K Number
K113086

Validate with FDA (Live)

Device Name
ACQA
Date Cleared
2012-01-13

(87 days)

Product Code
Regulation Number
892.1200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACQA system is a software on GE Advantage Windows workstation computers that allow the user to create averaged CT images for attenuation correction of the whole body or myocardial perfusion PET images, and to register CT images with myocardial perfusion PET images, and save registered CT images for attenuation correction of myocardial perfusion PET images. The PET images corrected with the CT images of ACQA are not a replacement of the original PET images, and they serve as a secondary data set to assist the physicians read the original PET images.

Device Description

ACQA is a software application installed on GE Advantage Windows workstation computers which allows physicians and healthcare professionals (1) to create averaged CT images for attenuation correction of the whole body oncology or myocardial perfusion PET images, and (2) to register the averaged or helical CT images with their corresponding myocardial perfusion PET images, and save the registered CT images for attenuation correction of the myocardial perfusion PET images. Use of this system is limited to qualified, licensed healthcare providers (radiologists, nuclear cardiologists, or nuclear medicine physicians) trained in the use of nuclear medicine imaging devices.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study findings for the ACQA system, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary for the ACQA system is a general summary and does not include specific quantitative acceptance criteria or detailed performance metrics. However, it does state a general conclusion:

Acceptance Criteria (Implied)Reported Device Performance
Device meets specifications"The tests with both laboratory and clinical data sets included in the submission passed the requirements and met specifications."
Functionality (Attenuation Correction)"create averaged CT images for attenuation correction of the whole body oncology or myocardial perfusion PET images"
Functionality (Image Registration)"register the averaged or helical CT images with their corresponding myocardial perfusion PET images, and save the registered CT images for attenuation correction of the myocardial perfusion PET images."
Safety & Effectiveness (Assisted Reading)"The PET images corrected with the CT images of ACQA are not a replacement of the original PET images, and they serve as a secondary data set to assist the physician read the original PET images."

2. Sample Size Used for the Test Set and Data Provenance

The 510(k) summary mentions "both laboratory and clinical data sets" were used for testing. However, it does not specify the sample size for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

The 510(k) summary does not provide information on the number of experts used to establish ground truth or their qualifications for the test set.

4. Adjudication Method for the Test Set

The 510(k) summary does not provide information on the adjudication method used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The 510(k) summary does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it quantify an effect size for human readers improving with AI vs. without AI assistance. The device is described as assisting physicians, but direct comparison of physician performance with and without the tool is not detailed.

6. Standalone (Algorithm Only) Performance Study

The 510(k) summary indicates that the ACQA system performs specific functions (creating averaged CT images and registering images) as a software application. The statement "The tests with both laboratory and clinical data sets included in the submission passed the requirements and met specifications" implies some form of standalone performance evaluation of these technical functionalities. However, it doesn't explicitly detail a standalone performance study in terms of traditional diagnostic metrics (e.g., sensitivity, specificity) for a clinical diagnosis, as the output is described as "secondary data set to assist the physician read the original PET images."

7. Type of Ground Truth Used

The 510(k) summary does not explicitly state the type of ground truth used. Given the device's function (attenuation correction and image registration), the "ground truth" would likely relate to the accuracy of these technical tasks rather than a clinical diagnosis. This could involve:
* Image registration accuracy metrics: Comparing the alignment achieved by ACQA to a manually confirmed or gold-standard registration.
* Attenuation correction accuracy: Comparing PET images corrected by ACQA to those corrected by established methods or phantoms with known attenuation properties.

8. Sample Size for the Training Set

The 510(k) summary does not specify the sample size for the training set.

9. How the Ground Truth for the Training Set Was Established

The 510(k) summary does not provide information on how the ground truth for the training set was established. Given the nature of the device (software for image processing, not a diagnostic AI system predicting disease), it's possible that the "training set" in the context of this device refers to data used for algorithm development and testing of technical parameters, rather than a clinical ground truth labeling.

{0}------------------------------------------------

K113086

510(K) Summary as per 21 CFR 807.92(c)

JAN 1 3 2012

Image /page/0/Picture/3 description: The image shows the logo for Texas Medical Imaging Consultants (TMIC). The logo consists of the letters "TMIC" in large, bold, sans-serif font. Below the letters, in a smaller font, are the words "TEXAS MEDICAL IMAGING CONSULTANTS."

