Leak Resistance Integrity
  • Resistance to Cuff Herniation .
  • Cuff Symmetry .
  • Resistance to tube collapse ●
  • Air Flow Resistance ●
  • Thermal Safety
  • Determination of effective inside diameter
  • Tracheal Seal
  • Determination of the bronchial segment
  • Cadaver Test
  • MR Environment Non-Clinical Test

Comparison to the Predicate Devices

The intended use of the DLVTTM is similar to the intended use of Well LEAD Endobronchial Tubes, previously cleared under K092886. The indication for use of the DLVTTM system is similar to the indication of ETView Tracheoscopic Ventilation Tube (TVTTM), previously cleared under K082420.

Both the DLVT™ and the TVT™ systems integrate imaging and lighting components into the wall of a tube for patient ventilation. The structure, the material, the dimensions and the sterilization method of the DLVT™ are similar to the cleared Well LEAD Endobronchial Tubes. The light source, the imaging sensor, the video transfer and format, the display and the power supply of the DLVT™ System are identical to the cleared ETView TVT™

{3}------------------------------------------------

The minor differences between the DLVT™ System and its predicate devices do not raise any new questions of safety or efficacy. Moreover, performance testing demonstrated that the DLVT™ is as safe and effective as the predicate devices. Thus, the DLVT™ System is substantially equivalent to Well LEAD Endobronchial Tubes and ETView TVT™.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY 1 8 2012

ETView Ltd. C/O Mr. Yoram Levy Osite 31 Haavoda Street Binvamina ISREAL 30500

Re: K113576

Trade/Device Name: Double Lumen Video Tracheoscope (DLVTTM) System Regulation Number: 21 CFR 868.5740 Regulation Name: Tracheal/Bronchial Differential Ventilation Tube Regulatory Class: II Product Code: CBI Dated: April 19, 2012 Received: April 24, 2012

Dear Mr. Levy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 – Mr. Levy

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related.adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours,

for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

T VIEW Airway Management

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

Device Name:

Double Lumen Video Tracheoscope (DLVTTM) System

Indications for Use:

The ETView Double Lumen Video Tracheoscope (DLVT™) is intended to isolate the left or right lung of a patient for intensive care or surgery, one lung ventilation or one lung anesthesia. The DLVT"M System is indicated for verifying tube placement and repositioning.

Prescription Use AND/OR X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-off)

Division of Anesthesiology, General Hospital, Infection Control and Dental Devi 510(k) Number

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:_____________________________

1-2

Double Lumen Video Tracheoscope (DLVT™) System – 510k Notification

§ 868.5740 Tracheal/bronchial differential ventilation tube.

(a)
Identification. A tracheal/bronchial differential ventilation tube is a device used to isolate the left or the right lung of a patient for anesthesia or pulmonary function testing.(b)
Classification. Class II (performance standards).