K Number
K113350

Validate with FDA (Live)

Device Name
BRIVO NM 615
Manufacturer
Date Cleared
2011-12-05

(21 days)

Product Code
Regulation Number
892.1200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GE Brivo NM 615 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy.

The GE Brivo NM 615 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring.

The GE Brivo NM 615 system may include signal analysis and display equipment. patient and equipment supports, components and accessories. The system may be used for patients of all ages.

Device Description

The BrivoNM615 is an all-purpose single detector nuclear imaging system intended for general nuclear medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The main components of the Brivo NM 615 system are: NM Gantry with Single detector head, patient table, Remote Control Unit and NM operation console. The Brivo NM 615 is a modification of its predicate device Discovery NM 630 using only single head detector instead of Dual Head detector like in the Discovery NM 630

AI/ML Overview

The GE Healthcare Brivo NM 615 is a modification of its predicate device, the Discovery NM 630. The Brivo NM 615 uses a single head detector instead of a dual-head detector.

1. Acceptance Criteria and Reported Device Performance
The submission does not explicitly state quantitative acceptance criteria or detailed reported device performance in the format of a table. Instead, it relies on a declaration of substantial equivalence to its predicate device.

Acceptance CriteriaReported Device Performance
Substantially equivalent to predicate device Discovery NM 630 (K111445) in terms of safety, effectiveness, and performance."GE Healthcare considers the Brivo Nm 615 to be as safe, as effective, and performance is substantially equivalent to the predicate device Discovery NM 630 (K111445)."
Compliance with voluntary standards."The Brivo NM615 and its applications comply with voluntary standards as detailed in Section 9, of this premarket submission."
No new safety questions or risks identified due to modifications."Engineering testing and standards compliance testing were successfully conducted and did not raise any new safety questions or identify any new risks."

2. Sample Size for Test Set and Data Provenance
The submission states that no clinical studies were required to support substantial equivalence for the Brivo NM 615. Therefore, there is no specific "test set" in the context of a clinical performance study. The evaluation was based on non-clinical engineering and standards compliance testing.

3. Number of Experts and Qualifications for Ground Truth
Not applicable, as no clinical studies were performed. The determination of "ground truth" was based on established engineering principles and compliance with voluntary standards.

4. Adjudication Method for Test Set
Not applicable, as no clinical studies were performed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was conducted. The submission specifically states: "The subject of this premarket submission. Brivo NM 615. did not require clinical studies to support substantial equivalence."

6. Standalone (Algorithm Only) Performance Study
Not applicable. The Brivo NM 615 is a hardware device (a SPECT system), not an algorithm or AI solution. The evaluation was for the entire system's performance, not just an algorithm.

7. Type of Ground Truth Used
The concept of "ground truth" in the clinical sense is not directly applicable here since no clinical studies were performed. The "ground truth" for the non-clinical evaluation was based on:

  • Engineering design specifications.
  • Performance metrics defined by voluntary standards.
  • The established safety and effectiveness profile of the predicate device (Discovery NM 630).

8. Sample Size for Training Set
Not applicable, as this is a hardware device submission, not an AI/algorithm submission requiring a training set for model development.

9. How Ground Truth for Training Set was Established
Not applicable.

{0}------------------------------------------------

DEC - 5 2011

Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. There are four decorative swirls around the outside of the circle, positioned at roughly equal intervals.

GE Healthcare 510(k) Premarket Notification Submission

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

November 9, 2011 Date: Submitter: GE Healthcare GE Healthcare, GE Medical Systems Israel, Functional Imaging 4 HAYOZMA St TIRAT HACARMEL, 30200, ISRAEL Primary Contact Person: Eli Werner Regulatory Affairs Leader GE Healthcare GE Healthcare, GE Medical Systems Israel, Functional Imaging (972) 4-8563666

(972) 4-8577664

Secondary Contact Person:

John Jaeckle Regulatory Affairs Manager, MI&CT GE Healthcare, 262-424-9547

Device: Trade Name: Brivo NM 615

Single Photon Emission Computed Tomography (SPECT ) Common/Usual Name: Classification Names: 21CFR 829.1200 Product Code: KPS

Predicate Device(s): Discovery NM 630 , K111445

The BrivoNM615 is an all-purpose single detector nuclear Device Description: imaging system intended for general nuclear medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The main components of the Brivo NM 615 system are: NM Gantry with Single detector head, patient table, Remote Control Unit and NM operation console. The Brivo NM 615 is a modification of its predicate device Discovery NM 630 using only single head detector instead of Dual Head detector like in the Discovery NM 630

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. The letters and the border are in black, while the background is white. There are some white accents around the border, giving it a slightly stylized appearance.

GE Healthcare 510(k) Premarket Notification Submission

Intended Use:

The Brivo NM 615 is an emission computed tomography system intended to detect the location and distribution of gamma ray photon emitting radionuclides in the body and to produce crosssectional images through computer reconstruction of the data. The Brivo NM 615 system is intended for General Nuclear Medicine imaging procedures using variety of scanning modes supported by various acquisition types.

This generic type of device may include signal analysis and display equipment, patient and equipment supports, and accessories.

Indications for Use:

The GE Brivo NM 615 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy.

The GE Brivo NM 615 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring.

The GE Brivo NM 615 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may be used for patients of all ages.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined in a stylized, cursive font. The letters are enclosed within a circular border. The logo is black and white.

GE Healthcare 510(k) Premarket Notification Submission

Technology:

technology as its predicate devices.

The Brivo NM 615 employs the same fundamental scientific

Determination of Substantial Equivalence:

Summary of Non-Clinical Tests:

The Brivo NM615 and its applications comply with voluntary standards as detailed in Section 9, of this premarket submission. The modifications from the predicate Discovery NM 630 system were completed in accordance with GE's quality management system design controls. Engineering testing and standards compliance testing were successfully conducted and did not raise any new safety questions or identify any new risks. The following quality assurance measures were applied to the development of the system:

  • . Risk Analysis
  • Requirements Reviews .
  • . Design Reviews
  • . Testing on unit level (Module verification)
  • . Integration testing (System verification)
  • Performance testing (Verification) .
  • . Safety testing (Verification)
  • . Simulated use testing (Validation)

Summary of Clinical Tests:

The subject of this premarket submission. Brivo NM 615. did not require clinical studies to support substantial equivalence.

GE Healthcare considers the Brivo Nm 615 to be as safe, as Conclusion: effective, and performance is substantially equivalent to the predicate device Discovery NM 630 (K111445).

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Eli Werner Regulatory Affairs Leader GE Healthcare GE Medical Systems Israel, Functional Imaging 4 Hayozma Street 30200 TIRAT HACARMEL ISRAEL

DEC - 5 2011

Re: K113350

Trade/Device Name: Brivo NM 615 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS & JAK Dated: November 9, 2011 Received: November 14, 2011

Dear Mr. Werner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, ' Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours.

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular border. The letters and the border are white, set against a black background, creating a high-contrast design. The logo is simple, recognizable, and represents the General Electric brand.

GE Healthcare 510(k) Premarket Notification Submission

Indications for Use Form

510(k) Number (if known): _ |< // 3350

Device Name: Brivo NM 615

Indications for Use:

The GE Brivo NM 615 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy.

The GE Brivo NM 615 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring.

The GE Brivo NM 615 system may include signal analysis and display equipment. patient and equipment supports, components and accessories. The system may be used for patients of all ages.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

M.A.D.O.K.

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K113350

Page 1 of 1

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.