K Number
K113135

Validate with FDA (Live)

Device Name
UNISTRIP 1
Date Cleared
2013-11-06

(744 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Unistrip1™ Test Strips are used with the OneTouch "Ultra", OneTouch" Ultra2", OneTouch" UltraMini and One-Touch UltraSmart meters purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The Unistrip1™ is meant for self-testing of blood . glucose as an aid to monitor the effectiveness of diabetes control.

They are for single patient use only and should not be shared.

They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm.

Testing is done outside the body (in vitro diagnostic use).

They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control.

Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

UniStrip1™ Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch Ultra meter). Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly). Alternative site testing (AST) testing should only be done during steady-state times (when glucose is not changing rapidly).

Device Description

The UniStrip1™ Test Strips are used with the OneTouch® Ultra®, OneTouch® Ultra2®, OneTouch® UltraMini® and QneTouch® UltraSmart® blood glucose monitoring systems. The UniStrip1™ test Strips is intended for use outside the body (in vitro diagnostic use) by people with diabetes using the OneTouch® Ultra®, OneTouch® Ultra2®, OneTouch® UltraMini® and OneTouch® UltraSmart® blood glucose monitoring systems as an aid to monitor the effectiveness of diabetes control. The UniStrip1™ Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch Ultra meter).

The UniStrip1™ Test Strip can only be used on the OneTouch®Ultra®, OneTouch®Ultra2®, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before October 2012. UniStrip1™ Test Strips are only for use with calibration code 49.

AI/ML Overview

The UniStrip1™ Test Strips are intended for use with specific OneTouch® blood glucose monitoring systems to measure glucose in capillary whole blood. The device manufacturer, UniStrip Technologies, LLC, conducted several non-clinical studies to demonstrate the safety and effectiveness of the UniStrip1™ Test Strips.

1. Table of Acceptance Criteria and Reported Device Performance:

The provided document does not explicitly state numerical acceptance criteria for each test. Instead, it lists the types of tests conducted (e.g., NCCLS EP5-A Precision Test, ISO 15971 System Accuracy Evaluation) and concludes that "no new issues of safety and effective have been raised." This implies that the device results met the established performance standards of these internationally recognized test protocols. For example, for an ISO 15971 System Accuracy Evaluation, typical acceptance criteria would involve a certain percentage of results falling within specific deviation ranges from a reference method (e.g., within ±15 mg/dL for glucose concentrations < 100 mg/dL and within ±15% for glucose concentrations ≥ 100 mg/dL). However, these specific numerical criteria and the detailed performance results are not provided in this summary.

Therefore, the table below will list the tests performed, and for "Reported Device Performance," it will indicate that the tests were conducted and their outcomes were deemed satisfactory by the manufacturer for substantial equivalence.

Acceptance Criteria (Implied by Test Standard)Reported Device Performance (Summary)
NCCLS EP5-A Precision TestTest conducted; results supported safety and effectiveness
NCCLS EP6-A Linearity TestTest conducted; results supported safety and effectiveness
ISO 15971 and EP7-A2 Interference TestsTests conducted; results supported safety and effectiveness
ISO 15971 System Accuracy EvaluationTest conducted; results supported safety and effectiveness
ISO 15971 and ISO 5725-1 Comparison TestTests conducted; results supported safety and effectiveness
ISO 15971 Altitude TestTest conducted; results supported safety and effectiveness
Volume Verification StudyStudy conducted; results supported safety and effectiveness
EN 13640 Operation Conditions Tests for Test StripTests conducted; results supported safety and effectiveness
EN 13640 Storage Conditions Test for Test StripTest conducted; results supported safety and effectiveness
ISO 15971 Control Solution testTest conducted; results supported safety and effectiveness
ISO 159712 and ISO 5725-1 User EvaluationTests conducted; results supported safety and effectiveness
Error Code Comparison TestsTests conducted; results supported safety and effectiveness

2. Sample Size Used for the Test Set and the Data Provenance:

The document does not explicitly state the sample sizes used for each of the non-clinical tests (e.g., for the ISO 15971 System Accuracy Evaluation). It also does not specify the country of origin of the data or whether the tests were retrospective or prospective. It only lists the types of tests performed.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

This information is not provided in the document. The studies listed are primarily non-clinical laboratory tests, which typically rely on reference methods (e.g., laboratory analyzers) rather than expert consensus for ground truth establishment.

