(136 days)
MD-6000 Bladder Scanner is a portable battery-operated ultrasonic equipment of pulse reflection. It is intended to measure the volume of bladder filled with urine.
The MD-6000 is intended to be used only by qualified medical professionals. Contraindications for the MD-6000 are fetal use and use on pregnant patients.
MD-6000 Bladder Scanner is a portable battery-operated ultrasonic equipment of pulse reflection. It utilizes ultrasonic distance measuring principle to calculate the bladder volume.
It makes mechanical sector scan by 2.5MHz ultrasonic wave and recognizes the reflected wave of the front and back wall of bladder to get the area information of bladder section. It calculates the volume of bladder by volume integral algorithm on the basis of the area information of 12 reference planes which are changed automatically with an interval of 15°.
MD-6000 Bladder Scanner has a Pre-Scan function, which shows the real-time B mode image for sectional plane of bladder. The Pre-Scan function helps a user to locate the bladder easily and get more accurate results.
Built-in thermal printer provides convenience of printing data.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Device: MD-6000 Bladder Scanner
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Measuring Range | 20-999 ml |
| Measuring Accuracy | ±15% |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: 10 volunteers.
- Data Provenance: Not explicitly stated whether the data was retrospective or prospective. The study mentions "collected and evaluated" and "validation to the measuring accuracy of bladder volume to 10 volunteers," which suggests a prospective collection for this specific validation. The country of origin is implied to be China, as the submitter, MEDA Co., Ltd., is located in Tianjin, P.R. China.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Number of Experts: "Professional doctors" were entrusted to make clinical effectiveness tests. The exact number of doctors is not specified, nor are their specific qualifications (e.g., years of experience or specialty).
4. Adjudication Method for the Test Set:
- The document does not describe a specific adjudication method like 2+1 or 3+1 for establishing ground truth from multiple experts. It states that "professional doctors" conducted clinical effectiveness tests, and a comparison was made between the device's measurement and the actual volume of excreted urine. This implies a direct comparison rather than an expert consensus adjudication process.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:
- No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done or reported in this summary. The device's primary function described is measuring bladder volume, not assisting human interpretation of complex medical images.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Yes, a standalone performance evaluation was done. The "validation to the measuring accuracy of bladder volume to 10 volunteers" involved comparing the device's measurement directly to the "actual volume of excreted urine." This assesses the device's performance in isolation.
7. The Type of Ground Truth Used:
- The ground truth for the device's accuracy validation was "actual volume of excreted urine." This is a direct, objective measurement rather than an expert consensus or pathology report.
8. The Sample Size for the Training Set:
- The document does not specify a sample size for a training set. The descriptions focus on the validation study. Ultrasonic bladder scanners typically rely on algorithms derived from physical principles (ultrasonic distance measuring principle, volume integral algorithm) rather than learning from a large training dataset in the same way modern AI algorithms do.
9. How the Ground Truth for the Training Set Was Established:
- As a training set is not explicitly mentioned (or likely not applicable in the modern sense of AI/ML for this type of device based on the description), there is no information on how its ground truth was established. The device utilizes "ultrasonic distance measuring principle to calculate the bladder volume" and a "volume integral algorithm," which are pre-determined mathematical models rather than learned from a training set with ground truth labels.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image is a logo that appears to be for the MFDA. The logo is an oval shape with a map of China in the center. The letters "MFDA" are written on the right side of the map. The logo is black and white.
MEDÁ CO... LT
4113304
MAR 23 2012
Al-5: 510(k) Summary
510(k) SUMMARY
1) Submitter Information
Name: MEDA. Co., Ltd. Address: F2C, F3D, F4C, F5, Building C2, Xinmao Science Skill Park, Huayuan Industry Development Area, Tianjin 300384, P. R. China Phone: +86-22-83713808 Fax: +86-22-83713880
Contact person: KAI CHEN Medtech International, Inc. and United States Designated Agent of MEDA Co., Ltd. Address: 13505 Broadfield Drive, Potomac, MD 20854 Tel: 240 888 4001 Fax: 301 251 2881 Drkaichen@gmail.com
Date Prepared: October 9, 2011
2) Device Information
Trade Name: MD-6000 Bladder Scanner Common Name: Diagnostic Ultrasound System with Accessories Classification Name: Ultrasonic Pulsed Echo Imaging System Diagnostic Ultrasound Transducer Regulation Number: 892.1560, 892.1570 Product Code: IYO, ITX
3) Predicate Devices
For identification of predicate device, we retrieved the legally marketed similar products from FDA's official website and according to the published 510(k) Summary, we selected the following device which has the same intended use and working principle, and is equivalent in structure (including accessories) and technical characteristics as the predicate device:
Manufacturer: Mcube Technology Co., Ltd. Device: CUBEscan /BioCon-500 510(k) Number: K091518
Add: F2C, F3D, F4C, F5, Building C2, Xinmao Science Skill Park, Huayuan Ind. Dev. Area, Tianjin 300384, P. R. China URL: http://www.MEDA.com.cn Tel: +86-22-83713808 E-mail: export(@meda.com.cn Fax: +86-22-83713880
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with an oval shape. Inside the oval, there is a stylized eye-like design. The letters "MEDA" are visible within the eye design, positioned towards the right side. The logo is presented in black and white.
