(86 days)
DiaCheck Smart Blood Glucose Monitoring System: The DiaCheck Smart Blood Glucose Monitoring System is intended for the quantitative rno Blacmost of glucose in fresh capillary whole blood samples from the fingertips by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The DiaCheck Smart Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. This system is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates. DiaCheck Smart Blood Glucose Test Strips: The DiaCheck Smart Blood Glucose Test Strips are for use with the DiaCheck Smart Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples from the fingertip. The Test Strip is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates. DiaCheck Glucose Control Solution: The DiaCheck Control solutions are intended for use with DiaCheck Blood Glucose The DiaCheck Blood Glucose Test Strips as a quality control check to verify that the meter and test strips are working together properly and that the test is being performed correctly.
DiaCheck Superior Blood Glucose Monitoring System: The DiaCheck Superior Blood Glucose Monitoring System is intended for the rne Black our of glucose in fresh capillary whole blood samples from the fingertips by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The DiaCheck Superior Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. This system is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates. DiaCheck Superior Blood Glucose Test Strip: The DiaCheck Superior Blood Glucose Test Strips are for use with the DiaCheck rne Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples from the fingertip. The Test Strip is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates. DiaCheck Glucose Control Solution: The DiaCheck Control solutions are intended for use with DiaCheck Blood Glucose The DiaCheck Blood Glucose Test Strips as a quality control check to verify that the meter and test strips are working together properly and that the test is being performed correctly.
Not Found
Please note that the provided document is a 510(k) clearance letter from the FDA, not a full study report or clinical trial documentation. Therefore, much of the detailed information typically found in such reports (like sample sizes for specific training/test sets, expert qualifications, adjudication methods, or MRMC study results) is not present. This letter primarily focuses on the FDA's determination of substantial equivalence based on the submitted materials, rather than providing a detailed breakdown of the underlying studies.
However, I can extract the acceptance criteria and the reported performance as presented in this document, and infer what can be from the context.
Acceptance Criteria and Device Performance for Glucose Test System
The DiaCheck Smart Blood Glucose Monitoring System and DiaCheck Superior Blood Glucose Monitoring System are intended for quantitative measurement of glucose in fresh capillary whole blood samples from fingertips. Based on the provided FDA 510(k) clearance, the acceptability of the device is determined by its substantial equivalence to legally marketed predicate devices, implying that its performance meets the regulatory standards for accuracy and reliability established for such systems.
The specific acceptance criteria for blood glucose monitoring systems typically involve accuracy metrics compared to a laboratory reference method. While these exact criteria are not explicitly detailed in the FDA clearance letter, common standards (e.g., ISO 15197) for blood glucose meters require that for glucose concentrations ≥ 4.2 mmol/L (75 mg/dL), 95% of the measured values should fall within ±15% of the reference value, and for glucose concentrations < 4.2 mmol/L (75 mg/dL), 95% of the measured values should fall within ±0.83 mmol/L (±15 mg/dL) of the reference value.
The document indicates that the device has met the requirements for substantial equivalence, implying its performance aligns with these or similar established standards for blood glucose monitoring.
1. Table of Acceptance Criteria and the Reported Device Performance
| Acceptance Criteria (Inferred from regulatory standards for blood glucose meters, e.g., ISO 15197) | Reported Device Performance (Inferred from FDA's substantial equivalence determination) |
|---|---|
| Accuracy relative to a reference method: | The FDA has determined the device to be substantially equivalent to legally marketed predicate devices. This implies that the DiaCheck Superior Blood Glucose Monitoring System's performance for quantitative glucose measurement in capillary whole blood meets the established accuracy and precision requirements of the predicate devices and regulatory standards. The letter itself does not provide specific performance statistics (e.g., % within ±15% of reference, mean bias, etc.) but rather acknowledges that the submitted data demonstrated this equivalence. |
| - For glucose concentrations ≥ 4.2 mmol/L (75 mg/dL): 95% of measured values within ±15% of reference. | |
| - For glucose concentrations < 4.2 mmol/L (75 mg/dL): 95% of measured values within ±0.83 mmol/L (±15 mg/dL) of reference. | |
| Precision (repeatability/intermediate precision): (Typically assessed through coefficients of variation) | (Not detailed in the provided document, but implicitly met for substantial equivalence.) |
| Interfering substances: No significant interference from common endogenous or exogenous substances. | (Not detailed in the provided document, but implicitly met for substantial equivalence.) |
| Operating range: Measurement capabilities across a specified glucose concentration range for intended use. | (Not detailed in the provided document, but implicitly met for substantial equivalence within the specified indications for use.) |
2. Sample size used for the test set and the data provenance
The provided FDA 510(k) clearance letter does not specify the exact sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). These details would typically be found in the manufacturer's submission documentation, which is not included here.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not available in the provided FDA clearance letter. For glucose monitoring systems, ground truth is typically established using a laboratory reference method (e.g., YSI analyzer) and does not typically involve expert consensus for individual readings.