4. Adjudication Method for the Test Set:

This information is not provided in the document. Given that the studies were non-clinical laboratory tests, adjudication methods like 2+1 or 3+1 (commonly used for subjective interpretations by multiple human readers) would not be applicable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No clinical studies were conducted." The studies performed were non-clinical laboratory and bench tests.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, all the listed studies were standalone performance evaluations of the UniStrip1™ Test Strips in conjunction with the specified blood glucose meters. There was no human-in-the-loop component in these non-clinical tests; they evaluated the device's inherent measurement capabilities.

7. The Type of Ground Truth Used:

The ground truth for the non-clinical tests would typically be established using reference laboratory methods or calibrated instruments as defined by the standards (e.g., ISO 15971, NCCLS EP documents). For instance, in a system accuracy evaluation, blood glucose values would be determined by a sophisticated laboratory analyzer, and these values would serve as the "ground truth" against which the UniStrip1™ Test Strips' measurements are compared.

8. The Sample Size for the Training Set:

The concept of a "training set" is relevant for machine learning algorithms. Blood glucose test strips are typically electrochemical or photometric devices, and while they involve internal calibration curves, they generally do not involve a "training set" in the sense of a machine learning algorithm. The document does not mention any artificial intelligence or machine learning components that would necessitate a training set. Therefore, this information is not applicable and not provided.

9. How the Ground Truth for the Training Set Was Established:

As explained in point 8, the concept of a "training set" as it pertains to machine learning is not applicable to this device. Therefore, the method for establishing ground truth for a training set is not provided.

{0}------------------------------------------------

K11313:5

510(k) Summary

A summary of 510k safety and effectiveness information in accordance with 21 CFR 807.92

Date Prepared:November 5, 2013
Sponsor Information:UniStrip Technologies, LLC.
Sponsor Address:2701-A Hutchison McDonald RoadCharlotte, NC 28269
Sponsor Contact:Richard Admani
Sponsor Tel Number:1-704-285-6400
Sponsor Fax Number:1-704-285-6495
Device Name:UniStrip1™ Test Strips
Common/Usual Name:Blood Glucose Test Strips
Classification Name:Glucose, Oxidase, Glucose
Regulation Number:21 CFR 862.1345
Product Code:NBW, CGA
Device Class:Class II
Review Panel:Clinical Chemistry

NOV 062013

Predicate Device: Manufacturer

Lifescan, Inc.

Brand Name
OneTouch® Test Strips

510(k) Number K923544

Device Description

The UniStrip1™ Test Strips are used with the OneTouch® Ultra®, OneTouch® Ultra2®, OneTouch® UltraMini® and QneTouch® UltraSmart® blood glucose monitoring systems. The UniStrip1™ test Strips is intended for use outside the body (in vitro diagnostic use) by people with diabetes using the OneTouch® Ultra®, OneTouch® Ultra2®, OneTouch® UltraMini® and OneTouch® UltraSmart® blood glucose monitoring systems as an aid to monitor the effectiveness of diabetes control. The UniStrip1™ Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch Ultra meter).

The UniStrip1™ Test Strip can only be used on the OneTouch®Ultra®, OneTouch®Ultra2®, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before October 2012. UniStrip1™ Test Strips are only for use with calibration code 49.

Indications for Use

The Unistrip1™ Test Strips are used with the OneTouch "Ultra", OneTouch" Ultra2", OneTouch" UltraMini and One-Touch UltraSmart meters purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The Unistrip1™ is meant for self-testing of blood . glucose as an aid to monitor the effectiveness of diabetes control.

