K Number
K113508

Validate with FDA (Live)

Device Name
DACRYOCATH
Date Cleared
2012-08-31

(277 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The lacrimal duct catheter is intended for use during dilation of the obstructed nasolacrimal duct.

The lacrimal duct catheter is indicated for use during dilation of the obstructed nasolacrimal duct in the following populations:

  • a. The 2mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct obstruction in patients over 12 months of age and under 30 months of age.
  • b. The 3 mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct in children over 30 months of age.
  • c. The 5mm catheter is indicated for use in adults during dilation of a lacrimal duct obstruction or blocked dacryocystorhinostomy ostium as a result of the following: functional or complete nasolacrimal duct obstruction, dacryocystitis, or failed dacryocystorhinostomy.
Device Description

The lacrimal duct catheter is a sterile, single use, non-pyrogenic disposable balloon catheter consisting of a semi-flexible stainless steel hypotube core and polyethylene terephthalate balloon tubing. The balloon is designed to inflate to a known diameter and length at the specific pressure. Markings are present 10 and 15 mm proximal to the beginning of the working portion of the balloon, which helps indicate when the balloon is placed correctly in the lacrimal system. The overall length of the catheter is 6 inches (15.24 cm) long. There is an opening on the distal end of the catheter hypotube to accommodate irrigation solutions. The Y-hub on the proximal end of the catheter has a luer port for inflation of the balloon catheter (labeled "inflation" with a red band) and a second luer port for irrigation through the balloon catheter (labeled "irrigation"). The balloon catheter is available in a 2 mm and 3 mm inflated diameter. The 3 mm balloon has a length of 15 mm and a deflated profile of approximately 1.1 mm. The 2 mm balloon has a length of 15 mm and a deflated profile of approximately 1.0 mm. The balloon has a 5 mm inflated diameter and a length of 10 mm. The deflated profile of the 5 mm balloon is approximately 1.2 mm.

AI/ML Overview

The provided text describes the DacryoCath Lacrimal Duct Balloon Catheter and its non-clinical performance testing for substantial equivalence to a predicate device. There is no clinical study or clinical performance data described in the provided document that would establish acceptance criteria or demonstrate the device meets such criteria in a clinical setting.

The document focuses on non-clinical performance tests to show similar technological characteristics and safety.

Here's a breakdown of the information requested, based only on the provided text:

1. A table of acceptance criteria and the reported device performance:

Test CategoryAcceptance Criteria/MethodologyReported Device Performance
BiocompatibilityISO10993 (Cytotoxicity, Sensitization, Irritation)PASS
ISO Guinea Pig Maximization Sensitization TestDid not elicit a sensitization response.PASS
ISO Intracutaneous Reactivity TestMet the requirements of ISO Intracutaneous Reactivity.PASS
Rabbit Pyrogen Test (Material Mediated) - ISONon-pyrogenic and meets ISO10993-11 guidelines.PASS
ISO Acute Systemic Injection TestMet the requirements of the ISO Acute Systemic Injection Test.PASS
MEM Elution GLP Report (Cytotoxicity)Determines cytotoxicity of extractable substances; cell monolayers examined and scored. All acceptance criteria met.PASS
Functional Performance
Inflation/Deflation TimeInflation and deflation of balloons using conventional techniques within a specified time.PASS
Fatigue TestingRepeatability of balloon inflation without failure using recommended inflation pressure.PASS
Rupture TestingBalloons will not burst at or below the maximum recommended burst pressure.PASS
Tensile TestingBond strength at joining locations can withstand tensile forces greater than clinical use.PASS
ISO 594-1 Testing (Luer Taper)Gauging, Liquid Leakage, Air Leakage, Separation Force. (Tests conducted within ISO 594-1 guidance)PASS
Sterilization & Packaging
Packaging ValidationTransportation (ASTM D169), Seal Peel Test, Dye Migration Test (ASTM F1929), Aerosol Challenge Test, Accelerated Aging Test (3.3 weeks @ 55± 4°C simulating 0.5 year shelf life).PASS
EO Sterilization ValidationMicrobiological challenge utilizing half-cycle (overkill) method per ISO11135.PASS

