K Number
K112997

Validate with FDA (Live)

Date Cleared
2011-12-17

(71 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HD-IVUS Ultrasound Imaging System (Console) is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.

Device Description

The HD-IVUS Ultrasound Imaging System is a minimally invasive intravascular ultrasound diagnostic imaging tool when used in conjunction with an intravascular ultrasound catheter. The System is a medical device for use by or on the order of a physician. The HD-IVUS Ultrasound Imaging System is comprised of a System Console with a touch screen monitor, a Roll Stand, a Patient Interface Module (PIM), a Power Supply, and an Intravascular Ultrasound Catheter (510(k) cleared and sold separately). The Catheter emits sound energy from a transducer at the distal tip of the catheter, which is guided into the coronary arteries of the heart. Sound waves that reflect from the inner vascular tissues are received bv the transducer and sent to the System Console where a high resolution, cross-sectional image is displayed in real time. The technique provides for in-vivo visualization of the coronary artery lumen, coronary artery wall morphology, and devices (such as stents) at or near the surface of the coronary artery wall.

AI/ML Overview

The provided text describes the 510(k) Summary for the Silicon Valley Medical Instruments (SVMI) HD-IVUS Ultrasound Imaging System. It focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a study proving device performance against those criteria as would be expected for an AI/CADe device.

Specifically, this document is a 510(k) summary for an imaging system, not an AI or CADe device. Therefore, it does not include acceptance criteria in terms of clinical performance metrics (like sensitivity, specificity, AUC) or a dedicated clinical study to prove the device meets such criteria. Instead, the "performance criteria" are focused on regulatory compliance and technical equivalence to a predicate device.

However, I can extract the information that is present and indicate where information is not available due to the nature of the submission.


1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Standard/Test/FDA Guidance)Reported Device Performance (Results Summary)
Electromagnetic CompatibilityIEC 60601-1-2Compliant with IEC 60601-1-2
Electrical Safety and Acoustic SafetyIEC 60601-1, IEC 60601-1-4, and IEC 60601-2-37Compliant with IEC 60601-1, IEC 60601-1-4, and IEC 60601-2-37
Acoustic OutputDoes not exceed Track 3 limits (Predicate device specified)Does not exceed Track 3 limits
Intended UseEquivalent to predicate device (Boston Scientific iLab System)Intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures. (Identical to predicate's intended use description, though phrasing differs slightly)
Signal TypeDoppler, Echo (Predicate)Doppler, Echo
Catheter Mechanical DrivePatient Interface Module (PIM) (cleared with ultrasound system)Motor Drive Unit (MDU) (cleared with ultrasound system) - Note: This is a design difference, not a performance criterion per se, but part of the equivalence claim.
Signal ConductorCoaxial wire traveling down catheter (Predicate)Coaxial wire traveling down catheter
Signal Generator/ Receiver MaterialsPZT crystal transducer (Predicate)PZT crystal transducer
Ultrasound Frequency40 MHz, 60 MHz ± 10% (Predicate)10 - 40 MHz ± 10% (Predicate specified 40 MHz, 60 MHz ± 10%; New device states 10-40MHz ± 10%. This is a significant difference in range, but the summary does not explain why it's considered equivalent.)
Method of UseIntravascular (Predicate)Intravascular
User InterfacePC-based system with keyboard and touch screen and mouse (Predicate)PC-based system with keyboard and touch screen and mouse
Software InterfaceCustom GUI (Predicate)Custom GUI

2. Sample size used for the test set and the data provenance

  • Not Applicable. This document describes a 510(k) submission for an imaging device (ultrasound system), not an AI/CADe device that would typically involve a "test set" of patient data for performance evaluation. The "studies" mentioned refer to technical tests for regulatory compliance (EMC, electrical/acoustic safety) and comparison with a predicate device's technical specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. See point 2. No clinical "ground truth" establishment by experts for a test set is described.

4. Adjudication method for the test set

  • Not Applicable. See point 2.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI/CADe device and no MRMC study is mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This device is an ultrasound imaging system, not an algorithm, and is intended for use by a physician.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not Applicable. The "ground truth" in this context is the compliance with established electrical, acoustic, and electromagnetic standards, and the technical specifications of the legally marketed predicate device used for substantial equivalence comparison.

8. The sample size for the training set

  • Not Applicable. This is an imaging device, not an AI/machine learning algorithm that requires a "training set."