They are for single patient use only and should not be shared.

They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm.

Testing is done outside the body (in vitro diagnostic use).

They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control.

Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

{1}------------------------------------------------

UniStrip1™ Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch Ultra meter). Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly). Alternative site testing (AST) testing should only be done during steady-state times (when glucose is not changing rapidly).

Similarities to Predicate Device
CharacteristicPredicate DeviceNew Device
Reagent CompositionGlucose Oxidase (Aspergillus niger)Same as predicate
Specimen TypeFresh capillary whole bloodSame as predicate
Sample Volume1.0 µLSame as predicate
Control SolutionUse OneTouch Ultra Control SolutionUse Prodigy Control Solution
Detecting Range20-600 mg/dLSame as predicate
CalibrationVarious codesCode 49 Only
Altitude UseUp to 10,000 feetSame as predicate
Storage TemperatureLess than 86°F (30° C)39°-86°F (4°-30°C)
Open Vial Shelf Life6 months after opening90 days after opening

Predicate Product Comparison Similarities to Predicate Device

Differences with Predicate Device

CharacteristicPredicate DeviceNew Device
Control SolutionUse OneTouch Ultra Control SolutionUse Prodigy Control Solution
CalibrationVarious codesCode 49 Only
Storage TemperatureLess than 86°F (30° C)39°-86°F (4°-30°C)
Open Vial Shelf Life6 months after opening90 days after opening

Safety and Effectiveness Tests and Studies Non-Clinical

The following tests and studies were conducted to ensure the UniStrip1™ Test Strips were safe and effective when using the UniStrip1™ Test Strip with the following blood glucose meters -OneTouch®Ultra®, OneTouch®Ultra2®, OneTouch® UltraMini® and One-Touch® UltraSmart®.

0 NCCLS EP5-A Precision Test

0 NCCLS EP6-A Linearity Test

ඹ ISO 15971 and EP7-A2 Interference Tests

₪ ISO 15971 System Accuracy Evaluation

₪ ISO 15971 and ISO 5725-1 Comparison Test

0 ISO 15971 Altitude Test

@ Volume Verification Study

₪ EN 13640 Operation Conditions Tests for Test Strip

@ EN 13640 Storage Conditions Test for Test Strip

@ ISO 15971 Control Solution test

ISO 159712 and ISO 5725-1 User Evaluation

@ Error Code Comparison Tests

{2}------------------------------------------------

Clinical Studies No clinical studies were conducted.

Conclusion

Based on the Predicate Product Comparison Table of similarities, standard and FDA guidance tests and study results UniStrip Technologies, LLC concludes that no new issues of safety and effective have been raised in this original 510(k) submission for the UniStrip1™ Test Strips to be used with the OneTouch®Ultra®, OneTouch®Ultra2®, OneTouch® UltraMini® and One-Touch® UltraSmart® blood glucose meters purchased before October 2012.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three wave-like lines below the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 6, 2013

UniStrip Technologies, LLC Richard Admani Chief Operations Officer 2701-A Hutchison McDonald Rd. . CHARLOTTE, NC 28269

Re: K113135

Trade/Device Name: UniStrip I Test Strip Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW. CGA Dated: November 01, 2013 Received: November 04, 2013

Dear Mr. Admani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2-Mr. Admani

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.hum.

Sincerely yours,

Carol C. Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health

Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K113135

Device Name: UniStrip1 Test Strip

Indications for Use:

The Unistrip 1 Test Strips are used with the OneTouch® Ultra®, OneTouch®Ultra2®, OneTouch® UltraMini® and One-Touch® UltraSmart® meters purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The Unistrip I is meant for self-testing of blood glucose as an aid to monitor the effectiveness of diabetes control.

Thev are for single patient use only and should not be shared.

They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm.

Testing is done outside the body (in vitro diagnostic use).

They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control.

Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

UniStrip1 Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch®Ultra® meter). Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly).

Over-The-Counter Use AND/OR x Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Stayce Beck

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k)

Page 1 of 1

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.