MEDA CO., LTD
4) Device Description
MD-6000 Bladder Scanner is a portable battery-operated ultrasonic equipment of pulse reflection. It utilizes ultrasonic distance measuring principle to calculate the bladder volume.
It makes mechanical sector scan by 2.5MHz ultrasonic wave and recognizes the reflected wave of the front and back wall of bladder to get the area information of bladder section. It calculates the volume of bladder by volume integral algorithm on the basis of the area information of 12 reference planes which are changed automatically with an interval of 15° .
MD-6000 Bladder Scanner has a Pre-Scan function, which shows the real-time B mode image for sectional plane of bladder. The Pre-Scan function helps a user to locate the bladder easily and get more accurate results.
Built-in thermal printer provides convenience of printing data.
5) Intended Use
MD-6000 Bladder Scanner is intended to measure the volume of bladder filled with urine.
The MD-6000 is intended to be used only by qualified medical professionals. Contraindications for the MD-6000 are fetal use and use on pregnant patients.
6) Technological Characteristics
a) Safety
Electrical, mechanical, environmental safety testing according to standard IEC 60601-1:1988+A1+A2 and IEC 60601-2-37:2007 was performed. EMC testing was conducted in accordance with standard IEC 60601-1-2:2007. All test results meet the requirements of the standard.
The safety standards performed by MD-6000 Bladder Scanner are identical with that of the predicate product.
b) Characteristics
Both MD-6000 and its predicate device are intended to measure the volume of bladder, and their Pre-Scan functions are real-time images for sectional plane of bladder. The measuring range of MD-6000 is 20-999ml and measuring accuracy is 15%, which are substantially equivalent with that of the predicate device.
The main differences of MD-6000 Bladder Scanner and its predicate device are: apart from the slight difference in ultrasound frequency (2,5MHz/2.8MHz), the monitor, housing, battery and battery charger of MD-6000 afe in different model, but the safety tests showed that they all meet the requirements.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with an oval shape resembling an eye. Inside the oval, there's a globe-like design with continents visible. The text "MEDA" is written on the right side of the globe. The logo is in black and white.
MEDA CO., LTD
While there are some differences between the MD-6000 and its predicate device, they do not affect the safety and effectiveness of the device.
7) Brief Discussion of nonclinical Tests
The safety tests that are based on FDA-recognized standards were conducted bv TÜV SÜD Laboratories, and all results comply with the requirements of standards.
The software and essential performance have passes verification and validation, and the results comply with the requirements.
8 ) Brief Discussion of Clinical Evaluation and Validation
The clinical data of Bladder Scanner have been collected and evaluated.
As part of the design validation, we have made validation to the measuring accuracy of bladder volume to 10 volunteers, with the method of comparing the measuring value and the actual volume of excreted urine. Moreover, we have entrusted professional doctors to make clinical effectiveness tests in medical environments. The clinical test results showed that MD-6000 Bladder Scanner met the requirements of intended clinical applications.
No adverse effect was found during the collection and evaluation of clinical data and clinical validation.
9) Conclusions
The results of nonclinical tests as well as clinical evaluation and validation demonstrate that the MD-6000 Bladder Scanner is equivalent in safety, effectiveness and performance to the legally marketed predicate device.
MEDA CO., LTD will update and include in this summary any other information deemed reasonably necessary by the FDA.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an abstract human figure with three arms or extensions, positioned to the right of the department's name. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the graphic.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
MAR 2 3 2012
MEDA. Co., Ltd. % Mr. Kai Chen United States Designated Agent Medtech International. Inc. 13505 Broadfield Drive POTOMAC MD 20854
Re: K113304
Trade/Device Name: MD-6000 Bladder Scanner Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO and ITX Dated: March 1, 2012 Received: March 1, 2012
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Al-4: Indications for Use Statement
MD-6000 Bladder Scanner
Indications for Use Statement
510(k) Number (if known): K113304
Device Name: MD-6000 Bladder Scanner
Indications For Use:
MD-6000 Bladder Scanner is a portable battery-operated ultrasonic equipment of pulse reflection. It is intended to measure the volume of bladder filled with urine.
The MD-6000 is intended to be used only by qualified medical professionals. Contraindications for the MD-6000 are fetal use and use on pregnant patients.
| Prescription Use | √ | AND/OR | Over-The-Counter Use | |
|---|---|---|---|---|
| (Part 21 CFR 801 Subpart D) | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
(Division Sign-Off)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K113304
N/A