K Number
K113570

Validate with FDA (Live)

Date Cleared
2012-02-22

(82 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Integra CUSA NXT Inferior Forward Bone Tip is an accessory to the CUSA NXT Ultrasonic Surgical Aspirator System that is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable, including Neurosurgery, Gastrointestinal and affiliated organ surgery, Urological surgery, Plastic and Reconstructive surgery, General surgery, Orthopedic surgery, Gynecological surgery, Thoracic surgery, Laparoscopic surgery and Thoracoscopic surgery.

Device Description

The Integra CUSA NXT Inferior Forward Bone Tip (Inferior Forward Bone Tip) will attach to the Selector 24 kHz Neuro Short Handpiece (1523000M7) and will be controlled by the CUSA NXT Ultrasonic Surgical Aspirator System (CUSA NXT System). The CUSA NXT System is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies and removes unwanted tissue. The device allows the selective dissection of target tissue while preserving vessels, ducts and other delicate structures. The system consists of a console which provides control and power functions, a surgical handpiece which provides ultrasonic mechanical energy, a single use tip, and a suction/irrigation system (manifold tubing and cooling water canister).
The Inferior Forward Bone Tip has an abrasive surface that is oriented inferiorly and distally at the distal end of the tip. This configuration facilitates the use of the system in procedures where it is necessary for practitioners to orient the surgical tip in the bottom dead center or inferior orientation.
The Inferior Forward Bone Tip consists of a titanium tip with a titanium nitride coating, silicone flue and an ultem shroud.
The Inferior Forward Bone Tip will be supplied sterile and is intended for single use.

AI/ML Overview

The Integra CUSA NXT™ Inferior Forward Bone Tip is an accessory to an ultrasonic surgical aspirator system. The provided text outlines the performance criteria and the non-clinical tests conducted to demonstrate the device meets these criteria. No studies involving human readers or comparative effectiveness are described, as this is a device for surgical aspiration, not an AI or imaging diagnostic tool.

Here's the information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Test)Reported Device Performance (Result)
Electromechanical Test (measures frequency, stroke, lateral movement, and power)Passed frequency, stroke, lateral movement, and quiescent power acceptance criteria.
Lateral Load Test (applies a lateral load on the vibrating tip to evaluate robustness)Performance of the tips was not affected after the application of the lateral load.
Dry Flue Test (checks the effect of ultrasonically vibrating a surgical tip without the presence of irrigation)Performance of the tips was not affected when operated with no irrigation for the time specified.
Accelerated Stress Bone Cutting (tests the effect of bone cutting for extended periods of time)No breakage of the abrasive surface occurred.
Measurement of Thermal Rise During Ultrasonic Aspiration of Representative TissueThermal rise in tissue field during tissue removal was found to be less than stated in the product specification.
BiocompatibilityNo additional testing performed, as materials, manufacturing, and sterilization are the same as the predicate device, implying satisfactory existing biocompatibility.

2. Sample size used for the test set and the data provenance

The document describes non-clinical bench testing. There is no test set of data in the context of imaging or AI analysis. The tests were performed on the device itself. Data provenance, country of origin, and retrospective/prospective status are not applicable in this context.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a physical surgical device, and testing involves engineering and technical evaluations, not expert-derived ground truth from medical images or diagnoses.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are relevant for subjective evaluations, typically in clinical studies or expert reviews of data, which are not described here.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI or imaging device. No MRMC study was conducted or is relevant to the device's function.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the physical device testing, the "ground truth" is established by engineering and performance specifications and direct measurement against those specifications, as well as comparison to the performance of the predicate device (K111741). For example, the thermal rise should be less than specified, and the tip should not break.

8. The sample size for the training set

Not applicable. This is not an AI device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI device.

{0}------------------------------------------------

510(k) Summary

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

807.92(a)(1) - Submitter Information
NameIntegra Burlington MA, Inc.
Address22 Terry AvenueBurlington, MA 01803
Phone number609-936-5583
Fax number609-275-9447
Establishment RegistrationNumber1222895
Name of contact personLindsay Mignone
Date preparedNovember 30, 2011
807.92(a)(2) - Name of device
Trade or proprietary nameIntegra™ CUSA NXT™ Inferior Forward Bone Tip
Common or usual nameUltrasonic Surgical Aspirator
Classification nameInstrument, Ultrasonic Surgical
Classification panelGeneral and Plastic Surgery
RegulationUnclassified
Product Code(s)LFL
807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed
CUSA NXT Bone Tips (K111741)
807.92(a)(4) - Device description
The Integra CUSA NXT Inferior Forward Bone Tip (Inferior ForwardBone Tip) will attach to the Selector 24 kHz Neuro Short Handpiece(1523000M7) and will be controlled by the CUSA NXT UltrasonicSurgical Aspirator System (CUSA NXT System). The CUSA NXTSystem is an ultrasonically vibrating surgical device which, incombination with irrigation and aspiration, fragments, emulsifies andremoves unwanted tissue. The device allows the selective dissection oftarget tissue while preserving vessels, ducts and other delicatestructures. The system consists of a console which provides controland power functions, a surgical handpiece which provides ultrasonicmechanical energy, a single use tip, and a suction/irrigation system(manifold tubing and cooling water canister).
The Inferior Forward Bone Tip has an abrasive surface that is orientedinferiorly and distally at the distal end of the tip. This configurationfacilitates the use of the system in procedures where it is necessary forpractitioners to orient the surgical tip in the bottom dead center orinferior orientation.

