K Number
K113475

Validate with FDA (Live)

Date Cleared
2012-03-20

(119 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ESTECH Cobra Adhere XL2 is intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when connected directly to the Estech Electrosurgical Unit (ESU).

The ESTECH Cobra Adhere XL 2 may be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external pacemaker or recording device.

Device Description

The Estech COBRA Adhere XL 2 is a sterile, single use device intended for the ablation of The Esteon CODIA Cranel to deliver RF energy to the target tissue via electrodes in the device when connected to the COBRA Electrosurgical Unit (ESU).

When the COBRA Adhere XL 2 is connected to an auxiliary temporary external pacemaker or recording device it can be used to provide temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery. The COBRA Adhere XL 2 must be disconnected from the ESU and connected to a temporary external pacemaker or recording device using the accessory cable provided.

AI/ML Overview

The provided text is a 510(k) summary for the Estech COBRA Adhere XL 2, a surgical device for ablation of cardiac tissue. It describes the device, its intended use, and compares it to predicate devices to establish substantial equivalence.

However, the document does NOT contain information about specific acceptance criteria or a detailed study proving the device meets acceptance criteria, as one would find in a clinical trial report or a performance study summary with quantified results.

The text focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and general non-clinical testing. It states that "Appropriate testing, included in this application, has been performed to ensure that the ESTECH COBRA Adhere XL 2 meets its product specifications" and that "In-vitro and in-vivo testing demonstrate that the COBRA Adhere XL2 is able to pace, sense, and stimulate and ablate cardiac tissue as safely and effectively as the predicate devices."

Without quantitative performance metrics, sample sizes for test sets, ground truth details, expert qualifications, or comparative effectiveness studies, it is impossible to complete a table of acceptance criteria versus reported device performance or address most of the requested points.

Therefore, the following points cannot be extracted from the provided text:

  • A table of acceptance criteria and the reported device performance
  • Sample sizes used for the test set
  • Data provenance for the test set
  • Number of experts used to establish the ground truth for the test set and their qualifications
  • Adjudication method for the test set
  • Information on a multi-reader multi-case (MRMC) comparative effectiveness study
  • Information on a standalone (algorithm only) performance study
  • Type of ground truth used for the test set
  • Sample size for the training set
  • How the ground truth for the training set was established

The document focuses on non-clinical testing for safety and effectiveness in establishing substantial equivalence to prior devices, rather than a detailed clinical performance study with specific acceptance criteria as you've requested.

{0}------------------------------------------------

KIT3445

SECTION 5: 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

MAR 2 0 2012

Device Information:

CategoryComments
Sponsor:ESTECH, Inc.2603 Camino RamonSuite 100San Ramon, CA 94583Tel: 925-543-2110
Correspondent:Tamer IbrahimVice PresidentESTECH, Inc
Contact Information:Tel: 925-543-2110Fax: 925-866-7117
Device Common Name:Surgical Device, For Ablation of CardiacTissue
Device Proprietary Name:Estech COBRA Adhere XL 2
Device Classification:Class II, OCL (21 CFR 878.4400)

Predicate Device Information:

Predicate Devices:Estech COBRA Adhere XL (K051749)Estech COBRA Bipolar Recording Electrode(K051749)
Predicate Device Manufacturers:Endoscopic Technologies, Inc (dba Estech)
Predicate Device Common Name:Surgical Device, For Ablation of CardiacTissue
Predicate Device Classification:21 CFR 878.4400
Predicate Device Classification Number:Class II, OCL

Predicate Device Information:

Predicate Devices:Multifunctional Linear Pen (K100501)
Predicate Device Manufacturers:AtriCure, Inc.
Predicate Device Common Name:Surgical Device, For Ablation of CardiacTissue
Predicate Device Classification:21 CFR 878.4400
Predicate Device Classification Number:Class II, OCL

b. Date Summary Prepared

21 November 2011 (Updated 22Feb2012)

c. Description of Device

The Estech COBRA Adhere XL 2 is a sterile, single use device intended for the ablation of The Esteon CODIA Cranel to deliver RF energy to the target tissue via electrodes in the device when connected to the COBRA Electrosurgical Unit (ESU).

{1}------------------------------------------------

When the COBRA Adhere XL 2 is connected to an auxiliary temporary external pacemaker or recording device it can be used to provide temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery. The COBRA Adhere XL 2 must be disconnected from the ESU and connected to a temporary external pacemaker or recording device using the accessory cable provided.

d. Intended Use

The ESTECH COBRA Adhere XL 2 is intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when connected directly to the Estech Electrosurgical unit (ESU).

