K Number
K113076

Validate with FDA (Live)

Device Name
CAAS A-VALVE
Date Cleared
2012-01-30

(105 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CAAS A-Valve has been developed to support the interventionalist during or in preparation of treatment of the aortic root. Based on a set of angiographic X-ray images an analysis is performed:

  • To assist in C-arm projection selection to optimize visualization during treatment;
  • To calculate dimensions of the aortic root corrected for out-of-plane magnification and foreshortening errors.
Device Description

The CAAS A-Valve is a stand-alone software application intended to run on a PC with a Windows operating system. The CAAS A-Valve is designed for objective, accurate and reproducible assessment of the aortic root geometry from a set of angiographic X-ray images from different projections.

On each of the 2D images the aortic root is segmented. The segmented 2D aortic root contours in each image are used to generate a 3D reconstruction of the aortic root. A number of analysis results can be calculated:

  1. Optimal C-arm projection to optimize visualization during treatment;
  2. Dimensions of the aortic root.
    Results are corrected for out-of-plane magnification and foreshortening errors.
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CAAS A-Valve, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The provided text does not explicitly state acceptance criteria in a quantitative table format. However, it does state the device's purpose and performance claims, which can be inferred as the basis for its perceived efficacy.

Acceptance Criteria (Inferred from Intended Use/Description)Reported Device Performance
Objective, accurate, and reproducible assessment of the aortic root geometry from angiographic X-ray images.The CAAS A-Valve is designed for objective, accurate, and reproducible assessment of the aortic root geometry from a set of angiographic X-ray images.
Generation of 3D reconstruction of the aortic root from segmented 2D contours.The segmented 2D aortic root contours in each image are used to generate a 3D reconstruction of the aortic root.
Calculation of optimal C-arm projection for visualization during treatment.A number of analysis results can be calculated: Optimal C-arm projection to optimize visualization during treatment.
Calculation of aortic root dimensions, corrected for out-of-plane magnification and foreshortening errors.A number of analysis results can be calculated: Dimensions of the aortic root. Results are corrected for out-of-plane magnification and foreshortening errors.
Support for interventionalists during or in preparation for aortic root treatment, including C-arm positioning assistance.CAAS A-Valve has been developed to support the interventionalist during or in preparation of treatment of the aortic root. CAAS A-Valve quantifies dimensions of the aortic root and assists in C-arm positioning based on a set of angiographic images. Indications for use: To assist in C-arm projection selection to optimize visualization during treatment; To calculate dimensions of the aortic root corrected for out-of-plane magnification and foreshortening errors.
Substantial equivalence to predicate devices (CAAS QxA 3D and IC-Pro)."The testing reported in this 510(k) establishes that CAAS A-Valve is [substantially equivalent] to a combination of predicate devices and is safe and effective for its intended use." "The intended use and technological characteristics of CAAS A-Valve are [substantially equivalent...]" (The full phrase is cut off, but the intent is clear from the title "Substantial Equivalence").

2. Sample Size Used for the Test Set and Data Provenance:

The provided text does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It simply states: "Performance Data: CAAS A-Valve is developed, tested, validated and produced under the same Quality Assurance system applicable to the development and production of products currently marketed by Pie Medical Imaging." This is a general statement about their development process, not specific study details.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

The provided text does not specify the number of experts used or their qualifications for establishing ground truth for the test set.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:

The provided text does not specify any adjudication method for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Improvement:

The provided text does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor does it provide an effect size for human reader improvement with AI assistance. The focus of the 510(k) summary is on the device's standalone capabilities and its equivalence to predicate devices.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done:

Yes, the information suggests that a standalone performance evaluation was done. The device is described as a "stand-alone software application" that performs "objective, accurate and reproducible assessment." The "Performance Data" section implies that the device itself was "tested" and "validated" to meet its intended functions, which primarily involve automated analysis without explicit mention of human input affecting the core measurements. The stated intended uses ("To assist in C-arm projection selection," "To calculate dimensions") are outputs of the algorithm.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.):

The provided text does not explicitly state the type of ground truth used for validation. Given the nature of aortic root geometry and dimensions derived from angiographic images, it is highly probable that the ground truth would have been established through expert consensus or measurements performed by highly experienced cardiologists or radiologists using established manual or semi-automated techniques on the same angiographic images. However, this is an inference, not a direct statement.