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: The document does not specify the exact sample sizes for each individual test. It mentions "balloons of each diameter and length" for rupture testing, and "test article" for others. No specific numbers are provided for these non-clinical tests.
  • Data Provenance: Not applicable for non-clinical lab tests. The tests refer to ISO and ASTM standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. These were non-clinical, laboratory-based engineering and biological safety tests. Ground truth in this context refers to the defined parameters and expected outcomes of the standardized tests, not expert interpretation of clinical data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods are typically used for clinical result interpretation, particularly when expert consensus or review is needed. These are objective engineering and biological tests with pass/fail criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This document does not describe an MRMC study or any clinical comparative effectiveness study involving human readers or AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This device is a physical medical device (balloon catheter) intended for manual use by a clinician, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the non-clinical tests, the "ground truth" is defined by the specific requirements and methodologies of the international and national standards (ISO, ASTM, USP) referenced for each test. For example, for biocompatibility, the ground truth is the absence of cytotoxicity, sensitization, irritation, etc., as determined by the specified assays. For functional tests, it's meeting predefined performance specifications (e.g., burst pressure, inflation time).

8. The sample size for the training set

  • Not applicable. This document describes the testing of a physical medical device. There is no mention of a "training set" as would be relevant for machine learning algorithms.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set described.

{0}------------------------------------------------

K113508

AUG 3 1 2012

510(k) SUMMARY

Suite 246

Armadillo Medical, LLC,

5363 Balboa Blvd.

Encino, CA 91316 TEL: 855.255.3310

510(k) Owner

Contact person

FAX: 818.783.9059 Robyn Scopis Regulatory Consultant to Armadillo Biomedical Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606 TEL: 949.422.3853 FAX: 949.552.2821 EMAIL: robyn@regulatoryspecialists.com

Date summary was prepared

August 20, 2012

Name of device Common Name Classification Name Regulation Product Code Predicate

DacryoCath Lacrimal duct balloon catheter Lacrimal Stents and Intubation Sets Pre-Amendment OKS K935233

Description

The lacrimal duct catheter is a sterile, single use, non-pyrogenic disposable balloon catheter consisting of a semi-flexible stainless steel hypotube core and polyethylene terephthalate balloon tubing. The balloon is designed to inflate to a known diameter and length at the specific pressure. Markings are present 10 and 15 mm proximal to the beginning of the working portion of the balloon, which helps indicate when the balloon is placed correctly in the lacrimal system. The overall length of the catheter is 6 inches (15.24 cm) long. There is an opening on the distal end of the catheter hypotube to accommodate irrigation solutions. The Y-hub on the proximal end of the catheter has a luer port for inflation of the balloon catheter (labeled "inflation" with a red band) and a second luer port for irrigation through the balloon catheter (labeled "irrigation"). The balloon catheter is available in a 2 mm and 3 mm inflated diameter. The 3 mm balloon has a length of 15 mm and a deflated profile of approximately 1.1 mm. The 2 mm balloon has a length of 15 mm and a deflated profile of approximately 1.0 mm.

{1}------------------------------------------------

K//3508

Description continued:

The balloon has a 5 mm inflated diameter and a length of 10 mm. The deflated profile of the 5 mm balloon is approximately 1.2 mm.

Intended Use

The lacrimal duct catheter is intended for use during dilation of the obstructed nasolacrimal duct.

Indications for Use

The lacrimal duct catheter is indicated for use during dilation of the obstructed nasolacrimal duct in the following populations:

  • a. The 2mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct obstruction in patients over 12 months of age and under 30 months of age.
  • b. The 3 mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct in children over 30 months of age.
  • c. The 5mm catheter is indicated for use in adults during dilation of a lacrimal duct obstruction or blocked dacryocystorhinostomy ostium as a result of the following: functional or complete nasolacrimal duct obstruction, dacryocystitis, or failed dacryocystorhinostomy.

Technological Characteristics

The predicate and the DacryoCath were compared in the following areas and found to have similar technological characteristics and to be equivalent:

Material Characteristics

Both the Predicate and DacryoCath are made of Stainless Steel and Polyethylene.

Design Characteristics

Both the Predicate and DacryoCath are designed to be used as lacrimal duct balloon catheters.

Operating Characteristics

Both the Predicate and DacryoCath are operated by placing the catheter into the canaliculus, inflate the balloon to 8atm with 10cc sterile saline, and inflate for 90seconds to dilate the obstructed nasolacrimal duct.