The ESTECH COBRA Adhere XL 2 may be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device.

e. Comparison to Predicate Device and Summary of Technological Characteristics

The ESTECH COBRA Adhere XL2 has the identical intended use and the same technological characteristics as the predicate devices. There have been no significant changes in technological characteristics with respect to the predicate device including materials, design, energy sources, or other features from the predicate device.

The COBRA Adhere XL 2 and predicate COBRA Adhere XL both comprise a vacuum stabilizer to maintain contact between the ablation probe and target tissue to be ablated. These vacuum te hilizers are flexible. The COBRA Adhere XL 2 further comprises an internal ribeage which gives the device the ability to conform into various shapes while providing the flexible suction stabilizer with structural integrity. The predicate COBRA Adhere XL has a vacuum stabilizer with a centrally located series of electrodes. The COBRA Adhere XL 2 has a vacuum stabilizer with a laterally located series of electrodes. As the vacuum stabilizers form a seal with the tissue, while is drawn into the device creating contact between tissue to be ablated and the electrodes.

Both the COBRA Adhere XL2 and predicate COBRA Adhere XL include temperature controlled Doar are of endicate and application devices each have multiple active RF electrodes that are individually selectable and controlled by the ESU (generator). Predicate and application devices each have two (2) thermocouples attached to each active RF electrode. The COBRA Adhere XL2 also comprises an integrated indifferent electrode. With the COBRA Adhere XL 2, RF energy is directed from the active electrode to the integrated indifferent electrode or to the indifferent electrode placed on the patient's back, depending on the mode the device is in, as selected by the surgeon. The surgeon depresses the button on the device handle and the switch in serveted by the bargeon. I am enables or disables the integrated indifferent electrode.

The COBRA Adhere XL 2 also provides transient pacing/sensing capability when connected to an auxiliary temporary external pacemaker or recording device like the predicate devices.

{2}------------------------------------------------

f. Summary of Nonclincal Testing

Appropriate testing, included in this application, has been performed to ensure that the ESTECH COBRA Adhere XL 2 meets its product specifications. The data demonstrates substantial equivalence to the predicate devices. In-vitro and in-vivo testing demonstrate that the COBRA Adhere XL2 is able to pace, sense, and stimulate and ablate cardiac tissue as safely and effectively as the predicate devices.

The patient contacting materials of the ESTECH COBRA Adhere XL 2 have been tested in accordance with ISO 10993: Biological Evaluation of Medical Device to ensure biocompatibility. The safety of the electrical design has been tested per and is in conformance with the pertinent sections of IEC 60601-1-2, (Second Edition, 2001), Medical Electrical Equipment – Part 1-2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility – Requirements and Tests and IEC 60601-2-2: Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment.

g. Summary of Substantial Equivalence

Estech concludes that the COBRA Adhere XL 2 is substantially equivalent to the predicate devices. The indications for use, basic overall function, materials, packaging, and sterilization methods are identical or substantially equivalent.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white circular seal for the Department of Health & Human Services USA. The seal features the department's logo, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the perimeter of the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAR 2 0 2012

Endoscopic Technologies, Inc. C/O Mr. Tamer Ibrahim 2603 Camino Ramon, Suite 100 San Ramon, CA 94583

Re: K113475

Trade/Device Name: ESTECH Cobra Adhere XL 2 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: OCL Dated: February 21, 2012 Received: February 22, 2012

Dear Mr. Ibrahim:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 310(x) premailer is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the cherosure) to regally manatiment date of the Medical Device Amendments, or to commence prior to May 20, 1970, the enaounter with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). allu Cosmetic Act (Act) that do not required to the general controls provisions of the Act. The You may, therefore, market me device, basjever to assements for annual registration, listing of general controls provisions of the fielding, and prohibitions against misbranding and devices, good manufacturing practice, tabeling, and information related to contract liability adulteration. Please note: CDNT docs not overance increases and misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (Scc above) finitials affecting your device can be may be subject to additional controls. "Internal Regulations, "In addition, FDA may lound in the Code of I each. Regarders
publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualled of a bays of a be requirements of the Act
that FDA has made a determination that your device complies with on any must that FDA has made a decernmanch and Joan Seeel by other Federal agencies. You must or any Federal statutes and regulations daminding, but not limited to: registration and listing (21
comply with all the Act's requirements, including, but not issuesting of m comply with all the Act 3 requirements, incruaing, our access of the subsiting of medical

{4}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 1100 for regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ronn D. Zuckerman, M.D.

Bram D. Zackerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K113475

Device Name: ESTECH Cobra Adhere XL 2

Indications For Use:

The ESTECH Cobra Adhere XL2 is intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when connected directly to the Estech Electrosurgical Unit (ESU).

The ESTECH Cobra Adhere XL 2 may be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external pacemaker or recording device.

Prescription Use __ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) NumberA13475
Page 1

of 1

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.