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

KI12997

DEC 1 7 2011

Traditional 510(k) (new device) Silicon Valley Medical Instruments, Inc. 510(k) Summary CONFIDENTIAL

Image /page/0/Picture/3 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters SVMII in a bold, sans-serif font. The words "Silicon Valley Medical Instruments Inc." are printed in a smaller font below the letters.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

Submitter Information
Name:Silicon Valley Medical Instruments, Inc.
Address:47697 Westinghouse Drive, Suite 101, Fremont, CA 94539-7401 USA
Phone Number:510-897-4695
Fax Number:510-226-1230
Establishment Registration Number:N/A - Not yet registered.
Contact PersonRichard E. Anderson; VP, RA/QA (richarda@svmii.com)
Date Prepared7 October 2011
Name of Device
Trade or Proprietary NameHD-IVUS Ultrasound Imaging System
Common or Usual NameSystem, Imaging, Pulsed Doppler, Ultrasonic (IYN); System, Imaging, Pulsed Echo, Ultrasonic (IYO)
Classification NameClass II
Classification PanelCardiovascular
Regulation21 CFR 892.1550 (IYN); 21 CFR 892.1560 (IYO)
Product Code(s)IYN; IYO
Legally Marketed Device(s) to which Equivalence is ClaimedBoston Scientific Corporation's iLab Ultrasound Imaging System (Cleared 14Jul05 via K051679)
Reason for SubmissionNew device
Device DescriptionThe HD-IVUS Ultrasound Imaging System is a minimally invasive intravascular ultrasound diagnostic imaging tool when used in conjunction with an intravascular ultrasound catheter. The System is a medical device for use by or on the order of a physician. The HD-IVUS Ultrasound Imaging System is comprised of a System Console with a touch screen monitor, a Roll Stand, a Patient Interface Module (PIM), a Power Supply, and an Intravascular Ultrasound Catheter (510(k) cleared and sold separately). The Catheter emits sound energy from a transducer at the distal tip of the catheter, which is guided into the coronary arteries of the heart. Sound waves that reflect from the inner vascular tissues are received by the transducer and sent to the System

{1}------------------------------------------------

Traditional 510(k) (new device) Silicon Valley Medical Instruments, Inc. 510(k) Summary CONFIDENTIAL

Image /page/1/Picture/1 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters "SVMI" in a bold, sans-serif font. The words "Silicon Valley Medical Instruments Inc." are printed in a smaller font below the letters. The letters and words are all in black.

Console where a high resolution, cross-sectional image is displayed in real time. The technique provides for in-vivo visualization of the coronary artery lumen, coronary artery wall morphology, and devices (such as stents) at or near the surface of the coronary artery wall. System Console The HD-IVUS system features a point-of-care system console that is mounted to a roll stand. The System Console is the central component that is responsible for: managing the generation and display of IVUS images . · storing and exporting the image data · providing the main user interface · providing analysis tools (i.e., area measurements, linear measurements, and annotations) The System Console receives, processes, displays, and records ultrasound image data from the transducer in the Kodama HD-IVUS catheter. The IVUS images are displayed on a high-resolution, 19inch, flat panel, touch screen monitor. The monitor serves as the graphical user interface (GUI) for operating the system. All system information and controls are located on the touch screen monitor. The HD-IVUS system has internal storage for 25 studies. Each study can have up to eight loops, with a maximum of 9000 frames per loop. Study data is saved in DICOM format and can be exported to DVD, using the integrated DVD drive, or to a USB storage device, using one of the two USB ports. Roll Stand . The variable height roll stand supports the System Console and enables the console to be attached to a mobile stand that can be rolled up to the patient bed. Patient Interface Module (PIM) The Patient Interface Module (PIM) provides the electromechanical interface between the catheter and the system console. The PIM provides the mechanical interface to secure the catheter, as well as the mechanical energy to rotate the catheter's imaging assembly. The PIM also provides the electrical interface that transmits the

{2}------------------------------------------------

Traditional 510(k) (new device) Silicon Valley Medical Instruments, Inc. 510(k) Summary CONFIDENTIAL

·

Image /page/2/Picture/1 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters SVMII in a bold, sans-serif font. The words "Silicon Valley Medical Instruments Inc." are written in a smaller font below the letters SVMII. The letters are in black, and the background is white.