.

{1}------------------------------------------------

nitride coating, silicone flue and an ultem shroud.The Inferior Forward Bone Tip consists of a titanium tip with a titanium
for single use.The Inferior Forward Bone Tip will be supplied sterile and is intended
807.92(a)(5) Intended use of the device
Indications for useThe CUSA NXT Inferior Forward Bone Tip is an accessory to theCUSA NXT Ultrasonic Surgical Aspirator System that is indicated foruse in surgical procedures where fragmentation, emulsification andaspiration of soft and hard (e.g. bone) tissue is desirable, includingNeurosurgery, Gastrointestinal and affiliated organ surgery, Urologicalsurgery, Plastic and Reconstructive surgery, General surgery,Orthopedic surgery, Gynecological surgery, Thoracic surgery,Laparoscopic surgery and Thoracoscopic surgery.The indications for use for the CUSA NXT system have not changed
due to the addition of the Inferior Forward Bone Tip.
predicate807.92(a)(6) Summary of the technological characteristics of the device compared to the
CharacteristicCUSA NXT Inferior Forward TipCUSA NXT Bone Tips (K111741)
Console used with tipCUSA NXT Ultrasonic SurgicalCUSA NXT Ultrasonic Surgical
Aspirator SystemAspirator System
ApproximateFrequency ofOperation24 kHz24 kHz
Max Stroke (inches)0.01200.0120
Tips Delivered AsSterile / Single UseSterile / Single Use
ShroudUses a curved shroud, packagedwith tipUses a curved shroud, packaged withtip
Vibration of TipLongitudinalLongitudinal
Design of distal endProtrusion with 10° inverse conicalProtrusion with 10° inverse conical(Forward)Protrusion with 10° conical (Reverse)
Abrasive surface oriented inferior tosurgical tip when bentAbrasive surface oriented superior tosurgical tip when bent
Pre-aspiration holesYesYes
Inner Diameter(inches)0.0780.078
Working Length(mm)8888 (Forward)80 (Reverse)
Bend Angle20°20°
Bend Radius (in)5.6843.15
Material
TipTitanium 6AL4V Grade 5Titanium 6AL4V Grade 5
FlueSiliconeSilicone
ShroudUltemUltem

CUSA NXT inferior Forward Bone Tip 510(k)

{2}------------------------------------------------

807.92(b)(1-2) Nonclinical tests submitted
TestResult
Electromechanical Test - measuresfrequency, stroke, lateral movement,and powerPassed frequency, stroke, lateral movement, andquiescent power acceptance criteria.
Lateral Load Test - applies a lateral loadon the vibrating tip to evaluaterobustnessPerformance of the tips was not affected after theapplication of the lateral load.
Dry Flue Test - checks the effect ofultrasonically vibrating a surgical tipwithout the presence of irrigationPerformance of the tips was not affected when operatedwith no irrigation for the time specified.
Accelerated Stress Bone Cutting - teststhe effect of bone cutting for extendedperiods of timeNo breakage of the abrasive surface occurred.
Measurement of Thermal Rise DuringUltrasonic Aspiration of RepresentativeTissueThermal rise in tissue field during tissue removal was foundto be less than stated in the product specification.
BiocompatibilitySince the modified device uses materials that have thesame chemical formulations, same manufacturing andsame sterilization processes as in the predicate device,additional testing was not performed.
807.92(b)(3) Conclusions drawn from non-clinical data
Testing confirmed that the performance of the Inferior Forward Bone Tip meets the product

Testing confirmed that the performance of the Interior Forward Bone Tips. Therefore, the modification resulted in a device that performs the same as the predicate device.

:

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three heads, representing the department's commitment to health, human services, and science. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle symbol.

FEB 2 2 2012

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Integra Burlington MA, Incorporated % Ms. Lindsay Mignone 22 Terry Avenue Burlington, Massachusetts 01803

Re: K113570

Trade/Device Name: Integra CUSA NXT Inferior Forward Bone Tip Regulatory Class: Unclassified Product Code: LFL Dated: November 30, 2011 Received: December 2, 2011

Dear Ms. Mignone:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Lindsay Mignone

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you acontra spov.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Mark A. Mulkern

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K113570

Device Name: Integra CUSA NXT Inferior Forward Bone Tip

Indications for Use:

The Integra CUSA NXT Inferior Forward Bone Tip is an accessory to the CUSA NXT Ultrasonic The mical Aspirator System that is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable, including emulomounon and adpiractinal and affiliated organ surgery, Urological surgery, Plastic and Neconstructive surgery, General surgery, Orthopedic surgery, Gynecological surgery, Thoracic surgery, Laparoscopic surgery and Thoracoscopic surgery.

X PRESCRIPTION USE _ (Part 21 CFR 801 Subpart D) . AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil Keller for Nixon

(Division Sign-Off) Division of Surgical. Orthopedic, and Restorative Devices

510(k) Number K113570

CUSA NXT Inferior Forward Bone Tip 510(k)

04-002

N/A