Intended Use

Both the Predicate and DacryoCath have the same intended use: The lacrimal duct catheter is intended for use during dilation of the obstructed nasolacrimal duct

2

{2}------------------------------------------------

Technological Characteristics Similar technological characteristics continued: Balloon Length -3 mm Predicate - 15mm DacryoCath - 15mm Balloon Length - 5mm Predicate - 10mm DacryoCath - 10mm

The predicate and the DacryoCath were compared in the following areas and found to have minor different technological characteristics. The following differences have been determined. through non-clinical performance testing, to not have any impact on the safety or efficacy of the DacryoCath when used as indicated:

15/13508

Length of Catheter Predicate - 24cm DacryoCath - 15.24cm Balloon Length (2mm Balloon only) Predicate - 13mm DacryoCath - 15mm Irrigation Port -Predicate - None Available DacryoCath - Yes

The following non-clinical performance tests were conducted:

Biocompatibility to ISO10993

CytotoxicityPASS
SensitizationPASS
IrritationPASS

ISO Guinea Pig Maximization Sensitization Test

The test article did not elicit a sensitization response under the conditions of this assay.

ISO Intracutaneous Reactivity Test

The requirements of ISO Intracutaneous Reactivity have been met by the test article.

Rabbit Pyrogen Test (Material Mediated) - ISO

3

The USP 0.9% Sodium Chloride for Injection (NaCl) extract of the test article, Lot #23054 - lacrimal balloon catheters, was evaluated for its potential to produce a pyrogenic response when tested in New Zealand White rabbits. Based on the criteria of the protocol, the test article is considered non-pyrogenic and meets the requirements of the Pyrogen Test, ISO10993-11 guidelines.

{3}------------------------------------------------

Non-clinical performance tests continued:

ISO Acute Systemic Injection Test

The requirements of the ISO Acute Systemic Injection Test have been met by the test article.

MEM Elution GLP Report

The Minimal Essential Media (MEM) Elution test was designed to determine the cytotoxicity of extractable substances. An extract of the test article was added to cell monolayers and incubated. The cell monolayers were examined and scored based on the degree of cellular destruction. All test method acceptance criteria were met.

Inflation/Deflation Time PASS

Inflation/Deflation Time testing was conducted to show that the inflation and deflation of the balloons using conventional techniques can be accomplished within a specified time.

Fatigue Testing

Fatigue testing was conducted to determine the repeatability of balloon inflation without failure using the recommended inflation pressure.

Rupture Testing

PASS

PASS

15113508

Rupture testing was done on balloons of each diameter and length. Test results show that the balloons will not burst at or below the maximum recommended burst pressure.

Tensile Testing

PASS

Tensile testing was done to test the bond strength at locations where joining methods are used for bonding components of the catheter. Testing demonstrated that all bonds can withstand tensile forces greater than those that may be experienced during clinical use.

ISO 594-1 Testing

PASS

ISO 594-1 testing was conducted for the 6% (Luer) taper for syringes. Under this guidance, the following tests were conducted:

Gauging Liquid Leakage Air Leakage Separation Force

Stress Cracking

Packaging ValidationPASS
TransportationASTM D169PASS
Seal Peel TestPASS
Dye Migration TestASTM F1929PASS
Aerosol Challenge TestPASS
Accelerated Aging TestPASS
3.3 weeks @ 55± 4°C
Simulating 0.5 year shelf life

{4}------------------------------------------------

K113508

Non-clinical performance tests continued:

EO Sterilization Validation

PASS A microbiological challenge utilizing the half-cycle (overkill) method, using a biological indicator challenge following ISO11135.

Conclusions from non-clinical performance data

After performing non-clinical performance studies, the data shows that the DacryoCath is substantially equivalent to the predicate as a lacrimal duct balloon catheter.

Clinical performance data

A review of the published peer-reviewed literature was conducted to review appropriate uses, adverse events, and clinical experience with this device type.

5

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird-like figure with flowing lines, suggesting movement or flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 3 1 2012

Armadillo Biomedical, LLC c/o Ms. Robyn Scopis Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606

Re: K113508

Trade/Device Name: DacryoCath Regulation Number: Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Product Code: OKS Dated: August 20, 2012 Received: August 29, 2012

Dear Ms. Scopis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{6}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

for

Sincerely yours,

Kesia Alexander

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: DacryoCath by Armadillo Biomedical, LLC

Indications for Use:

The lacrimal duct catheter is indicated for use during dilation of the obstructed nasolacrimal duct in the following populations:

  • b. The 2mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct obstruction in patients over 12 months of age and under 30 months of age.
  • b. The 3 mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct in children over 30 months of age.
  • c. The 5mm catheter is indicated for use in adults during dilation of a lacrimal duct obstruction or blocked dacryocystorhinostomy ostium as a result of the following: functional or complete nasolacrimal duct obstruction, dacryocystitis, or failed dacryocystorhinostomy.

× AND/OR Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

CLCLy

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K113508

1 Page 1 of

N/A