4. Adjudication method for the test set
Not applicable, as ground truth for this type of device is established by a reference laboratory method rather than human expert consensus requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable as the DiaCheck Superior Blood Glucose Monitoring System is a standalone diagnostic device for quantitative bodily fluid measurement, not an AI-assisted diagnostic imaging or interpretation tool for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, this device is inherently a standalone diagnostic system. Its performance evaluation would have been conducted as a standalone system (algorithm/device only) measuring glucose levels, with human involvement primarily in operating the device and collecting samples as per instructions for use. The FDA's substantial equivalence determination confirms that the device itself meets performance criteria.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For blood glucose monitoring systems, the ground truth is established using a laboratory reference method (e.g., YSI Glucose Analyzer) which is considered a highly accurate and precise method for measuring glucose concentration in blood samples.
8. The sample size for the training set
The provided FDA 510(k) clearance letter does not specify a training set size. While in vitro diagnostic devices like blood glucose meters may involve internal optimization or calibration using various samples, the concept of a "training set" in the context of typical machine learning algorithms with explicitly defined sizes is not directly applicable or typically reported in this manner for these types of devices in regulatory documents.
9. How the ground truth for the training set was established
Ground truth for any internal development, calibration, or training (if applicable) for the DiaCheck Blood Glucose Monitoring System would have been established using laboratory reference methods (e.g., YSI Glucose Analyzer) to provide accurate glucose concentration values for the samples used. This ensures the device is calibrated against a highly reliable standard.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple and iconic, representing the HHS's role in promoting the health and well-being of the nation.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Delbio Incorporation c/o Nicky Pan 3F. No. 252, Shangying Road Guishan Industrial Zone Taoyuan County China (Taiwan) 33341
FEB - 8 2012
K113343 Re:
R1100 15 DIACHECK SUPERIOR BLOOD GLUCOSE MONITORING SYSTEM
Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW, CGA, JJX Dated: December 9, 2011 Received: January 9, 2012
Dear Nicky Pan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Amendinens, or to act resol and metic Act (Act) that do not require approval of a premarket the Federal I vou, Drag, and Coou may, therefore, market the device, subject to the general approvin approvisions of the Act. The general controls provisions of the Act include controls provisions of the samual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III If your device to such additional controls. Existing major regulations affecting (1 Nr.), It may of says. It Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. your device can be found in Therther announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not I least be advised that I DTF o lessination that your device complies with other requirements mean that I DA mas made a statutes and regulations administered by other Federal agencies. of the Act of any I ouctures and submitted including, but not limited to: registration in rou intest compt) was t 807); labeling (21 CFR Parts 801 and 809); medical device and insting (21 CF R F a v 80 - ), device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
Signature
Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DELBio Incorporation Special 510(k)
DiaCheck Smart Blood Glucose Monitoring System DiaCheck Superior Blood Glucose Monitoring System
Indications for Use
510(k) Number (if known): K113343
Device Name: DiaCheck Smart Blood Glucose Monitoring System
Indications for Use:
DiaCheck Smart Blood Glucose Monitoring System:
The DiaCheck Smart Blood Glucose Monitoring System is intended for the quantitative rno Blacmost of glucose in fresh capillary whole blood samples from the fingertips by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
The DiaCheck Smart Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. This system is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates.
DiaCheck Smart Blood Glucose Test Strips:
The DiaCheck Smart Blood Glucose Test Strips are for use with the DiaCheck Smart Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples from the fingertip. The Test Strip is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates.
DiaCheck Glucose Control Solution:
The DiaCheck Control solutions are intended for use with DiaCheck Blood Glucose The DiaCheck Blood Glucose Test Strips as a quality control check to verify that the meter and test strips are working together properly and that the test is being performed correctly.
Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use x (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) |
|---|
| ------------------------------------------------------------------- |
| Division Sign-Off |
|---|
| Office of In Vitro Diagnostic Device |
| Evaluation and Safety |
| 510(k) K 11 3343 |
Page 1 of 2
{3}------------------------------------------------
DELBio Incorporation . Special 510(k)
DiaCheck Smart Blood Glucose Monitoring System DiaCheck Superior Blood Glucose Monitoring System
Indications for Use
510(k) Number (if known): K113343
Device Name: DiaCheck Superior Blood Glucose Monitoring System
Indications for Use:
DiaCheck Superior Blood Glucose Monitoring System:
The DiaCheck Superior Blood Glucose Monitoring System is intended for the rne Black our of glucose in fresh capillary whole blood samples from the fingertips by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
The DiaCheck Superior Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. This system is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates.
DiaCheck Superior Blood Glucose Test Strip:
The DiaCheck Superior Blood Glucose Test Strips are for use with the DiaCheck rne Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples from the fingertip. The Test Strip is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates.
DiaCheck Glucose Control Solution:
The DiaCheck Control solutions are intended for use with DiaCheck Blood Glucose The DiaCheck Blood Glucose Test Strips as a quality control check to verify that the meter and test strips are working together properly and that the test is being performed correctly.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _______________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)_K 11 334
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.