i

signal from the integrated ultrasound generator to the catheter andreceives the return signal. When placed in a sterile bag, the PIMcan be situated near the patient in the sterile field. The PIMfeatures an imaging on/off button and a recording on/off button,which enable the physician to start or stop imaging and recordingfrom within the sterile field. An orange LED light on the PIMindicates active recording.
Power Supply
The system includes a universal, medical grade, external powersupply that enables direct connection to the hospital power source.
Intravascular Ultrasound Catheter (510(k) cleared and soldseparately)
The Kodama Intravascular Ultrasound Catheter is a minimallyinvasive intravascular ultrasound coronary imaging catheter. Thecatheter emits acoustic energy from a transducer at its distal tip,which is guided into the coronary arteries of the heart. Soundwaves that are reflected from vascular tissues are received by thetransducer and sent through the PIM to the system console. Thecatheter can be operated at two different frequencies, 40 MHz or60 MHz, depending on user preference. An integrated telescopeallows the imaging of multiple regions of interest in a singleprocedure by advancing or retracting the imaging assembly withoutmoving the catheter sheath. (NOTE: Refer to the Directions for Usesupplied with the Kodama Intravascular Ultrasound Catheter foradditional information.)
Intended Use of the DeviceThe HD-IVUS Ultrasound Imaging System (Console) is intended forultrasound examination of coronary intravascular pathology only.
Indications for UseThe HD-IVUS Ultrasound Imaging System (Console) is intended forultrasound examination of coronary intravascular pathology only.Intravascular ultrasound imaging is indicated in patients who arecandidates for transluminal coronary interventional procedures.
Summary of the Technological Characteristics of the Device Compared to the Predicate Device
CharacteristicPredicateNew Device
(Boston Scientific Corporation'siLab Ultrasound Imaging System– Cleared 14Jul05 via K051679)(SVMI's HD-IVUS UltrasoundImaging System)

·

{3}------------------------------------------------

Traditional 510(k) (new device) Silicon Valley Medical Instruments, Inc. 510(k) Summary CONFIDENTIAL

Image /page/3/Picture/1 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters SVM in a bold, sans-serif font. The words "Silicon Valley Medical Instruments Inc." are printed in a smaller font below the letters.

Signal TypeDoppler, EchoDoppler, Echo
Catheter Mechanical DrivePatient Interface Module (PIM)(cleared with ultrasound system)Motor Drive Unit (MDU)(cleared with ultrasound system)
Signal ConductorCoaxial wire traveling downcatheterCoaxial wire traveling down catheter
Signal Generator/ ReceiverMaterialsPZT crystal transducerPZT crystal transducer
Acoustic OutputDoes not exceed Track 3 limitsDoes not exceed Track 3 limits
Ultrasound Frequency40 MHz, 60 MHz ± 10%10 - 40 MHz ± 10%
Method of UseIntravascularIntravascular
User InterfacePC-based system with keyboardand touch screen and mousePC-based system with keyboard andtouch screen and mouse
Software InterfaceCustom GUICustom GUI
Intended UseThe HD-IVUS Ultrasound ImagingSystem (Console) is intended forultrasound examination ofcoronary intravascular pathologyonly. Intravascular ultrasoundimaging is indicated in patientswho are candidates fortransluminal coronaryinterventional proceduresThe iLab Ultrasound Imaging Systemis intended for ultrasoundexaminations of intravascularpathology. Intravascular ultrasoundis indicated in patients who arecandidates for transluminalinterventional procedures such asangioplasty and atherectomy.
Performance Data
Summary of Non-clinical tests Conducted for Determination of Substantial Equivalence
Performance Test Summary - New Device
CharacteristicStandard/Test/FDA GuidanceResults Summary
ElectromagneticCompatibilityIEC 60601-1-2.Compliant with IEC 60601-1-2.
Electrical Safety andAcoustic SafetyIEC 60601-1, IEC 60601-1-4, andIEC 60601-2-37.Compliant with IEC 60601-1, IEC60601-1-4, and IEC 60601-2-37.

Conclusions Drawn from Design Verification/Validation and Non-clinical Data The HD-IVUS Ultrasound Imaging System is substantially equivalent in design and technology to the predicate device(s) with regard to intended use.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo of the Department of Health & Human Services USA. The logo consists of a circular seal on the left, containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the seal is a stylized emblem resembling an abstract human figure. To the right of the seal, the word "DEPARTMENT" is printed in bold, large font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Richard E. Anderson VP, RA/QA Silicon Valley Medical Instruments, Inc. 47697 Westinghouse Drive, Suite 101 FREMONT CA 94539-7401

DEC 1 7 2011

Re: K112997

Trade/Device Name: HD-IVUS Ultrasound Imaging System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN and IYO Dated: October 7, 2011 Received: October 7, 2011

Dear Mr. Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{5}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice medical device returns as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirements as set form in the quality -device as described in your Section 510(k) premarket will anow you to begin marketing your intial equivalence of your device to a legally marketed noticate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and If you desire spositional weeks your viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to under the MDK regulation (21 Of RT Pur 005), product of and the for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may other butter generarers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number: K112997

Device Name: SVMI's HD-IVUS Ultrasound Imaging System

Indications for Use:

:

Silicon Valley Medical Instrument's HD-IVUS Ultrasound Imaging System (Console) is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Aruld D'Aun

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) Number: K112997

Page 1 of 1

{7}------------------------------------------------

Appendix G: Example Diagnostic Ultrasound Indications For Use Format

SVMI HD-IVUS Console System: SVMI Kodama Catheter Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalImaging& OtherFetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Specify)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)P
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

  • Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.