SMITH & NEPHEW RADIOPAQUE TRIAL NECKS

K113039 · Smith & Nephew, Inc. · MEH · Oct 5, 2012 · Orthopedic

Device Facts

Record IDK113039
Device NameSMITH & NEPHEW RADIOPAQUE TRIAL NECKS
ApplicantSmith & Nephew, Inc.
Product CodeMEH · Orthopedic
Decision DateOct 5, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2

Intended Use

Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity.

Device Story

Radiopaque Trial Necks are surgical instruments used during hip arthroplasty to determine the correct modular neck implant for Smith & Nephew SMF Hip Stem and MDF Revision Hip systems. Devices are manufactured from polyphenylsulfone (PPSU) blended with 6% barium sulfate, providing radiopacity to ensure visibility on X-rays if dislodged during trialing. Various offsets and anteversion angles are color-coded for identification. Used in clinical settings by surgeons during hip replacement procedures; the trial necks assist in intraoperative sizing and positioning. By confirming the appropriate implant size before final placement, the device helps optimize joint reconstruction and patient outcomes.

Clinical Evidence

No clinical data was required to support the safety and effectiveness of the devices. Biocompatibility testing was conducted in accordance with ISO 10993:2009.

Technological Characteristics

Materials: Polyphenylsulfone (PPSU) with 6% barium sulfate radiopacifier. Form factor: Modular trial necks in various offsets and anteversion angles. Connectivity: None. Sterilization: Non-sterile (reusable).

Indications for Use

Indicated for individuals undergoing primary and revision hip surgery due to trauma, inflammatory joint disease (e.g., rheumatoid arthritis), or noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant, nonunion, femoral neck/trochanteric fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity correction) where other treatments have failed.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113039 Page 1/2 5 2012 # Smith & Nephew, Inc. Summary of Safety and Effectiveness Radiopaque Trial Necks # Date of Summary: 05/29/2012 # Contact Person and Address Natalie P. Williams Sr. Regulatory Affairs Specialist Smith & Nephew, Inc. 7135 Goodlett Farms Parkway Cordova, TN 38016 (901) 399-5161 Name of Device: Radiopaque Trial Necks Common Name: Surgical Instrumentation Device Classification Name and Reference: 21 CFR 888.3353 - Hip joint metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prosthesis; 21 CFR 888.3358 - Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Device Class: II Panel Code: Orthopaedics/87 LPH, MEH Predicate Devices: SMF Hip Stem (K080625), MDF Revision Hip System (K081124), Promos Reverse Shoulder System Trial Inserts (K081016) #### Device Description Subject of this Traditional 510(k) are the Smith & Nephew Radiopaque Trial Necks. The Radiopaque Trial Necks are intended to be used to determine the correct modular neck implant option for the femoral stems developed as part of the Smith & Nephew SMF Hip Stem System and MDF Revision Hip System cleared via premarket notifications K080625 and K081124, respectively. The subject devices are manufactured from polyphenylsulfone (PPSU) with 6% of the radiopacifier barium sulfate, which makes them visible on an x-ray in the instance that they become dislodged during the trialing process. Colorants are blended with the PPSU resin to differentiate the various radiopaque trial necks. The subject devices are available in the options provided in Table 1 below. | Component Description | Color | |-----------------------------------------|---------| | Radiopaque Trial Neck Standard Offset | Natural | | Radiopaque Trial Neck High Offset | Natural | | Radiopaque Trial Neck Left Anteverted | Green | | Radiopaque Trial Neck Right Anteverted | Red | | Radiopaque Trial Neck High Offset +10MM | Natural | | | | | Table 1: Radiopaque Trial Neck Options | | | | |--|--|--|----------------------------------------|--|--|--| |--|--|--|----------------------------------------|--|--|--| #### Intended Use Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion, femoral neck {1}------------------------------------------------ K 113039 Page 2/2 fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteolomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. ## Performance Data Biocompatibility testing has been conducted on the subject devices according to ISO 10993:2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process. A review of the testing has demonstrated that there are no new issues related to the safety or effectiveness of the Radiopaque Trial Necks. Clinical data was not needed to support the safely and effectiveness of the devices. ### Substantial Equivalence Information The Radiopaque Trial Necks have the same indications for use as the Smith & Nephew SMF Hip Stem System and MDF Revision Hip System. The material and sterilization of the subject devices is substantially equivalent to the Trial Inserts included in the Smith & Nephew Promos Reverse Shoulder System cleared by K081016. The design features of the Radiopaque Trial Necks have been compared to the previously cleared devices in Table 2 below. | Feature | Subject Radiopaque<br>Trial Necks - K113039 | Predicate SMF<br>Hip Stem System<br>- K080625 | Predicate MDF<br>Revision Hip Stem<br>System - K081124 | Promos Reverse<br>Shoulder System<br>Trial Inserts -<br>K081016 | |-----------------------------|---------------------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------| | Same Indications<br>for Use | Yes | Yes | Yes | No | | Implant/Instrument | Hip Trial Instruments | Hip Implants | Hip Implants | Shoulder Trial<br>Instruments | | Material | Polyphenylsufone (PPSUI)<br>with 6% Barium Sulfate<br>(Color Added) | Ti-6Al-4V - Stem<br>Cobalt Chrome -<br>Modular Necks | Ti-6Al-4V - Stem and<br>Modular Sleeves<br>Cobalt Chrome -<br>Modular Necks | Polyphenylsufone<br>(Color added) | | Sterilization | Non-sterile | Gamma<br>Sterilization | Gamma Sterilization | Non-sterile | | Single Use or<br>Reusable | Reusable | Single Use | Single Use | Reusable | #### Table 2: Comparison to Predicate Devices #### Conclusion This Traditional 510(k) premarket notification is being submitted to request clearance for the Smith & Nephew Radiopaque Trial Necks. Based on the similarities to the predicate devices and acceptable biocompatibility testing, the subject devices are considered substantially equivalent to the devices currently marketed under premarket notifications K080625, K081124, and K081016. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT . 5 2012 Smith & Nephew, Incorporated % Ms. Natalie P. Williams Sr. Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova, Tennessee 38016 Re: K113039 Trade/Device Name: Radiopaque Trial Necks Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prosthesis; Regulatory Class: Class II Product Code: LPH, MEH Dated: October 4, 2012 Received: October 5, 2012 Dear Ms. Williams: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 – Ms. Natalie P. Williams CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, } Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K113039 Indications for Use 510(k) Number (if known): Device Name: Smith & Nephew Radiopaque Trial Necks ## Indications for Use: Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Prescription Use Over-The-Counter Use × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ht (Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices Page 1 of A 510(k) Number K113039
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%