General Information:

A. Submitted by:Texas Medical Imaging Consultants
4027 Underwood St.
Houston, TX 77025
Tel: 713-409-3801

Tinsu Pan Contact Person:

Date Summary Prepared: 8/25/2011

  • Device Trade Name: ACQA B.
    Classification Code: KPS

System, Emission Computed Tomography (per 21 CFR 892 1200) Classification Name: Accessory to Emission Computed Tomography System

  • C. Predicate Devices:
    GE Healthcare Technologies (formerly GE Medical Systems) - GE Discovery ST-K041543 GE Medical Systems -- Advantage 4D option - K032915

  • D. Device Description:
    ACQA is a software application installed on GE Advantage Windows workstation computers which allows physicians and healthcare professionals (1) to create averaged CT images for attenuation correction of the whole body oncology or myocardial perfusion PET images, and (2) to register the averaged or helical CT images with their corresponding myocardial perfusion PET images, and save the registered CT images for attenuation correction of the myocardial perfusion PET images. Use of this system is limited to qualified, licensed healthcare providers (radiologists, nuclear cardiologists, or nuclear medicine physicians) trained in the use of nuclear medicine imaging devices.

  • Intended Use of the Device: ﻧﻨ
    The ACQA system is a software on GE Advantage Windows workstation computers that allow the user to create averaged CT images for attenuation correction of the whole body or myocardial perfusion PET images, and to register CT images with myocardial perfusion PET images, and save registered CT images for attenuation correction of the myocardial perfusion PET images. The PET images corrected with the CT images of ACQA are not a replacement of the original PET images, and they serve as a secondary data set to assist the physician read the original PET images.

  • F. Comparison of Technical Characteristics to Predicate Device:
    The ACQA system and its predicates, the GE DST (K041543) and the GE Advantage 4D-option (K032915) utilize the same type of data sets for analysis and data processing. The ACQA, applicable to all GE PET/CT scanners, accomplishes the tasks of registration of CT and PET, and calculation of averaged CT. The same function of ACQA is only supported on the GE PET/CT

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for TMIC, which stands for Texas Medical Imaging Consultants. The logo is in black and white, with the letters "TMIC" in large, bold font. Below the letters, the words "TEXAS MEDICAL IMAGING CONSULTANTS" are written in a smaller font. The logo is simple and professional, and it is likely used to represent the company on its website, marketing materials, and other communications.

scanners with Dimension console, and the Advantage 4D option. A summary of differences is provided in the following table.

FunctionalityGE Discovery STGE Advantage 4D optionACQA
1. Attenuation correction ofPET data with averaged CTdata2. Registration of CT andPET data1. Calculation of averagedCT data with the VarianTMrespiratory gatinghardware1. Attenuation correctionof PET data with averagedCT data2. Registration of CT andPET data3. No VarianTM respiratorygating hardware isneeded.
Applicable PET/CTsGE PET/CT with DimensionconsoleAll GE PET/CT scannersAll GE PET/CT scanners
Advantages over ACQABoth data processing anddata correction can beperformed on theDimension consoleCan produce 10 phases of4DCT images for RadiationOncology applicationData processing on the GEAdvantage Windowsworkstation, andattenuation correction ofPET data on the PET/CTconsole.
ACQA Advantages overGE predicate devicesNo averaged CT data isproduced. The user canperform either attenuationcorrection by averaged CTdata or registration ofregular CT with PET data,and cannot use bothoptions.Can produce averaged CTdata. However, 4DCToption is typically installedfor radiation oncology,not for cardiology.ACQA is the only softwarethat does not require theDimension console or4DCT option (installationof both costs > $200k.) tocompute averaged CT datafor attenuation correctionof the PET data andregistration of the PETdata with either averagedCT or helical CT data.
Large installed base ofAdvantage WindowsworkstationOver 10,000 Advantage Windows workstations have been installed worldwide. It isvery likely that a clinic/hospital which wants to use the ACQA software is already withan Advantage Windows workstation. In this case, ACQA software can be installedwithout any addition of hardware.
  • G. Performance data: The tests with both laboratory and clinical data sets included in the submission passed the requirements and met specifications.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Tinsu Pan Chief Technology Officer Texas Medical Imaging Consultants 4027 Underwood Street HOUSTON TX 77025

Re: K113086

Trade/Device Name: ACQA System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS & JAK Dated: October 12, 2011 Received: October 18, 2011

Dear Mr. Pan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

JAN 1 3 2012

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary S Patel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

い。

The ACQA system is a software on GE Advantage Windows workstation computers that allow the user to create averaged CT images for attenuation correction of the whole body or myocardial perfusion PET images, and to register CT images with myocardial perfusion PET images, and save registered CT images for attenuation correction of myocardial perfusion PET images. The PET images corrected with the CT images of ACQA are not a replacement of the original PET images, and they serve as a secondary data set to assist the physicians read the original PET images.

Mary Pastel

ation and Safety Office of In

510K K113086

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.