8. The Sample Size for the Training Set:

The provided text does not specify the sample size for the training set.

9. How the Ground Truth for the Training Set Was Established:

The provided text does not specify how the ground truth for the training set was established.

{0}------------------------------------------------

510(k) SummaryCAAS A-Valve
Submitter NamePie Medical Imaging BV
Submitter AddressBecanusstraat 13 D; 6216 BX Maastricht, The Netherlands
Contact Person:Florie Daniels, Product Registration CoordinatorJAN 30 2012
Phone Number+31 43 32 81 328
Fax Number+31 43 32 81 329
Email AddressFlorie.Daniels@pie.nl
Preparation Date22 December 2011
Trade NameCAAS A-Valve
Common NameCardiovascular Angiography Analysis SystemCAAS A-Valve
Regulation ClassClass II (21 CFR, part 892.2050, LLZ)
Predicate DevicesCAAS QxA 3D, cleared under K100292IC-Pro, cleared under K110256
Device DescriptionThe CAAS A-Valve is a stand-alone software application intended to run on a PCwith a Windows operating system. The CAAS A-Valve is designed for objective,accurate and reproducible assessment of the aortic root geometry from a set ofangiographic X-ray images from different projections.On each of the 2D images the aortic root is segmented. The segmented 2D aorticroot contours in each image are used to generate a 3D reconstruction of the aorticroot. A number of analysis results can be calculated:1. Optimal C-arm projection to optimize visualization during treatment;2. Dimensions of the aortic root.Results are corrected for out-of-plane magnification and foreshortening errors.
Intended UseCAAS A-valve has been developed to support the interventionalist during or inpreparation of treatment of the aortic root. CAAS A-Valve quantifies dimensionsof the aortic root and assists in C-arm positioning based on a set of angiographicimages. The software is used by or under supervision of a cardiologist orradiologist.Indications for use:CAAS A-Valve has been developed to support the interventionalist during or inpreparation of treatment of the aortic root. Based on a set of angiographic X-rayimages an analysis is performed:- To assist in C-arm projection selection to optimize visualization duringtreatment;- To calculate dimensions of the aortic root corrected for out-of-planemagnification and foreshortening errors.
Performance DataCAAS A-Valve is developed, tested, validated and produced under the sameQuality Assurance system applicable to the development and production ofproducts currently marketed by Pie Medical Imaging.
Substantial EquivalenceThe intended use and technological characteristics of CAAS A-Valve are

{1}------------------------------------------------

.

The testing reported in this 510(k) establishes that CAAS A-Valve is substantial equivalent to a combination of predicate devices and is safe and effective for its intended use. :

:

ﻟﺴﻴﺮ ﺍﻟﻤ

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Ms. Florie Daniels Product Registration Coordinator Pie Medical Imaging BV Becanusstraat 13D 6216 BX MAASTRICHT LIMBURG THE NETHERLANDS

Re: K113076

Trade/Device Name: CAAS A-Valve Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 22, 2011 Received: December 28, 2011

Dear Ms. Daniels:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

JAN 3 0 2012

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours.

Mary Patel

Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K113076

Device Name: CAAS A-Valve

Indications for Use:

CAAS A-Valve has been developed to support the interventionalist during or in preparation of treatment of the aortic root. Based on a set of angiographic X-ray images an analysis is performed:

To assist in C-arm projection selection to optimize visualization during treatment; —

  • To calculate dimensions of the aortic root corrected for out-of-plane magnification and foreshortening errors.
    Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mary Pastel

Division Sign-C Office of In Vitro Diagnostic Device Evaluation and S

510(k) K113076

N/A